ALOKA LISENDO 880 is intended for use by trained personnel (doctor, sonographer, etc.) for the diagnostic ultrasound evaluation of Fetal, Abdominal, Pediatric, Small Organ (Spec.), Intra-operative (Spec.), Neonatal Cephalic, Musculo-skel. (Convent./Superfic.), Peripheral vessel, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (card), Adult Cephalic clinical applications. The Modes of Operation are B mode, PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler), Color Doppler, Power Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging.
Device Story
The ALOKA LISENDO 880 is a diagnostic ultrasound system used by clinicians (doctors, sonographers) in clinical settings. It utilizes ultrasound transducers to transmit energy into the body and detect reflected echoes. A computer system processes these signals to generate images or Doppler data, which are displayed on a video monitor. The device supports multiple modes, including B-mode, M-mode, Pulsed/Continuous Wave Doppler, Color/Power Doppler, Tissue Doppler Imaging, and 3D/4D imaging. Clinicians use the output to perform diagnostic evaluations and fluid flow analysis. The system aids in clinical decision-making by providing real-time visualization of internal structures and hemodynamics, facilitating procedures such as biopsies and general diagnostic assessments.
Clinical Evidence
No clinical data was required or provided. Substantial equivalence is supported by bench testing, including acoustic output measurements, biocompatibility, electromagnetic compatibility, and electrical/mechanical safety testing in accordance with applicable standards.
Technological Characteristics
Diagnostic ultrasound system with convex, linear, and sector transducers. Features include B, M, PW, CW, Color/Power Doppler, TDI, and 3D/4D imaging. Connectivity includes DICOM (SR, QR). Biocompatibility per ISO 10993-1. System is designed for re-usable probes with defined cleaning/disinfection protocols. Operates under Track 3 FDA limits.
Indications for Use
Indicated for diagnostic ultrasound evaluation of fetal, abdominal, pediatric, small organ, intra-operative, neonatal cephalic, musculoskeletal, peripheral vessel, cardiac (adult/pediatric), trans-esophageal, and adult cephalic applications in adult and pediatric patients.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2016
Hitachi, Ltd. % Ms. Angela Van Arsdale RA/QA Manager Hitachi Aloka Medical America, Inc. 10 Fairfield Boulevard WALLINGFORD CT 06492-5903
Re: K162583
Trade/Device Name: ALOKA LISENDO 880 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 29, 2016 Received: September 30, 2016
Dear Ms. Van Arsdale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162583
Device Name ALOKA LISENDO 880
### Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
ALOKA LISENDO 880 is intended for use by trained personnel (doctor, sonographer, etc.) for the diagnostic ultrasound evaluation of Fetal, Abdominal, Pediatric, Small Organ (Spec.), Intra-operative (Spec.), Neonatal Cephalic, Musculo-skel. (Convent./Superfic.), Peripheral vessel, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (card), Adult Cephalic clinical applications.
