CODMAN VersaTru Standard Disposable Non-Stick Bipolar Forceps and CODMAN VersaTru Slim Disposable Non-Stick Bipolar Forceps are intended for use in electrosurgery for coagulation of tissue.
Device Story
Single-use, sterile, disposable electrosurgical bipolar forceps; connects to external electrosurgical bipolar generator via bipolar cord. Device delivers electrical current from generator to distal tips to coagulate tissue. Used in clinical settings by surgeons. Output is thermal energy for tissue coagulation; assists in achieving hemostasis during surgery. Benefits include non-stick tip properties and disposable design to ensure sterility and performance consistency.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including lesion size, resistivity, weight, non-stick performance, thermal effects, mechanical integrity, dimensional verification, and ergonomics validation. Biocompatibility, electrical safety (IEC 60601-1, 60601-2-2), and EMC testing (IEC 60601-1-2) were performed.
Technological Characteristics
Bayonet-style bipolar forceps; plated aluminum tips; polyamide coating. Energy source: external electrosurgical generator. Connectivity: wired connection to generator. Sterilization: gamma irradiation (SAL 10^-6). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 10993, ISO 11137.
Indications for Use
Indicated for use in electrosurgery for coagulation of tissue in patients requiring surgical intervention.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Synergetics Disposable Spetzler Malis Standard Bipolar Forceps (K121426)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 26, 2016
Medos International SARL Ms. Jennifer Siu Regulatory Affairs Specialist Chemin-Blanc 38 2400 LeLocle, Switzerland
Re: K162469
Trade/Device Name: CODMAN VersaTru Standard Disposable Non-Stick Bipolar Forceps. CODMAN VersaTru Slim Disposable Non-Stick Bipolar Forceps Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 1, 2016 Received: September 2, 2016
Dear Ms. Siu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162469
Device Name
CODMAN VersaTru Standard Disposable Non-Stick Bipolar Forceps CODMAN VersaTru Slim Disposable Non-Stick Bipolar Forceps
Indications for Use (Describe)
CODMAN VersaTru Standard Disposable Non-Stick Bipolar Forceps and CODMAN VersaTru Slim Disposable Non-Stick Bipolar Forceps are intended for use in electrosurgery for coagulation of tissue.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | <span> <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">☒</span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">☐</span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K162469 – 510(k) Summary
#### Codman & Shurtleff, Inc. I. Submitter 325 Paramount Drive Raynham, MA 02767
On behalf of: Medos International SARL Chemin-Blanc 38 2400 LeLocle, Switzerland
Contact: Jennifer Siu Phone: (508) 828-3288 Fax: (508) 977-6979
Date of Submission: September 1, 2016
# II. Device
| II. Device | | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| | Device Proprietary Name | CODMAN VersaTru™ Disposable Non-Stick Bipolar Forceps |
| | Common Name | Disposable Bipolar Forceps |
| | Classification Name | Electrosurgical, Cutting & Coagulation & Accessories<br>(21 CFR 878.4400) |
| | Regulatory Classification | II |
| | Product Code | GEI |
| III. Predicate Device | The predicate device for this submission is the Synergetics™ Disposable<br>Spetzler™ Malis® Standard Bipolar Forceps (K121426), which was cleared<br>on July 10, 2012. | |
| IV. Device Description | The CODMAN VersaTru Disposable Non-Stick Bipolar Forceps (VersaTru<br>Bipolar Forceps) are single use, disposable, sterile electrosurgical devices, for<br>use in electrosurgery for the coagulation of tissue. The forceps are part of an<br>electrosurgical system consisting of an electrosurgical bipolar generator and<br>bipolar cord attached to the proximal end of the forceps, to provide power and<br>deliver electrical current from the generator to the distal tips of the forceps. | |
| V. Indications for Use | CODMAN VersaTru Disposable Non-Stick Bipolar Forceps are intended for<br>use in electrosurgery for coagulation of tissue. | |
{4}------------------------------------------------
# VI. Comparison to Predicate Device
The VersaTru Bipolar Forceps is substantially equivalent to the predicate device, the Synergetics Disposable Spetzler Malis Standard Bipolar Forceps. The subject device has the same indications for use and clinical utility, and similar design principles, materials, and packaging as the predicate device. The table below provides a comparison between the subject device and the predicate device.
