K162424 · Biomet, Inc. · HRS · Oct 12, 2016 · Orthopedic
Device Facts
Record ID
K162424
Device Name
Ulna Plating System
Applicant
Biomet, Inc.
Product Code
HRS · Orthopedic
Decision Date
Oct 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The System is indicated for fixation of fractures, osteotomies and non-unions of the ulna, particularly in osteopenic bone.
Device Story
The Ulna Plating System consists of low-profile, anatomically contoured metallic plates and associated screws used for orthopedic fracture fixation. The system is intended for use by surgeons in clinical settings to stabilize ulna fractures and non-unions. The device functions by allowing surgeons to perform compression plating techniques to stabilize fractures, followed by the use of locking screws to provide neutralization and stability, particularly in osteopenic bone. The system includes both locking and non-locking holes to accommodate various screw types. The implants are provided non-sterile and must be steam-sterilized by the healthcare facility prior to surgical implantation. By providing rigid fixation, the system facilitates bone healing and restores structural integrity to the ulna.
Clinical Evidence
No clinical tests were provided. Substantial equivalence is based on non-clinical performance testing, including construct testing, 4-point bend testing per ASTM F382, and galvanic corrosion engineering evaluations.
Technological Characteristics
Plates: Type II anodized Titanium Alloy (ASTM F136). Screws: Titanium Alloy (ASTM F136) or Cobalt Chromium Alloy (ASTM F1537). Instrumentation: Stainless steel (ASTM F899). Design: Low-profile, anatomically contoured plates with locking and non-locking holes. Sterilization: User-performed steam sterilization. Connectivity: None.
Indications for Use
Indicated for fixation of fractures, osteotomies, and non-unions of the ulna, particularly in patients with osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Locking Anatomic and Composite Plate System (K083843)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 12, 2016
Biomet Inc. Ms. Julie Largent Regulatory Affairs Specialist 56 East Bell Drive Warsaw, Indiana 46582
Re: K162424
Trade/Device Name: Ulna Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: August 29, 2016 Received: August 30, 2016
Dear Ms. Largent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Mark N. Melkerson -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The letters are uniformly sized and spaced, contributing to the legibility of the name.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) TBD
K162424
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Device Name Ulna Plating System
Indications for Use (Describe)
The System is indicated for fixation of fractures and non-unions of the ulna, particularly in osteopenic bone.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a white "Z" inside of it, followed by the text "ZIMMER BIOMET" in gray. The logo is simple and modern, and it is likely used to represent the company's brand.
## Ulna Plating System | Traditional 510(k)
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Ulna Plating System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet Inc. |
|-----------------|------------------------------------------------------------------------------------------------------------------------------|
| | 56 East Bell Drive |
| | PO Box 587 |
| | Warsaw, IN 46581 |
| | Establishment Registration Number: 1825034 |
| Contact: | Julie Largent |
| | Regulatory Affairs Specialist |
| | 305-269-6391 |
| Date: | October 5, 2016 |
| Subject Device: | Trade Name: Ulna Plating System |
| | Common Name: Plate, Fixation, Bone |
| | Classification Name:<br>HRS- Single/multiple component metallic bone fixation appliance<br>and accessories (21 CFR 888.3030) |
## Legally marketed devices to which substantial equivalence is claimed:
- Locking Anatomic and Composite Plate System K0838432
- Distal Volar Radius Plating System K112345
#### Device Description
The Ulna Plating System is comprised of two new low profile plates anatomically contoured for the ulna, existing screws and both system specific and general instrumentation. The plates, locking screws and non-locking screws are manufactured from titanium alloy Ti6Al4V ELI per ASTM F136. All the plates are Type II anodized. The plates incorporate both locking and nonlocking holes that allow the surgeon to stabilize the fracture by the use of compression plating techniques and then obtain neutralization of the construct with the locking screws. The locking screws construct can protect the compression screws while providing relative stability across extensive comminution of in osteopenic bone. The plate material and the inclusion of both locking and non-locking holes are design features also included in the predicate device cleared in K083843. The plates are designed to accept existing 2.7mm locking screws, 2.7mm non-
<sup>1</sup> K083843 includes two families of plates. The subject device is claiming substantial equivalence to the Locking Composite Plate due to the similar size, shape and function.
<sup>2</sup> See Section 14 for a listing of existing screws used in the Ulna Plating System.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is in a dark gray color.
locking screws, and 2.7mm multidirectional screws. The non-locking and locking screws are manufactured from Titanium Alloy per ASTM F-136 and are color anodized. The multidirectional screws are manufactured from Cobalt Chromium Alloy per ASTM 1537. The predicate also includes Cobalt Chromium (ASTM F1537) multidirectional screws. The Ulna Plating System will offer the surgeon several system specific instruments as well as various general instruments to facilitate the installation of the implants.
## Intended Use and Indications for Use
The System is indicated for fixation of fractures, osteotomies and non-unions of the ulna, particularly in osteopenic bone.
## Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: The Intended Use is similar to the Intended Use cleared in K083843.
- Indications for Use: The Indications for Use is similar to the Indications for Use cleared in K083843.
- Materials: The new Ulna Shortening System plates are manufactured from Type II anodized Titanium Alloy (Ti-6Al-4V ELI ASTM F136). The existing 2.7mm locking and non-locking screws included in the system are also manufactured from Titanium Alloy (Ti-6Al-4V ELI ASTM F136) and are either Type II anodized or color anodized. The system also contains existing 2.7mm multi-directional screws that are manufactured from Cobalt Chromium Alloy (ASTM F1537). Titanium alloys as well as implant grade cobalt chromium are commonly used materials in orthopeadic implants and are both used in the predicate device cleared via K083843. The patient contact instrumentation is comprised of stainless steel that meets ASTM F-899.
- Design Features: The design features for the Ulna Shortening System are similar to ● those in currently marketed devices cleared K083843. The design differences have not identified any issues that would impact the safety and effectiveness of the devices.
- . Sterilization: The implants and instruments are offered to the user in the non-sterile configuration. The non-sterile implants and instruments will be required to be steam sterilized by the user prior to use. The non-sterile packaging configuration is the same as the predicate devices currently marketed and cleared via K083843.
## Summary of Performance Data (Nonclinical and/or Clinical)
- Non-Clinical Tests
- o Non-clinical performance testing included construct testing as well as 4-point bend testing per ASTM F382 to determine substantial equivalence of the Ulna Plating System. A Galvanic Corrosion engineering evaluation was also completed and was found to meet the acceptance criteria. Results indicate that the subject plating system is substantially equivalent to legally marketed devices.
- Clinical Tests
- No clinical tests are provided for basis of substantial equivalence. o
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Zimmer Biomet logo. The logo consists of a blue circle with a white "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is in a light gray color.
## Ulna Plating System | Traditional 510(k)
## Substantial Equivalence Conclusion
The Ulna Plating System has shown to be substantially equivalent to the predicate devices. Results of non-clinical tests/engineering evaluation and the similarities with legally marketed predicated devices indicate the device will perform within the intended uses and no new issues of safety and effectiveness have been raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.