Historical clinical case library (QT Library); Patient medical records/imaging archives (Mammography and Handheld Ultrasound)
Retrospective clinical cases from a library were used to conduct a Visual Grading Assessment (VGA) to compare the image quality of the QT Ultrasound Breast Scanner-1 against X-ray mammography and handheld ultrasound.
VGA Review with QT Ultrasound vs. Xray Mammography (XRM); Retrospective paired-reader, paired-subject Visual Grading Assessment (VGA)
20 subjects (22 cases); Sample Size: 22 cases
X-ray mammography (XRM)
Image quality scores of ten anatomical breast structures
VGA Review with QT Ultrasound vs. HandHeld Ultrasound (HHUS); Retrospective paired-reader, paired-subject Visual Grading Assessment (VGA)
17 cases; Sample Size: 17 cases
Handheld ultrasound (HHUS)
Image quality scores of ten anatomical breast structures
Indications for Use
The QT Ultrasound Breast Scanner-1 is for use as an ultrasonic imaging system to provide reflection-mode and transmission-mode images of a patient's breast. The device is not intended to be used as a replacement for screening mammography.
Device Story
Automated, software-controlled ultrasonic imaging system for whole-breast scanning. Patient lies prone with breast suspended in warm water bath; transducer arrays rotate vertically around breast to acquire data. System uses reflection-mode and transmission-mode ultrasound to generate tomographic, speed-of-sound, and B-scan images. Components include patient scanning system (tank, transducers, water management), operator console, and viewer console. Radiologists review coronal, axial, and sagittal images on viewer console; probe and region-of-interest (ROI) tools allow querying of speed-of-sound values (m/s) to aid diagnostic evaluation. Device provides adjunct imaging to assist in characterizing breast tissue and lesions across various breast densities.
Clinical Evidence
Visual Grading Assessment (VGA) studies compared QT Ultrasound to X-ray mammography (22 cases, 20 subjects) and handheld ultrasound (17 cases). Four to five board-certified radiologists scored image quality of ten anatomical structures. Results indicated QT Ultrasound images were equivalent or superior to comparators in >80-90% of cases for most features. Representative clinical cases (n=16) demonstrated capability across various breast densities (fatty to extremely dense) and lesion types (cysts, solid benign, solid malignant).
Technological Characteristics
Automated ultrasonic imaging system; broadband transducer arrays; reflection and transmission (speed-of-sound) modes. Patient interface: prone position, water bath coupling. Connectivity: DICOM-compatible viewer console. Software-controlled image reconstruction (3D). Electrical safety per AAMI ES60601-1; EMC per IEC 60601-1-2; acoustic output per IEC 60601-2-37 (Track 1).
Indications for Use
Indicated for ultrasonic breast exams in patients requiring diagnostic imaging of breast tissue. Not for use as a replacement for screening mammography.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 6, 2017
QT Ultrasound, LLC % John C. Klock, M.D. Chief Executive Officer and Managing Director 3 Hamilton Landing, Suite 160 NOVATO CA 94949
Re: K162372
Trade/Device Name: QT Ultrasound Breast Scanner-1 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: May 31, 2017 Received: June 1, 2017
Dear Dr. Klock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162372
Device Name
QT Ultrasound Breast Scanner-1
Indications for Use (Describe)
The QT Ultrasound Breast Scanner-1 is for use as an ultrasonic imaging system to provide reflection-mode and transmission-mode images of a patient's breast. The device is not intended to be used as a replacement for screening mammography.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Diagnostic Ultrasound Indications for Use
System: QT Ultrasound Breast Scanner-1
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | Other |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|----------|-------|
| General<br>(Track 1 only) | Specific (Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal Imaging<br>& Other | Small Organ (Breast) | N1 | | | | | | N2 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | |
| Vessel | Other (Specify) | | | | | | | |
N=New Indication; P = previously cleared by FDA; E = added under this appendix
1 – Reflection
2 – Transmission (Speed of Sound)
Additional Comments: QT Ultrasound Breast Scanner-1 is intended for ultrasonic breast exams
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for QT ultrasound. The logo consists of the letters "QT" in a light green color, with the word "ultrasound" in a light gray color to the right of the letters. The logo is simple and modern, and the colors are soft and calming.
