K162343 · Westmed, Inc. · CCK · Dec 14, 2016 · Anesthesiology
Device Facts
Record ID
K162343
Device Name
Westmed Gas Sampling Cannula with O2 delivery
Applicant
Westmed, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Dec 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Westmed nasal cannula styles are intended to deliver supplemental oxygen to patients and provide a means to sample expired gases. Cumulative duration of use < 24 hours.
Device Story
The Westmed Gas Sampling Cannula is a disposable, single-patient-use device designed to deliver supplemental oxygen while simultaneously sampling expired gases for monitoring. The device features a channeled or split nasal cannula design; one channel delivers oxygen to the nares, while the other samples exhaled gases. Certain configurations include an additional oral sampling port positioned near the mouth to capture exhaled gases from mouth-breathing patients. The device connects to standard oxygen tubing and gas sampling lines. It is intended for use in hospital, sub-acute, and pre-hospital settings by healthcare providers. The sampled gas is directed to an external gas analyzer (not part of this device) for monitoring end-tidal CO2 and waveforms. By providing a dual-function interface, the device allows for continuous respiratory monitoring and oxygen therapy, facilitating clinical decision-making regarding patient ventilation and oxygenation status.
Clinical Evidence
Bench testing only. Comparative performance testing evaluated CO2 sampling accuracy and waveform fidelity across three breathing rates, two CO2 concentrations (2% and 5%), and three oxygen flow rates (1, 3, and 5 lpm). Mechanical testing included luer fitting integrity, flow rates, bond strength, and kink resistance. Biocompatibility testing was conducted per ISO 10993-5 and ISO 10993-10, covering cytotoxicity, sensitization, and intracutaneous irritation. No clinical data was required or provided.
Technological Characteristics
Materials: Flexible PVC tested to ISO 10993-5 and ISO 10993-10. Design: Channeled/split nasal cannula with integrated oxygen delivery and gas sampling ports; optional oral sampling extension. Dimensions: Sampling tubing ID 0.060", OD 0.109". Connectivity: Standard luer fittings for connection to external gas analyzers. Energy source: None (passive). Sterilization: Not specified.
Indications for Use
Indicated for patients requiring supplemental oxygen and/or expired gas monitoring, including adults and pediatrics (> 10 kg).
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 14, 2016
Westmed, Inc. c/o Paul Dryden Consultant 5580 S Nogales Hwy Tucson, Arizona 85706
Re: K162343
Trade/Device Name: Westmed Gas Sampling Cannula with O2 delivery Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK, CAT Dated: November 10, 2016 Received: November 14, 2016
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K162343
#### Device Name
Westmed Gas Sampling Cannula with O2 delivery
Indications for Use (Describe)
The Westmed nasal cannula styles are intended to deliver supplemental oxygen to patients and provide a means to sample expired gases. Cumulative duration of use < 24 hours.
## Environment of use:
Hospital, sub-acute, and pre-hospital settings
### Patient population:
Patients requiring supplemental oxygen and / or expired gas monitoring, adult to pediatrics (> 10 kg.).
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement on last page.
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 4
| Date Prepared: | 13-Dec-2016 |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Westmed, Inc.<br>5580 S Nogales Hwy<br>Tuscon, AZ 85706<br>Tel - (520) 294-7987 | |
| Official Contact: | Diana Upp, VP RA/QA |
| Proprietary or Trade Name: | Westmed Gas Sampling Cannula with O2 delivery |
| Common/Usual Name: | analyzer, gas, carbon-dioxide, gaseous phase (accessories) |
| Classification Name: | analyzer, gas, carbon-dioxide, gaseous phase (accessories)<br>CCK - CFR 868.1400<br>Class II |
| Additional Product Code: | cannula, nasal, oxygen<br>CAT |
| Predicate Device:<br>Reference Device: | K010024 – Oridion<br>K011050 - Oridion |
Device Description: Westmed has designed several 0%CO2 nasal cannula / exhaled gas sampling devices. They include:
- . Nasal cannula style that can provide supplemental O2 and sample exhaled gases at only the nares and
- Nasal cannula style that can provide supplemental O2 and sample exhaled gases at both ● the nares and at the mouth
The device configurations include nasal cannula that have a division to deliver oxygen through one nare and to sample exhaled gases through the other. There is also an oral sampling style which is positioned near the mouth to sample exhaled gases if the patient is a mouth breather.
