K162235 · Laclede, Inc. · NUC · Mar 24, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K162235
Device Name
LUVENA VAGINAL MOISTURIZER & LUBRICANT
Applicant
Laclede, Inc.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Mar 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Luvena Vaginal Moisturizer and Lubricant is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms, but is not compatible with polyurethane condoms.
Device Story
Luvena Vaginal Moisturizer and Lubricant is a non-sterile, aqueous-based, cranberry-flavored gel; supplied in single-use EVA/PE copolymer plastic applicator tubes. Intended for personal vaginal use to provide moisture and lubrication during intimate sexual activity. Device is OTC; user-applied. Formulation provides physical lubrication; does not contain contraceptives or spermicides. Compatibility with natural rubber latex and polyisoprene condoms confirmed via ASTM D7661-10; incompatible with polyurethane condoms. Healthcare providers use the product to assist patients experiencing vaginal dryness; output is the physical application of the gel to the vaginal area to reduce friction and improve comfort.
Clinical Evidence
Bench testing only. Biocompatibility established per ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity). Physical testing included antimicrobial preservative effectiveness (USP <51>), condom compatibility (ASTM D7661-10), and shelf-life stability (ICH-Q1A/Q1E). Microbiological testing performed per USP <61>, <62>, and <1111>.
Indicated for vaginal application in individuals requiring moisturization and lubrication to enhance comfort during intimate sexual activity and supplement natural lubrication. Compatible with natural rubber latex and polyisoprene condoms; contraindicated for use with polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
Luvena Enhanced Personal Moisturizer and Luvena Enhanced Personal Lubricant (K150841)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right, with a design element below them that resembles flowing water or fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2017
Laclede, Inc. Michael Pellico President 2103 E. University Dr. Rancho Dominguez, CA 90220
Re: K162235
> Trade/Device Name: Luvena Vaginal Moisturizer and Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: February 21, 2017 Received: February 24, 2017
Dear Michael Pellico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Charles Viviano -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. There is a faint, light blue watermark in the background that is partially obscured by the text.
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
#### K162235
Device Name
Luyena Vaginal Moisturizer and Lubricant
Indications for Use (Describe)
Luyena Vaginal Moisturizer and Lubricant is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms, but is not compatible with polyurethane condoms.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
# 510 (k) Summary
| Submission By: | Laclede, Inc. |
|----------------|--------------------------------------------------------|
| Address: | 2103 E. University Drive<br>Rancho Dominguez, Ca 90220 |
| Telephone: | 310-605-4280 |
| Facsimile: | 310-605-4288 |
| Contact Name: | Michael A. Pellico, President |
| Email: | mpellico@laclede.com |
Date 510(k) Summary Prepared: March 23, 2017
| Trade Name: | Luvena Vaginal Moisturizer & Lubricant |
|----------------------|----------------------------------------|
| Common Name: | Personal Lubricant |
| Classification Name: | Condom (21 C.F.R. 884.5300) |
| Classification: | Class II |
| Product Code: | NUC (lubricant, personal) |
## Primary Predicate Device:
Luvena Enhanced Personal Moisturizer and Luvena Enhanced Personal Lubricant: K150841, Laclede, Inc. This predicate device has not been subject to any design related recall.
## Device Description:
Luvena Vaginal Moisturizer & Lubricant is non-sterile, translucent, non-staining, non-greasy, aqueous-based, cranberry-flavored gel and is packed in EVA/PE copolymer plastic applicator tube. Each application tube is single-use. It is for personal vaginal use as need for moisturizing and lubrication during intimate sexual activity.
This device is not a contraceptive or spermicide, nor does it contain any such component. It is compatible with natural rubber latex and polyisoprene condoms and not compatible with polyurethane condoms as demonstrated in condom compatibility testing conducted according to ASTM D7661-10.
## Device Specifications:
The device specifications for the Luvena Vaginal Moisturizer and Lubricant include appearance, odor, pH, viscosity, osmolality, Total Aerobic Microbial Count ((TAMC) (USP<61> and <1111>)), Total Yeast and Mold Count ((TYMC) (USP<61> and <1111>)), and absence of pathogenic organisms ((Pseudomonus aeruginosa, Staphylococcus aureus, Candida albicans (USP<62>)). These specifications were evaluated during the length of the proposed shelf life.
