YES oil-based Personal Lubricant, YES Water-based Personal Lubricant, YES Vaginal Moisturizer
Applicant
Biotech Lubricants, Ltd.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Jun 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
YES® oil-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® oil-based Personal Lubricant is not compatible with natural rubber latex, polyurethane or polyisoprene condoms. YES® water-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® water-based Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. YES® Vaginal Moisturizer: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® Vaginal Moisturizer is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
Device Story
Personal lubricants/moisturizers for vaginal/penile application; intended to supplement natural lubrication and enhance comfort during sexual activity. Oil-based variant (sunflower oil/shea butter) and water-based variants (aloe vera/plant gums) provided in HDPE tubes. Applied topically by user. Oil-based product incompatible with latex/polyurethane/polyisoprene condoms; water-based products compatible with latex/polyisoprene but not polyurethane. Bench testing confirms biocompatibility and condom compatibility per ASTM D7661-10. Non-sterile, OTC use.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (acute systemic toxicity) demonstrated non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic results. Condom compatibility tested per ASTM D7661-10.
Technological Characteristics
Non-sterile, topical lubricants. Oil-based: sunflower oil, shea butter, beeswax. Water-based/Moisturizer: water, aloe vera, flax seed extract, gums. Packaging: HDPE tubes. Specifications: USP <61>/<62> for microbial limits. Condom compatibility per ASTM D7661-10. No software or electronic components.
Indications for Use
Indicated for vaginal and penile application to moisturize and lubricate, enhancing ease and comfort of intimate sexual activity and supplementing natural lubrication. Contraindications: YES oil-based is not compatible with natural rubber latex, polyurethane, or polyisoprene condoms. YES water-based and Vaginal Moisturizer are not compatible with polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Lifestyles Luxe Premium Personal Lubricant (K122477)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three human profiles facing right, stacked on top of each other, with flowing lines extending from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 9, 2017
Biotech Lubricants Ltd % Karen Cornelius Senior Associate Maetrics BioCity Nottingham, Pennyfoot Street Nottingham, NG1 1GF United Kingdom
K162569 Re:
> Trade/Device Name: YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, YES® Vaginal Moisturizer Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: May 9, 2017 Received: May 11, 2017
Dear Karen Cornelius:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K162569
#### Device Name
YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, YES® Vaginal Moisturizer
Indications for Use (Describe) Device Name: YES® oil-based Personal Lubricant
Indications for Use:
A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® oil-based Personal Lubricant is not compatible with natural rubber latex, polyurethane or polyisoprene condoms.
Device Name: YES® water-based Personal Lubricant
Indications for Use:
A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® water-based Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
Device Name: YES® Vaginal Moisturizer
Indications for Use:
A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® Vaginal Moisturizer is compatible with natural rubber latex and polvisoprene condoms. This product is not compatible with polyurethane condoms.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
| Submitter: | Biotech Lubricants Limited<br>Ramshill<br>Petersfield<br>GU31 4AT<br>United Kingdom<br>Phone: +44 (0) 3450 94 11 41<br>Email: sarah@biotechlubricants.com |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Karen Cornelius<br>Maetrics<br>BioCity Nottingham, Pennyfoot Street<br>Nottingham, NG1 1GF<br>United Kingdom<br>Phone: +44 (115) 921 6200<br>Email: kcornelius@maetrics.com |
| Date Prepared: | June 9, 2017 |
| Device Proprietary Names: | YES® oil-based Personal Lubricant<br>YES® water-based Personal Lubricant<br>YES® Vaginal Moisturizer |
| Device Common Name: | Personal Lubricant |
| Classification: | Condom (21 CFR 884.5300) |
| Regulatory Class: | II |
| Product Code: | NUC (lubricant, personal) |
| Predicate Device: | K124044 Aloe Cadabra Personal Lubricant (Seven Oaks Ranch, Inc.)<br>K122477 Lifestyles Luxe Premium Personal Lubricant (Ansell Healthcare Products LLC) |
The predicate devices have not been subject to a design-related recall.
