WAVi™ Headset and WAVi™ eSoc™ Single Use Electrode Contacts
Applicant
Wavi Co.
Product Code
GXY · Neurology
Decision Date
Apr 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The WAVi Headset is intended for use in routine clinical and research settings where rapid placement of a number of EEG electrodes is desired.
Device Story
WAVi Headset is an EEG electrode positioning system; facilitates rapid, uniform placement of electrodes per international 10-20 system. Device inputs: electrophysiological brain signals acquired via 0.9% Normal Saline-soaked eSoc electrode contacts. Device transforms inputs by conducting signals through embedded wire harness to 32-pin connector port for transmission to external EEG data collection device. Used in healthcare facilities/hospitals; operated by clinicians/researchers. Output: raw EEG signals for recording/evaluation by healthcare provider. Benefits: enables consistent, rapid EEG setup (under 20 minutes). Non-invasive, non-sterile, non-radiation emitting.
Clinical Evidence
Bench testing only. Comparison of WAVi Headset and Electro-Cap performed on three test subjects using a Lexicor Neurosearch-24 Brain mapper. Spectral shape and maximum frequency at 10/20 EEG locations were consistently similar between devices. Quality assurance testing included visual inspection, dimensional verification, and individual electrode resistance testing.
Technological Characteristics
Full head cap system; 22 contacts (19 recording channels). Materials: EVA (headset), Nylon 6/6 (101), and pure tin (electrodes). Conductive medium: 0.9% Normal Saline. Connectivity: 32-pin connector port, standard ribbon cable, lead wires. Non-sterile. 10-20 system placement. Standalone passive electrode array.
Indications for Use
Indicated for use in routine clinical and research settings for rapid placement of EEG electrodes in adults and children.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, creating a flowing, abstract design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2017
WAVi, Co. David Jones FDA Consultant 3535 S Irving St. Englewood, Colorado 80110
Re: K162460
Trade/Device Name: WAVi™ Headset and WAVi™ eSoc™ Single Use Electrode Contacts Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: March 22, 2017 Received: March 31, 2017
Dear Mr. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely.
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162460
Device Name
WAVi™ Headset and WAVi™ eSocTM Single Use Electrode Contacts
Indications for Use (Describe)
The WAV Headset is intended for use in routine clinical and research settings where rapid placement of a number of EEG electrodes is desired.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Submitted by: | WAVi Co.<br>3535 S. Irving Street<br>Englewood, CO 80110<br>804-909-2389 | |
|----------------------|---------------------------------------------------------------------------------------------------------|-------------------|
| Contact Person: | David Jones<br>281-989-8515 | |
| Date Prepared: | April 28, 2017 | |
| Proprietary Name: | WAVi™ Headset and WAVi™ eSoc™ Single Use Electrode<br>Contacts | |
| Model Numbers: | WH-100: XS Headset<br>WH-200: S Headset<br>WH-300: M Headset<br>WH-400: L Headset<br>WH-500: XL Headset | |
| Common Name: | EEG 10-20 Electrode Headset and Electrodes | |
| Classification: | Class II: | 21 CFR § 882.1320 |
| Classification Name: | Cutaneous Electrode – GXY | |
| Predicate Devices: | Electro-Cap™ System (K112319)<br>Electro-Cap™ Intl., Inc.<br>1011 West Lexington Rd.<br>Eaton, OH 45320 | |
#### Device Description:
The WAVI™ Headset is an EEG electrode positioning system used to quickly place the electrodes in a uniform and consistent manner in accordance with the international standard Ten-Twenty System (10-20) to acquire electrophysiological EEG signals from an individual to a suitable EEG data collection device.
The device consists of the WAVi™ Headset, WAVi™ eSoc™ Single Use and Tin Electrode Contacts; the head set comes in five models/sizes (XS, S, M, L, XL). This device is portable, nonsterile, non-invasive, non-radiation emitting, point-of-care use device for use in healthcare facilities and hospitals.
Device characteristics include the eSoc™ Single Use Electrode Contacts which are soaked in 0.9% Normal Saline which is unique as it serves as the electro-conductive material and patient contact allowing the brain's electrical signals to be read through an EEG data collection device. Typical set-up and procedure time is less than twenty minutes.
The device does not contain software, biologics, drugs, coatings or any claim of sterility.
{4}------------------------------------------------
The WAVI™ Headset, eSoc™ Single Use and Ear Electrode Contacts are made from wellestablished medical grade materials; the Headset is made from a proprietary EVA material with tin plated ring electrode ports for placement of the Nylon 101 WAVi™ eSoc™ Single Use Electrode Contacts, and two ear electrodes are also made from Tin. A wire harness is embedded between two layers of EVA and is attached to each electrode port. The wire harness exits the headset to a 32-pin connector port.
