K162537 · Prismatik Dentalcraft, Inc. · EBF · Nov 16, 2016 · Dental
Device Facts
Record ID
K162537
Device Name
CAD/CAMouflage Milling Block
Applicant
Prismatik Dentalcraft, Inc.
Product Code
EBF · Dental
Decision Date
Nov 16, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3690
Device Class
Class 2
Indications for Use
CAD/CAMouflage Milling Block is indicated as an indirect restorative for both anterior and posterior restorations, including occlusal surfaces. The CAD/CAMouflage Milling Block is made for fabricating temporary and permanent restorations such as inlays, onlays, veneers and full crown restorations.
Device Story
CAD/CAMouflage Milling Block is a nanohybrid composite material used for fabricating indirect dental restorations; anterior/posterior teeth; inlays, onlays, veneers, full crowns. Material is processed via CAD/CAM milling (grinding, shaping, cutting) using a mandrel for manipulation, which is removed prior to final placement. Device is used in dental clinics/labs by dental professionals. Output is a custom-milled dental restoration; provides structural repair for carious lesions or defects. Benefits include restoration of tooth function and aesthetics.
Clinical Evidence
Bench testing only. Evaluated per ISO 4049:2009 and ISO 10477:2004(E). Tests included flexural strength, water sorption, solubility, radio opacity, shade stability, diametral strength, and compressive strength. Biocompatibility data relied on the predicate device.
Technological Characteristics
Nanohybrid composite resin material. Non-sterile, single-use. Complies with ISO 4049:2009 and ISO 10477:2004(E). Form factor is a milling block with a removable mandrel for processing.
Indications for Use
Indicated for patients requiring indirect anterior or posterior dental restorations, including occlusal surfaces, inlays, onlays, veneers, and full crowns.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 16, 2016
Prismatik Dentalcraft, Inc. Lisa Maloney Sr. Manager, Regulatory Affairs 2212 Dupont Drive, Suite P Irvine, California 92612
Re: K162537
Trade/Device Name: CAD/CAMouflage Milling Block Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: Class II Product Code: EBF Dated: October 14, 2016 Received: October 17, 2016
Dear Ms. Lisa Maloney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| Form Approved: OMB No. | 0910-0120 |
|------------------------|--------------------------|
| Expiration Date: | January 31, 2017 |
| | See PRA Statement below. |
| 510(k) Number (if known) | |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | CAD/CAMouflage Milling Block |
| Indications for Use (Describe) | CAD/CAMouflage Milling Block is indicated as an indirect restorative for both anterior and posterior restorations, including occlusal surfaces. The CAD/CAMouflage Milling Block is made for fabricating temporary and permanent restorations such as inlays, onlays, veneers and full crown restorations. |
| Type of Use (Select one or both, as applicable) | <div> <span> <svg height="12" width="12"> <rect height="12" style="fill:black;" width="12"></rect> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> <div> <span> <svg height="12" width="12"> <rect height="12" style="fill:white;stroke:black;stroke-width:1" width="12"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
. The
Armaria
lər Bərək Qaraq Qaran Qaran Qaran Qaran Qara Qaran Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara Qara
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for PRISMATIK DENTALCRAFT, INC. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in a sans-serif font.
# 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
### SUBMITTER INFORMATION A.
| Company Name: | Prismatik Dentalcraft, Inc. |
|-------------------------------|-----------------------------------------------------------------|
| Company Address: | 2212 Dupont Dr., Suite P.<br>Irvine, CA 92612 |
| Company Phone: | 949-440-2636 |
| Company FAX: | 949-553-0924 |
| Facility Registration Number: | 3005477956 |
| Primary Contact Person: | Aliva Das, (949) 225-1249<br>Regulatory Affairs Associate |
| Secondary Contact Person | Lisa Maloney, (949) 440-2631<br>Sr. Manager, Regulatory Affairs |
| Date Summary Prepared: | October 14, 2016 |
#### B. DEVICE IDENTIFICATION
| Trade/Proprietary Name: | CAD/CAMouflage Milling Block |
|-------------------------|------------------------------|
| Common Name: | Tooth Shade Resin Material |
| Classification Code: | Class II, 21 CFR 872.3690 |
| Classification Name: | Tooth Shade Resin Material |
| Product Code: | EBF |
| Panel: | Dental |
{4}------------------------------------------------
#### C. IDENTIFICATION OF PREDICATE DEVICE
| Trade/Proprietary Name: | CAD/CAMouflage Milling Block |
|----------------------------------|------------------------------|
| Predicate Device: | CAD/CAMouflage Milling Block |
| Predicate Clearance Number/Date: | K160425; July 15, 2016 |
#### D. DEVICE DESCRIPTION
The CAD/CAMouflage Milling Block is a nanohybrid composite material that is designed as an indirect restorative for both anterior and posterior restorations including occlusal surfaces. The CAD/CAMouflage Milling Block is made for fabricating indirect restorations such as inlays, onlays, veneers and full crown restorations. The mandrel is solely to aid in the manipulation of the material during processing (grinding, shaping, cutting, etc.). It is removed prior to the material (subject device) being put into final placement. As such, there are no implications for device performance or safety, since it is removed, and it has no patient contact.
