ACUSON SC2000 Diagnostic Ultrasound System

K162221 · Siemens Medical Solutions Inc. Ultrasound Group · IYN · Sep 6, 2016 · Radiology

Device Facts

Record IDK162221
Device NameACUSON SC2000 Diagnostic Ultrasound System
ApplicantSiemens Medical Solutions Inc. Ultrasound Group
Product CodeIYN · Radiology
Decision DateSep 6, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac. Neo-natal and Fetal Cardiac. Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system also supports catheters which are intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Vascular Imaging Applications and Analysis The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck: superficial and arteries in the arms and legs and abdomen: and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Superficial Imaging Applications The system transmits ultrasound energy into varis of the body of adult patients creating 2D (B). Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. Intraoperative Imaging Applications The system transmits ultrasound energy into varis of the body of adult patients creating 2D (B), Color Dopbler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures. Transcranial Imaging Applications The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Device Story

The Acuson SC2000 is a mobile, software-controlled diagnostic ultrasound system. It transmits ultrasound energy into the body to acquire primary or secondary harmonic echo data. The system supports various imaging modes, including B-mode, M-mode, Pulsed/Continuous Wave Doppler, Color Doppler, Color Power Doppler, Color Velocity Imaging, and 3D/4D imaging. It can synchronize image capture with patient ECG. The device is operated by clinicians in clinical settings to visualize anatomy and blood flow, and to perform measurements and calculations. These outputs are used adjunctively with other medical data to evaluate pathology and guide intraoperative procedures. The system supports intra-cardiac and intraluminal catheters for visualization of cardiac/great vessel anatomy and other devices. Benefits include non-invasive diagnostic imaging and procedural guidance.

Clinical Evidence

No clinical data required; substantial equivalence established via bench testing and technological comparison.

Technological Characteristics

Multi-purpose mobile ultrasound system. Transducer types: Matrix, Phased, Linear, Curved arrays. Modes: B, M, PWD, CWD, Color Doppler, Power Doppler, Color Velocity, 3D/4D. Connectivity: Wireless, DICOM. Standards: AAMI/ANSI 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62366, IEC 62304, IEC 60601-2-18, NEMA UD-2, NEMA UD-3, ISO 10993-1.

