← Product Code [MQX](/productcode/MQX) · K162566

# AIK Sterile Acupuncture Needles for Single Use (K162566)

_Changchun Aik Medical Devices Co., Ltd. · MQX · May 11, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162566

## Device Facts

- **Applicant:** Changchun Aik Medical Devices Co., Ltd.
- **Product Code:** [MQX](/productcode/MQX.md)
- **Decision Date:** May 11, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5580
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The AIK Sterile Acupuncture Needles for Single Use is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

## Device Story

Sterile, single-use acupuncture needles; stainless steel (304) body; metal handle; optional plastic guide tube. Used by qualified acupuncture practitioners in clinical settings. Device pierces skin to perform acupuncture. Biocompatible materials (polydimethylsiloxane coating); sterilized via ethylene oxide (EO). Performance verified against ISO 17218 (hardness, breakage resistance, puncture performance, corrosion resistance).

## Clinical Evidence

Bench testing only. No clinical data. Performance evaluated via dimensional analysis, biocompatibility (cytotoxicity, intracutaneous reactivity, sensitization, pyrogen, systemic toxicity, hemodialysis), and mechanical testing (drawing strength, hardness, breakage resistance, puncture performance, corrosion resistance) per ISO 17218, ISO 10993, and ISO 11135.

## Technological Characteristics

Needle body: 304 stainless steel (06Cr19Ni10). Handle: stainless steel. Coating: polydimethylsiloxane. Guide tube: PVC. Dimensions: 0.14mm-0.35mm diameter; 13mm-75mm length. Sterilization: Ethylene oxide (EO) per ISO 11135. Biocompatibility: ISO 10993-1. Mechanical performance: ISO 17218.

## Regulatory Identification

An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.

## Predicate Devices

- SMC Acupuncture Needle ([K141473](/device/K141473.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and support. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 11, 2017

CHANGCHUN AIK MEDICAL DEVICES CO., LTD Mr. Xingwang Li General Manager No. 25, Karen Industrial Park, Jiutai Economic Development Zone Changchun, Jilin 130507 CHINA

Re: K162566

Trade/Device Name: AIK Sterile Acupuncture Needles for Single Use Regulation Number: 21 CFR 880.5580 Regulation Name: Acupuncture needle Regulatory Class: Class II Product Code: MQX Dated: September 8, 2016 Received: April 10, 2017

Dear Mr. Xingwang Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Xingwang Li

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tina Kiang
-s

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Change Control Table, Change History

# Change Control Table

| Version | Document Author     | Document Approver   | Date Approved |
|---------|---------------------|---------------------|---------------|
| 1.00    | Name, Title, Office | Name, Title, Office | MM/DD/YYYY    |

Complete Change Control Table (all versions) retained in SWIFT Docs.

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K162566

Device Name

AIK Sterile Acupuncture Needles for Single Use

Indications for Use (Describe)

The AIK Sterile Acupuncture Needles for Single Use is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

| Type of Use (Select <i>one or both</i> , as applicable) |  |
|---------------------------------------------------------|--|
|---------------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## K162566

## 510(k) Summary

| Submitter:           | CHANGCHUN AIK MEDICAL DEVICES CO., LTD<br>No. 25, Karen Industrial Park, Jiutai Economic Development Zone<br>Changchun, Jilin Province, 130507 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:             | Mr. Li, Xingwang<br>General Manager<br>Phone: (86)-13364688867 /<br>(86)-18616806984<br>Email: aik2000@163.com                                 |
| Date Prepared:       | May, 4 2017                                                                                                                                    |
| Device Trade Name:   | AIK Sterile Acupuncture Needles for Single Use                                                                                                 |
| Device Common Name:  | Acupuncture Needle                                                                                                                             |
| Classification Name: | Acupuncture Needle                                                                                                                             |
| Class:               | II                                                                                                                                             |
| Regulation Number:   | 880.5580                                                                                                                                       |
| Panel:               | General Hospital and Personal Use Devices                                                                                                      |
| Product Code:        | MQX                                                                                                                                            |

#### 1. Predicate Device:

| Device                 | Company | Product Code | 510(k) Number |
|------------------------|---------|--------------|---------------|
| SMC Acupuncture Needle | SMC.    | MQX          | K141473       |

## 3. Indications for Use:

The AIK Sterile Acupuncture Needles for Single Use is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

{5}------------------------------------------------

## 3. Product Description:

The Acupuncture Needles are manufactured from stainless steel and sterilized with ethylene oxide gas. Acupuncture needle is made up of needle handle and needle body. The body of needle uses 304 stainless steel wire (06Cr19Ni10). The handle is a body coil with stainless steel wire, the spiral loop of which is arranged symmetrically without gaps.

