FORTEZZA Spinal Stabilization System

K162576 · Osso Surgical · NKB · Nov 10, 2016 · Orthopedic

Device Facts

Record IDK162576
Device NameFORTEZZA Spinal Stabilization System
ApplicantOsso Surgical
Product CodeNKB · Orthopedic
Decision DateNov 10, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The FORTEZZA Spinal Stabilization System is intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor: pseudarthrosis; and/or failed previous fusion. In addition, when used as a pedicle screw fixation system, the FORTEZZA Spinal Stabilization System is intended for skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral, L5-S1 vertebra, who are receiving fusion by autogenous bone graft only, who are having the device attached to the lumbar and sacral spine (levels may be from L3 to the sacrum/ilum), who are having the device removed after the attainment of a solid fusion. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the FORTEZZA Spinal Stabilization System are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The FORTEZZA Spinal Stabilization System is intended to be used with autograft and/or allograft. Pedicle screw fixation is limited to a posterior approach.

Device Story

FORTEZZA Spinal Stabilization System; posterior, non-cervical spinal fixation system. Components include titanium alloy multi-axial screws, straight rods, and pre-bent rods. Surgeon-assembled construct; tailor-made for individual patient anatomy. Used in OR by orthopedic surgeons/neurosurgeons to provide rigid stabilization as adjunct to spinal fusion. Fixation achieved via posterior approach. Benefits include stabilization of spinal segments to facilitate fusion in patients with degenerative, traumatic, or deformity-related spinal conditions. System used with autograft/allograft.

Clinical Evidence

Bench testing only. Static compression bending, static torsion, and fatigue compression bending performed per ASTM F1717-14 and FDA guidance. Results demonstrate subject device performance meets or exceeds predicate devices.

Technological Characteristics

Materials: Titanium alloy. Components: Multi-axial screws, straight and pre-bent rods (5.5mm). Fixation: Posterior, non-cervical. Testing standards: ASTM F1717-14. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients and pediatric patients with adolescent idiopathic scoliosis requiring posterior, non-cervical spinal fixation as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis (including severe L5-S1 grades 3-4), trauma, spinal stenosis, spinal curvatures, tumor, pseudarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 10, 2016 OSSO Surgical % Ms. Kimberly Strohkirch Executive VP MRC-X, LLC 3416 Roxee Run Cove Bartlett, Tennessee 38133 Re: K162576 Trade/Device Name: FORTEZZA Spinal Stabilization System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI Dated: September 13, 2016 Received: September 15, 2016 Dear Ms. Strohkirch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K162576 Device Name FORTEZZA Spinal Stabilization System #### Indications for Use (Describe) The FORTEZZA Spinal Stabilization System is intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor: pseudarthrosis; and/or failed previous fusion. In addition, when used as a pedicle screw fixation system, the FORTEZZA Spinal Stabilization System is intended for skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral, L5-S1 vertebra, who are receiving fusion by autogenous bone graft only, who are having the device attached to the lumbar and sacral spine (levels may be from L3 to the sacrum/ilum), who are having the device removed after the attainment of a solid fusion. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the FORTEZZA Spinal Stabilization System are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The FORTEZZA Spinal Stabilization System is intended to be used with autograft and/or allograft. Pedicle screw fixation is limited to a posterior approach. Type of Use (Select one or both, as applicable) | <span style="font-family: Arial, sans-serif;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16" x="0" y="0"></rect> <path d="M2 2 L14 2 L14 14 L2 14 Z" style="stroke:black; stroke-width:1; fill:none"></path> <path d="M3 3 L13 3 L13 13 L3 13 Z" style="stroke:black; stroke-width:1; fill:none"></path> <path d="M4 4 L12 4 L12 12 L4 12 Z" style="stroke:black; stroke-width:1; fill:none"></path> <path d="M5 5 L11 5 L11 11 L5 11 Z" style="stroke:black; stroke-width:1; fill:none"></path> <path d="M6 6 L10 6 L10 10 L6 10 Z" style="stroke:black; stroke-width:1; fill:none"></path> <path d="M7 7 L9 7 L9 9 L7 9 Z" style="stroke:black; stroke-width:1; fill:none"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: Arial, sans-serif;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16" x="0" y="0"></rect> <path d="M2 2 L14 2 L14 14 L2 14 Z" style="stroke:black; stroke-width:1; fill:none"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary FORTEZZA Spine Stabilization System November 9, 2016 | Company: | OSSO Surgical<br>2150 West 43rd Street<br>Chicago, IL 60609 | |--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Primary Contact: | Kimberly Strohkirch<br>Phone: 901-361-2037<br>strohkirch@memphisregulatory.com | | Company Contact: | Basheer Hassan<br>Phone: 847-513-1774<br>basheerhassan@backbonesurgical.com | | Trade Name: | FORTEZZA Spinal Stabilization System | | Common Name: | Pedicle Screw Spinal System | | Classification: | Class III | | Regulation Number: | 21 CFR 888.3070 (Orthosis, Spondylolisthesis Spinal Fixation; Orthosis Spinal Pedicle Fixation) | | Panel: | 87- Orthopedic | | Product Code: | NKB, OSH, MNH, MNI | | Predicate Devices: | Primary Predicate Device:<br>K160466 – OrthoPediatrics Response 5.5 Spinal System<br><br>Additional Predicate Devices:<br>K013962 – Medtronic EQUATION™ Fixation System<br>K122332 – Captiva Spine TowerLOX Pedicle Screw System | {4}------------------------------------------------ ### Device Description: The FORTEZZA Spine Stabilization System consists of a variety of shapes and sizes of screws and 5.5 rods that can be rigidly locked in a variety of configurations, with each construct being tailor-made for the individual case. Fixation is provided via a posterior approach. The components are made from titanium alloy. Components of the system include straight and pre-bent rods, multi-axial screw. ### Indications for Use: The FORTEZZA Spine Stabilization System is intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion. In addition, when used as a pedicle screw fixation system, the FORTEZZA Spine Stabilization System is intended for skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral, L5-S1 vertebra, who are receiving fusion by autogenous bone graft only, who are having the device attached to the lumbar and sacral spine (levels may be from L3 to the sacrum/liium), who are having the device removed after the attainment of a solid fusion. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the FORTEZZA Spine Stabilization System are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. The FORTEZZA Spine Stabilization System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach. ### Substantial Equivalence: The FORTEZZA Spinal Stabilization System is substantially equivalent to the cleared predicate device, the OrthoPediatrics Response 5.5 Spine System (K160466) because it has similar indications for use, is composed of the same materials, and has similar technological characteristics. Differences between the subject and predicate devices include sizes of screws and rods. The differences in technological characteristics do not raise different questions of safety and effectiveness. In addition, the FORTEZZA Spinal Stabilization System meets or exceeds the predicate devices in ASTM F1717 bench tests. ### Performance Testing: Static compression bending, static torsion and fatigue compression bending were completed for the FORTEZZA Spinal Stabilization System according the ASTM F1717-14 and Guidance for Industry and FDA Staff: Spinal System 510(k)s issued May 3, 2004. Performance testing demonstrates that the subject device meets or exceeds performance of predicate devices demonstrating that the subject devices are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...