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with 21 CFR 807.92 the following summary of information is provided:
### 1.0 SUBMITTER INFORMATION
| Submitted By | QT Ultrasound LLC®<br>3 Hamilton Landing, Suite 160<br>Novato, CA 94949 |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Contact Information | John Klock, MD<br>Phone: (415) 842-7242<br>Fax: (415) 234-6511<br>Email: john.klock@qtultrasound.com |
| Date of Submission | August 23, 2016 |
| 510(k) Number | K162372 |
| DEVICE INFORMATION | |
| Trade / Proprietary Name | QT Ultrasound Breast Scanner-1 |
| Common Name | System, Imaging, Pulsed Echo Ultrasonic<br>Transducer, Ultrasonic, Diagnostic |
| Classification Name and<br>Regulation Number | 21CFR §892.1560 Ultrasonic pulsed echo imaging system<br>21CFR §892.1570 Diagnostic ultrasonic transducer |
| Product Codes | 90-IYO, 90-ITX |
#### 3.0 PREDICATE DEVICE
2.0
The predicate device is identified as the SoftVue System manufactured by Delphinus Medical Technologies. SoftVue received market clearance under 510(k) number K142517.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "QT ultrasound". The logo is in a light green and gray color scheme. The letters "QT" are in green, while the word "ultrasound" is in gray.
#### 4.0 DEVICE DESCRIPTION
The QT Ultrasound Breast Scanner-1 is an automated software-controlled ultrasonic imaging system that performs a standardized scan of the whole breast. The OT Ultrasound Breast Scanner-1 is comprised of a Patient Scanning System, Operator Console and Viewer Console.
The Patient Scanning System consists of a patient support table, scan tank, water management system, ultrasound transducer arrays and all associated image processing electronics. The scan tank is centered below a patient's breast and contains the ultrasound transducer arrays. The transducer arrays include a set of three reflection transducers that transmit pulsed ultrasound plane waves into targeted tissues using the water bath in the scan tank as coupling medium. An additional transmitter and receiver array pair collect the ultrasound energy to provide speed of sound values.
During scanning, a patient lies prone on the examination table with the breast suspended in a warm water bath maintained near skin temperature. Images are automatically acquired on a pendant breast positioned with the nipple as a point of reference. The transducer arrays rotate about a vertical axis to circle the breast in the coronal plane. The array is then translated vertically and the scanning process is repeated until the entire breast is scanned, allowing B-scan images to be constructively combined into tomographic, speed of sound and reflection ultrasound images.
The QT Ultrasound Breast Scanner-1 outputs the images to the QTviewer which allows the images to be stored until they are reviewed on a Viewer Console. Coronal, axial and sagittal images are generated for review by the radiologist. Speed of sound images may be queried by the Probe and Region of Interest (ROI) tools provided in the Viewer Console. These tools provide speed of sound values in meters/sec. to aid in diagnostic evaluation of the breast.
#### 5.0 INTENDED USE
The OT Ultrasound Breast Scanner – 1 is for use as an ultrasonic imaging system to provide reflection-mode and transmission-mode images of a patient's breast. The device is not intended to be used as a replacement for screening mammography.
#### PREDICATE DEVICE COMPARISON 6.0
The OT Ultrasound Breast Scanner-1 is substantially equivalent to the Delphinus Medical Technologies SoftVue System cleared by the FDA in K142517. QT Ultrasound claims
{6}------------------------------------------------
substantial equivalence because the proposed device has an equivalent intended use, operating principles, and physical and operational specifications as compared to the predicate device. The QT Ultrasound Breast Scanner - 1 and the predicate device utilize B-mode grayscale ultrasound images to achieve their intended use. Both the OT Ultrasound Breast Scanner-1 and the predicate SoftVue device are table-top systems that have automatic scanning transducers to image breast tissue.
The specific details regarding similarities and differences between the QT Ultrasound Breast Scanner-1 and the SoftVue device have been identified and explained in Comparison Tables provided in Section 5.0 of this submission. A brief summary of the similarities and differences between the OT Ultrasound Breast Scanner-1 and the SoftVue device is included below.
### Similarities
- Both systems use an automated transducer to acquire images of a patient's breast. ●
- Both systems use broadband transducers.
- o Both systems acquire and process B-mode grayscale images of a patient's breast.
- o Both systems acquire and process grayscale speed of sound images of a patient's breast.
- Both systems position the patient in a prone position lying on their examination table with the patient's breast in a pendulous position within an imaging chamber.