The device can only be used by qualified practitioners of acupunctures. Qualified practitioners should already be knowledgeable on how to use the acupuncture needle and when not to use acupuncture in certain patient populations and medical conditions.

Handle of acupuncture needle is a plain handle. Types of acupuncture needle (sterile acupuncture needle for single use) are: acupuncture needle without a guide tube and acupuncture needle with a guide tube.

Image /page/5/Figure/4 description: The image shows a diagram with labels. There are two sections, one with a straight line labeled with 'Φd' and 'l1', and another with a coiled line labeled with 'l2'. The entire diagram is labeled with 'L'.

## Product Specification

Diameter specification of acupuncture needle: Ø 0.14mm-Ø 0.35mm; length specification of needle body: 13mm-75mm.

| Gauge (#) | Diameter | Lengths of needle body |      |      |      |      |      |      |      |
|-----------|----------|------------------------|------|------|------|------|------|------|------|
| 42G       | 0.14mm   | 13mm                   | 15mm | 25mm |      |      |      |      |      |
| 40G       | 0.16mm   | 13mm                   | 15mm | 25mm | 30mm |      |      |      |      |
| 38G       | 0.18mm   | 13mm                   | 15mm | 25mm | 30mm | 40mm |      |      |      |
| 36G       | 0.20mm   | 13mm                   | 15mm | 25mm | 30mm | 40mm | 50mm |      |      |
| 34G       | 0.22mm   | 13mm                   | 15mm | 25mm | 30mm | 40mm | 50mm | 60mm |      |
| 32G       | 0.25mm   | 13mm                   | 15mm | 25mm | 30mm | 40mm | 50mm | 60mm | 75mm |
| 30G       | 0.30mm   | 13mm                   | 15mm | 25mm | 30mm | 40mm | 50mm | 60mm | 75mm |
| 28G       | 0.35mm   | 13mm                   | 15mm | 25mm | 30mm | 40mm | 50mm | 60mm | 75mm |

## 4. Non-Clinical Testing

Performance testing was conducted to evaluate and characterize the performance of the AIK Sterile Acupuncture Needles. Non-clinical data included dimensional evaluation, and performance requirements evaluation per international standard, ISO 17218. Material performances are evaluated per international standard ISO 10993-1. In detail, the needle body needle handle, and guide tube were tested by tests of in vitro cytotoxicity, intracutaneous reactivity/irritation, skin sensitization and pyrogen. And the needle body were further tested by tests of hemodialysis study, and systemic toxicity study. All biocompatibility tests results show that the subject device is biocompatibility safe.

{6}------------------------------------------------

The primary package performance are evaluated per international standard ISO 11607. EO sterilization process are evaluated per international standard ISO 11135. And the EO and ECH Residuals Testing in accordance with ISO10993-7.

The EO Sterilization Validation, Sterility Testing, Packaging Validation, and Long-Term Stability Testing were also performed per relevant standards.

All the test results demonstrate AIK Sterile Acupuncture Needles for Single Use meet the requirements of its pre-defined acceptance criteria and intended uses. For example, the function performance tests including:

- Appearance and cleanliness tests
- Drawing strength test
- Needle body hardness test
- Resistance to breakage of needle body test
- Intensity and puncture performance of needle tip test
- Sterile Test
- Resistance to corrosion test

#### 5. Substantial Equivalence Determination

The indications for use and technology characteristics of the proposed AIK Sterile Acupuncture Needles for Single Use, and the substantial equivalence to the predicate devices have been demonstrated via data collected in design verifications. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its indications for use. The subject device is substantial equivalent to the predicate device.

The comparison to predicate devices as below table.

