← Product Code [DAP](/productcode/DAP) · K162227

# STA® - Liatest® D-Di (K162227)

_Diagnostica Stago · DAP · Dec 10, 2016 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162227

## Device Facts

- **Applicant:** Diagnostica Stago
- **Product Code:** [DAP](/productcode/DAP.md)
- **Decision Date:** Dec 10, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.7320
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The STA® - Liatest® D-Di kit is an immuno-turbidimetric assay for the quantitative determination of D-dimer in venous plasma (3.2% sodium citrate) for use on STA-R®, STA Compact® and STA Satellite® analyzers by professional laboratory personnel. The STA® - Liates® D-Di is intended for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT) in outpatients suspected of PE or DVT.

## Device Story

STA® - Liatest® D-Di is an in vitro diagnostic immuno-turbidimetric assay for quantitative D-dimer measurement in venous plasma. Used on STA-R®, STA Compact®, and STA Satellite® analyzers by professional laboratory personnel in hospital or clinical laboratories. Principle: plasma sample mixed with suspension of latex microparticles coated with mouse monoclonal anti-human D-dimer antibodies (8D2 and 2.1.16); antigen-antibody reaction causes agglutination, increasing turbidity; change in turbidity measured via photometry (540 nm). Increase in absorbance correlates to D-dimer level. Results used alongside clinical pretest probability (PTP) models to exclude PE and DVT in outpatients. Negative results help rule out VTE, potentially reducing need for imaging procedures.

## Clinical Evidence

Prospective multi-center study (n=980) evaluating DVT exclusion in outpatients with low/moderate PTP. Primary endpoints: sensitivity, specificity, NPV, PPV. Results: Sensitivity 100% (LCL 95.8%), Specificity 55.2% (LCL 51.9%), NPV 100% (LCL 99.3%), PPV 17.5% (LCL 14.2%). Comparator: Imaging and 3-month follow-up.

## Technological Characteristics

Immuno-turbidimetric assay; reagents: Tris buffer and microlatex particles coated with monoclonal anti-human D-dimer antibodies (8D2, 2.1.16) stabilized with bovine albumin. Measurement: photometric absorbance at 540 nm. Analytical range: 0.27–10 μg/mL (FEU). Standard: CLSI H59-A. Automated processing on specific STA-series analyzers.

## Predicate Devices

- VIDAS® D-Dimer Exclusion™ Assay ([K040882](/device/K040882.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K162227

B. Purpose for Submission:
To expand the intended use of the previously cleared STA® - Liatest® D-Di to include the exclusion of deep vein thrombosis (DVT)

C. Measurand:
D-dimer (μg/mL FEU)

D. Type of Test:
Quantitative

E. Applicant:
Diagnostica Stago

F. Proprietary and Established Names:
STA® - Liatest® D-Di

G. Regulatory Information:
1. Regulation section:
21 CFR 864.7320, Fibrinogen/fibrin degradation products assay
2. Classification:
Class II
3. Product code:
DAP, Fibrinogen and fibrin split products, antigen, antiserum, control
4. Panel:

{1}

Hematology (81)

## H. Intended Use:

1. Intended use:

Same as indication(s) for use below.

2. Indication(s) for use:

The STA® - Liatest® D-Di kit is an immuno-turbidimetric assay for quantitative determination of D-dimer in venous plasma (3.2% sodium citrate) for use on STA-R®, STA Compact® and STA Satellite® analyzers by professional laboratory personnel. The STA® - Liatest® D-Di is intended for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT) in outpatients suspected of PE or DVT.

3. Special conditions for use statement(s):

For prescription use only

4. Special instrument requirements:

STA-R® (K983460)

STA Compact® (K093167)

STA Satellite® (K082248)

## I. Device Description:

The STA® - Liatest® D-Di kit consists of two liquid reagents: 6 x 5 mL vials of ready-for-use Tris buffer and 6 x 6 mL vials of a suspension of microlatex particles coated with two different mouse monoclonal anti-human D-dimer antibodies (8D2 and 2.1.16) stabilized with bovine albumin.

## J. Substantial Equivalence Information:

1. Predicate device names:

VIDAS® D-Dimer Exclusion™ Assay

2. Predicate 510(k) numbers:

K040882

3. Comparison with predicate:

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device STA® - Liatest® D-Di | Predicate VIDAS® D-Dimer Exclusion  |
|  Intended Use | The STA® - Liatest® D-Di is an immuno-turbidometric assay for the quantitative determination of D-dimer in venous plasma (in 3.2% sodium citrate) for use on STA®, STA Compact® and STA Satellite® analyzers by professional laboratory personnel. The STA® - Liatest® D-Di is intended for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT) in outpatients suspected of PE or DVT. | VIDAS® D-dimer Exclusion™ assay is an automated quantitative test for use on the VIDAS analyzers for the immunoenzymatic determination of fibrin degradation products (FbDP) in citrated human plasma using the ELFA techniques (Enzyme Linked Fluorescent Assay). VIDAS® D-Dimer Exclusion assay is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep venous thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE.  |
|  Clinical cut-off for exclusion of DVT | 0.5 μg/mL (FEU) | 500 ng/mL (FEU)  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device STA® -Liatest® D-Di | Predicate VIDAS® D-dimer Exclusion  |
|  Assay methodology | Immuno-turbidimetric | Enzyme Linked Fluorescent Assay (ELFA)  |
|  Test Principle | Immuno-turbidimetric method based on the measurement of light absorbance (at 540 nm) produced by a suspension of microlatex particles coated with specific mouse anti-human D-dimer monoclonal antibodies. | Two-step enzyme immunoassay sandwich method with a final fluorescent detection step (ELFA).  |
|  Analytical measurement range | 0.27–10 μg/mL (FEU) | 45–10,000 ng/mL (FEU)  |

