K162459 · Microprobes For Life Science, Inc. Dba Neuron · GZL · Jun 15, 2017 · Neurology
Device Facts
Record ID
K162459
Device Name
AccuPoint Electrode
Applicant
Microprobes For Life Science, Inc. Dba Neuron
Product Code
GZL · Neurology
Decision Date
Jun 15, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1330
Device Class
Class 2
Indications for Use
Neuron AccuPoint™ Electrodes are intended for use in intraoperative recording of single unit neuronal activity or intraoperative stimulation of neural elements in the brain.
Device Story
AccuPoint Electrode is a depth electrode for intraoperative recording/stimulation of single unit neuronal activity; used in stereotactic neurosurgery to assist in target localization for implant placement. Device consists of a tungsten or platinum-iridium core needle insulated with Parylene C, housed within a stainless steel and polyimide cannula. Features include a sliding mechanism for tip extension/retraction, a stop collar for frame placement, and a reference contact. Connects to third-party MER/microstimulation equipment via gold pin connectors. Clinician operates device within a stereotaxic frame; output is electrical signal data displayed on external equipment to guide surgical decision-making. Benefits include precise neural mapping and accurate electrode placement.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1, cytotoxicity), electrode impedance, continuity, insulation integrity, and dynamic impedance testing to verify performance and safety.
Technological Characteristics
Depth electrode; tungsten or platinum-iridium (80/20) core; Parylene C insulation; stainless steel and polyimide cannula; gold pin connectors. Dimensions: 235.24-291 mm length, 0.200-0.250 mm diameter. Impedance: 0.5-1 MΩ. Sterilization: Ethylene Oxide (EO). Non-pyrogenic. Standalone device connecting to third-party MER/stimulation systems.
Indications for Use
Indicated for intraoperative recording of single unit neuronal activity or intraoperative stimulation of neural elements in the brain during neurosurgical procedures.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 15, 2017
Microprobes For Life Science, Inc. dba Neuron Julie Stephens President Regulatory Resources Group, Inc. 111 Laurel Ridge Drive Alpharetta, Georgia 30004
Re: K162459
Trade/Device Name: Accupoint Electrode Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: May 15, 2017 Received: May 16, 2017
Dear Ms. Stephens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162459
Device Name AccuPoint™ Electrode
Indications for Use (Describe)
Neuron AccuPoint™ Electrodes are intended for use in intraoperative recording of single unit neuronal activity or intraoperative stimulation of neural elements in the brain.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# MicroProbes for Life Science, Inc. dba Neuron Traditional 510(k) - AccuPoint™ Electrode
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements under 21 CFR 807.92.
| Submitted By: | MicroProbes for Life Science, Inc. dba Neuron<br>18247-D Flower Hill Way<br>Gaithersburg, MD 20879<br>Phone: (301) 330-9788 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Julie Stephens, President/Consultant<br>Regulatory Resources Group, Inc. |
| Date Prepared: | June 8, 2017 - Revised |
Device Name and Classification:
| Trade/Proprietary Name: | AccuPoint™ Electrode | Common Name: | Depth Electrode |
|-------------------------|----------------------|---------------|-----------------|
| Classification Name: | Depth Electrode | Product Code: | GZL |
| Regulation: | 21 CFR 882.1330 | Class: | II |
#### Legally Marketed Predicate Device:
FHC, Inc. - FHC microTargeting Electrode, 510(k) # K033173 & K991522 Reference device: Alpha Omega - Sterile Disposables for MER, 510(k) # K120098
#### Device Description:
The AccuPoint™ Electrode is used for monitoring, recording, or microstimulation during intraoperative procedures to map the electrical activity of single unit neurons at subsurface levels of the brain. Intraoperative microelectrode recording (MER) and microstimulation equipment is utilized with depth electrodes during stereotactic surgery to assist in localization for further applications such as permanent implant placement.
The MER or microstimulation equipment which connects to the AccuPoint Electrode are supplied by other manufacturers and is not sold by MicroProbes for Life Science, Inc. The AccuPoint Electrode is produced with standard connectors to connect with other manufacturers' cables that then connect to MER and microstimulation equipment.
The AccuPoint Electrode is a long, extremely thin metal needle insulated with a Parylene C coating. This needle will be composed of either tungsten or a platinum-iridium alloy, depending on the preference of the clinician. The tip of this needle is exposed using either an electric arc or laser process, to create an electrode. The tip exposure length can be carefully controlled to imbue the electrode with a particular impedance. This core needle electrode is situated within a stainless steel and polyimide cannula, with the electrode and cannula being employed together as an integrated unit. The electrode will be capable of sliding within the cannula. allowing the clinician to extend the tip during use, or retract the tip back within the cannula during handling and storage. The rear portion of the cannula will be equipped with a "stop collar", or a thick stainless steel ring which assures proper placement of the electrode within the surgical frame. A laver of stainless steel within the cannula will serve as a second electrode or "reference contact". The electrode may or may not be equipped with an outer stainless steel shield. The electrode will be equipped with two gold pin connectors on the back end, which allow the electrode to be connected to the cable leading to the MER or microstimulation equipment possessed by the surqeon. The model number of each electrode configuration was chosen to be analogous to the comparable predicate electrode model from one of two other manufacturers (FHC and Alpha Omega), allowing clinicians to easily adopt the AccuPoint
{4}------------------------------------------------
# MicroProbes for Life Science, Inc. dba Neuron Traditional 510(k) - AccuPoint™ Electrode
Electrode with minimal confusion. Only a length within 5 cm of the AccuPoint Electrode comes into direct contact with patient tissue.