- Both systems position the patient's breast in a fluid environment to eliminate the need for breast compression and facilitate the transmission of ultrasound waves.
## Differences
The differences between the QT Ultrasound Breast Scanner-1 and SoftVue System are listed in Table 2-1.
| Table 2-1. Differences Between the QT Ultrasound Breast Scanner-1 and SoftVue System | |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristic | Discussion |
| Display of Speed of<br>Sound Information | The QT Ultrasound Breast Scanner-1 provides grayscale speed of<br>sound images that may be queried by the Probe and/or ROI tools<br>available in the QTviewer. Both systems use this information as an<br>aid / reference information for diagnostic evaluation of the breast. |
| 3-D volume Image<br>Acquisition | Although, QT image acquisition is performed in 3D and SoftVue is<br>acquired in 2D, this technological characteristic does not raise any |
{7}------------------------------------------------
| Table 2-1. Differences Between the QT Ultrasound Breast Scanner-1 and SoftVue System | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristic | Discussion |
| | different questions of safety and effectiveness compared to the<br>predicate device. |
| Image Reconstruction | Although, QT image reconstruction is performed in 3D and<br>SoftVue is reconstructed in 2D, this technological characteristic<br>does not raise any different questions of safety and effectiveness<br>compared to the predicate device. |
The differences noted between the QT Ultrasound Breast Scanner-1 and the predicate device do not present any new or different questions related to safety and effectiveness.
### 7.0 SUMMARY OF NON-CLINICAL TESTING
The function and performance of the QT Ultrasound Breast Scanner-1 has been evaluated through non-clinical design verification and validation testing. Testing included system performance and simulated use tests. When applicable, non-clinical testing was conducted per the standards listed in Table 2-2.
| Table 2-2. Testing Performed | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of Testing | Tests Performed |
| Electrical Safety<br>AAMI ES60601-1:2005/(R)2012 And<br>A1:2012 Medical electrical equipment -<br>Part 1: General requirements for basic<br>safety and essential performance | All applicable electrical, basic safety and essential<br>performance tests.<br>Testing was conducted by Intertek, an independent<br>testing laboratory, located in Menlo Park, CA. |
| Electromagnetic Compatibility<br>IEC 60601-1-2 Edition 3: 2007-03<br>Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance - Collateral<br>Standard: Electromagnetic disturbances -<br>Requirements and tests | All applicable testing pertaining to<br>electromagnetic compatibility.<br>Testing was conducted by Intertek, an independent<br>testing laboratory, located in Menlo Park, CA. |
| Table 2-2. Testing Performed | |
| Type of Testing | Tests Performed |
| Usability<br>IEC 62366 Edition 1.1 2014-01 -<br>Medical devices- Application of usability<br>engineering to medical devices<br>IEC 60601-1-6 Edition 3.1 2013-10 -<br>Medical electrical equipment Part 1-6<br>General requirements for safety -<br>Collateral Standard: Usability | All applicable testing pertaining to usability.<br>Testing was conducted by Intertek, an independent<br>testing laboratory, located in Menlo Park, CA. |
| Acoustic Output<br>IEC 60601-2-37 Edition 2.0 2007<br>Medical electrical equipment - Part 2-37:<br>Particular requirements for the basic<br>safety and essential performance of<br>ultrasonic medical diagnostic and<br>monitoring equipment.<br>NEMA UD 2-2004 (R2009) Acoustic<br>Output Measurement Standard for<br>Diagnostic Ultrasound Equipment –<br>Revision 3 | All applicable testing pertaining to the<br>requirements for the safety of ultrasonic medical<br>diagnostic and monitoring equipment and to<br>demonstrate compliance with the "Acoustic<br>Output Measurement Standard for Diagnostic<br>Ultrasound Equipment".<br>Testing was conducted by QT Ultrasound and<br>witnessed by Intertek, an independent testing<br>laboratory, located in Menlo Park, CA. The QT<br>Ultrasound Breast Scanner-1 meets all Track 1<br>acoustic output requirements. The results of<br>acoustic output testing are listed in Table 2-3<br>below. |
| Software Development<br>IEC 62304:2006 (First Edition) - Medical<br>device software - Software life cycle<br>processes | Internal procedures for software life cycle<br>management were used for software development,<br>verification / validation and configuration control |
| Software Verification and Validation | Software was tested at the module and system<br>levels to ensure that it met the software's design<br>and intended use requirements. All requirements<br>were met and no new issues of safety or<br>effectiveness compared to the predicate device<br>were raised. |