## Device Description:
YES® oil-based Personal Lubricant: This non-sterile and oil-based personal lubricant consists primarily of Sunflower oil and Shea butter, and is not compatible with condoms. It is neither a contraceptive nor a spermicide. The formulation is a cream-colored thick paste that melts at body temperature. The product is packaged in a high density polyethylene (HDPE) tube with a flip-top cap. The specifications for the YES® oil-based Personal Lubricant are shown below:
| Parameter | Specification |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Color | Opaque paste |
| Odor | Slight beeswax and/or cocoa odor |
| Viscosity | N/A: measuring viscosity is not feasible, since changes in phase during measurement lead to differences in viscosity readings |
| Total yeast/mold count | <10 cfu/g (USP<61>) |
| Total aerobic mesophilic count | <100 cfu/g (USP<61>) |
| Absence of Pathogenic bacteria:<br>Staphylococcus aureus, Candida | Absent in 1g (USP<62>) |
{4}------------------------------------------------
| albicans, and gram negative bacteria | |
|--------------------------------------|--|
| (e.g., Pseudomonas aeruginosa) | |
YES® water-based Personal Lubricant: This non-sterile and water-based personal lubricant contains Aloe vera and other components, and is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. It is neither a contraceptive nor a spermicide. The formulation is a lightly opaque, non-greasy liquid gel. The product is packaged in a HDPE tube with a flip-top cap, and a HDPE single-use applicator. The specifications for the YES® water-based Personal Lubricant are shown below:
| Parameter | Specification |
|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Color | Opaque gel |
| Odor | Slight-mild, delicate linseed / leather note |
| Viscosity | 2,000 - 20,000 cP |
| pH | 3.70 - 4.50 |
| Osmolality | 100 - 400 mOsm/kg |
| Antimicrobial Effectiveness | At least 2.0 log reduction from the initial count at 14 days; no increase from the 14 days count at 28 days; no increase from the initial calculated count at 14 and 28 days (USP <51>) |
| Total yeast/mold count | <10 cfu/g (USP<61>) |
| Total aerobic mesophilic count | <100 cfu/g (USP<61>) |
| Absence of Pathogenic bacteria:<br>Staphylococcus aureus, Candida albicans, and gram negative bacteria (e.g., Pseudomonas aeruginosa) | Absent in 1g (USP<62>) |
YES® Vaginal Moisturizer: This non-sterile and water-based personal lubricant contains Aloe vera and other components, and is compatible with natural rubber latex, and polyisoprene condoms. This product is not compatible with polyurethane condoms. It is neither a contraceptive nor a spermicide. The formulation is a lightly opaque, non-greasy liquid gel. The product is packaged in a HDPE tube with a flip-top cap, and a HDPE single-use applicator. The specifications for the YES® Vaginal Moisturizer are shown below:
| Parameter | Specification |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Color | Opaque gel |
| Odor | Slight-mild, delicate linseed / leather |
| Viscosity | 2,000 – 20,000 cP |
| pH | 3.50–4.50 |
| Osmolality | 250-400 mOsm/kg |
| Antimicrobial Effectiveness | At least 2.0 log reduction from the initial count at 14 days; no increase from the 14 days count at 28 days; no increase from the initial calculated count at 14 and 28 days (USP <51>) |
| Total yeast/mold count | <10 cfu/g (USP<61>) |
| Total aerobic mesophilic count | <100 cfu/g (USP<61>) |
| Absence of Pathogenic bacteria: | Absent in 1g (USP<62>) |
{5}------------------------------------------------
| Staphylococcus aureus, Candida<br>albicans, and gram negative<br>bacteria (e.g., Pseudomonas<br>aeruginosa) | |
|-------------------------------------------------------------------------------------------------------------|--|
|-------------------------------------------------------------------------------------------------------------|--|
### Indications for Use:
YES® oil-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® oil-based Personal Lubricant is not compatible with natural rubber latex, polyurethane or polyisoprene condoms.
