K162365 · Terumobct, Inc. · KCY · Feb 16, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162365
Device Name
T-Cuff
Applicant
Terumobct, Inc.
Product Code
KCY · General, Plastic Surgery
Decision Date
Feb 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5910
Device Class
Class 1
Indications for Use
T-Cuff is indicated for use in donor populations during apheresis procedures as an alternative method to maintain pressure on the arm and obtain venipuncture. The T-Cuff operates similar to a pneumatic tourniquet intended to partially restrict blood flow on the upper arm to result in optimal venous access during apheresis collection procedures.
Device Story
T-Cuff is a non-sterile, manual pneumatic tourniquet used in blood banking/apheresis facilities. It consists of a polyurethane bladder, cotton/polyester sheath, PVC bulb, and stainless steel pressure relief valve. Operators (nurses/technicians) apply the cuff to the donor's upper arm. The donor squeezes a rubber bulb to inflate the system. The device operates in two modes: Venipuncture (up to 120 mmHg) and Collection (20-40 mmHg). The pressure relief valve mechanically regulates air pressure to maintain these ranges. By restricting blood flow, the device assists in vein selection and maintains consistent venous pressure during collection. It requires no external energy source. The device helps optimize blood flow, potentially reducing operator interventions and procedure time during apheresis.
Clinical Evidence
Prospective, IRB-approved study on the Trima Accel Automated Blood Collection system. Evaluated impact on operator interventions and procedure time using the T-Cuff. Results showed decreased operator interventions without increasing procedure time. Collected products met FDA standards for residual white blood cells (< 5.0 × 10^6) and platelet yield (≥ 3.0 × 10^11). No adverse events attributed to the device.
Technological Characteristics
Manual pneumatic tourniquet; no energy source. Materials: polyurethane bladder, cotton/polyester sheath, PVC bulb, non-latex EPDM tubing, stainless steel pressure relief valve. Mechanical pressure regulation via spring-loaded plunger in relief valve. Two operation modes: Venipuncture (max 120 mmHg) and Collection (20-40 mmHg). Non-sterile. No electromechanical components.
Indications for Use
Indicated for donor populations undergoing apheresis procedures to maintain arm pressure and facilitate venipuncture/venous access.
Regulatory Classification
Identification
A pneumatic tourniquet is an air-powered device consisting of a pressure-regulating unit, connecting tubing, and an inflatable cuff. The cuff is intended to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 16, 2017
Terumo BCT, Inc. Ms. Nithya Rajan Specialist, Regulatory Affairs 10811 W. Collins Ave. Lakewood, Colorado 80215
Re: K162365
Trade/Device Name: T-Cuff Regulation Number: 21 CFR 878.5910 Regulation Name: Pneumatic tourniquet Regulatory Class: Class I Product Code: KCY Dated: January 16, 2017 Received: January 17, 2017
Dear Ms. Rajan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OM용%18″19-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K162365
Device Name T-Cuff
#### Indications for Use (Describe)
T-Cuff is indicated for use in donor populations during apheresis procedures as an alternative method to maintain pressure on the arm and obtain venipuncture. The T-Cuff operates similar to a pneumatic tourniquet intended to partially restrict blood flow on the upper arm to result in optimal venous access during apheresis collection procedures.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for TerumoBCT. The logo is written in a bold, green font. The letters are large and easy to read.
