The EVS 4343 and EVS 4343G Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Device Story
EVS 4343/4343G are wired flat-panel digital X-ray detectors; capture projection radiographic images in digital format. Indirect conversion principle: scintillator (CsI for EVS 4343; GoS for EVS 4343G) converts X-rays to visible light; optical sensor/TFT array converts light to electrical charges; signal amplified and digitized. Integrates with X-ray systems via Auto Trigger, Sync Trigger, or USB SW modes. Used in clinical settings by radiology staff to replace film/screen systems. Output is digital image file for viewing, printing, or storage. Benefits include rapid image acquisition, elimination of chemical film processing, and digital workflow integration.
Clinical Evidence
Single-blinded clinical concurrence study conducted per CDRH guidance. Confirmed EVS 4343G provides images of equivalent diagnostic capability to predicate devices. Clinical images provided as supplemental evidence to laboratory performance data.
Indicated for general radiographic diagnosis of human anatomy in patients requiring projection radiography. Not intended for mammography applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Carestream DRX-1 System w/ DRX Plus Detectors (K150766)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 21, 2016
DRTECH Corporation % Choul-Woo Shin Suite No. 2, 310 Floor, 29, Dunchon-daero541 beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, 13230 REPUBLIC OF KOREA
Re: K162555
Trade/Device Name: EVS 4343. EVS 4343G Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: December 2, 2016 Received: December 6, 2016
Dear Choul-Woo Shin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162555
Device Name EVS 4343, EVS 4343G
#### Indications for Use (Describe)
The EVS 4343 and EVS 4343G Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radio graphic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
# 1. Date Prepared [21 CFR 807.92(a) (1)]
09/08/2016
# 2. Submitter's Information [21 CFR 807.92(a) (1)]
| • | Name of Sponsor: | DRTECH Corporation |
|---|-----------------------|-----------------------------------------------------|
| • | Address: | Suit No. 2, 3 Floor, 29, Dunchon-daero541 beon-gil, |
| | | Jungwon-gu, Seongnam-si, Gyeonggi-do, 13230 |
| | | Republic of Korea |
| • | Contact Name: | Choul-Woo Shin |
| • | Telephone No.: | + 82-31-779-7783 |
| • | Fax No .: | + 82-31-779-7790 |
| • | Email Address : | cwshin@drtech.co.kr |
| • | Registration Number: | 3005172103 |
| • | Name of Manufacturer: | Same as Sponsor |
# 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| • | Trade Name: | EVS 4343, EVS 4343G |
|---|----------------------------|-----------------------------------|
| • | Common Name: | Digital Flat Panel X-ray Detector |
| • | Classification Name: | Stationary X-ray System |
| • | Classification Panel: | Radiology |
| • | Classification Regulation: | 21 CFR 892.1680 |
| • | Product Code: | MQB |
| • | Device Class: | II |
| • | 501(k) Number: | K162555 |
| • | | |
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
| Parameter | Predicate Device |
|------------------------------|--------------------------------------------------------|
| • 510(k) Number: | K142475 K150766 |
| • Applicant: | DRTECH Corporation Carestream Health, Inc. |
| • Trade Name: | Carestream DRX-1 System w/ DRX Plus Detectors EVS 4343 |
| • Classification Name: | Stationary X-ray System |
| • Classification Panel: | Radiology |
| • Classification Regulation: | 21 CFR 892.1680 |
| • Product Code: | MQB |
| • Device Class: | II |
{4}------------------------------------------------
### 5. Description of the Modified Device [21 CFR 807.92(a) (4)]
<Modification>
- The EVS 4343 was removed the function of wireless communication to transfer data.
- . Addition of EVS 4343G: This is to notify that EVS 4343 / EVS 4343G are same Hardware, Software and components. But scintillator layer are different. Scintillator is a phosphor that produces scintillations.
| Model Name | Scintillator layer |
|------------|-----------------------------|
| EVS 4343 | CsI (Cesium Iodide) |
| EVS 4343G | GoS (Gadolinium Oxysulfide) |
The EVS 4343(G) is a wired flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EVS 4343(G) differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.
#### 6. Intended Use [21 CFR 807.92(a)(5)]
The intended use has not changed as a result of the modification and is as follows:
The EVS 4343 and EVS 4343G Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
#### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
The EVS 4343(G) Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.
The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.
Comparisons with the predicate, devices show the technological characteristics of the EVS 4343(G) to be same to the predicate devices. The EVS 4343(G) is functionally identical to the predicate devices.