YES® water-based Personal Lubricant: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® water-based Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
YES® Vaginal Moisturizer: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. YES® Vaginal Moisturizer is compatible with natural rubber latex, and polyisoprene condoms. This product is not compatible with polyurethane condoms.
| | YES® oil-based Personal Lubricant | | Comparison |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K162569 | K122477 | |
| Device Name | YES® oil-based personal<br>lubricant | LifeStyles Luxe Premium<br>Personal Lubricant | |
| Manufacturer | Bio-tech Lubricants Ltd | Ansell Healthcare Products,<br>LLC | |
| Classification | 884.5300 | 884.5300 | |
| Product code | NUC | NUC | |
| Description | Personal Lubricant | Personal Lubricant | Same |
| Indications for use | A personal lubricant, for<br>vaginal and penile<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is not<br>compatible with natural<br>rubber latex, polyurethane<br>or polyisoprene condoms. | LifeStyles Luxe Premium<br>Personal Lubricant Is a<br>personal lubricant, for penile<br>and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate<br>sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural<br>rubber latex, polyurethane,<br>and polyisoprene condoms. | Different: The subject and<br>predicate devices indication<br>for use statements differ in<br>regards to compatibility<br>with condoms. Differences<br>noted do not represent a<br>new intended use. |
| Oil-based lubricant | YES | YES | Same: the subject and<br>predicate devices are both<br>oil-based lubricants;<br>however, the specific oils in<br>the formulation are |
### Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices:
{6}------------------------------------------------
| YES® oil-based Personal Lubricant | Comparison | | |
|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ingredients | Sunflower seed oil, Shea<br>butter, Sweet almond oil,<br>Bees wax, Cocoa seed<br>butter, Vitamin E | Not known | Different: The difference in<br>ingredients do not raise<br>different questions of S &<br>E. |
| Biocompatible | YES | YES | Same |
| Non-sterile | YES | YES | Same |
| Contains<br>preservatives | NO | NO | Same |
| Microbial limits<br>(TAMC, TYMC, and<br>absence of<br>pathogenic bacteria) | See product specifications<br>table<br>(USP <61> & <62>) | Meets the requirements of<br>USP <61> & <62> | Different: the microbial<br>limits of the predicate are<br>not known; however this<br>does not raise different<br>questions of S & E as the<br>microbial limits<br>specifications are within the<br>accepted ranges for these<br>types of devices. |
| Condom<br>Compatibility | Not compatible with natural<br>rubber latex, polyurethane,<br>or polyisoprene condoms. | Compatible with natural<br>rubber latex, polyurethane,<br>and polyisoprene condoms. | Different: The condom<br>compatibility for the subject<br>and predicate devices are<br>different; However, this<br>difference does not<br>represent a new intended<br>use. |
As noted in the table above the subject and predicate devices indications are similar in that they are both used to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. However, they do differ in regards to condom compatibility as the subject device is not compatible with condoms, while the predicate is condom compatible. Differences in condom compatibility between the subject and predicate devices do not represent a new intended use. In regards to technology, the subject and predicate devices are similar in that they are both non-sterile and are oil-based. Differences do exist as described in the table above (e.g., product formulations, specifications for microbial limits, condom compatibility, etc.). The differences identified do not raise different questions of safety and effectiveness as discussed in the table above
| YES® water-based Personal Lubricant | Comparison | | |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K162569 | K124044 | |
| Device name | YES® water-based<br>Personal Lubricant | Aloe Cadabra Personal<br>lubricant | |
| Manufacturer | Bio-tech Lubricants Ltd | Seven Oaks Ranch, Inc. | |
| Classification | 884.5300 | 884.5300 | |
| Product code | NUC | NUC | |
| Description | Personal Lubricant | Personal Lubricant | Same |
| Indications for use | A personal lubricant, for<br>vaginal and penile<br>application, intended to<br>moisturize and lubricate, to | Aloe Cadabra Lubricant and<br>Aloe Cadabra<br>Flavored/Scented Lubricants<br>are personal lubricants for | Same: The subject and<br>predicate devices have the<br>same indication for use,<br>and therefore have the. |
