← Product Code [IYE](/productcode/IYE) · K162577

# Dosimetry Check Version 5 Release 1 (K162577)

_Math Resolutions, LLC · IYE · Nov 18, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162577

## Device Facts

- **Applicant:** Math Resolutions, LLC
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Nov 18, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Dosimetry Check quality control software uses the radiation fields that are measured with media such as x-ray film, electronic portal imaging devices (EPID), diode or ion chamber arrays, or in the case of TomoTherapy, a fan line detector array, and provides a theoretical calculation. Dosimetry Check computes the dose distribution using the patient specific CT or other image set or alternately a phantom that is likewise scanned, to calculate the reconstructed dose that is then compared to the plan dose. The results reported can include the computed percent difference at specific points as compared to the patient specific radiation treatment plan. Dosimetry Check does not provide any conclusions regarding the comparisons and does not provide any criteria to be used for interpreting the results. The experienced radiological physicist can reevaluate his patient specific radiation treatment plan in accordance with his clinical judgment. This product is not a treatment planning system and is not to be used as one. This product only checks the applied dose based on the measurement of each x-ray field applied to the patient and provided in an exported file, and a theoretical calculation. The product can only be used with photons (high energy x-rays) and cannot be used with electron or proton therapy. This product does not provide any quality assurance that the fields are in fact correctly aligned with the patient anatomy as planned. In addition, the product may be used to display the above dose on other fused image sets which could provide additional supportive quality information to the user regarding the correctness of treatment.

## Device Story

Standalone software for radiation therapy quality assurance. Inputs: measured radiation field data (x-ray film, EPID, diode/ion chamber arrays, or TomoTherapy fan beam detector) and patient-specific CT/image sets. Operation: computes reconstructed dose distribution using pencil beam or collapsed cone convolution/superposition algorithms; compares results against planned dose. Used in clinical settings by radiological physicists. Output: dose distribution maps, point-dose percent differences, and fused image displays. Clinical impact: provides independent verification of treatment plan delivery; assists physicists in evaluating plan correctness. Does not control delivery devices or provide clinical interpretation criteria.

## Clinical Evidence

Bench testing only. Internal validation compared results of the new collapsed cone (CC) algorithm against the previously cleared pencil beam (PB) algorithm. Regression testing performed to ensure system stability. No clinical data provided.

## Technological Characteristics

Standalone software; runs on Windows/Linux. Requires OpenGL-capable graphics card (24-bit true color, depth buffer); supports optional stereoscopic 3D displays (Nvidia Quadro FX). Utilizes GPU for computation. Algorithms: Pencil Beam and Collapsed Cone (CC) convolution/superposition. Connectivity: file-based import of radiation field measurements and treatment plans.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Dosimetry Check Version 4 Release 1 ([K132605](/device/K132605.md))
- Mobius3D ([K140660](/device/K140660.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 18, 2016

Math Resolutions, LLC % Mr. Wendel Renner President 5975 Gales Lane COLUMBIA MD 21045

Re: K162577

Trade/Device Name: Dosimetry Check Version 5 Release 1 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 17, 2016 Received: September 20, 2016

Dear Mr. Renner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D'Hara

For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162577

### Device Name

Dosimetry Check Version 5 Release 1

### Indications for Use (Describe)

Dosimetry Check quality control software uses the radiation fields that are measured with media such as x-ray film, electronic portal imaging devices (EPID), diode or ion chamber arrays, or in the case of TomoTherapy, a fan line detector array, and provides a theoretical calculation. Dosimetry Check computes the dose distribution using the patient specific CT or other image set or alternately a phantom that is likewise scanned, to calculate the reconstructed dose that is then compared to the plan dose. The results reported can include the computed percent difference at specific points as compared to the patient specific radiation treatment plan.

Dosimetry Check does not provide any conclusions regarding the comparisons and does not provide any criteria to be used for interpreting the results. The experienced radiological physicist can reevaluate his patient specific radiation treatment plan in accordance with his clinical judgment.