| System Verification and Performance | System verification testing was conducted to<br>ensure that the QT Ultrasound Breast Scanner-1<br>met design requirements. In addition, the<br>following system performance characteristics are<br>reported:<br>• Measurement Range / Accuracy<br>• Spatial Resolution<br>• Contrast Resolution / Contrast to Noise Ratio<br>• Speed of Sound Uniformity and Accuracy |
| Table 2-2. Testing Performed | |
| Type of Testing | Tests Performed |
| | All requirements were met and no new issues of<br>safety or effectiveness compared to the predicate<br>device were raised. |
| Biocompatibility<br>ISO 10993-1:2009/(R)2013 - Biological<br>evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk<br>management process | Biocompatibility testing was conducted to ensure<br>that the patient contacting materials in the QT<br>Ultrasound Breast Scanner-1 met design<br>requirements. All requirements were met and no<br>new issues of safety or effectiveness compared to<br>the predicate device were raised. |
| ISO 10993-5:2009/(R)2014 - Biological<br>evaluation of medical devices - Part 5:<br>Tests for in vitro cytotoxicity | |
| ISO 10993-10:2010/(R)2014 - Biological<br>evaluation of medical devices - Part 10:<br>Tests for irritation and skin sensitization | |
| ISO 10993-11:2006/(R)2010 - Biological<br>evaluation of medical devices - Part 11:<br>Tests for systemic toxicity | |
| Cleaning Procedures<br>Scan Tank | Microbial analysis and particulate testing of the<br>water in the scan tank was conducted. All<br>requirements were met and no new issues of safety<br>or effectiveness compared to the predicate device<br>were raised. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in a light gray color. The letters "QT" are larger than the word "ultrasound".
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for QT ultrasound. The logo is in a light green color. The letters "QT" are larger than the word "ultrasound".
Acoustic output testing per IEC 60601-2-37 Edition 2.0 2007 was conducted by QT Ultrasound LLC and witnessed by Intertek, an independent testing laboratory, located in Menlo Park, California. The QT Ultrasound Breast Scanner-1 meets all Track 1 acoustic output requirements. The results of acoustic output testing are listed in Table 2-3. A copy of the full test report is provided in Section 6.0 of this submission.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for QT ultrasound. The logo consists of the letters "QT" in a light green color, with the word "ultrasound" in gray next to it. The font is sans-serif and the overall design is simple and modern.
| Table 2-3. Global Maximum Acoustic Output Values | | | | | | |
|--------------------------------------------------|------------------------------|-----------------------------------------|------|-------|-------|-------------|
| Feature | Track 1<br>Exposure<br>Level | QT Ultrasound Breast<br>Scanner-1 Level | | | | Pass / Fail |
| | | Trans | 0 | 1 | 2 | |
| Max. Mechanical Index (MI) | 1.9 | 0.145 | .98 | 1.07 | 1.186 | Pass |
| Max. ISPTA (mW/cm²) | 94 mW/cm2 | 0.79 | .713 | 0.995 | 1.63 | Pass |
The results of safety, performance and verification / validation testing demonstrate that the QT Ultrasound Breast Scanner-1 successfully meets the requirements of its intended use.
### SUMMARY OF CLINICAL TESTING 8.0
### VGA Review with QT Ultrasound vs. Xray Mammography (XRM)
QT Ultrasound LLC performed a Visual Grading Assessment (VGA) review of 22 cases from 20 subjects using four independent board-certified radiology readers. This was a paired-reader, paired-subject evaluation comparing the image quality of QT Ultrasound to X-ray mammography (XRM). The four readers independently scored the image quality of ten anatomical breast structures with XRM and OT Ultrasound during separate reading sessions, along with an overall image quality rating. The readers were provided the mammograms (CC and MLO views) and corresponding QT Ultrasound DICOM studies (Speed of Sound and Reflection images displayed in the coronal, axial, and sagittal planes) for each case.
### VGA Review with OT Ultrasound vs. HandHeld Ultrasound (HHUS)
A similar VGA review of 17 cases was performed using 5 independent board-certified radiology readers. This was a paired-reader, paired-subject evaluation comparing the image quality of QT Ultrasound to HHUS. The five readers independently scored the image quality of ten anatomical breast structures with HHUS and QT Ultrasound during separate reading sessions, along with an overall image quality rating. The readers were provided the available targeted HHUS images (static DICOM images) and corresponding QT Ultrasound DICOM studies (Speed of Sound and Reflection images displayed in the coronal, axial, and sagittal planes) for each case.