# 510(k) Summary
#### I. SUBMITTER
Terumo BCT, Inc. 10811 West Collins Ave. Lakewood, Colorado 80215 Phone: 720-480-6702
Contact Person: Nithya Rajan Title: Specialist, Regulatory Affairs Phone: 303-231-4402 303-231-4756 Fax:
Date Prepared: February 16, 2017
#### II. DEVICE
T-Cuff® Trade Name of Device: Common or Usual Name: Inflatable tourniquet Classification Name: Pneumatic tourniquet In accordance with 21 CFR 878.5910 the classification for this device Regulatory Class: is Class I medical device Product Code: KCY
#### III. PREDICATE DEVICE
### Table 1: Predicate Device Information
| Device | Product<br>Classification | Trade Name Of<br>Predicate Device | Manufacturer and<br>510(k) Holder | 510(k) Clearance<br>Number |
|-----------|---------------------------|-----------------------------------|-----------------------------------|----------------------------|
| Predicate | KCY | EZ Vein inflatable<br>tourniquet | Dominion Medical<br>Devices, LLC | K112874 |
#### IV. DEVICE DESCRIPTION
## A. Device Identification
The T-Cuff is a pneumatic tourniquet used as an alternative method to maintain consistent pressure on the arm and blood flow at the access site during apheresis procedures. This device is composed of 4 major components and was developed to enhance venous access during venipuncture and help donors maintain a consistent pressure on the arm during apheresis procedures. The catalog number for the T-Cuff is 71252.
## B. Device Characteristics
T-Cuff is designed to maintain a consistent pressure when applied to the donor's bicep with consistent squeezing of the inflatable bulb. During the apheresis process, the device operates in two modes: Venipuncture and Collection. The mode of operation is changed by adjusting the pressure relief valve cap. During Venipuncture mode, application of T-Cuff to the donor's upper arm prior to obtaining venipuncture allows the T-Cuff to inflate to assist with vein selection. During Collection mode, the T-Cuff maintains a consistent pressure of 20-40 mmHg (millimeter
Terumo BCT. Inc 10811 West Collins
Europe, Middle East and Af
Ikaroslaan 41
1930 Zaventem
Belgium
Phone: +32.2.715.05.90
Fax: +32.2.721.07.70
Terumo BCT (Asia Pacific) Lt
Room 3903-3903A, 39/F
ACE Tower, Windsor House
311 Gloucester Road
Causeway Bay, Hong Kong
Phone: +852.2283.0700
Fax: +852.2576.1311
Terumo BCT Latin America S.A
La Pampa 1517 – 12th Floor
C1428DZE
Buenos Aires
Argentina
Phone: +54.11.5530.5200
Fax: +54.11.5530.5201
Terumo BCT Japan, Inc.
20-14, 3-chrome
Higashi Gotanda, Shinagawa-ku
Tokyo 141-0022
apan
Phone: +81.3.6743.7890
Fax: +81.3.6743.9800
{4}------------------------------------------------
of mercury) on the donor's upper arm to help facilitate optimal blood flow during apheresis procedures. T-Cuff is a non-sterile, non-single use device.
# C. Device Description
The T-Cuff is a pneumatic tourniquet used as an alternative method to maintain pressure on the arm and, aid in the facilitation of blood flow during apheresis procedures. It operates in two modes, Venipuncture and Collection. During Venipuncture mode, the T-Cuff is inflated to higher pressure (no greater than 120 mmHg) as venipuncture is performed. Once Venipuncture is complete, the device maintains a steady, lower, consistent pressure (20 - 40 mmHg) during Collection mode. The donor squeezes a rubber bulb in continuous intervals to distribute air through the system and, facilitate optimal blood flow during apheresis procedures. The T-Cuff does not contain an energy source. Primary mechanism of action is obtained through the influx of air into the system and, by a modulating pressure relief valve that maintains defined pressure ranges.
# D. Environment of Use
The T-Cuff is intended for use in a blood banking environment or other similar facility. The operation of T-Cuff is performed by trained operators who manage the collection of blood products on apheresis devices. Operators of the T-Cuff have a variety of backgrounds and professional training. Primary users are expected to be nurses or qualified laboratory technicians who manage the apheresis process.
# E. Materials of Use
T-Cuff does not come into contact with blood or other bodily fluids. The device is applied at the start of the apheresis procedure to un-broken, intact skin and, comprised of a polyurethane bladder encased in a cotton/polyester sheath. The bulb squeezed by the operator/donor is made of PVC and the tube connecting the bulb to the bladder is a non-latex EPDM. The contact surface of the pressure relief valve is stainless steel.