The subject device integrates with X-ray system through three modes such as Auto Trigger mode (AT), Sync. Trigger mode and USB SW mode. The Auto Trigger Mode (AT) is available for acquiring images without any connection to X-ray generator interface cable is not required. The Sync. Trigger mode is the most common and recommended exposure mode at B2B scopes. User can achieve the high quality images with Sync. Tigger Mode. The USB SW Mode is the most common and recommended exposure mode at a retrofit scope. User can achieve high quality images with USB SW Mode.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "DRTECH" in blue font. The letters are all capitalized and sans-serif. There is a small blue square above and to the left of the "D" in "DRTECH". A cursor is pointing at the "D".
# 8. Hardware and Software Requirements
- X-ray System Requirement
| Contents | Requirements | |
|-----------|-----------------|------------------------------------------------------|
| Generator | Power frequency | 30kHz ~ 240kHz |
| | KV | 40kVp ~ 150kVp |
| | mA Range | 10mA ~ 1000mA |
| | Exposure Time | 0.001~10sec |
| | mAs Range | 0.1~1000mAs |
| | Accuracy | $\pm$ 5% |
| Bucky | Operating Type | moving<br>Stepping Motor<br>Spring<br>CAM Motor type |
| | Static(Fixed) | |
| | Trey size (mm) | 460 x 460 x 15 or higher |
| Grid | Ratio | 5:1, 6:1, 8:1, 10:1, 12:1, 15:1 |
| | Line | 85 ~ 215 Line |
| | SID | 100 ~ 180 cm |
* The EVS 4343(G) detector is not compatible with trays that are less than 460 mm x 15 mm.
# - Software Requirement
The EVS 4343(G) detector is compatible with Econsole1 (K152172).
# 9. Substantial Equivalence [21 CFR 807.92(b)]
| Parameter | Subject Device | Predicate Device | Remark | | |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(K) Number | Unknown | K142475 | K150766 | - | |
| Manufacturer | DRTECH<br>Corporation | DRTECH<br>Corporation | Carestream<br>Health, Inc. | - | |
| Model Name | EVS 4343, EVS<br>4343G | EVS 4343 | Carestream DRX-1<br>System w/ DRX<br>Plus Detectors | - | |
| Classification<br>Name | Stationary X-ray System | | | same | |
| Classification Panel | Radiology | | | same | |
| Classification<br>Regulation | 21 CFR 892.1680 | | | same | |
| Product Code | MQB | | | same | |
| Device Class | Class II | | | same | |
| Intended Use | The EVS 4343 and<br>EVS 4343G Digital<br>X-ray detector is<br>indicated<br>for digital imaging<br>solution designed for<br>providing general | The EVS 4343<br>Digital X-ray<br>detector is<br>indicated<br>for digital imaging<br>solution designed<br>for providing | The device is<br>intended to<br>capture for<br>display<br>radiographic<br>images of human | same | |
| | | radiographic<br>diagnosis of human<br>anatomy. This device<br>is intended to replace<br>film or screen based<br>radiographic systems<br>in all general purpose<br>diagnostic<br>procedures. This<br>device is not<br>intended for<br>mammography<br>applications | general<br>radiographic<br>diagnosis of<br>human anatomy.<br>This device is<br>intended to<br>replace film or<br>screen based<br>radiographic<br>systems in all<br>general purpose<br>diagnostic<br>procedures. This<br>device is not<br>intended for<br>mammography<br>applications | anatomy<br>including both<br>pediatric and<br>adult patients.<br>The device is<br>intended for use<br>in general<br>projection<br>radiographic<br>applications<br>wherever<br>conventional<br>screen-film<br>systems or CR<br>systems may be<br>used. Excluded<br>from the<br>indications for<br>use are<br>mammography,<br>fluoroscopy, and<br>angiography<br>applications. | |
| Design | Panel<br>Shape | Square Panel | Square Panel | Rectangular Panel | same |
| | Detector<br>Size | 17" X 17" | 17" X 17" | 13" X 17" | same |
| | Dimensions | 460(W) x 460(L) x<br>15(H) | 460(W) x 460(L)<br>x 15(H) | - | same |
| | Pixel<br>Pitch | 140μm | 140μm | 139µm | same |
| | Image<br>Size | 3,072 x 3,072 | 3,072 x 3,072 | - | same |
| Materials<br>Scintillator | | TFT -amorphous<br>Silicon | TFT -amorphous<br>Silicon | TFT -amorphous<br>Silicon | same |
| | | EVS 4343: CsI<br>EVS 4343G: GOS | EVS 4343: CsI | GOS | same |
| Performance | DQE | EVS 4343:<br>43.9 % at 1 lp/mm<br>16.5 % at 3 lp/mm<br>EVS 4343G:<br>23.6 % at 1 lp/mm<br>5 % at 3 lp/mm | 28 % at 1 lp/mm<br>13 % at 3 lp/mm | 29.7 % at 1 lp/mm<br>5.0 % at 3 lp/mm | similarity |