The device includes a length of standard oxygen tubing and gas sampling lines.
## Indications for Use:
The Westmed nasal cannula styles are intended to deliver supplemental oxygen to patients and provide a means to sample expired gases. Cumulative duration of use < 24 hours.
### Environment of use:
Hospital, sub-acute, and pre-hospital settings
## Patient population:
Patients requiring supplemental oxygen and / or expired gas monitoring, adult to pediatrics (> 10 kg.).
### Substantial Equivalence Discussion -
Tables 1 and 2 below compare the key features of the proposed devices with the identified predicates and demonstrate that the proposed device can be found to be substantially equivalent.
{4}------------------------------------------------
| Attribute | Predicate<br>Oridion<br>K010024 | Proposed<br>Model 0581 |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | To sample exhaled gas via a nasal cannula and<br>simultaneously provide supplemental oxygen<br>near the nose and mouth for inhalation | The Westmed nasal cannula styles are intended to<br>deliver supplemental oxygen to patients and provide<br>a means to sample expired gases. Cumulative<br>duration of use < 24 hours |
| Environments of use | Hospitals, sub-acute, pre-hospital settings | Environment of use:<br>Hospital, sub-acute, and pre-hospital settings |
| Patient population | Patient requiring supplemental oxygen and / or<br>sampling of expired gases<br>Adult to pediatric | Patient population:<br>Patients requiring supplemental oxygen and / or<br>expired gas monitoring, adult to pediatrics (>10 kg.) |
| Duration of use | Single patient use, disposable | Single patient use, disposable |
| Basic components | Channeled / split nasal cannula<br>Oxygen tubing<br>Gas sampling line | Channeled / split nasal cannula<br>Oxygen tubing<br>Gas sampling line |
| Sampling tubing specs | Not provided | ID - 0.060" / OD - 0.109"<br>Length - 2"<br>Luer fitting - Male / Female<br>Connected to customer supplied gas sampling line |
| Technological<br>characteristics / design | Split nasal cannula with sampling in one nares<br>and oxygen delivery in the other | Split nasal cannula with sampling in one nares and<br>oxygen delivery in the other |
| Materials | Flexible PVC for the cannula, oxygen tubing and<br>gas sampling line | Tested to ISO 10993-5 and 10993-10<br>Flexible PVC for the cannula, oxygen tubing and<br>gas sampling line |
| Performance testing | | Comparison End-tidal CO2 results and waveform at<br>various settings<br>Tensile strength of connections<br>Luer fitting testing<br>Age testing - Environmental<br>Mechanical testing |
## Table 1 – Comparison - O2 delivery / CO2 sampling Cannula style
## Table 2 – Comparison - O2 delivery / CO2 sampling Oral / Nasal Cannula style
| Attribute | Reference<br>Oridion<br>K011050 | Proposed<br>Model 0430 |
|---------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Used whenever the physician needs to measure<br>the CO2 in a patient's breathing in a non-<br>intubated patient | The Westmed nasal cannula styles are intended to<br>deliver supplemental oxygen to patients and provide<br>a means to sample expired gases. Cumulative<br>duration of use < 24 hours |
| Environments of use | Hospitals, sub-acute, pre-hospital settings | Environment of use:<br>Hospital, sub-acute, and pre-hospital settings |
| Patient population | Patient requiring supplemental oxygen and / or<br>sampling of expired gases | Patient population:<br>Patients requiring supplemental oxygen and / or<br>expired gas monitoring, adult to pediatrics (>10<br>kgs). |
| Multiple sizes | Adult to pediatric | Adult to pediatric |
| Duration of use | Single patient use, disposable | Single patient use, disposable |
{5}------------------------------------------------
### 510(k) Summary Page 3 of 4
| Attribute | Reference<br>Oridion<br>K011050 | Proposed<br>Model 0430 |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Basic components | Channeled / split nasal cannula<br>Mouth sampling option<br>Oxygen tubing<br>Gas sampling line | Channeled / split nasal cannula<br>Mouth sampling option<br>Oxygen tubing<br>Gas sampling line |
| Sampling tubing specs | Not provided | ID – 0.060" / OD –0.109"<br>Length –7 ft.<br>Luer fitting – Male / Female |
