K162394 · Capintec, Inc. · KPT · Mar 8, 2017 · Radiology
Device Facts
Record ID
K162394
Device Name
CRC-77tHR
Applicant
Capintec, Inc.
Product Code
KPT · Radiology
Decision Date
Mar 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1360
Device Class
Class 2
Indications for Use
The CRC-77tHR is intended to be used by qualified radiopharmacists, radiochemists, radiation physicists, or other appropriately trained personnel in isotope production sites to measure a wide range of measure a wide range of high activity radiopharmaceuticals and radioactive materials, including high energy beta and gamma emitters. The device is not intended for use with brachytherapy sources. This device is also used in numerous research applications for measurement of high activity levels of radioactive materials.
Device Story
CRC-77tHR is a radionuclide dose calibrator used in isotope production and research facilities. It utilizes a sealed reentrant well ionization chamber as a detector to measure high-activity radiopharmaceuticals and radioactive materials. The system includes a readout unit with a touch screen interface. The device operates by measuring ionization current, which is converted to activity levels. The electrometer board is located in the readout unit to protect electronics from radiation. Compared to the predicate, the CRC-77tHR features lower gas fill pressure in the chamber, reducing sensitivity to allow for the measurement of higher activity levels (e.g., Tc-99m up to 400 Curies, F-18 up to 130 Curies). It is operated by trained personnel (radiochemists/physicists) in a laboratory setting. The output provides activity measurements, which are used by staff to verify radiopharmaceutical doses or material activity. The device does not have direct patient contact.
Clinical Evidence
Bench testing only. No clinical or animal studies performed. Performance data includes electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and usability (IEC 60601-1-6). Technical specifications validated include measurement range (up to 400 Ci for Tc-99m), energy range (15 keV to 3 MeV), electrometer accuracy (+/- 2%), linearity (< 2%), and reproducibility (+/- 1%). Software verification and validation conducted per FDA guidance.
Technological Characteristics
Radionuclide dose calibrator; sealed reentrant well ionization chamber detector; touch screen readout. Electrometer board located in readout unit. Energy range: 15 keV to 3 MeV. Connectivity: cable-connected chamber and optional printer. Software life cycle: IEC 62304. Safety standards: IEC 60601-1 (Ed. 3), IEC 60601-1-2 (Ed. 4), AAMI ES60601-1, CSA C22.2#60601-1.
Indications for Use
Indicated for qualified radiopharmacists, radiochemists, radiation physicists, or trained personnel in isotope production/research sites to measure high activity radiopharmaceuticals and radioactive materials (beta/gamma emitters). Not for brachytherapy sources.
Regulatory Classification
Identification
A radionuclide dose calibrator is a radiation detection device intended to assay radionuclides before their administration to patients.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2017
Capintec, Inc. % Ms. Mary Yusko VP, Product Development and Regulatory Affairs 7 Vreeland Road FLORHAM PARK NJ 07932
Re: K162394
Trade/Device Name: CRC-77tHR Regulation Number: 21 CFR 892.1360 Regulation Name: Radionuclide Dose Calibrator Regulatory Class: II Product Code: KPT Dated: December 19, 2016 Received: December 21, 2016
Dear Ms. Yusko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162394
Device Name
CRC-77tHR
Indications for Use (Describe)
The CRC-77tHR is intended to be used by qualified radiochemists, radiation physicists, or other appropriately trained personnel in isotope production sites to measure a wide range of measure a wide range of high activity radiopharmaceuticals and radioactive materials, including high energy beta and gamma emitters. The device is not intended for use with brachytherapy sources. This device is also used in numerous research applications for measurement of high activity levels of radioactive materials.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CAPINTEC, INC. The logo consists of the letters "CII" stacked on top of each other inside of a square. Below the square is the text "CAPINTEC, INC."
RE: Summary of Safety and Effectiveness Information for the Capintec CRC-77tHR
### I. SUBMITTER
Capintec, Inc. 7 Vreeland Road Florham Park, NJ 07932 Phone 201-825-9500 201-825-1336 FAX
Contact Person: Mary Anne Yusko Date Prepared: February 12, 2017
- II. DEVICE Name of Device: CRC-77tHR
Common or Usual Name: Radionuclide Dose Calibrator Regulatory Class II, Product Code: KPT (90) 21 CFR 892.1360
- III. PREDICATE DEVICE CRC -55t Dose Calibrator series, which received FDA clearance under 510(k) number K101452. This predicate device has not been subject to a design related recall. The predicate devices, Capintec dose calibrator lines, upon which the CRC-77tHR is based, have a long history of over 30 years of safe, reliable, and effective use in the field.