| | MTF | EVS 4343: 35 % at<br>2.0 lp/mm | 35 % at 2.0 lp/mm | 17.4% at 2 lp/mm | same |
| | | EVS 4343G: 34 % at<br>2.0 lp/mm | | | same |
| | Resolution | 3.5 lp/mm | 3.5 lp/mm | 3.5 lp/mm | same |
| | Anatomical Sites | General Radiography | General Radiography | General Radiography | same |
| | Power Supply | 100~240V~, 50/60 Hz | 100~240V~, 50/60 Hz | 100~240V~, 50/60 Hz | same |
| Communication<br>Interface | | Wired: Ethernet | Wireless:<br>2.4g/802.11g/n<br>5g/802.11a/n<br>Wired: Ethernet | - | different |
{6}------------------------------------------------
# IDRTECH
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DR TECH. The logo is in blue and features a square above the letters 'DR'. The letters 'TECH' are in a slightly lighter shade of blue than the 'DR'. A mouse cursor is visible near the left side of the logo.
When compared to the predicate device (K142475 and K150766), the EVS 4343(G) presented in this submission has the same:
- Intended Use
- Technological characteristics
- Operating principle
- · Design features
- · Performance (DQE, MTF, Resolution)
- Materials Scintillator
A few differences are as follows
- Communication Method
There are no significant differences between the EVS 4343(G) and the predicate device(s) that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
## 10. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
The non-clinical performance testing constrains that the main physical values for comparison of Xray devices like DQE and MTF are basically equal or worth than the predicate device as following table:
| Parameter | Modified Device | Predicate Device | | Remark |
|-----------|-------------------------------------------------------|------------------------------------|---------------------------------------|--------------------|
| | | K142475 | K150766 | |
| DQE | EVS 4343:<br>43.9 % at 1<br>lp/mm<br>16.5% at 3 lp/mm | 28 % at 1 lp/mm<br>13 % at 3 lp/mm | 29.7 % at 1 lp/mm<br>5.0 % at 3 lp/mm | same<br>similarity |
| | EVS 4343G:<br>23.6 % at 1<br>lp/mm<br>5 % at 3 lp/mm | | | |
| MTF | EVS 4343: 35 % at 2.0 lp/mm | 35 % at 2.0 lp/mm | 17.4% at 2.0 lp/mm | same |
{8}------------------------------------------------
| EVS 4343G:<br>34 % at 2.0<br>lp/mm | same |
|------------------------------------|------|
|------------------------------------|------|
The EVS 4343(G) complies with the following international and FDA-recognized consensus standards:
| AAMI ANSI ES60601-1: | Medical Electrical Equipment -- Part 1: General Requirements for Basic<br>Safety And Essential Performance (IEC 60601-1:2005, Mod) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2: | Medical Electrical Equipment - Part 1-2: General Requirements For Basic<br>Safety And Essential Performance - Collateral Standard: Electromagnetic<br>Compatibility - Requirements And Tests (Edition 3) |
| ISO 14971: | Medical Devices - Application of Risk Management to Medical Devices.<br>(General I (QS/RM)) |
| IEC 62220-1: | Medical electrical equipment - Characteristics of digital X-ray imaging<br>devices - Part 1: Determination of the detective quantum efficiency |
| NEMA PS 3.1 - 3.20: | Digital Imaging and Communications in Medicine (DICOM) Set |
### 11. Summary of Clinical Data [21 CFR 807.92(b)(2)]
A single-blinded concurrence study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted, and the study confirmed that the new x-ray detectors EVS 4343G provide images of equivalent diagnostic capability to the predicate devices (K142475 and K150766) and its results demonstrate substantial equivalence.
Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note X-ray detector technology that is identical to the predicate device) but they provide further evidence in addition to the laboratory performance data to show that the subject device works as intended.
#### 12. Conclusion [21 CFR 807.92(b)(3)]
The modified EVS 4343(G) is substantially equivalent to the currently marketed and predicate device (EVS 4343, K142475) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recogmized consensus standards, IEC60601-1, IEC 60601-1-2, IEC6220-1 and IEC 62133 and the clinical test, which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.
The results of these tests demonstrate that EVS 4343(G) meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.