Indications for Use

Indicated for adult, pediatric, neonatal, and fetal patients requiring diagnostic ultrasound imaging of cardiac, vascular, musculoskeletal, abdominal, and cranial structures, including intraoperative and intra-cardiac/intraluminal visualization.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three overlapping shapes. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 6, 2016 SIEMENS MEDICAL SOLUTIONS INC. ULTRASOUND GROUP % Mr. MARK JOB RESPONSIBLE THIRD PARTY OFFICIAL 1394 25TH STREET, NW BUFFALO MN 55313 Re: K162221 Trade/Device Name: Acuson SC2000 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, OBJ Dated: August 4, 2016 Received: August 8, 2016 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert Oaks Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162221 ### Device Name Acuson SC2000TM Diagnostic Ultrasound System ### Indications for Use (Describe) The SC2000 ultrasound imaging system is intended for the following applications: Cardiac. Neo-natal and Fetal Cardiac. Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system also supports catheters which are intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Vascular Imaging Applications and Analysis The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck: superficial and arteries in the arms and legs and abdomen: and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Superficial Imaging Applications The system transmits ultrasound energy into varis of the body of adult patients creating 2D (B). Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. Intraoperative Imaging Applications The system transmits ultrasound energy into varis of the body of adult patients creating 2D (B), Color Dopbler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures. Transcranial Imaging Applications The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. {3}------------------------------------------------ Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ 510(k) Number (if known): #### Device Name: SC2000 Diagnostic Ultrasound System Ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------|------------------------------| | Ophthalmic | | | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | P* | P | | P | | Abdominal | P | P | P | P | P | P | P | | P* | P | P | | | Intraoperative<br>Abdominal | P | P | P | P | P | P | P | | P* | P | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P* | P | P | P | | Small Organ<br>(specify) | | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | P | P | | P* | P | | | | Cardiac | P | P | P | P | P | P | P | | P* | P | P | P | | Trans-esophageal | P | P | P | P | P | P | | | P* | | P | | | Transrectal | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | | Intra-Luminal | P | P | P | P | P | P | P | | P* | | | P | | Peripheral Vessel | P | P | P | P | P | P | P | P | P* | P | | | | Laparoscopic | | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | P | P | P | P | P | P | P | P | P* | P | | | | Musculo-skeletal<br>Superficial | P | P | P | P | P | P | P | P | P* | P | | | | Other (Neonatal<br>Cardiac) | P | P | P | P | P | P | P | | P* | P | | | | Other (Intra-<br>Cardiac) | P | P | P | P | P | P | P | | P* | | | P | | Other<br>(Abdominal<br>Vascular) | P | P | P | P | P | P | P | | P* | P | | | N=new indication. P = Previously Cleared in 510(k) K142628 ### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {5}------------------------------------------------ 510(k) Number (if known): CW2 Device Name: Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude) | Color<br>Velocity | Combined<br>(Specify) | Other:<br>Harmonic | |----------------------------------|---|---|---|-----|-----|------------------|----------------------|-------------------|-----------------------|--------------------| | | | | | | | | Doppler | Imaging | | Imaging | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | P | | | | | | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | P | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | | | | | P | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N=new indication. Previously Cleared in 510(k) K142628 #### Additional Comments: (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {6}------------------------------------------------ ### 510(k) Number (if known): Device Name: 4Z1c Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|--| | Ophthalmic | | | | | | | | | | | | | | Fetal | | P | P | P | P | P | | | P* | P | P | | | Abdominal | | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | | Pediatric | | P | P | P | P | P | | | P* | P | P | | | Small Organ<br>(specify) ** | | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | | | P* | P | P | | | Trans-esophageal | | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | | Other (specify) | | | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K 142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {7}------------------------------------------------ 510(k) Number (if known): Device Name: 4V1c Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | P * | P | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P * | P | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | P | P | | P * | P | | Cardiac | P | P | P | P | P | P | P | | P * | P | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (Neonatal<br>Cardiac) | P | P | P | P | P | P | P | | P * | P | N=new indication. Previously Cleared in 510(k) K142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {8}------------------------------------------------ ### 510(k) Number (if known): Device Name: V5Ms Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------|--| | Ophthalmic | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | | | P* | | P | | | Small Organ<br>(specify) | | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | | | P* | | P | | | Trans-esophageal | P | P | P | P | P | P | | | P* | | P | | | Transrectal | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | | Other (specify) | | | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K 142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {9}------------------------------------------------ ### 510(k) Number (if known): Device Name: ୧୯.IHD Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | P* | P | | Abdominal | P | P | P | P | P | P | P | | P* | P | | Intraoperative<br>Abdominal | P | P | P | P | P | P | P | | P* | P | | Intraoperative | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P* | P | | Small Organ<br>(specify) | P | P | P | P | P | P | P | | P* | P | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | P | P | P | P | P | P | P | | P* | P | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other<br>(Abdominal<br>Vascular) | P | P | P | P | P | P | P | | P* | P | N=new indication. P = Previously Cleared in 510(k) K142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {10}------------------------------------------------ ### 510(k) Number (if known): Device Name: ### Z6Ms Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ<br>(specify) | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | | | P* | P | P | | Trans-esophageal | P | P | P | P | P | P | | | P* | P | P | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (specify) | | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K 142628 ### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {11}------------------------------------------------ ### 510(k) Number (if known): Device Name: V7M Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | P | P | P | P | P | P | P | | P* | P | P | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P* | | P | | Small Organ<br>(specify) | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | P | | P* | | P | | Trans-esophageal | | P | P | P | P | P | P | | P* | | P | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (specify) | | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {12}------------------------------------------------ ### 510(k) Number (if known): Device Name: 8V3 Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | P * | P | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P * | P | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | P | | P * | P | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (Neonatal<br>Cardiac) | P | P | P | P | P | P | P | | P * | P | N=new indication. Previously Cleared in 510(k) K142628 ### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {13}------------------------------------------------ 510(k) Number (if known): Device Name: 9L4 Indications for Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | P | P | P | P | | P | P | P | P* | P | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | P | P | P | P | | P | P | P | P* | P | | Musculo-skeletal<br>Superficial | P | P | P | P | | P | P | P | P* | P | | Other (specify) | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {14}------------------------------------------------ ### 510(k) Number (if known): Device Name: 10V4 Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | P * | P | | Abdominal | P | P | P | P | P | P | P | | P * | P | | Intraoperative<br>Abdominal | P | P | P | P | P | P | P | | P * | P | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P * | P | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | P | | P * | P | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (Neonatal<br>Cardiac) | P | P | P | P | P | P | P | | P * | P | N=new indication. Previously Cleared in 510(k) K142628 ### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {15}------------------------------------------------ 510(k) Number (if known): Device Name: AcuNav 8F and 10F Ultrasound Catheter Intended Use: Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | | P | P | P | P | P | P | | P* | | | | Small Organ<br>(specify) ** | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | P | | P* | | | | Trans-esophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | | P | P | P | P | P | P | | P* | | | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (Intra-<br>Cardiac) | | P | P | P | P | P | P | | P* | | | N=new indication. P = Previously Cleared in 510(k) K142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {16}------------------------------------------------ 510(k) Number (if known): Device Name: Intended Use: ACUSON AcuNav™ V 10F Ultrasound Catheter Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P* | | P | | Small Organ<br>(specify) ** | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | P | | P* | | P | | Trans-esophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | P | P | P | P | P | P | P | | P* | | P | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (Intra-<br>Cardiac) | P | P | P | P | P | P | P | | P* | | P | N=new indication. P = Previously Cleared in 510(k) K 142628 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED) {17}------------------------------------------------ 510(k) Number (if known): Device Name: SoundStar 10F Ultrasound Catheter Intended Use: Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------| | Ophthalmic | | | | | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...