K162598 · Bisco, Inc. · EBC · May 23, 2017 · Dental
Device Facts
Record ID
K162598
Device Name
HAPI Seal
Applicant
Bisco, Inc.
Product Code
EBC · Dental
Decision Date
May 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3765
Device Class
Class 2
Indications for Use
Pit and fissure sealant
Device Story
HAPI Seal is a light-cured, fluoride-containing, methacrylate-based resin sealant; applied to enamel pits and fissures of teeth to prevent decay. Used in dental clinics by dental professionals. Device is applied to tooth surface and cured using a dental light source for 20 seconds. Provides physical barrier and fluoride release to benefit patient dental health.
Clinical Evidence
Bench testing only. Evaluated per ISO 6874:2015 (Depth of Cure), ISO 29022:2013 (Shear Bond Strength), ISO 4049:2009 (Flexural Strength), and ANSI/ADA Spec No. 27 (Flexural Modulus, Compressive Strength). Biocompatibility assessed via ISO 7405:2008 and ISO 10993-1, including cytotoxicity and oral toxicity study (10 mice, 14 days), showing no toxicity.
Technological Characteristics
Methacrylate-based resin containing silica filler and fluoride. Light-cured (20-second cure time). Complies with ISO 6874:2015, ISO 4049:2009, ISO 29022:2013, and ANSI/ADA Spec No. 27. Form factor is a dental sealant material.
Indications for Use
Indicated for sealing enamel pits and fissures of teeth.
Regulatory Classification
Identification
A pit and fissure sealant and conditioner is a device composed of resin, such as polymethylmethacrylate, intended for use primarily in young children to seal pit and fissure depressions (faults in the enamel) in the biting surfaces of teeth to prevent cavities.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an emblem featuring a stylized depiction of three human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 23, 2017
Bisco, Inc. Ryan Hobson Regulatory Affairs Product Registration Coordinator 1100 West Irving Park Rd. Schaumburg, Illinois 60193
Re: K162598
Trade/Device Name: HAPI Seal Regulation Number: 21 CFR 872.3765 Regulation Name: Pit And Fissure Sealant And Conditioner Regulatory Class: Class II Product Code: EBC Dated: May 23, 2017 Received: May 23, 2017
Dear Ryan Hobson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/1/Picture/6 description: The image shows the signature of Mary S. Runner -S, followed by the signature of Tina Kiang, Ph.D. Above the signature of Mary S. Runner -S, the word "Sincerely" is written. To the right of the signature is a digital signature with the date 2017.05.23 12:25:35-04'00'.
Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162598
Device Name HAPI Seal
Indications for Use (Describe) Pit and fissure sealant
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BISCO. The logo consists of five connected hexagons, alternating between red and blue. Below the hexagons is the word "BISCO" in a bold, sans-serif font, with the letters in gray and black.
## K162598
# 510(k) SUMMARY
Bisco, Inc. 1100 W. Irving Park Road Schaumburg IL, 60193
Contact Person:
Applicant:
Ryan Hobson Tel: 847-534-6143 Fax: 847-534-6143
Date Prepared:
19 May 2017
Trade Name: Common Name: Product Code: Classification/Name: HAPI Seal Pit & Fissure Sealant ЕВС Pit and fissure sealant and conditioner Class II per 21 CFR 872.3765
### Predicate Devices:
HAPI Seal is substantially equivalent to:
Clinpro by 3M Company, K992326
### Indications for Use:
Pit and fissure sealant
### Description of Applicant Device:
HAPI Seal is a light-cured, fluoride containing, resin-based pit & fissure sealant designed for sealing the enamel pits and fissures of teeth.
### Comparison of Technological Characteristics:
#### Indications for use
HAPI Seal indications for use are identical to that of the predicate device and summarized in the table below:
| Clinpro<br>K992326 | HAPI Seal |
|-------------------------|-------------------------|
| Pit and fissure sealant | Pit and fissure sealant |
All components of HAPI Seal are based upon industry standard chemistry. The chemical composition of HAPI Seal is similar to Clinpro, methacrylate based chemistry, and is summarized in the table below:
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text 'K162598' in a bold, sans-serif font. The text is arranged horizontally, with each character clearly legible. The color of the text is black, contrasting with the white background.
