STERISPINE CC

K162531 · Safe Orthopaedics · OVE · Dec 29, 2016 · Orthopedic

Device Facts

Record IDK162531
Device NameSTERISPINE CC
ApplicantSafe Orthopaedics
Product CodeOVE · Orthopedic
Decision DateDec 29, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sterispine™ CC is intended to be used as an intervertebral body fusion cage as a standalone system used with bone screws provided and requires no additional supplementary fixation systems. It is inserted between the vertebral bodies into the disc space from levels C2 to T1 for the treatment of cervical degenerative disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The device system is designed for use with autograft bone to facilitate fusion. Sterispine™ CC is intended to be used at one level. The cervical cage is to be used in a skeletally mature patient who has had six weeks of non-operative treatment prior to implantation of the cage.

Device Story

Intervertebral body fusion cage system; used as standalone device for cervical fusion. System comprises PEEK cage, Tantalum marker, and Titanium fixation screws. Implanted into disc space between C2-T1 levels; requires autograft bone to facilitate fusion. Used by surgeons in clinical settings. Provides mechanical stability to vertebral bodies; eliminates need for additional supplementary fixation. Benefits patient by promoting spinal fusion and alleviating discogenic neck pain.

Clinical Evidence

Bench testing only. Performed static/dynamic compression, compression shear, and torsion (ASTM F2077); subsidence (ASTM F2267); expulsion (in-house protocol); and wear testing (ASTM F1877).

Technological Characteristics

Cervical cage: VESTAKEEP i4R (PEEK, ASTM F2026) with Tantalum marker (ASTM F560). Fixation screws: Titanium Ta6V Eli (ASTM F136). Available in 6 heights (5-10mm) and 2 diameters (12-14mm). Supplied sterile with single-use surgical instruments.

Indications for Use

Indicated for skeletally mature patients with cervical degenerative disc disease (neck pain of discogenic origin) at one level from C2 to T1, following six weeks of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 29, 2016 Safe Orthopaedics Sophie Rehault OA/RA Director Parc des Bellevues Allée R. Luxemburg - Bat. Californie 95610 Eragny sur Oise, France Re: K162531 Trade/Device Name: Sterispine™ CC Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: November 30, 2016 Received: December 2, 2016 Dear Ms. Rehault: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. ## Caroline Rhim -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K162531 Device Name STERISPINETM CC ### Indications for Use (Describe) The Sterispine™M CC is intended to be used as an intervertebral body fusion cage as a standalone system used with bone screws provided and requires no additional supplementary fixation systems. It is inserted between the vertebral bodies into the disc space from levels C2 to T1 for the treatment of cervical degenerative disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The device system is designed for use with autograft bone to facilitate fusion. Sterispine™ CC is intended to be used at one level. The cervical cage is to be used in a skeletally mature patient who has had six weeks of non-operative treatment prior to implantation of the cage. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <span> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for SafeOrthopaedics. The logo features a stylized letter 'S' in a light green color, with a small dot to the right of the 'S'. Below the 'S', the words 'SafeOrthopaedics' are written in a smaller font size, with the 'e' in 'Safe' also in light green. # 5.510(k) summary ### 510(k) Summary | 510k | Traditional | |------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitted by | Safe Orthopaedics | | | Parc des Bellevues | | | Allée R. Luxemburg - Bat. Californie<br>95610 Eragny sur Oise - FRANCE | | | Contacts | | s.rehault@safeorthopaedics.com | | | +33 (0) 1 34 21 50 00 | | | Isabelle DRUBAIX Regulatory contact | | | idee-consulting@nordnet.fr | | | +33 (0)3 21 05 64 23 | | | Date Prepared | 1st September 2016 | | Common Name | Intervertebral body fusion device | | Trade Name | SteriSpine™ CC | | Classification Name | Intervertebral body fusion device | | Class | II | | Product Code | OVE | | CFR section | 888.3080 | | Device panel | Orthopedic | | Legally marketed<br>predicate devices | Primary predicate: Stalif-C manufactured by Centinel | | | Spine cleared under K120819 | | | Other predicate: PEEK PREVAIL CERVICAL INTERBODY | | | DEVICE manufactured by Medtronic Sofamor Danek | | | cleared under K073285. | | STERISPINET™ CC | | | Indications for use | The Sterispine™ CC is intended to be used as an<br>intervertebral body fusion cage as a standalone system<br>used with bone screws provided and requires no<br>additional supplementary fixation systems. It is inserted<br>between the vertebral bodies into the disc space from<br>levels C2 to T1 for the treatment of cervical degenerative<br>disc disease (defined as neck pain of discogenic origin<br>with degeneration of the disc confirmed by history and<br>radiographic studies). The device system is designed for<br>use with autograft bone to facilitate fusion. Sterispine™<br>CC is intended to be used at one level. | | | The cervical cage is to be used in a skeletally mature<br>patient who has had six weeks of non-operative<br>treatment prior to implantation of the cage | | Description of the device | Sterispine™ CC system includes one cervical cage and<br>four cervical fixations (two cervical fixations screws and<br>two rescue screws) but only two screws are used within<br>the cervical cage. Implantable components of<br>Sterispine™ CC are made of Titanium Ta6V Eli grade<br>conforming to ASTM F136 for screws and the cervical<br>cage is made of VESTAKEEP i4R (PEEK) conforming to the<br>ASTM F2026 with one Tantalum marker conforming to<br>ASTM F560 to visualize the position of the cervical cage<br>in the disc space. The Sterispine™ CC range of products<br>is supplied sterile with a sterile single-use set of surgical<br>instruments. | | Summary of<br>Technological<br>Characteristics | The purpose of this 510(k) submission is to obtain<br>clearance for cervical cages available in 6 heights (from<br>5mm to 10mm) and 2 diameters (12 & 14mm). The<br>width is 14.5mm for a diameter of 12mm and for a height<br>from 5 to 8mm. The width is 16mm for a diameter of<br>14mm and heights from 5 to 10mm. | | Discussion of Testing | Testing in compliance with: FDA's "Class II Special<br>Controls Guidance Document: Intervertebral Body Fusion<br>Device" was performed for the Sterispine™ CC and<br>demonstrated substantially equivalent performance to<br>the identified predicate devices. The following non-clinical<br>tests were conducted on Sterispine™ CC: Static<br>compression, Static compression shear, Static torsion,<br>Dynamic compression, Dynamic compression shear,<br>Dynamic torsion according to ASTM F2077, Subsidence<br>according to ASTM F2267, Expulsion according to in-<br>house protocol and wear testing according to ASTM<br>F1877 | | Conclusion | Based on the design features, the use of established well known materials, feature comparisons, indications for use, and results of the mechanical testing, the Sterispine™ CC has demonstrated substantial equivalence to the identified predicate devices. | {4}------------------------------------------------ ## Traditional 510k STERISPINE™ CC Image /page/4/Picture/1 description: The image shows the logo for Safe Orthopaedics. The logo features a stylized letter 'S' in a light green color, with a dot to the right of the 'S'. Below the 'S', the words 'SafeOrthopaedics' are written in a sans-serif font, with the 'e' in 'Safe' colored light green to match the 'S' above. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Safe Orthopaedics. The logo features a stylized, cursive letter "S" in a light green color. Below the "S" are the words "SafeOrthopaedics," with "Safe" in a slightly darker green and "Orthopaedics" in a lighter shade. ## Traditional 510k STERISPINE™ CC
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...