| Technological<br>characteristics / design | Split nasal cannula with sampling in one nares<br>and oxygen delivery in the other and a separate<br>gas sampling placed near the mouth | Split nasal cannula with sampling in one nares and<br>oxygen delivery in the other and a separate gas<br>sampling placed near the mouth |
| Materials | Flexible PVC for the cannula, oxygen tubing and<br>gas sampling line | Tested to ISO 10993-5 and 10993-10<br>Flexible PVC for the cannula, oxygen tubing and<br>gas sampling line |
| Performance testing | | Comparison End-tidal CO2 results and waveform at<br>various settings<br>Tensile strength of connections<br>Luer fitting testing<br>Age testing - Environmental<br>Mechanical testing |
Tables 1 and 2 above compare the key features of the proposed Westmed Oxygen delivery / CO2 sampling cannula style devices with the identified predicate and reference, K010024 and K011050 - Oridion Nasal cannula sampling, and it demonstrates that the proposed devices can be found to be substantially equivalent.
### Discussion of Differences -
## Indications for Use -
The indications for use are similar for the proposed device when compared to the identified predicate and reference - K010024 and K011050 - Oridion Nasal cannula sampling. Discussion - Each device is indicated for use delivering supplemental oxygen and sampling expired gases.
## Technology and construction -
The design, fabrication, shape, size, etc. are equivalent to the identified predicate and reference -K010024 and K011050 - Oridion Nasal cannula sampling. The materials are similar: flexible PVC for the nasal cannula, oxygen tubing and gas sampling line.
Discussion – The design incorporates several styles of split / channeled nasal cannula where oxygen and expired gases are provided through the various ports within the cannula. For the oral / nasal style, there is an extra part which is placed near the mouth to sample expired gases which may be exhaled by the patient.
## Environment of Use -
The environments of use are similar to identified predicate and reference - K010024 and K011050 - Oridion Nasal cannula sampling.
Discussion - The environments of use are not specifically disclosed in the predicate 510(k) Summary, but literature would support the listed environments of use. There are no differences in the requirements of each environment of use which would raise any new safety concerns when compared to the identified predicate and reference – K010024 and K011050 – Oridion Nasal cannula sampling.
{6}------------------------------------------------
## Patient Population -
The patient population of pediatrics to adults is equivalent to the identified predicate and reference - K010024 and K011050 - Oridion Nasal cannula sampling. Discussion - The patient populations are equivalent to the identified predicate and reference -K010024 and K011050 - Oridion Nasal cannula sampling.
### Non-Clinical Testing Summary -
### Bench Testing -
We performed comparative testing which evaluated the ability to measure a breath, measure a gas, and breathe waveforms under several simulated breathing conditions.
The tests included:
- . Comparative CO2 sampling and waveform performance at 3 different breathing rates. CO2 concentrations (2% and 5%), and oxygen flow rates (1, 3, and 5 1pm).
- . Environmental and Age testing
- Mechanical testing ●
- o Luer Fitting
- o Flow
- Strength of bonds O
- Resistance to kinks O
The results presented show that the proposed Westmed Oxygen delivery / CO2 sampling cannula style devices performed equivalent to the predicates.
All testing demonstrated that the proposed devices are substantially equivalent to the identified predicate and reference - K010024 and K011050 - Oridion Nasal cannula sampling.
## Biocompatibility -
We have performed ISO 10993 testing on the component materials of the proposed devices which is considered as Externally Communicating and Surface Contact with the patient which means the following tests are required.
- Cytotoxicity ●
- Sensitization
- . Intracutaneous Irritation
## Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate can be found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.