K. Standard/Guidance Document Referenced (if applicable):

CLSI document H59-A, Quantitative D-dimer for the Exclusion of Venous Thromboembolic Disease; Approved Guideline

L. Test Principle:

A suspension of latex microparticles, coated by covalent bonding with monoclonal antibodies specific for D-dimer is mixed with the test plasma for which the D-dimer level is to be

{3}

assayed. An antigen-antibody reaction takes place, leading to agglutination of the latex microparticles which causes an increase in turbidity of the reaction medium. This increase in turbidity is reflected by an increase in absorbance, the latter being measured photometrically. The increase in absorbance is a function of the D-dimer level present in the test sample expressed in fibrinogen equivalent units (FEU).

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

In premarket notification K964728 the following studies were performed with STA - Liatest® D-Di: precision/reproducibility, limit of detection, linearity, interference, and stability studies. Additionally, instrument-specific analytical performance of STA - Liatest® D-Di in combination with the STA-R® and STA Satellite was evaluated in premarket notifications K983460 and K082248, respectively. The established analytical performance characteristics cleared in K964728, K983460 and K082248 remain unchanged; therefore, additional performance studies were not required to support substantial equivalence in this premarket notification.

### 2. Comparison studies:

In premarket notification K964728, the performance of STA - Liatest® D-Di compared to Asserachrom D-Di Kit (K862156) was evaluated. The established performance characteristics cleared in K964728 remain unchanged; therefore, additional comparison studies were not required to support substantial equivalence in this premarket notification.

### 3. Clinical studies:

The STA® - Liatest® D-Di assay was evaluated on the STA-R® (K983460), STA-R® Evolution (K093001), STA Compact® (K093167), and STA Compact Max® (K130090) analyzers in a multi-center study to validate exclusion of DVT in outpatients with a low or moderate pre-test probability (PTP) score. In total, 1,219 consented patients presenting to the emergency department or outpatient clinic with suspected venous thromboembolism (VTE) were prospectively recruited. Overall, samples were collected from a total of 16 sites; 9 in the United States and 7 in France, Italy, Spain and Canada. The Wells' score clinical PTP assessment model was used to categorize patients as low, moderate, or high probability for suspected DVT. Patients with a high PTP score were excluded from the study.

Of the 1,219 patients, 254 were excluded from analysis; of which 158 had a high PTP score and 96 were excluded on the basis of the defined exclusion criteria, for a total of 980 patients included in the data analysis. The sensitivity, specificity, negative predictive value (NPV) and positive predictive value (PPV) with the lower bound of two-sided 95% confidence intervals (LCL) were obtained with the STA - Liatest® D-Di clinical cut-off of 0.50 µg/mL (FEU). The overall prevalence of DVT in the 980 patients was 8.4% (85 of 980) with 6.0% (22 of 369) in the U.S. population and 10.3% (63 of 611) in the Canadian and European populations.

4

{4}

|  Combined Sites | Imaging and 3-month follow-up |   |   | Predictive Value (Lower confidence limit)  |   |
| --- | --- | --- | --- | --- | --- |
|  STA - Liatest® D-Di | Positive | Negative | Total | NPV | 100.0% (LCL: 99.3%)  |
|  Positive | 85 | 401 | 486 | PPV | 17.5% (LCL: 14.2%)  |
|  Negative | 0 | 494 | 494 | Sensitivity | 100.0% (LCL: 95.8%)  |
|  Total | 85 | 895 | 980 | Specificity | 55.2% (LCL: 51.9%)  |
|  US Sites | Imaging and 3-month follow-up |   |   | Predictive Value (Lower confidence limit)  |   |
| --- | --- | --- | --- | --- | --- |
|  STA - Liatest® D-Di | Positive | Negative | Total | NPV | 100.0% (LCL: 98.2%)  |
|  Positive | 22 | 145 | 145 | PPV | 13.2% (LCL: 8.4%)  |
|  Negative | 0 | 202 | 202 | Sensitivity | 100.0% (LCL: 84.6%)  |
|  Total | 22 | 347 | 369 | Specificity | 58.2% (LCL: 52.8%)  |
|  Non-US Sites | Imaging and 3-month follow-up |   |   | Predictive Value (Lower confidence limit)  |   |
| --- | --- | --- | --- | --- | --- |
|  STA - Liatest® D-Di | Positive | Negative | Total | NPV | 100.0% (LCL: 98.7%)  |
|  Positive | 63 | 256 | 319 | PPV | 19.7% (LCL: 15.5%)  |
|  Negative | 0 | 292 | 292 | Sensitivity | 100.0% (LCL: 94.3%)  |
|  Total | 63 | 548 | 611 | Specificity | 53.3% (LCL: 49.0%)  |

# 4. Clinical cut-off:

The clinical cut-off was established at  $0.50~\mu \mathrm{g / mL}$  (FEU).

# 5. Expected values/Reference range:

In premarket notification K964728, the adult reference range for D-dimer using STA - Liatest  $\text{®}$  D-Di was established as  $&lt; 0.50~\mu \mathrm{g / mL}$  (FEU).

# N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{5}

6

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K162227](https://fda-staging.innolitics.com/device/K162227)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