The statistical analyses for both studies were performed under the direction of Nancy Obuchowski, PhD, Vice Chair of Quantitative Health Sciences at the Cleveland Clinic Foundation.
For the VGA study comparing QT Ultrasound and XRM, the results are as follows.
{11}------------------------------------------------
The median image quality score for each of the 10 anatomical features, as well as the overall image quality, was reported for each reader. The proportion of breasts where the image quality was rated better on QT Ultrasound than XRM or equivalent to XRM was reported for each feature. A 95% Confidence Interval (CI) for the proportion of breasts rated as equivalent or better image quality on QT Ultrasound was constructed for each feature using methods for clustered binary data, treating subject as the cluster. Similarly, the proportion of breasts where the image quality was rated better on QT Ultrasound than XRM was reported for each feature, along with its 95% CI.
| Table 2-4. Median Image Quality Scores by Modality, Reader, and Feature | | | | | | | | |
|-------------------------------------------------------------------------|---------------|-----|-----|-----|-----|-----|-----|-----|
| (1=excellent, 5=poor)* | | | | | | | | |
| | QT Ultrasound | | | | XRM | | | |
| Anatomical Feature | R1 | R2 | R3 | R4 | R1 | R2 | R3 | R4 |
| Skin (Overall) | 1.5 | 1.0 | 1.0 | 2.0 | 2.0 | 4.0 | 3.0 | 2.0 |
| Epidermis | 1.0 | 1.0 | 1.0 | 1.0 | 2.0 | 4.0 | 2.0 | 1.0 |
| Dermis | 2.0 | 2.0 | 2.0 | 2.0 | 5.0 | 5.0 | 5.0 | 5.0 |
| Hypodermis | 2.0 | 1.0 | 1.0 | 1.0 | 3.0 | 4.0 | 2.0 | 1.0 |
| Cooper's Ligament | 1.5 | 1.0 | 1.0 | 1.0 | 4.0 | 4.0 | 4.0 | 4.0 |
| Superficial Veins | 2.0 | 1.0 | 1.0 | 2.0 | 3.5 | 4.0 | 4.0 | 4.0 |
| Central Ducts Entering<br>Nipple | 2.0 | 2.0 | 2.0 | 3.0 | 3.5 | 5.0 | 4.5 | 4.0 |
| Intermediate or<br>peripheral Ducts | 2.0 | 2.0 | 1.0 | 2.0 | 5.0 | 5.0 | 5.0 | 5.0 |
| Terminal Duct Lobular<br>Units | 2.0 | 2.0 | 2.0 | 2.0 | 5.0 | 5.0 | 5.0 | 5.0 |
| Pectoralis Muscle<br>(Chest Wall) | 3.0 | 4.0 | 3.0 | 3.0 | 3.0 | 2.0 | 2.0 | 2.0 |
| Overall visualization<br>of breast anatomy | 2.0 | 2.0 | 1.0 | 2.0 | 4.0 | 4.0 | 4.0 | 4.0 |
* Ordinal Step Scale indicates how well the reader can see the anatomical breast structures. The definitions for the step scale are as follows:
- 1 = Excellent No limitations
- 2 = Good Minimal limitations
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, with the word "ultrasound" in a light gray color next to it. The font is sans-serif and the overall design is simple and modern.
3 = Sufficient - Moderate limitations with no substantial loss of information
- 4 = Restricted Relevant limitations with clear loss of information
Image /page/12/Figure/3 description: This line graph compares the readers' median score of QT Ultrasound and XRM across different features. The x-axis lists the features: Sk, Epi, Der, Hy, Coop, Ve, CenD, XLob, TerD, and Mus. The y-axis represents the readers' median score, ranging from 0 to 5. The XRM line generally has higher scores than the QT Ultrasound line, indicating a higher median score for XRM across most features.
- 5 = Poor Significant loss of information
Figure 1: Readers' median image quality score by modality (QT Ultrasound in blue and XRM in red) and breast feature: skin (overall), epidermis, dermis, Cooper's ligaments, superficial veins, central ducts entering nipple, intermediate or peripheral ducts (extra-lobular ducts), terminal duct lobular units and pectoralis muscle (chest wall).