| | YES® water-based Personal Lubricant | | Comparison |
| Water-based<br>lubricant | enhance the ease and<br>comfort of intimate sexual<br>activity and supplement<br>the body's natural<br>lubrication. YES® water-<br>based Personal Lubricant<br>is compatible with natural<br>rubber latex and<br>polyisoprene condoms.<br>This product is not<br>compatible with<br>polyurethane condoms. | penile and/or vaginal<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is compatible<br>with natural rubber latex and<br>polyisoprene condoms. This<br>product is not compatible<br>with polyurethane condoms. | same intended use. |
| Ingredients | Water, Aloe vera, Flax<br>seed extract, Guar gum,<br>Locust bean gum, Xanthan<br>gum, Sodium chloride,<br>Potassium sorbate, Citric<br>acid, Phenoxyethanol | Aloe vera | Different: The difference<br>in ingredients do not raise<br>different questions of S &<br>E. |
| Biocompatible | YES | YES | Same |
| Non-sterile | YES | YES | Same |
| Contains<br>preservatives | YES | YES | Same |
| Microbial limits<br>(TAMC, TYMC,<br>and absence of<br>pathogenic<br>bacteria) | See product specifications<br>table<br>(USP <61> & <62>) | Not known | Different: the microbial<br>limits of the predicate are<br>not known; however this<br>does not raise different<br>questions of S & E as these<br>specifications are within<br>the accepted range for<br>these types of devices. |
| Viscosity | 2,000 - 20,000 cP | Not known | Different: the viscosity of<br>the predicate is not known.<br>This does not raise<br>different questions of S &<br>E as the viscosity<br>specification is within the<br>accepted range for these<br>types of devices. |
| pH | 3.70 - 4.50 | Not known | Different: the pH of the<br>predicate is not known.<br>This does not raise<br>different questions of S &<br>E as the pH specification is<br>within the accepted range<br>for these types of devices. |
| Osmolality | 100 - 400 mOsm/kg | Not known | Different: the osmolality of<br>the predicate is not known.<br>This does not raise<br>different questions of S &<br>E as the osmolality<br>specification is within the<br>accepted range for these<br>types of devices. |
| | YES® water-based Personal Lubricant | | Comparison |
| Condom<br>Compatibility | Compatible with natural<br>rubber latex and<br>polyisoprene condoms.<br>Not compatible with<br>polyurethane condoms. | Compatible with natural<br>rubber latex and<br>polyisoprene condoms. Not<br>compatible with<br>polyurethane condoms. | Same |
{7}------------------------------------------------
{8}------------------------------------------------
As noted in the table above the subject and predicate devices have indication for use and intended use. In addition, the subject and predicate devices are similar in that they are both non-sterile, compatible with the same types of contain Aloe vera, and are water-based. Differences do exist as described in the table above (e.g., product formulations, specifications for microbial limits, viscosity, etc.). The differences identified do not raise different questions of safety and effectiveness as discussed in the table above.
| YES® Vaginal Moisturizer | | Comparison | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K162569 | K124044 | |
| Device Name | YES® VM (Vaginal<br>Moisturizer) | Aloe Cadabra Personal<br>lubricant | |
| Manufacturer | Bio-tech Lubricants Ltd | Seven Oaks Ranch, Inc. | |
| Classification | 884.5300 | 884.5300 | |
| Product code | NUC | NUC | |
| Description | Personal Lubricant | Personal Lubricant | Same |
| Indications for use | A personal lubricant, for<br>vaginal and penile<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>YES® Vaginal Moisturizer<br>is compatible with natural<br>rubber latex and<br>polyisoprene condoms. This<br>product is not compatible<br>with polyurethane condoms. | Aloe Cadabra Lubricant and<br>Aloe Cadabra<br>Flavored/Scented Lubricants<br>are personal lubricants for<br>penile and/or vaginal<br>application, intended to<br>moisturize and lubricate, to<br>enhance the ease and comfort<br>of intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural<br>rubber latex and polyisoprene<br>condoms. This product is not<br>compatible with polyurethane<br>condoms. | Same: The subject and<br>predicate devices have the<br>same indication for use, and<br>therefore have the same<br>intended use. |
| Water-based<br>lubricant | YES | YES | Same |
