← Product Code [LPN](/productcode/LPN) · K162597

# Clear Care Cleaning & Disinfecting Solution, AOCup Lens Case with AODisc (K162597)

_Alcon Laboratories, Inc. · LPN · Oct 19, 2016 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162597

## Device Facts

- **Applicant:** Alcon Laboratories, Inc.
- **Product Code:** [LPN](/productcode/LPN.md)
- **Decision Date:** Oct 19, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 886.5928
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

For use in simultaneous cleaning, daily protein removal, disinfecting, and storage of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) or rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses as recommended by your eye care practitioner.

## Device Story

System consists of AOCup lens case with integrated AODisc; used for contact lens care. User places lenses in baskets; fills cup with hydrogen peroxide solution. Platinum-coated disc neutralizes hydrogen peroxide into water and oxygen gas; ensures safe lens storage. Device used by patients at home. Modifications include colorant changes and minor basket geometry adjustments; principle of operation remains identical to predicate. Benefits include effective cleaning, protein removal, and disinfection of lenses.

## Clinical Evidence

Bench testing only. Evaluated via biocompatibility, physicochemical, design functionality, and microbiological tests. Results support substantial equivalence.

## Technological Characteristics

Specialized lens case with transparent cup, screw cap, lens baskets, and platinum-coated neutralization disc. Platinum catalyst facilitates chemical neutralization of hydrogen peroxide. Materials include plastic components and platinum coating. No software or electronic components.

## Regulatory Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

## Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

## Predicate Devices

- AOCup Lens Case with AODisc (PMA [P820040](/device/P820040.md))
- Clear Care Cleaning & Disinfecting Solution ([K003345](/device/K003345.md))
- Clear Care Cleaning & Disinfecting Solution ([K013512](/device/K013512.md))
- Clear Care Cleaning & Disinfecting Solution ([K022687](/device/K022687.md))
- Clear Care Cleaning & Disinfecting Solution ([K023455](/device/K023455.md))
- Clear Care Cleaning & Disinfecting Solution ([K030522](/device/K030522.md))
- Clear Care Cleaning & Disinfecting Solution ([K031521](/device/K031521.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 19, 2016

Alcon Laboratories, Inc. % Mr. Ralf Finke Senior Regulatory Specialist, Alcon Vision Care CIBA Vision GmbH Industriering 1 Grosswallstadt, Bavaria, DE 63868

Re: K162597

Trade/Device Name: Clear Care Cleaning & Disinfecting Solution, AOCup Lens Case with AODisc Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated: September 16, 2016 Received: September 19, 2016

Dear Mr. Finke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

Page 2 - Mr. Ralf Finke

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162597

Device Name

AOSEPT Clear Care Cleaning and Disinfecting Solution (AOCup Lens Case with AODisc)

Indications for Use (Describe)

For use in simultaneous cleaning, daily protein removal, disinfecting, and storage of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) or rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses as recommended by your eye care practitioner.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 8 510(K) SUMMARY

This summary document is being prepared in accordance with section 21 CFR 807.92(c).

## Submitter Information: A.

| Company:                | Alcon Laboratories, Inc.           |
|-------------------------|------------------------------------|
|                         | 6201 South Freeway                 |
|                         | Fort Worth, TX 76134-2099, USA     |
| Primary Contact Person: | Ralf Finke                         |
| Phone:                  | (Germany) +49 (6022) 240.515       |
| Fax:                    | (Germany) +49 (6022) 240.512       |
| Email:                  | ralf.finke@alcon.com               |
| Back up Contact Person: | Sherri Lakota                      |
| Phone:                  | (USA, Fort Worth, TX) 817-615-5472 |
| Fax:                    | (USA, Fort Worth, TX) 817-551-4630 |
| Email:                  | sherri.lakota@alcon.com            |
| Date Prepared:          | September 2016                     |

### B. Device Name:

| Trade Names:           | AOSEPT Clear Care Cleaning and Disinfecting Solution<br>(AOCup Lens Case with AODisc)                                                            |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:           | Contact Lens Accessory                                                                                                                           |
| Classification Name:   | Ophthalmic                                                                                                                                       |
| Device Classification: | Class II (21 CFR 886.5928) Soft (hydrophilic) contact lens<br>care products; (21 CFR 886.5918) Rigid gas permeable<br>contact lens care products |
| Product Code:          | LPN (Solution)                                                                                                                                   |

{4}------------------------------------------------

### C. Predicate Devices:

The legally marketed device(s) to which we are claiming substantial equivalence are:

| 510(k) Number                                                                                                                                     | Device                                                                                                          |
|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Previously approved in<br>PMA P820040,<br>Supplement S032, 23-Oct-<br>1995<br>(Reclassified to Class II in<br>1997)                               | AOCup Lens Case with AODisc                                                                                     |
| K003345, 26-Mar-2001;<br>K013512, 20-Dec-2001;<br>K022687, 19-Nov-2002;<br>K023455, 28-Feb-2003;<br>K030522, 12-Sep-2003;<br>K031521, 27-Jun-2003 | Clear Care Cleaning & Disinfecting Solution (contains the AOCup Lens Case with AODisc as part of the<br>system) |

### Device Description: D.

The two color AOCup Lens Case with AODisc is equivalent to the predicate device except modifications of colorants and minor modifications to basket geometry.

It is a specialized lens case consisting of a transparent cup with a connected unit of screw cap, lens holders (baskets) and platinum-coated neutralization disc. The platinum catalyst coating and shape of the plastic disc are designed to effectively neutralize the strong oxidizing agent hydrogen peroxide to harmless water and oxygen gas.

{5}------------------------------------------------

The modifications of colorants and basket geometry do not change any indications for use nor the basic technical principle of the device functions.

#### E. Indications for Use:

For use in simultaneous cleaning, daily protein removal, disinfecting, and storage of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) or rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses as recommended by your eye care practitioner.

#### F. Brief Summary of Nonclinical Test and Results:

The two color AOCup Lens Case with AODisc was evaluated in biocompatibility, physicochemical, design functionality and microbiological tests. Successful results of all nonclinical testing supported the substantial equivalence and therefore safety and efficacy of the modified AOCup Lens Case with AODisc to the existing product for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K162597](https://fda-staging.innolitics.com/device/K162597)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
