CardioChek Plus Test System, CardioChek Home Test System

K162282 · Polymer Technology Systems, Inc. D/B/A Pts Diagnostics · CGA · Dec 22, 2016 · Clinical Chemistry

Device Facts

Record IDK162282
Device NameCardioChek Plus Test System, CardioChek Home Test System
ApplicantPolymer Technology Systems, Inc. D/B/A Pts Diagnostics
Product CodeCGA · Clinical Chemistry
Decision DateDec 22, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The CardioChek Plus Test System is intended for multiple patient use in professional healthcare settings. The CardioChek Home Test System is intended for single-patient use.

Device Story

CardioChek Plus/Home Test System; point-of-care diagnostic analyzer; measures glucose, cholesterol, HDL from blood samples. Modifications include wireless/wired communication, software solutions, printer connectivity, MEMo chip update, battery change, multi-language support, and updated cleaning/disinfection instructions. Operated by healthcare professionals or patients; provides quantitative test results for clinical decision-making. System robustness validated for 11,000 cleaning/disinfection cycles using Super Sani-Cloth. Device maintains fundamental scientific technology of previously cleared predicates.

Clinical Evidence

Bench testing only. Verification and validation activities confirmed that the modified analyzers meet all predetermined acceptance criteria and perform as well as the predicate devices.

Technological Characteristics

Reflectance photometry; enzymatic dry-strip chemistry; lot-specific MEMo chip calibration; 4 AA battery power; wired/wireless connectivity; USB printer support; multi-language software; larger form factor than predicates.

Indications for Use

Indicated for quantitative determination of total cholesterol, HDL cholesterol, and glucose in capillary whole blood (Home system) or venous/capillary whole blood (Plus system). Home system intended for single-person use; Plus system intended for multiple-patient use in professional healthcare settings. Used for diagnosis/treatment of lipid, lipoprotein, and carbohydrate metabolism disorders, atherosclerosis, and liver/renal diseases.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Decision Summary To: THE FILE RE: DOCUMENT NUMBER k162282 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) PTS PANELS CHOL+HDL+GLU Panel Test Strips (k071507) PTS PANELS CHOL+HDL Panel Test Strips (k071593) Chol +Glu Test Panel (k041750) 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for the following modifications to the original device: a. Wireless communications capability (professional system) b. Software solutions capability c. Printer connectivity capability d. MEMo chip appearance e. Battery type f. Multiple language software capability g. Wired PC Communication h. Analyzer dimensions i. The addition of validated cleaning and disinfection instructions to the labeling 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics (including robustness of the system to repeated cleaning and disinfection and software verification and validation). 5. Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. {1} b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modifications. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. The CardioChek Plus Test System is intended for multiple patient use in professional healthcare settings. The CardioChek Home Test System is intended for single-patient use. Disinfection efficacy studies for this device were performed in k140068. The studies were performed by an outside commercial testing laboratory and demonstrated complete inactivation of duck hepatitis B virus (HBV) with the chosen disinfectant, Super Sani-Cloth (EPA Registration #9480-4). In this submission robustness studies were performed and demonstrated that there was no change in functionality, performance of glucose, cholesterol, and HDL, or external materials after 11,000 cleaning and disinfection cycles. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures. 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...