#### E. INDICATIONS FOR USE
CAD/CAMouflage Milling Block is indicated as an indirect restorative for both anterior and posterior restorations, including occlusal surfaces. The CAD/CAMouflage Milling Block is made for fabricating temporary and permanent restorations such as inlays, onlays, veneers and full crown restorations.
## SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE F. PROPOSED DEVICE AND THE PREDICATE DEVICES
Prismatik utilized the FDA's Guidance for Industrv and FDA Staff - Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions for comparing its CAD/CAMouflage Milling Blocks with the Primary Device: CAD/CAMouflage Milling Block (K160425) cleared on July 15, 2016. There are no changes in the technological characteristics of the proposed device, and the predicate device. Outlined below are the similarities and the substantial equivalency of the CAD/CAMouflage Milling Blocks and the cited predicate device:
{5}------------------------------------------------
# Comparison of Technological Characteristics of the Proposed Device and the Predicate Devices
| Attributes | Predicate Device: 510(k)<br>K160425<br>CAD/CAMouflage<br>Milling Block<br>Prismatik Dentalcraft, Inc. | Proposed Device:<br>CAD/CAMouflage Milling<br>Block | Similarities<br>&<br>Differences | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|--|
| FDA<br>Classification<br>Name | Tooth shade resin material | Tooth shade resin material | Same | |
| FDA Product<br>Code | EBF | EBF | Same | |
| FDA Device<br>Classification | Class II per 21 CFR 872.3690 | Class II per 21 CFR 872.3690 | Same | |
| Chemical<br>Composition | Composite Restorative Material | Composite Restorative Material | Same | |
| Final Vita Shades | A2 LT | A1 HT,A2 HT,A3 HT,A3.5 HT, B1<br>HT, A1 LT, BL LT, A3 LT, A3.5<br>LT, B1 LT, C2 LT, D2 LT, | Different | |
| Intended Use | Intended to restore carious<br>lesions or structural defects in<br>teeth | Intended to restore carious lesions<br>or structural defects in teeth | Same | |
| Indications | CAD/CAMouflage Milling<br>Block is indicated as an indirect<br>restorative for both anterior and<br>posterior restorations, including<br>occlusal surfaces. The<br>CAD/CAMouflage Milling<br>Block is made for fabricating<br>temporary and permanent<br>restorations such as inlays,<br>onlays, veneers and full crown<br>restorations. | CAD/CAMouflage Milling Block is<br>indicated as an indirect restorative<br>for both anterior and posterior<br>restorations, including occlusal<br>surfaces. The CAD/CAMouflage<br>Milling Block is made for<br>fabricating temporary and<br>permanent restorations such as<br>inlays, onlays, veneers and full<br>crown restorations. | Same | |
| Biocompatibility | Biocompatible | Biocompatible | Same | |
| Sterility | Non-Sterile | Non-Sterile | Same | |
| Single or<br>Multiple Use | Single use | Single use | Same | |
| Prescription/OTC<br>Use | Prescription | Prescription | Same | |
| Substantial<br>Equivalence | The proposed device has the same indications, intended use, biocompatibility, and<br>chemical composition, mechanical and chromatic properties to the predicate device. We<br>consider the proposed device to be substantially equivalent to the predicate device, and<br>believe any differences between the proposed and predicate devices will not pose new<br>safety or effectiveness concerns. The intended use of the modified device has not<br>changed as a result of the modification. | | | |
{6}------------------------------------------------
#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE G.
The comparison table above outlines and provides the similarities and substantial equivalency of the technological characteristics of the predicate device. CAD/CAMouflage Milling Block K160425 and the proposed device CAD/CAMouflage Milling Block. Prismatik believes that the comparative data presented, demonstrate that the CAD/CAMouflage Milling Blocks are essentially the same as currently marketed devices for the same intended use, indications for use, and supports our claim of substantial equivalence to predicate Class II devices under the classification of 21 CFR 872.3690 that have previously been found to be substantially equivalent. The proposed and predicate devices consist of the same Product Code: EBF, and are biocompatible for the same indications for use. The only change is to include the additional shades to the CAD/CAMouflage Milling Blocks.
#### SUMMARY OF NON-CLINICAL TESTING H.
Non-clinical test data was used to support the substantial equivalency. Specifically, the CAD/CAMouflage Milling Blocks were evaluated using the relevant FDA recognized standard: ISO 4049: 2009 Dentistry - Polymer-based restorative materials. In addition. Prismatik utilized the FDA's Guidance for Industry and FDA Staff - Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions. To provide evidence and assurance for the device safety and meeting the ISO 4049 and ISO 10477:2004(E) requirements, Prismatik performed various tests including the Flexural Strength, Water Sorption and Solubility, Radio Opacity, Shade Stability, Diametral Strength, Compressive Strength. Biocompatibility was provided in the predicate device.
#### I. CONCLUSION FROM THE NON-CLINICAL TESTING
The results of the above-described studies demonstrate that the CAD/CAMouflage Milling Blocks are substantially equivalent to the cleared Primary Device: CAD/CAMouflage Milling Blocks: K160425.
The modification for our CAD/CAMouflage Milling Block is to add additional shades as provided in the Comparison Table above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.