Image quality was scored on an ordinal rating scale: 1=Excellent, 2=Good, 3=Sufficient, 4=Restricted, 5=Poor.
The analysis of anatomical breast structures comparing QT Ultrasound images to XRM, demonstrated that the OT Ultrasound Breast Scanner-1 produces clinically-useful depictions of patient anatomy. The readers scored the image quality on the OT Ultrasound images as equivalent to or better than on XRM for each feature in more than 90% of breasts.
For the VGA review comparing OT Ultrasound vs. HHUS, the results are as follows:
Table 1 summarizes the median image quality scores from the results pooled over all readers, as well as the lowest median by any reader and highest median by any reader for each of the breast features. Note that readers used a score of 99 (i.e. anatomy not included on image) on both HHUS and QT Scan, but more often on HHUS. Figure 2 illustrates the readers' pooled median score for the two modalities. Except for epidermis where the median scores are equivalent, when the anatomy was visible on the image, the readers' median scores indicated superior image quality on the QT Scan.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in a light gray color. The text is simple and modern.
| | HHUS | | | QTUS | | |
|--------------------------------------------|------------------------------------------------------------------|-------|-------|--------------------------------------------------------------|-------|-------|
| | Median* | Min** | Max** | Median* | Min** | Max** |
| Skin (Overall) | 3.0<br>***(80% not scored<br>because not included on<br>image) | 3.0 | 3.0 | 2.0 | 1.0 | 2.0 |
| Epidermis | 1.0 | 1.0 | 2.0 | 1.0 | 1.0 | 2.0 |
| Dermis | 3.0 | 1.0 | 4.0 | 1.0 | 1.0 | 3.0 |
| Hypodermis | 2.0 | 1.0 | 3.0 | 1.0 | 1.0 | 2.0 |
| Cooper's Ligament | 3.0 | 2.0 | 4.0 | 1.0 | 1.0 | 1.0 |
| Superficial Vessels | 5.0<br>(18.8% not scored because<br>not included on image) | 2.0 | 5.0 | 1.0 | 1.0 | 2.0 |
| Central Ducts<br>Entering Nipple | 5.0<br>***(68.2% not scored<br>because not included on<br>image) | 2.0 | 5.0 | 2.0 | 1.0 | 5.0 |
| Intermediate or<br>peripheral Ducts | 5.0<br>(23.5% not scored because<br>not included on image) | 1.0 | 5.0 | 2.0 | 1.0 | 5.0 |
| Terminal Duct<br>Lobular Units | 5.0<br>(21.2% not scored because<br>not included on image) | 4.0 | 5.0 | 2.0<br>(3.5% not scored<br>because not included on<br>image) | 1.0 | 3.0 |
| | Median* | Min** | Max** | Median* | Min** | Max** |
| Pectoralis Muscle<br>(Chest Wall) | 3.0<br>(7.1% not scored because not<br>included on image) | 3.0 | 5.0 | 2.0<br>(8.2% not scored<br>because not included on<br>image) | 1.0 | 4.0 |
| Overall visualization<br>of breast anatomy | 3.0<br>***(61.2% not scored<br>because not included on<br>image) | 3.0 | 4.0 | 2.0 | 1.0 | 3.0 |
* median over 85 observations pooled from 5 readers
** minimum and maximum median of the 5 readers
***For HHUS, overall skin, central ducts and overall visualization of breast anatomy are features that have a majority of cases not scored due to targeted HHUS.
{14}------------------------------------------------
Image /page/14/Figure/1 description: The image is a line graph comparing the median scores of two different categories, QTUS and HHUS, across various features. The x-axis represents the features, labeled as Sk, Epi, Der, Hy, Coop, Ve, CenD, XLob, TerD, and Mus. The y-axis represents the median score, ranging from 0 to 5. The QTUS line is mostly flat at a score of 1, with increases at Ve and CenD to a score of 2, while the HHUS line fluctuates between 1 and 5.
Figure 2: Readers' median image quality score by modality (QT Scan in blue and HHUS in red) and breast feature: overall skin (SK), epidermis (Epi), dermis (Hy), Cooper's ligaments (Coop), superficial vessels (Ve), central ducts entering nipple (CenD), intermediate or peripheral ducts (XLob), terminal duct lobular units(TerD), and pectoralis muscle (Mus).