## Intended Use:
Luvena Vaginal Moisturizer and Lubricant is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with
{4}------------------------------------------------
natural rubber latex and polyisoprene condoms, but is not compatible with polyurethane condoms.
# Substantial Equivalence:
Luvena Vaginal Moisturizer & Lubricant have the same intended use as the predicate device. However, the subject moisturizer & lubricants and the predicate device have different technological characteristics – namely differences in viscosity, formulation and package type. These differences do not raise different questions of safety & effectiveness.
| Features | Luvena Vaginal<br>Moisturizer &<br>Lubricant | Luvena Enhanced Personal<br>Moisturizer and<br>Luvena Enhanced Personal<br>Lubricant<br>K150841 | Differences |
|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Water Based | Yes | Yes | Same, The subject and<br>predicate device are water<br>based |
| Water Soluble | Yes | Yes | Same, The subject and<br>predicate device are soluble in<br>water |
| Non Sterile | Yes | Yes | Same, The subject and<br>predicate device are non-sterile. |
| Flavoring | Cranberry | Cranberry | Same, The subject and<br>predicate device have same<br>flavoring. |
| Container Type | Single use Applicator | Bottle with pump | The subject and predicate<br>device are different size and<br>shape for OTC packaging, but<br>these differences have no<br>impact on the safety or efficacy<br>of the personal lubricant. |
| Condom<br>compatibility | Yes<br>ASTM D7661-10<br>Compatible with<br>natural rubber latex<br>and polyisoprene<br>condoms, and not<br>compatible with<br>polyurethane<br>condoms. | Yes<br>ASTM D7661-10<br>Compatible with natural<br>rubber latex and<br>polyisoprene condoms, and<br>not compatible with<br>polyurethane condoms. | Same, The subject and<br>predicate device have been<br>tested for condom compatibility<br>in accordance with ASTM<br>D7661-10. |
| pH | Yes<br>3.8 - 4.8 | Yes<br>3.8 - 4.8 | Same, The subject and predicate<br>device have same pH range. |
| Microbiological<br>Examination of<br>Non-sterile<br>Products Testing | USP <61>/<62> | USP <61>/<62> | Same, The subject and predicate<br>device have been tested for<br>Microbial Limit test in<br>accordance with USP <61> and<br><62> and have same<br>specifications as recommended<br>in USP<1111> for vaginal use |
| | | products. | |
| Body Location<br>Target area | Vaginal application | Penile and/or<br>Vaginal application | Same, Subject and predicate device are for Vaginal application. |
| Intended Use | Moisturize & Lubricate during sexual activity | Moisturize & Lubricate during sexual activity | Same, Subject and predicate device have same intended use. |
{5}------------------------------------------------
# Biocompatibilitv:
Biocompatibility of the final formulation of Luvena Vaginal Moisturizer & Lubricant was established through testing generally following ISO 10993, and includes cytotoxicity, sensitization, irritation, and acute systemic toxicity testing. The results demonstrate that the subject lubricant is biocompatible.
- · Acute Systemic Toxicity: ISO 10993-11:2006
- · Cytotoxicity: ISO 10993-5:2009
- · Vaginal Irritation Testing: ISO 10993-10:2010
- · Sensitization Testing Guinea Pig Maximization Test (GPMT): ISO 10993-10:2010.
# Physical & Other Testing:
The physical and other testing performed on Luvena Vaginal Moisturizer & Lubricant included an Antimicrobial Preservative Effectiveness Validation testing under USP < 5 > Category 2. Lubricant- Condom Compatibility testing per ASTM D7661 - 10 Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms and Shelf life testing. Luvena Vaginal Moisturizer and Lubricant passed the Antimicrobial Effectiveness test and were found to be compatible with natural rubber latex and polyisoprene condoms, but not compatible with polyurethane condoms. The results of accelerated time testing and real time testing according to ICH-Q1A (Stability testing of new drug substances and products) and ICH-Q1E (Evaluation of stability data) support the proposed shelf life for the subject lubricant.
# Conclusion:
The subject and predicate device have the same intended use and the different technological characteristics do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.