This product is not a treatment planning system and is not to be used as one. This product only checks the applied dose based on the measurement of each x-ray field applied to the patient and provided in an exported file, and a theoretical calculation. The product can only be used with photons (high energy x-rays) and cannot be used with electron or proton therapy. This product does not provide any quality assurance that the fields are in fact correctly aligned with the patient anatomy as planned. In addition, the product may be used to display the above dose on other fused image sets which could provide additional supportive quality information to the user regarding the correctness of treatment.

| Type of Use (Select one or both, as applicable)                            |
|----------------------------------------------------------------------------|
| <span> <span></span>Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
| <span> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary for Dosimetry Check Version 5 Release 1

| Submitter's Name<br>Address          | Math Resolutions, LLC<br>5975 Gales Lane<br>Columbia, MD USA 21045 |
|--------------------------------------|--------------------------------------------------------------------|
| Establishment<br>Registration Number | 1833314                                                            |
| Summary Date                         | 16 Nov 2016                                                        |
| Telephone Number                     | (410) 997-9578                                                     |
| Email                                | WDRenner@MathResolutions.com                                       |
| Contact Person                       | Wendel Dean Renner, President                                      |

| Name of the Device                  | Common or Usual Name                       | Classification Name                                                                        |
|-------------------------------------|--------------------------------------------|--------------------------------------------------------------------------------------------|
| Dosimetry Check version 5 release 1 | Standalone Software Quality Control System | Medical Charged-particle<br>Radiation Therapy Systems, 21 CFR 892.5050<br>Product Code IYE |

| Identification of the Legally Marketed Devices (Predicate Devices) |                   |
|--------------------------------------------------------------------|-------------------|
| Dosimetry Check Version 4 Release 1, K132605                       | Mobius3D, K140660 |

## Indications for Use

#### Dosimetry Check Version 5 Release 1 (Modified Device) ●

Dosimetry Check quality control software uses the x-ray radiation fields that are measured with media such as x-ray film, electronic portal imaging devices (EPID), diode or ion chamber arrays, or in the case of TomoTherapy, a fan line detector array, and provides a theoretical calculation. Dosimetry Check computes the dose and dose distribution using the patient specific CT or other image set or alternately a phantom that is likewise scanned, to calculate the reconstructed dose that is then compared to the plan dose. The results reported can include the computed percent difference at specific points as compared to the patient specific radiation treatment plan.

Dosimetry Check does not provide any conclusions regarding the comparisons and does not provide any criteria to be used for interpreting the results. The experienced radiological physicist can reevaluate his patient specific radiation treatment plan in accordance with his clinical judgment.

This product is not a treatment planning system and is not to be used as one. This product only checks the applied dose based on the measurement of each x-ray field applied to the patient and provided in an exported file, and a theoretical calculation. The product can only be used with photons (high energy x-rays) and cannot be used with electron or proton therapy. This product does not provide any quality assurance that the fields are in fact correctly applied to and correctly aligned with the patient anatomy as planned. In addition, the product may be used to display the above dose on other fused image sets which could provide additional supportive quality information to the user regarding the correctness of treatment.

{4}------------------------------------------------

#### . Dosimetry Check Version 4 Release 1. K132605 (Predicate Device)

Dosimetry Check is a standalone software product intended to be used by an experienced radiological physicist for quality control purposes only. Dosimetry Check is intended to check the correctness of x-ray treatment plans delivered from high energy charged-particle radiation therapy treatment machines by using a measurement of the applied radiation fields that are planned to be or have been applied to a patient, and computing the dose to the patient from the measured radiation fields. This product is to be used as a quality control check for the treatment planning system and delivery system.

Dosimetry Check quality control software uses the radiation fields that are measured with media such as x-ray film, electronic portal imaging devices (EPID), diode or ion chamber arrays, or in the case of TomoTherapy, a fan line detector array, and provides a theoretical calculation. Dosimetry Check computes the dose and dose distribution using the patient specific CT or other image set or alternately a phantom that is likewise scanned, to calculate the reconstructed dose that is then compared to the plan dose. The results reported can include the computed percent difference at specific points as compared to the patient specific radiation treatment plan.

Dosimetry Check does not provide any conclusions regarding the comparisons and does not provide any criteria to be used for interpreting the results. The experienced radiological physicist can reevaluate his patient specific radiation treatment plan in accordance with his clinical judgment.

This product is not a treatment planning system and is not to be used as one. This product only checks the applied dose based on the measurement of each x-ray field applied to the patient and provided in an exported file, and a theoretical calculation. This product does not provide any quality assurance that the fields are in fact correctly applied to and correctly aligned with the patient anatomy as planned. In addition, the product may be used to display the above dose on other fused image sets which could provide additional supportive quality information to the user regarding the correctness of treatment.