Image quality was scored on an ordinal rating scale: 1=Excellent, 2=Good, 3=Sufficient, 4=Restricted, 5=Poor.
Except for epidermis and pectoralis muscle, the readers scored the image quality on the QT Scan as equivalent or better than HHUS on each feature in more than 80% of breasts. Readers scored the image quality of epidermis and muscle as better on the QT Scan in more than 70% of breasts. On this initial study utilizing historical QT Library cases, the findings from the 5 study readers suggest that QT images are at least equivalent or better in defining the anatomical components of the breast than HHUS.
## Representative Clinical Cases
Sixteen clinical cases of QT Ultrasound images representative of different breast densities and lesion types, such as cancer in fatty breast, cancer in dense breast and cyst in fatty breast, were reviewed by board certified radiologist, Dr. Elaine Iuanow, M.D. and Chief Medical Officer. There was no grading performed on the Representative Clinical Cases. The cases were representative of the different types of cases seen in the clinical setting. Each case, consisting of Mammography, Hand Held Ultrasound and QT Ultrasound imaging, for each subject was reviewed.
The OT Ultrasound images were displayed in two modes. Speed of Sound (Transmission) and B-Mode (Reflection) images. Each imaging mode included coronal, axial and sagittal views. The cases included the span of clinical scenarios that the device would typically be used for.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in gray. The font is sans-serif and the overall design is simple and modern.
The breast density patterns include fatty, scattered, heterogeneously dense and extremely dense. The density was determined by visual inspection on the QT Ultrasound images. The clinical history for each case was provided. Clinical information, such as previous biopsies and/or imaging studies were used to confirm the identification of lesions in the QT Ultrasound images using clock position, appearance, size and location. The speed of sound values for any identified lesions were also provided. The lesion type was determined by histology (ground truth) for solid benign and solid malignant lesions. The ground truth for cyst lesions was determined by its appearance on handheld ultrasound (HHUS).
| Case # | Case ID | Breast Density Determined Visually | Lesion Type |
|--------|-----------|------------------------------------|-----------------|
| 1 | 901-065-L | Extremely Dense | Solid-Benign |
| 2 | 102007V | Heterogeneously Dense | * |
| 3 | 901-005-L | Fatty | Cyst |
| 4 | 102019V | Fatty | * |
| 5 | 901-020-R | Scattered | Cyst |
| 6 | 901-022-L | Scattered | Cyst |
| 7 | 901-008-R | Heterogeneously Dense | Cyst |
| 8 | 901-037-L | Scattered | Solid-Benign |
| 9 | 901-043-L | Heterogeneously Dense | Solid-Malignant |
| 10 | 901-044-L | Heterogeneously Dense | Solid-Malignant |
| 11 | 901-001-L | Scattered | Cyst |
| 12 | 901-029-R | Heterogeneously Dense | Cyst |
| 13 | 901-041-L | Fatty | Solid-Malignant |
| 14 | 901-019-R | Extremely Dense | Cyst |
| 15 | 102002V | Fatty | * |
| 16 | 901-034-R | Scattered | Solid-Benign |
### Table 2-6. Case summary by breast density and lesion type
* No lesion types are provided for volunteers as they were scanned for training only.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for QT Ultrasound. The logo consists of the letters "QT" in a light green color, followed by the word "ultrasound" in gray. The letters "QT" are larger than the word "ultrasound", and the "Q" is stylized with a curved tail.
The clinical cases demonstrated the QT Ultrasound Breast Scanner-1 is capable of imaging different breast densities and lesion types over the span of clinical scenarios that the device would typically be used for.
#### 9.0 CONCLUSION
The QT Ultrasound Breast Scanner-1 is substantially equivalent to the Delphinus SoftVue System with respect to intended use, principle of operation, design, performance and safety features. Both devices are intended as an adjunct to mammography and are not intended for screening purposes. The primary difference between the QT Ultrasound Breast Scanner-1 and the SoftVue System is the manner in which the qualitative speed of sound information is displayed. Although the method of display is different, the fundamental intended use is the same.
QT Ultrasound has demonstrated through verification and validation testing that the different technological characteristics between the SoftVue device and the QT Ultrasound Breast Scanner-1 do not raise new or different questions of safety and effectiveness. Therefore, it is the opinion of QT Ultrasound LLC that the QT Ultrasound Breast Scanner-1 is substantially equivalent to the predicate device identified in this submission that is currently cleared for market in the United States.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.