| Item                  | Proposed Device<br>(AIK Sterile Acupuncture Needles for<br>Single Use)                                                                                                                                       | Predicate Device<br>(SMC Acupuncture Needle)                                                                                                                                                                                              |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>use | Intended to pierce the skin in the<br>practice of acupuncture by<br>qualified practitioners of<br>acupuncture as determined by the<br>states.                                                                | Intended to pierce the skin in the<br>practice of acupuncture by<br>qualified practitioners of<br>acupuncture as determined by the<br>states                                                                                              |
| Applied types         | Needle with/without plastic guide<br>tube                                                                                                                                                                    | Needle with/without plastic guide<br>tube                                                                                                                                                                                                 |
| Needle tip shape      | Arrow                                                                                                                                                                                                        | Arrow                                                                                                                                                                                                                                     |
| Design                | Stainless steel needle with a steel,<br>aluminum or copper handle and a<br>plastic guide tube for user                                                                                                       | Stainless steel needle with a steel,<br>aluminum or copper handle and a<br>plastic guide tube for user                                                                                                                                    |
| Materials             | Needle: surgical stainless steel<br>(SUS304);<br>Handle: metal (steel);<br>Coating material:<br>Polydimethylsiloxane;<br>Guild Tube(not contact): PVC;<br>following ISO 10993-1 for<br>cytotoxicity test<br> | Needle: surgical stainless steel<br>(SUS304)<br>Handle: metal (steel or aluminum<br>tube or copper wire)<br>Coating material:<br>Polydimethylsiloxane;<br>Guild Tube(not contact): PVC;<br>following ISO 10993-1 for<br>cytotoxicity test |

#### Comparison to Predicate Devices

{7}------------------------------------------------

|                                     |                                                    | intracutaneous reactivity test;<br>hemodialysis study;<br>systemic toxicity study;<br>Pyrogen and Endotoxins study;                 | sensitization test<br>and intracutaneous reactivity test                                                                            |
|-------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Sterility                           |                                                    | Sterilized with ethylene oxide gas<br>with residuals at a validated dose<br>level                                                   | Sterilized with ethylene oxide gas<br>and or / Sterilized by Gamma<br>Irradiation                                                   |
| Available in<br>Needle<br>Diameters |                                                    | 0.14 mm, 0.16 mm, 0.18 mm, 0.20<br>mm, 0.22 mm, 0.25 mm, 0.30 mm,<br>0.35 mm                                                        | 0.16 to 0.70 mm                                                                                                                     |
| Available in<br>Needle Lengths      |                                                    | 13 mm, 15 mm, 25 mm, 30 mm, 40<br>mm, 50 mm, 60 mm, 75 mm                                                                           | 15-135 mm                                                                                                                           |
| Labeling                            |                                                    | Comply to 21 CFR 801                                                                                                                | Comply to 21 CFR 801                                                                                                                |
| Performance Requirements            | Appearance<br>and<br>Cleanliness                   | Examined under corrected-to-normal<br>vision, and x 10 magnification:<br>Comply to the requirements specified<br>is ISO 17218       | Examined under corrected-to-<br>normal vision, and x 10<br>magnification:<br>Comply to the requirements<br>specified is ISO 17218   |
|                                     | Drawing<br>strength                                | Needle solidly joined after the test<br>following ISO 17218, clause 5.3.2                                                           | Needle solidly joined after the test<br>following ISO 17218, clause 5.3.2                                                           |
|                                     | Hardness of<br>the needle<br>body                  | Hardness of needle body comply the<br>requirements specified in ISO 17218,<br>clause 5.3.3                                          | Hardness of needle body comply the<br>requirements specified in ISO 17218,<br>clause 5.3.3                                          |
|                                     | Resistance<br>to breakage<br>of the<br>needle body | The body of the needle is sufficient<br>resistance to breakage after test<br>following ISO 17218, clause 5.3.4                      | The body of the needle is sufficient<br>resistance to breakage after test<br>following ISO 17218, clause 5.3.4                      |
|                                     | Intensity<br>and<br>puncture<br>performance        | The tip of the needle has good<br>intensity and puncture performance<br>following the tests specified in ISO<br>17218, clause 5.3.5 | The tip of the needle has good<br>intensity and puncture performance<br>following the tests specified in ISO<br>17218, clause 5.3.5 |
|                                     | Resistance<br>to corrosion                         | No corrosion of the body of the<br>needle following the test specified in<br>ISO 17218, clause 5.3.6                                | No corrosion of the body of the<br>needle following the test specified in<br>ISO 17218, clause 5.3.6                                |

The subject device acupuncture needles are substantially equivalent to the predicate device listed in this 510(k) submission; that is, the acupuncture needle has the same intended use (i.e., indications for use) and is similar, and in some cases the same, in both design (e.g., materials, sizes) and performance. One difference is the available needle diameters and needle lengths. Another difference is that our acupuncture needle claims for EO sterilization only, and the predicate device claims for both EO sterilization and Gamma irradiation sterilization methods.

#### 6. Conclusion

Based on the indications for use, technological characteristics and performance testing, the AIK Sterile Acupuncture Needles for Single Use has demonstrated to be substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K162566](https://fda-staging.innolitics.com/device/K162566)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).