#### . Mobius3D, K140660 (Predicate Device)

Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.

## DEVICE DESCRIPTIONS

#### Dosimetry Check version 5 release 1 (Modified Device) .

Dosimetry Check is a standalone software product intended to be used by an experienced radiological physicist for quality control purposes only. Dosimetry Check is intended to check the correctness of x-ray treatment plans delivered from high energy charged-particle radiation therapy treatment machines by using a measurement of the applied radiation fields that are planned to be or have been applied to a patient, and computing the calculated dose to the patient from the measured radiation fields that is intended to be used as a quality control tool tothen be compared to the intended planned dose. This product is to be used as a quality control check for the treatment planning system and delivery system. Dosimetry Check is not a treatment planning system. It does not design or transmit instructions to a delivery device, nor does it control any other medical device. Being a software only quality control tool, Dosimetry Check never comes into contact with patients or persons. Dosimetry

{5}------------------------------------------------

Check is an analysis tool meant solely for quality assurance purposes when used by trained medical professionals.

{6}------------------------------------------------

### 510(k) Summary continued

#### ● Dosimetry Check version 4 release 1, K132605 (Predicate Device)

System 2100 for which 510(k) K993530 was cleared by the FDA on December 15, 1999 that is a medical image display system serves as a foundation that provides basic image display functionality for Dosimetry Check.

Dosimetry Check is a software program that will compute the dose and dose distribution to the patient from a measurement of the radiation fields that are applied to the patient. The dose so computed serves as a means to verify the correctness of the radiation treatment and to serve as a final sanity check. The radiation fields are measured with media such as x-ray film or electronic devices that will measure over the area of the field, such as electronic portal imaging devices (EPID), or diode or ion chamber arrays.

To extend Dosmetry Check to support the TomoTherapy machine, the device uses the data measured by the fan beam radiation detector that is part of the TomoTherapy machine. The detectors capture the radiation intensity periodically at predetermined gantry angles and couch positions, known as control points, from the treatment plan. The detector only measures the intensity across the center of the radiation beam in the transverse plane. A prior measured profile in the perpendicular longitudinal direction is then applied to complete the radiation field map. The radiation field map is then applied as a stationary beam at the center gantry angle and couch position for the integration period (between two control points), from which the dose to the patient is computed. The patient dose is then summed up from all such radiation field maps

#### ● Mobius3D, K140660 (Predicate Device)

Mobius3D is a software product used within a radiation therapy clinic for quality assurance and treatment plan verification. It is important to note that while Mobius3D operates in the field of radiation therapy, it is neither a radiation delivery device (e.g. a linear accelerator), nor is it a treatment planning system (TPS). Mobius3D cannot design or transmit instructions to a delivery device, nor does it control any other medical device. Mobius3D is an analysis tool meant solely for quality assurance (QA) purposes when used by trained medical professionals. Being a software-only OA tool, Mobius3D never comes into contact with patients

Mobius3D performs dose calculation verifications for radiation treatment plans by doing an independent calculation of radiation dose is initially calculated by a treatment planning system (TPS), which is a software tool that develops a detailed set of instructions (i.e. a plan) for another system (e.g. a linear accelerator) to deliver radiation to a patient: The dose calculation performed by Mobius3D uses a proprietary collapsed cone convolution superposition (COS) algorithm. Mobius3D also performs dose delivery quality assurance for radiation treatment plans by using the measured data recorded in a linear accelerator's delivery log files to calculate a delivered dose. This is presented to the end user in a software component of Mobius3D called MobiusFX. The MobiusFX component is available to users through licensing as an add-on to the core Mobius3D software features.