# F. Key Performance Specifications/Characteristics of the Device
The pressure relief valve regulates air pressure within the system and is mechanically controlled through the two operation modes. As the inflatable bulb is squeezed by the donor, air is pushed through the rubber tubing into the pressure relief valve. As pressure gradually builds within the system; the plunger within the relief valve compresses springs based on the operation mode. This allows for air to escape through the top of the pressure relief valve and, ensures a consistent pressure range is maintained based on the operation mode of the device.
#### V. INDICATIONS FOR USE
T-Cuff is indicated for use in donor populations during apheresis procedures as an alternative method to maintain pressure on the arm and obtain venipuncture. The T-Cuff operates similar to a pneumatic tourniquet intended to partially restrict blood flow on the upper arm to result in optimal venous access during apheresis collection procedures.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the text "Page 3 of 4" at the top, indicating it is part of a multi-page document. Below this, the logo "TERUMOBCT" is displayed in a bold, green font. The logo appears to be the name of a company or organization. The overall image is simple and focuses on conveying the page number and brand identity.
#### TECHNOLOGICAL COMPARISON VI.
Control of air pressure within the device is the technological principle for both the proposed and predicate devices. It is based on limiting pressure exerted on the donor's upper arm. The subject and predicate devices are based on the same technological elements as described below:
- o Squeeze bulb to circulate air within the system.
- o Pressure control valve to regulate and maintain a certain pressure level within the device.
- Inflatable cuff that compresses donor's forearm or bicep to gently restrict blood flow and increase venous pressure.
The following technological differences exist between the subject and predicate devices:
- The predicate device delivers one pressure mode while the proposed device offers two ● pressure modes.
- o The predicate device is primarily used for enhanced presentation of veins for access with a hypodermic needle while the proposed device allows for use during both venipuncture and apheresis collections.
#### VII. PERFORMANCE DATA
The following performance data are provided in support of the substantial equivalence determination.
## A. Biocompability Testing
The biocompatibility evaluation for the T-Cuff device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity ●
- Intracutaneous reactivity ●
- Sensitization ●
- USP Physico-chemical tests ●
# B. Electrical Safety and Electromagnetic Compatibility (EMC) Testing
T-Cuff does not contain any electromechanical components and thus did not require EMC testing.
## C. Software Verification and Validation Testing
The T-Cuff does not contain any software, therefore, software verification and validation testing was not performed.
## D. Sterility Testing
T-Cuff is a non-sterile device; therefore sterility testing was not performed
# E. Stability/Shelf Life Testing
The Pressure relief valve has been tested to be functional for a 3 year (3,000 apheresis procedure) shelf life. The cuff bladder was tested to be functional for a 1 year shelf life at which point, the customer has the option to replace the bladder component on an as-needed basis. Due
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the text "Page 4 of 4" at the top, indicating it is the last page of a document. Below that is the logo for TERUMOBCT in green. The logo is in a bold, sans-serif font and is easily readable. The overall image is clean and simple, with a focus on the text and logo.
to the low likelihood of time-dependent product degradation. a specific shelf-life time is not applicable.
# F. Clinical Studies
The design was validated through an IRB approved, prospective study on the Trima Accel® Automated Blood Collection system. The primary endpoint for this study was to determine if manual and/or automatic adjustments to the draw flow rate would decrease the number of operator interventions and overall procedure time. The T-Cuff was utilized for all donations in this study, and it was demonstrated that manual and automatic adjustment to the draw flow rate decreased the number of operator interventions and did not increase the overall procedure time. All products collected met the FDA regulations regarding the level of residual white blood cells (< 5.0 × 106) and platelet yield (≥ 3.0 × 1011). There was also no reduction in the quantity (volume and yield) or the quality (residual white blood cells) of products collected compared to historical controls. No reported AE's were attributed to the T-Cuff.
#### VIII. CONCLUSIONS
Based on data collected and analyzed during Bench and Clinical testing, the T-Cuff has demonstrated to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.