| Ingredients | Water, Aloe vera, Flax seed<br>extract, Guar gum, Locust<br>bean gum, Xanthan gum,<br>Sodium chloride, Potassium<br>sorbate, Citric acid,<br>Phenoxyethanol | Aloe vera | Different: The difference in<br>ingredients do not raise<br>different questions of S & E |
| Biocompatible | YES | YES | Same |
| Non-sterile | YES | YES | Same |
| Contains<br>preservatives | YES | YES | Same |
| Microbial Limits | See product specifications | Not known | Different: the microbial |
{9}------------------------------------------------
| | YES® Vaginal Moisturizer | | Comparison |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (TAMC, TYMC, and<br>absence of<br>pathogenic bacteria) | table<br>(USP <61> & <62>) | | limits of the predicate are<br>not known; however this<br>does not raise different<br>questions of S & E as these<br>specifications are within the<br>accepted range for these<br>types of devices. |
| Viscosity | 2,000 - 20,000 cP | Not known | Different: the viscosity of<br>the predicate is not known.<br>This does not raise different<br>questions of S & E as the<br>viscosity specification is<br>within the accepted range<br>for these types of devices |
| pH | 3.50-4.50 | Not known | Different: the pH of the<br>predicate is not known. This<br>does not raise different<br>questions of S & E as the pH<br>specification is within the<br>accepted range for these<br>types of devices. |
| Osmolality | 250-400 mOsm/kg | Not known | Different: the osmolality of<br>the predicate is not known.<br>This does not raise different<br>questions of S & E as the<br>osmolality specification is<br>within the accepted range<br>for these types of devices. |
| Condom<br>Compatibility | Compatible with natural<br>rubber latex and<br>polyisoprene condoms. Not<br>compatible with<br>polyurethane condoms. | Compatible with natural<br>rubber latex and polyisoprene<br>condoms. Not compatible<br>with polyurethane condoms. | Same |
As noted in the table above the subject and predicate devices have the same indication for use and intended use. In addition, the subject and predicate devices are similar in that they are both non-sterile, compatible with the same types of contain Aloe vera, and are water-based. Differences do exist as described in the table above (e.g., product formulations, specifications for microbial limits, viscosity, etc.). The differences identified do not raise different questions of safety and effectiveness as discussed in the table above.
### Performance Testing:
### Biocompatibility:
Testing was conducted to assess the biocompatibility of the YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant and YES® Vaginal Moisturizer. Testing included assessments for Cytotoxicity (ISO 10993-5:2009(R)2014), Vaginal Irritation (ISO 10993-10:2010(R)2014), Sensitization (ISO 10993-10:2010(R)2014), and Acute Systemic Toxicity (ISO 10993-11:2006).
The results of testing demonstrated that they are non-cytotoxic, non-irritating, non-sensitizing and nonsystemically toxic.
{10}------------------------------------------------
Condom Compatibility:
YES® oil-based Personal Lubricant: Condom compatibility testing was performed for polyurethane condoms per ASTM D7661-10. The results indicate that the device is not compatible with polyurethane condoms. Additionally, compatibility testing was not performed for natural rubber latex or polyisoprene condoms; therefore, the device is not compatible with natural rubber latex or polyisoprene condoms.
YES® Vaginal Moisturizer and YES® water-based Personal Lubricant: Condom compatibility testing was performed per ASTM D7661-10. The results indicated that the devices are compatible with natural rubber latex and polyisoprene condoms, but are not compatible with polyurethane condoms.
Shelf Life and Microbial Limits:
YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, and YES® Vaginal Moisturizer have a 9-month shelf life. The results of shelf-life testing demonstrated that the subject lubricants met all device specifications as shown in the Device Description section for the duration of the 9-month shelf-life.
#### Conclusion:
YES® oil-based Personal Lubricant, YES® water-based Personal Lubricant, and YES® Vaginal Moisturizer have the same intended use and similar technological characteristics as the predicate devices. The results of the performance testing conducted on the subject demonstrate that they are as safe and effective as the predicate devices and support substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.