| Comparison of the Predicate and Subject Device | | |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| | Predicate Device:<br>Synergetics Disposable<br>Spetzler Malis Standard<br>Bipolar Forceps<br>(K121426) | Subject Device:<br>CODMAN VersaTru<br>Disposable Non-Stick Bipolar<br>Forceps<br>(This Submission) |
| FDA Product Code | GEI | Same as predicate |
| Classification | Class II - 21 CFR 878.4400 | Same as predicate |
| Classification Name | Electrosurgical, Cutting &<br>Coagulation & Accessories | Same as predicate |
| Indications for Use | Sterile, single use, for use in<br>electrosurgery for coagulation of<br>tissue | Intended for use in<br>electrosurgery for coagulation of<br>tissue |
| Single Use | Yes | Same as predicate |
| Forceps Design | Bayonet Style | Same as predicate |
| Product Line | Standard | Standard<br>Slim |
| Forceps Length | 7", 8", 9" | Standard: 7", 8", 9"<br>Slim: 8", 9" |
| Forceps Tip Size | 0.5 mm, 1.0 mm, 1.5 mm | Same as predicate |
| Forceps Color | Yellow | Blue |
| Forceps Tips | Plated aluminum | Plated aluminum |
| Coating | Polyvinylidene fluoride | Polyamide |
| Electrical Safety<br>Testing | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | Same as predicate |
| Packaging<br>Materials<br>(Sterile Barrier) | PETG tray<br>Tyvek lid | Same as predicate |
| Packaging Unit | 5 units per box | 1 unit per box |
| Sterility<br>Assurance Level<br>(SAL) | 10-6 | Same as predicate |
| Sterilization<br>Method | Ethylene oxide | Gamma irradiation |
| Shelf Life | 5 years | 1 year |
#### VII. Performance Data
The following performance data has been provided in support of the substantial equivalence determination. All testing was performed on final sterile devices unless otherwise specified.
# Bench Testing
Performance bench testing was conducted in alignment with FDA's guidance
{5}------------------------------------------------
Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements.
| Performance Bench Test Results | |
|------------------------------------|------------|
| Test | Conclusion |
| Lesion Size | Pass |
| Resistivity | Pass |
| Weight Test | Pass |
| Non-Stick Test | Pass |
| Thermal Effects Lesion Study | Pass |
| Mechanical Test | Pass |
| Dimensional Verification | Pass |
| Physical Characterization | Pass |
| Functional Testing (after transit) | Pass |
| Ergonomics Validation | Pass |
# Electromagnetic Compatibility/ Electrical Safety Testing
EMC and electrical safety testing were conducted in accordance with IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, IEC 60601-1:2005/A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, and IEC 60601-2-2:2009/C1:2014 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. The VersaTru Bipolar Forceps passed all EMC and electrical safety testing.
# Sterilization
The VersaTru Bipolar Forceps is sterilized using a validated gamma irradiation sterilization cycle. The sterilization cycle has been validated to ensure a sterility assurance level (SAL) of 10-6 in accordance with ISO 11137-1:2006/Amd1:2013 and ISO 11137-2:2013, Sterilization of health care products – Radiation.
# Shelf-Life Testing
Shelf-life testing was conducted in accordance with FDA's guidance document Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements. VersaTru Bipolar Forceps was subjected to accelerated aging. The aging studies established that the device and packaging remain functional and maintain sterility for up to 1 year.
# Biocompatibility Testing
Biocompatibility testing was conducted in accordance with ISO 10993-1:2009/AC:2010, Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process, and FDA's guidance documents, Use of International Standard ISO 10993-1, "Biological
{6}------------------------------------------------
evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued June 16, 2016 and Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016.
| Biocompatibility Test Results | | |
|--------------------------------------------------------------------------------------|-------------------------|--|
| Test | Conclusion | |
| Cytotoxicity Study using the ISO Elution Method<br>(1X MEM)<br>ISO 10993-5 | PASS<br>Non-cytotoxic | |
| ISO Guinea Pig Maximization Sensitization Test<br>ISO 10993-10 | PASS<br>Non-sensitizing | |
| ISO Intracutaneous Study in Rabbits<br>ISO 10993-10 | PASS<br>Non-irritating | |
| ISO Systemic Toxicity Study in Mice<br>ISO 10993-11 | PASS<br>Non-toxic | |
| USP Rabbit Pyrogen Study, Material Mediated<br>ISO 10993-11<br>General Chapter <151> | PASS<br>Non-pyrogenic | |
| ASTM Hemolysis Study (Extract Only)<br>ISO 10993-4<br>ASTM F756 | PASS<br>Non-hemolytic | |
| Chemical Characterization of Extractables<br>ISO 10993-18 | PASS | |
| Toxicology Risk Assessment<br>ISO 10993-17 | PASS | |
# Animal Studies
No animal studies were performed as appropriate verification and validation of the new device was achieved based on the comparison to the predicate device and from the results of the bench testing, biocompatibility evaluation, and electrical/safety testing.
# Clinical Studies
No clinical studies were performed as appropriate verification and validation of the new device was achieved based on the comparison to the predicate device and from the results of the bench testing, biocompatibility evaluation, and electrical/safety testing.
VIII. Conclusion
Based upon the intended use, design, function, materials, comparison to the predicate device, and testing conducted, it is concluded that the subject device, VersaTru Bipolar Forceps, is substantially equivalent to the predicate device, Synergetics Disposable Spetzler Malis Standard Bipolar Forceps (K121426), and therefore does not raise different issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.