{7}------------------------------------------------

# Device Comparison Table

| Features                                                                                                                                                                                                                                    | Predicate<br>Dosimetry Check<br>Version 4 Release 1<br>K132605                                                                                                                                                                                                                                                                                         | Predicate<br>Mobius3D<br>K140660           | Modified Device<br>Dosimetry Check<br>Version 5<br>Release 1                                                                                                                                                                                                                                                                                                                                        |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pre-treatment images                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                    | No                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Exit images                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                    | No                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Pencil beam dose algorithm                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                    | No                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Convolution/superposition "collapsed cone"<br>algorithm (CC). Additional technical comparison<br>information regarding this added CC algorithm is<br>provided above.                                                                        | No                                                                                                                                                                                                                                                                                                                                                     | Yes                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Compute dose to patient                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                    | Yes                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Compare to planning system dose                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                    | Yes                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Used for verifying the correctness of radiation<br>therapy treatments                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                    | Yes                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| TomoTherapy: Uses a line in the transverse plane<br>through the radiation field measurement provided to<br>Dosimetry Check. A prior measured longitudinal<br>profile is applied to each detector signal to complete<br>the radiation field. | Yes                                                                                                                                                                                                                                                                                                                                                    | No, uses<br>linac log<br>files             | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Generates a report as described in the Dosimetry<br>Check manual using either the auto-report feature,<br>or the user may construct their own report using the<br>evaluate tools.                                                           | Yes                                                                                                                                                                                                                                                                                                                                                    | Provides a<br>report in<br>their<br>format | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Installed by downloading the software from the<br>Math Resolutions web site at<br>http://www.mathresolutions.com/downlprg.htm                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                    | unknown                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Photons (x-ray)                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                    | Yes                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Electrons                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                                                                                                                     | unknown                                    | No                                                                                                                                                                                                                                                                                                                                                                                                  |
| Protons                                                                                                                                                                                                                                     | No                                                                                                                                                                                                                                                                                                                                                     | unknown                                    | No                                                                                                                                                                                                                                                                                                                                                                                                  |
| Ability to use measured radiation field for<br>conventional linac in a pretreatment dry run without<br>the patient and measured radiation fields taken<br>during treatment.                                                                 | yes                                                                                                                                                                                                                                                                                                                                                    | No, uses<br>linac log<br>files             | yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Ability to use the TomoTherapy detector data<br>measured in a pretreatment dry run without the<br>patient and the detector data taken during treatment                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                    | No, uses<br>linac log<br>files             | Yes                                                                                                                                                                                                                                                                                                                                                                                                 |
| Features                                                                                                                                                                                                                                    | Predicate<br>Dosimetry Check<br>Version 4 Release 1<br>K132605                                                                                                                                                                                                                                                                                         | Predicate<br>Mobius3D<br>K140660           | Modified Device<br>Dosimetry Check<br>Version 5<br>Release 1                                                                                                                                                                                                                                                                                                                                        |
| Operating Systems                                                                                                                                                                                                                           | Microsoft Windows<br>XP, Windows Vista,<br>Windows 7, 10,and<br>Ubuntu 14.04 LTS<br>(Linux)                                                                                                                                                                                                                                                            | unknown                                    | Microsoft<br>Windows XP,<br>Windows Vista,<br>Windows 7,<br>10,and Ubuntu<br>14.04 LTS<br>(Linux)                                                                                                                                                                                                                                                                                                   |
| Hardware                                                                                                                                                                                                                                    | - Open GL capable<br>graphics card is<br>required with 24<br>true color and a<br>depth buffer.<br>- For added<br>stereoscopic three<br>dimensional<br>displays, an Nvidia<br>Quadro fx card that<br>supports stereo is<br>needed with a single<br>monitor capable of<br>120 Hertz refresh<br>rate or the Planar<br>Mirror System with<br>two monitors. | Graphics<br>processing<br>unit             | - Open GL<br>capable graphics<br>card is required<br>with 24 true<br>color and a depth<br>buffer.<br>- For added<br>stereoscopic<br>three<br>dimensional<br>displays, an<br>Nvidia Quadro fx<br>card that supports<br>stereo is needed<br>with a single<br>monitor capable<br>of 120 Hertz<br>refresh rate or the<br>Planar Mirror<br>System with two<br>monitors.<br>- Graphics<br>processing unit |

{8}------------------------------------------------

{9}------------------------------------------------

## 510(k) Summary continued

## Description and Conclusion of Testing

## Testing

Testing consisted of internal testing with and without the Graphic Processing Unit for the comparison of the results obtained using the Pencil Beam (PB) algorithm in the 510(k) cleared version of Dosimetry Check to the results obtained using the Convolution/Superposition "Collapsed Cone" Algorithm (CC) added in Dosimetry Check Version 5 Release 1 as well as regression testing. Successful testing was accomplished with internal testing. No issues were identified that required external beta testing.

## Conclusion:

The successful testing demonstrates the safety and effectiveness of the Dosimetry CheckVersion 5 Release 1 when used for the defined indications for use demonstrates that the device for which this 510(k) is submitted perform as well as or better than the legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K162577](https://fda-staging.innolitics.com/device/K162577)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