The Modes of Operation are B mode, PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler), Color Doppler, Power Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Device Name: ALOKA LISENDO 880
| Clinical Application | | | | Mode of Operation | | | | |
|---------------------------|------------------------------------|----|----|-------------------|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| Fetal Imaging<br>& Other | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | Pb | Pb | Pb | Pb | Pb | Pb | Pb |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | Pd | Pd | Pd | Pd |
| | Neonatal Cephalic | P | P | P | P | P | P | P |
| | Adult Cephalic | P | P | P | P | P | P | P |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | Pg | Pg | Pg | Pg | Pg | Pg | Pg |
| | Musculo-skel. (Convent.) | P | P | P | P | P | P | P |
| | Musculo-skel. (Superfic.) | P | P | P | P | P | P | P |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | P | P |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | P | P |
| | Trans-esophageal (Adult/Pediatric) | Pg | Pg | Pg | Pg | Pg | Pg | Pg |
| | Other (spec.) | | | | | | | |
| Peripheral | Peripheral vessel | P | P | P | P | P | P | P |
| Vessel | Other (spec.) | | | | | | | |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |
|--------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------|
N = new indication; P = previously cleared in K134016, K142368
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography). Tissue Doppler Imaging, Free Angular M-mode, 3D Imaging
Additional Comments:
Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Includes imaging for guidance of trans-vaginal biopsy. Subscript "f":
Subscript "g": For Adult and pediatric patients
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 2 of 17
{4}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: C251 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|--|-------------------|----|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | | P | P | P | P | P | P | P |
| Fetal Imaging<br>& Other | Abdominal | | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | | Pd | Pd | Pd | Pd | Pd | Pd | Pd |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (Wound) | | | | | | | | |
| | Other (Gynecological) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (spec.) | | | | | | | | |
N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 3 of 17
{5}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: C252 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|------------------------------------|----|----|-----|-----|------------------|--------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| Fetal Imaging<br>& Other | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | Pd | Pd | Pd | Pd |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | |
| Vessel | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 4 of 17
{6}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: C35 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| Fetal Imaging<br>& Other | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | Pd | Pd | Pd | Pd |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 5 of 17
{7}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: C42K Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application Mode of Operation B M PWD CWD Color Other** General Specific Combined* (Track I only) (Tracks I & III) Doppler (Spe (Spec.) Ophthalmic Ophthalmic Fetal Fetal Imaging Abdominal & Other Intra-operative (Spec.) Pb Pb Pb Pb Pb Pb Intra-operative (Neuro.) Laparoscopic Pediatric Small Organ (Spec.) Pd Pd Pd Pd Pd Pd Neonatal Cephalic P P P P P P Adult Cenhalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (Wound) Other (Gynecological) Cardiac Adult Cardiac Cardiac Pediatric Trans-esophageal (Adult/Pediatric) Other (spec.) Peripheral vessel Peripheral Vessel Other (spec.) N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Free Angular M-mode
Additional Comments:
Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Includes imaging for guidance of trans-rectal biopsy Subscript "e":
Includes imaging for guidance of trans-vaginal biopsy. Subscript "f":
Subscript "g": For Adult and pediatric patients
Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 6 of 17
{8}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: L34 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| Fetal Imaging<br>& Other | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | Pd | Pd | Pd | Pd |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | P | P | P | P |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | P | P |
| | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode
Additional Comments:
Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).
Includes imaging of organs and structures exposed during surgery and laparoscopic procedures). Subscript "b":
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 7 of 17
{9}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: L441 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | | Mode of Operation | | | | |
|---------------------------|------------------------------------|----|----|-------------------|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| Fetal Imaging<br>& Other | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | Pd | Pd | Pd | Pd |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | P | P | P | P |
| | Musculo-skel. (Superfic.) | P | P | P | P | P | P | P |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | P | P |
| | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016, K142368
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode
Additional Comments:
Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).
Subscript "b": Includes imaging of organs and structures exposed during surgery
- (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
- Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Includes imaging for guidance of trans-vaginal biopsy. Subscript "f":
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 8 of 17
{10}------------------------------------------------
# Device Name: ALOKA LISENDO 880 Transducer: L64
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation B M PWD CWD Color Other** General Specific Combined* (Track I only) (Tracks I & III) Doppler (Spe (Spec.) Ophthalmic Ophthalmic Fetal Fetal Imaging Abdominal Pa Pa Pa Pa Pa Pa Pa & Other Intra-operative (Spec.) Intra-operative (Neuro.) Laparoscopic Pediatric P P P P P P P Small Organ (Spec.) Pd Pd Pd Pd Pd Pd Pd Neonatal Cephalic Adult Cenhalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skel. (Convent.) P P P P P P P P Musculo-skel. (Superfic.) P P P P P P Intra-luminal Other (Wound) Other (Gynecological) Cardiac Adult Cardiac Cardiac Pediatric Trans-esophageal (Adult/Pediatric) Other (spec.) Peripheral vessel P P P P P P P Peripheral Vessel Other (spec.)