Image /page/4/Picture/1 description: The image shows the BISCO company logo. The logo consists of a row of five connected hexagons, with the two outer hexagons being red and the middle hexagon being blue. The word "BISCO" is written in large, bold letters below the hexagons, with the letters being gray and black.
# 510(k) SUMMARY (continued)
| Chemical Composition | Clinpro | HAPI Seal |
|-----------------------|--------------------|--------------------|
| | K992326 | |
| Filler | Silica | Silica |
| Resin composition | Methacrylate Resin | Methacrylate Resin |
| Polymerization Method | light cured | light cured |
| lons Released | Fluoride | Fluoride |
| Physical Mechanical Property | Clinpro<br>K992326 | HAPI Seal |
|------------------------------|----------------------|----------------------|
| Depth of Cure | Meets ISO 6874:2015 | Meets ISO 6874:2015 |
| lons Released | Fluoride releasing | Fluoride releasing |
| Polymerization time | 20 second light cure | 20 second light cure |
| Bonding | Bonds to enamel | Bonds to enamel |
#### Performance Data:
HAPI Seal meets ISO 6874:2015 "Dentistry - Polymer-based pit and fissure sealants." Shear bond strength was tested based upon ISO 29022:2013 Dentistry - Adhesion - Notched-edge shear bond strength test. - Additionally, Light Curability, Ambient Light Sensitivity, Depth of Cure, Flexural Strength, Flexural Modulus, Cumulative Fluoride Release, Compressive Strength, Surface Hardness, and Shrinkage were tested and demonstrate HAPI Seal is substantially equivalent to the predicate device. A comparison of the physical/mechanical properties of HAPI Seal to the predicate device is provided in the table below.
| Physical / Mechanical Property | HAPI Seal |
|-----------------------------------------------------------|-----------------------------------------------------------------------------|
| Shear Bond Strength<br>(Modified ISO 29022) | HAPI Seal is equivalent to the predicate. |
| Light Curability (Cure time)<br>Ambient Light Sensitivity | HAPI Seal is equivalent to the predicate. |
| Depth of Cure<br>(ISO 6874:2015) | HAPI Seal meets the requirements of ISO 6874:2015. |
| Flexural Strength<br>(ISO 4049:2009) | HAPI Seal meets the requirements of ISO 4049:2009 for<br>Flexural Strength. |
| Flexural Modulus<br>(ANSI/ADA Spec No. 27) | HAPI Seal is equivalent to the predicate |
| Cumulative Fluoride Release | HAPI Seal is equivalent to the predicate. |
| Compressive Strength<br>(ANSI/ADA Spec No. 27) | HAPI Seal is equivalent to the predicate. |
| Surface Hardness<br>(Barcol-Colman 934-1) | HAPI Seal is equivalent to the predicate. |
| Shrinkage (Volumetric) | HAPI Seal is equivalent to the predicate. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text 'K162598' in a bold, sans-serif font. The text is black against a light, slightly textured background. The numbers and letters are evenly spaced and clearly legible.
Image /page/5/Picture/1 description: The image shows the logo for BISCO. The logo consists of a series of connected hexagons in red and blue. Below the hexagons is the word "BISCO" in large, bold, gray letters.
# 510(k) SUMMARY (continued)
### Biocompatibility:
An evaluation of biocompatibility was conducted using ISO 7405:2008 and ISO 10993-1 to determine the biocompatibility of HAPI Seal. Cytotoxicity and Oral Toxicity testing was completed. It is concluded from the biocompatibility evaluation and the results of Oral Toxicity Study (10 mice, 14 days) that HAPI Seal was not toxic in this test.
### Conclusion:
It is concluded from review of the predicate device indications, chemical composition, biocompatibility, and physical properties that HAPI is substantially equivalent to the predicate devices.
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.