K162472 · Varian Medical Systems, Inc. · IYE · Jan 19, 2017 · Radiology
Device Facts
Record ID
K162472
Device Name
TrueBeam-TrueBeam STx-Edge
Applicant
Varian Medical Systems, Inc.
Product Code
IYE · Radiology
Decision Date
Jan 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The TrueBeam TM system is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Device Story
Medical linear accelerator system; delivers photon, electron, and kV X-ray radiation for radiotherapy and stereotactic radiosurgery. Installed in radiation-shielded vault; operated by clinicians via control console outside treatment room. System integrates imaging (CBCT, 4D CBCT, iCBCT) and automated delivery features (HyperArc, respiratory gating) to target lesions/tumors. Software processes imaging data to guide beam delivery; enables precision treatment of various malignancies and benign conditions. Benefits include high-precision dose delivery to tumors while sparing surrounding healthy tissue; supports complex clinical workflows through automated exposure control and auxiliary interfaces.
Clinical Evidence
Bench testing only. Verification and validation testing performed per 21 CFR 820, ISO 13485, and ISO 14971. Software validation conducted per FDA guidance; classified as major level of concern. Biocompatibility evaluated per ISO 10993-1. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
Medical linear accelerator; photon/electron/kV X-ray beam source. Features: HyperArc, gated/4D/extended/iCBCT imaging, automatic exposure control. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-1, IEC 60825, IEC 61217, IEC 62274, IEC 62304, IEC 62366, IEC 60976. Biocompatibility per ISO 10993-1. Software: Major level of concern.
Indications for Use
Indicated for stereotactic radiosurgery and precision radiotherapy in adults and pediatric patients for lesions, tumors, and conditions including brain, spine, head and neck, thoracic, gynecologic, gastrointestinal, genitourinary, lymphoid, skin, and pediatric tumors, as well as benign diseases and metastases.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
TrueBeam Radiotherapy System and Accessories (K143224)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of four faces in profile, stacked on top of each other. The faces are stylized and appear to be connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2017
Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Varian Oncology Systems Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
Re: K162472
Trade/Device Name: Truebeam STx. and Edge Radiotherapy Delivery System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: December 22, 2016 Received: December 27, 2016
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162472
#### Device Name
TrueBeam, TrueBeam STx and Edge Radiotherapy Delivery System
#### Indications for Use (Describe)
The TrueBeam and Edge Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.
The TrueBeam and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck turnors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinomaly tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation and skull base tumors).
Type of Use (Select one or both, as applicable)
| | <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--|----------------------------------------------------------------------------|
| | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Premarket Notification [510(k)] Summary TrueBeam Radiotherapy Treatment System
The following information follows the format of 21 CFR 807.92.
| I. Submitter's Name: | Varian Medical Systems, Inc.<br>3120 Hansen Way C-260<br>Palo Alto, CA 94304<br><br>Contact Name: Peter J. Coronado<br>Phone: 650.424.5731<br>Fax: 650.842.5040<br>Date: December 2016 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| II. Trade Name: | TrueBeamTM/TrueBeam STXTM/EdgeTM |
| Common Name: | Linear accelerator radiation therapy system |
| Classification Name: | Medical charged-particle radiation therapy system<br>21 CFR 892.5050, Class II<br>Product Code: 90 IYE |
| III. Predicate Device: | TrueBeam Radiotherapy System and Accessories: K143224 |
| IV. Device Description: | The TrueBeamTM Radiotherapy Delivery System is a medical linear accelerator.<br>The system consists of two major components: 1) a photon, electron, and diagnostic kV X-ray radiation beam-producing component that is installed in a radiation-shielded vault in a healthcare facility and 2) a control console using the device software in an area located outside the treatment room. |
| V. Intended Use<br>Statement | The TrueBeam TM system is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. |
| Indications for Use<br>Statement | The TrueBeam, TrueBeam STx and Edge Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.<br><br>The TrueBeam and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney) |
{4}------------------------------------------------
lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation and skull base tumors).
# VI. Technological Characteristics:
Listed below is a comparison of significant changes to the predicate device.
| Feature | Cleared device | Device with change |
|--------------------------------------|----------------|--------------------|
| Hyperarc Automated Dynamic Beam | No | Yes |
| Gated CBCT | No | Yes |
| Online 4D CBCT | No | Yes |
| Extended Length CBCT | No | Yes |
| iCBCT | No | Yes |
| 4D CBCT reconstruction new algorithm | No | Yes |
| Automatic Exposure Control | No | Yes |
| Auxiliary Device Interface v3.0 | No | Yes |
| Single Camera for Respiratory Gating | No | Yes |
## VII. Summary of performance testing:
Hardware and software verification and validation testing was conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards listed below. Bench testing results showed passing criteria were met, device conformance to applicable requirements specifications and assured hazard safeguards functioned properly.
Software verification and validation bench testing were conducted. Results showed passing criteria were met and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
The biocompatibility evaluation for the patient-contact materials in this medical device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
Electrical safety and electromagnetic compatibility (EMC) testing was conducted on this medical device. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC.
Varian TrueBeam™, TrueBeam STx™ and Edge™ medical linear Standards accelerator models conform in whole or in part with the following FDA conformance:
{5}------------------------------------------------
| AAMI/ANSI ES60601-1: 2005/<br>(R)2012 & A1:2012; C1:2009/<br>(R)2012 & A2:2010/ (R)2012<br>ANSI/AAMI/ISO 10993-1: 2009 | IEC 60601-1-2:2007<br>IEC 60601-1-3: 2013<br>IEC 60601-1-6: 2013<br>IEC 60601-2-1: 2014<br>IEC 60825: 2007 | IEC 61217: 2011<br>IEC 62274: 2005<br>IEC 62304: 2006 IEC<br>62366:2015 IEC<br>60976: 2007 |
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
#### Conclusion: The results of verification, validation and safety standards bench testing demonstrate that the Varian Medical Systems TrueBeam™, TrueBeam STx™ and Edge™ medical linear accelerators are substantially equivalent to their predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.