### IV. DEVICE DESCRIPTION
The Capintec CRC-77tHR is dose calibrator which uses a touch screen user interface and sealed reentrant well chamber as the detector. The gas fill pressure has been lowered in order to extend the maximum activity range for isotope manufacturing and radiopharmaceutical production environments. A diagram of the system is shown below.
{4}------------------------------------------------
Image /page/4/Figure/0 description: The image shows a diagram of a CRC-77t Readout system. The system includes a CRC-77t Readout device, a dipper, a liner, an ionization chamber, a line cord, and an optional printer. The CRC-77t Readout device is connected to the ionization chamber and the optional printer via cables. The dipper and liner are separate components of the system.
#### V. INDICATIONS FOR USE
The CRC-77tHR is intended to be used by qualified radiopharmacists, radiochemists, radiation physicists, or other appropriately trained personnel in isotope production sites to measure a wide range of measure a wide range of high activity radiopharmaceuticals and radioactive materials, including high energy beta and gamma emitters. The device is not intended for use with brachytherapy sources. This device is also used in numerous research applications for measurement of high activity levels of radioactive materials.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE
DEVICE The CRC -77tHR series is identical to the predicate device, the CRC-55t series, in terms of scientific technology, principles of operations, radiation detectors, measurement processes, calibration, algorithms, calculations, user interface, and basic circuitry design. There are two basic differences between models. The first difference is that CRC-77tHR has a lower fill gas pressure, which reduces the sensitivity of the chamber in order to extend maximum activity which the device can measure. This change allows the 77tHR to be used in isotope and radiopharmaceutical production facilities which need to measure higher activity levels. The second difference is that the electrometer board, which is located in the base of the ionization chamber for the 55t models, has been relocated to the readout of the 77tHR. This was done to protect the electronics from radiation damage by removing them from the radiation field. Otherwise, the technical characteristics of the models are the same.
The predicate device includes brachytherapy measurements as an intended use. The CRC-77tHR, with lower fill pressure and reduced sensitivity, has not been validated for LDR measurements at this time. Therefore, brachytherapy source measurements were not included in the listed intended use. The removal of this intended use does not affect safety or effectiveness of the CRC-77tHR for its listed applications.
{5}------------------------------------------------
- VII. PERFORMANCE DATA The CRC-77tHR The following performance data were provided in support of the substantial equivalence determination
- Electrical safety, electromagnetic compatibility (EMC) and Usability Compliance IEC 60601-1-2 (2014):Ed. 4. Medical Electrical Equipment – Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
IEC 60601-1: Issued: 2005/12/15 Ed. 3 Medical Electrical Equipment – Part 1: General Requirements for Basic Safety & Essential Performance; Corr. 1: 2006, Corr. 2: 2007. Amd. 1: 2012
IEC 60601-1-6: Issue:2010/01/27 Ed:3 Medical electrical equipment – Part 1-6: General requirements for safety - Collateral Standard: Usability Review of usability per 60601-1 3rd Edition
AAMI ES60601-1: Issued: 2012/08/20 Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance, Amendment 1. Product safety evaluation to AAMI 60601-1 3rd Edition, Amendment 1
CSA C22.2#60601-1: Issued: 2014/03/01 Ed: 3 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
IEC 62304: Issue:2006/05/09 Ed:1 Medical Device Software - Software Life Cycle Processes
Technical Specifications:
Measurement Range: Maximum activity for Tc-99m is approximately 400 Curies Maximum activity for F-18 is approximately 130 Curies Resolution: .01 mCi (10 µCi) Minimum activity range > 100 µCi This model should not be used for activity measurements less than 100 µCi Energy Range: 15 keV to 3 MeV Electrometer Accuracy: + 2% NOTE: Overall accuracy is determined by the specific nuclide and sample configuration, geometry difference between the reference source and sample, and accuracies of the reference sources used for calibration. Electrometer Linearity: < 2% over a dynamic range of 10115 A to 106 A Reproducibility: + 1% in 24 hours with unit powered at all times.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The level of concern is moderate because a failure or latent design flaw could directly result in minor injury to the patient or operator. The level of concern is also moderate because a failure or latent flaw could indirectly result in minor injury to the patient through incorrect or delayed information or through the action of a care provider.
{6}------------------------------------------------
The CRC-77tHR was also evaluated at an isotope production site. Animal, clinical and biocompatibility studies are not applicable and were not performed. The CRC-77tHR is used in a laboratory bench top setting and has no direct patient contact.
## SUMMARY
Summary Based on documented performance, the CRC-77tHR was found to have a safety and effectiveness profile that is similar to the predicate device.
- CONCLUSIONS The data demonstrate that the CRC-77tHR device performs comparably to VIII. the predicate device at a higher activity range.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.