Accessories include stainless steel spacer and insertion tubes and stylets which are used during stereotaxic functional neurosurgical procedures to accurately guide a microelectrode or instrument into the brain. The insertion tubes with the stylet inserted are initially placed through the stereotaxic frame access locations for determining the estimated depth of the target location to then insert the depth electrode (AccuPoint Electrode).
The AccuPoint Electrodes and Accessories will be sold sterile, for single patient use, and will be disposable.
#### Indications for Use:
Neuron AccuPoint™ Electrodes are intended for use in intraoperative recording of single unit neuronal activity or intraoperative stimulation of neural elements in the brain.
| Characteristics | Device<br>K162459 | Predicate Device<br>K033173; K991522 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | AccuPoint™ Electrode | microTargeting® Electrode |
| Indications For Use | Neuron AccuPoint™ Electrodes are<br>intended for use in intraoperative<br>recording of single unit neuronal<br>activity or intraoperative stimulation of<br>neural elements in the brain. | The FHC microTargeting Electrodes<br>are intended for use in intraoperative<br>recording of single unit neuronal<br>activity or intra-operative stimulation<br>of neural elements in the brain. |
| Anatomical Locations of<br>Use | Use in operating room for<br>neurosurgical procedures where<br>recording of neuronal activity and<br>stimulation of brain neurons will aid in<br>the placement of depth electrodes | Use in operating room for<br>neurosurgical procedures where<br>recording of neuronal activity and<br>stimulation of brain neurons will aid in<br>the placement of depth electrodes |
| Device Materials (Body Contact) | | |
| All Electrodes<br>(Body Contact) | Stainless Steel Tubing<br>Polyimide Shielding<br>Parylene C Insulation<br>Loctite Medical Grade Adhesive | Stainless Steel Tubing<br>Polyimide Shielding<br>Epoxylite Insulation<br>Adhesive |
| • Tungsten Electrodes | Core Electrode - Tungsten | Core Electrode - Tungsten |
| • Platinum Iridium<br>Electrodes | Core Electrode - Platinum Iridium<br>[80% and 20%] Alloy | Core Electrode - Platinum Iridium<br>[80% and 20%] Alloy |
| Use with Other Medical<br>Devices | • Stereotaxic frame<br>• Microdrive system<br>• Connecting cables<br>• Direct connection to microelectrode<br>recording (MER) and<br>microstimulation equipment | • Stereotaxic frame<br>• Microdrive system<br>• Connecting cables<br>• Direct connection to microelectrode<br>recording (MER) and<br>microstimulation equipment |
| Use Condition Provided | Sterile, Single Patient Use<br>[Not sold Non-Sterile] | Sterile, Single Patient Use<br>• Also sold Non-Sterile |
| Method of Sterilization | Ethylene Oxide, EO<br>Non-pyrogenic [Electrodes and<br>Accessories] | Ethylene Oxide, EO<br>Non-pyrogenic [Electrodes]<br>Steam [Accessories] |
#### Substantial Equivalence Matrix:
{5}------------------------------------------------
| Characteristics | Device<br>(Proposed) | Predicate Device<br>K033173; K991522 |
|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Physical and Dimensional Attributes [Comparison of AccuPoint Models to Equivalent FHC Models] | | |
| • Overall Lengths | 235.24 to 291 mm<br>(9.26 to 11.46 in) | 235.24 to 291 mm<br>(9.26 to 11.46 in) |
| • Diameters<br>Core Electrode | 0.200 to 0.250 mm<br>(0.008 to 0.010 in) | 0.200 to 0.250 mm<br>(0.008 to 0.010 in) |
| • Electrode Tip Style | Straight Sharpened Point | Straight Sharpened Point |
| • Electrode Tip Taper | 10-15 degree tip angle | 10-15 degree tip angle |
| Electrode Tip<br>Impedance | 0.5 MΩ-Megohms (+/-30%) to<br>1 MΩ-Megohms (+/-20%) Measured<br>at 1 kHz or 220 Hz | 0.5 MΩ-Megohms (+/-30%) to<br>1 MΩ-Megohms (+/-20%) Measured<br>at 1 kHz or 220 Hz |
| Accessories | Within this Proposed 510(k)<br>• Insertion and Spacer Tubes<br>• Stylets<br>Tubes and Stylets made from 304<br>Stainless Steel<br>Length: Tip to Stop: 179 - 224 mm<br>(7.05 to 8.82 in)<br>Diameter: Varies by Model | 510(k) # K121950 and # K092562<br>• Insertion and Spacer Tubes<br>• Stylets<br>Tubes and Stylets made from 304<br>Stainless Steel<br>Length: Tip to Stop: 179 - 224 mm<br>(7.05 to 8.82 in)<br>Diameter: Varies by Model |
| Expiration Date | One (1) Year¹ | Three (3) Years |
Substantial Equivalence Matrix (Continued):
# Summary of Testing:
The biocompatibility risk assessment was completed as directed by FDA guidance under ISO 10993-1 biocompatibility requirements. Cytotoxicity testing was completed to demonstrate that the known biocompatible materials, the manufacturing processes, and the validated sterilization processes maintained compliance. Performance and safety testing completed for the AccuPoint Electrode included tests for electrode impedance, continuity and insulation integrity, and dynamic impedance to detect short circuiting between the inner and outer electrodes.
### Substantial Equivalence Conclusions:
The completion of the testing demonstrates that the AccuPoint Electrode has the same principles of operation, indications for use, and technological characteristics as the predicate devices.
<sup>1</sup> The Aging Study will be continuing to obtain a longer expiration date.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.