N = new indication; P = previously cleared in K134016, K142368
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Free Angular M-mode
Additional Comments:
Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures)
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Includes imaging for guidance of trans-rectal biopsy Subscript "e":
Includes imaging for guidance of trans-vaginal biopsy. Subscript "f":
Subscript "g": For Adult and pediatric patients
Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 9 of 17
{11}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: MXS1 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|----------------------|--------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | P | P | P | P | P | P | P | |
| Fetal Imaging<br>& Other | Abdominal | P | P | P | P | P | P | P | |
| | Intra-operative (Spec.) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P | |
| | Small Organ (Spec.) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | P | P | P | P | P | P | P | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (Wound) | | | | | | | | |
| | Other (Gynecological) | | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | P | P | P | P | |
| | Cardiac Pediatric | P | P | P | P | P | P | P | |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | P | P | |
| | Other (spec.) | | | | | | | | |
N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Dopler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode, 3D Imaging
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures)
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 10 of 17
{12}------------------------------------------------
#### Device Name: ALOKA LISENDO 880 Transducer: MXS2ESLL Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
intended use. Diagnostic ultrasound imaging of fluid flow analysis of the human body as follows.
| | Clinical Application | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| Fetal Imaging | Abdominal | | | | | | | |
| & Other | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Spec.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | Pg | Pg | Pg | Pg | Pg | Pg | Pg |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Adult/Pediatric) | Pg | Pg | Pg | Pg | Pg | Pg | Pg |
| | Other (spec.) | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | |
| Vessel | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016, K142368
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Dopler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode, 3D Imaging
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript g. For Adult and pediatric patients.
Subscript “h.” Includes imaging for Cavernous/Naso
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 11 of 17
{13}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: S121 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| Fetal Imaging | Abdominal | P | P | P | P | P | P | P |
| & Other | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | P | P | P | P | P | P | P |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | P | P |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | P | P |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral | Peripheral vessel | P | P | P | P | P | P | P |
| Vessel | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 12 of 17
{14}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: S31 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
intended use. Diagnostic ultrasound imaging of fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation |
|----------------------|-------------------|
|----------------------|-------------------|
| | Clinical Application<br>Mode of Operation | | | | | | | |
|---------------------------|-------------------------------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| Fetal Imaging<br>& Other | Abdominal | P | P | P | P | P | P | P |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | | | | | | | |
| | Neonatal Cephalic | P | P | P | P | P | P | P |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | P | P |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | P | P |
| | Trans-esophageal (Adult/Pediatric) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 13 of 17
{15}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: S3ESEL Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Intended use. Diagnostic ultrasound imaging of fluid flow analysis of the human body as follows.
| Clinical Application | | | Mode of Operation | | | | | |
|---------------------------|------------------------------------|----|-------------------|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| Fetal Imaging<br>& Other | Abdominal | | | | | | | |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Spec.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | Pg | Pg | Pg | Pg | Pg | Pg | Pg |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Wound) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Adult/Pediatric) | Pg | Pg | Pg | Pg | Pg | Pg | Pg |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N = new indication; P = previously cleared in K134016, K142368
* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW ** Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). Subscript "a":
Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures)
Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.
Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy. Subscript "d":
Subscript "e": Includes imaging for guidance of trans-rectal biopsy
Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.
For Adult and pediatric patients Subscript "g":
Subscript "h" Includes imaging for Cavernous/Non-Cavernous wounds
Prescription Use Only (Per 21 CFR 801.109)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Page 14 of 17
{16}------------------------------------------------
## Device Name: ALOKA LISENDO 880 Transducer: S42 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| Fetal Imaging<br>& Other | Abdominal | P | P | P | P | P | P | P |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.