### Intended Use:
The WAVi Headset is intended for use in routine clinical and research settings where rapid placement of a number of EEG electrodes is desired.
#### Technological Characteristics
| Manufacturer | WAVi Co. | Electro-Cap |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | WAVi Headset | Electro-Cap System (K780045) |
| Indication for Use | The WAVi Headset is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired. | The Electro-Cap is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired. |
| Environmental Use | Electrophysiological | Electrophysiological |
| Target Patient | Adults and Children | Adults and Children |
| Where Used | On the head | On the head |
| Anatomical Contact Sites | Patient's skin (scalp) | Patient's skin (scalp) |
| Number of Contacts | 22 | 2 to 256 |
| Sterile | No | No |
| Size of Cap | Various- Extra Small to Extra Large | Various- Extra Small to Large |
| Style of Cap | Full Head Cap | Full Head Cap |
| Cap Material | EVA | Spandex |
| Location of Wiring | Inside Cap | Inside Cap |
| Type of Cables | Standard Ribbon Cable and Lead Wires | Standard Ribbon Cable and Lead Wires |
| Type of Electrode Drop | Detachable and Non Detachable | Detachable and Non Detachable |
| Electrode Material | Nylon 6/6 (101) and Pure Tin | Pure Tin |
| Electrode Placement System | The International 10-20 System is used as a basis for the electrode placement. | The International 10-20 System is used as a basis for the electrode placement. |
| Number of Recording Channels | 19 | 19 |
| Electrode Positions Utilized | Fp1, Fp2, F7, F3, Fz, F4, F8, T3, C3, Cz, C3, T4, T5, P3, Pz, P4, T6, 01, 02, A1, A2 | Fp1, Fp2, F7, F3, Fz, F4, F8, T3, C3, Cz, C3, T4, T5, P3, Pz, P4, T6, 01, 02, A1, A2 |
| Type of Connectors | D-Sub Connectors, Touch Proof Din Sockets and Special Connectors to Match EEG Equipment and Computers | D-Sub Connectors, Touch Proof Din Sockets and Special Connectors to Match EEG Equipment and Computers |
| Biocompatibility Testing | None was conducted | None was conducted |
The following is a side-by-side of the WAVi device compared to the Predicate:
{5}------------------------------------------------
| Performance<br>Requirements | Needs to transmit<br>electrophysiological signals from an<br>individual to data collection devices.<br>Does not transmit electrical current,<br>nor are they intended to be used for<br>stimulation. | Needs to transmit<br>electrophysiological signals from an<br>individual to data collection devices<br>with a maximum impedance of<br>5K/Ohms. Does not transmit<br>electrical current, nor are they<br>intended to be used for stimulation. |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
The electrical activity of the brain is acquired through 0.9% Normal Saline soaked eSocTM inserted into the electrode ports and connected through the Headset's wiring harness to the Patient Cable which is connected to the EEG device where brain activity may be recorded for evaluation.
## Non-Clinical Testing:
The WAVI™ Headset and accessories are manufactured of medical grade materials including ethylene vinyl acetate (EVA), nylon 6/6 (101) and tin. Multiple 510 (k)'s have been cleared for each of these materials demonstrating no need for further biocompatibility testing.
The WAVI™ Headset and accessories' Quality Assurance testing includes visual inspection of the headset, label and labeling, dimensional verification, and individual electrode resistance of the components and finished product. The WA Vi™ Headset and accessories meet all performance specifications.
The WAVi™ Headset and the Electro-Cap™ System were compared for sizing, electrode placement and labeling, and was found to be substantially equivalent.
The WAVI™ Headset and the Electro-Cap™ were also compared side-by-side on three test subjects using a Lexicor Neurosearch-24 Brain mapper (K915820); the spectral shape and maximum frequency of the corresponding WAVi Headset and Electro-Cap spectrums at each of the 10/20 EEG locations were consistently similar within each subject.
#### Substantial Equivalence
The WAVi™ Headset, WAVi™ eSoc™ Single Use and Tin Ear Electrode Contacts are portable, non-sterile, non-invasive, non-radiation emitting, point of care, electroencephalogram (EEG) devices, and is intended for use in routine clinical and research settings where rapid placement of a number of EEG electrodes is desired.
All non-clinical tests demonstrate that the WAVI™ Headset and accessories are as safe, as effective, and perform as well as or better than the legally marketed predicate device.
The WAVi™ Headset is substantially equivalent to the predicate devices in the following manner:
- . Same intended use
- Same operating principle ●
- Same fundamental scientific technology
- Same or substantially equivalent materials, including headset and electrodes. .
There are no other substantial or significant differences between the WA Vi™ Headset and its' accessories and the predicate that would affect safety, effectiveness or performance, therefore the WAVi™ Headset and accessories' are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.