CADD® Infusion Disposables Portfolio with NRFit™ connectors
K162219 · Smiths Medical Asd, Inc. · PWH · Jul 20, 2017 · General Hospital
Device Facts
Record ID
K162219
Device Name
CADD® Infusion Disposables Portfolio with NRFit™ connectors
Applicant
Smiths Medical Asd, Inc.
Product Code
PWH · General Hospital
Decision Date
Jul 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
CADD Yellow Medication Cassette Reservoirs with NRFit™ connectors are designed for use only with NRFit™ components for delivery of regional anesthetics or narcotics, and are designed for use with CADD pumps (see CADD pump Operator's Manual for compatibility) and CADD Yellow Extension Sets with NRFit™ connectors. CADD Filling Adapter with male NRFit™ connector and female Luer is an accessory for use in filling of the CADD Medication Cassette Reservoirs with NRFit™ Connector. CADD Yellow Extension Sets with NRFit™ connectors are designed for use only with CADD Yellow Medication Cassette Reservoirs with NRFit™ connectors for the delivery of regional anesthetics or narcotics. CADD Yellow Administration Sets with NRFit™ connectors are designed for use only with NRFit™ components for delivery of regional anesthetics or narcotics and are designed for use with CADD pumps (see CADD pump Operator's Manual for compatibility). Male Yellow Cap with NRFit™ connector is intended for use as an accessory with the CADD Yellow Medication Cassette reservoirs. Female Yellow Cap with NRFit™ connector is intended for use as an accessory with the CADD Yellow Extension sets with NRFit™ connectors and the CADD Yellow Administration sets with NRFit™ connector. The CADD Infusion Adapter with male NRFit™ connector and female Luer is intended to be used by clinicians to convert administration sets with Luer connectors into permanently dedicated lines for use only with NRFit connectors. The NRFit™ connector with non-luer taper is intended for the injection of regional anesthetics or narcotics and may help reduce the risk of mis-connection or mis-injection.
Device Story
Portfolio of infusion disposables (cassettes, extension sets, administration sets, adapters, caps) for CADD pumps; utilizes ISO 80369-6 compliant NRFit™ connectors to prevent misconnection with standard Luer devices; color-coded yellow for neuraxial/regional anesthetic identification. Used by clinicians in clinical settings to deliver regional anesthetics or narcotics. Adapters convert existing Luer-based lines to dedicated NRFit™ lines. Devices ensure secure, leak-resistant fluid delivery; reduce risk of medication mis-injection; maintain compatibility with CADD pump systems.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 80369-6 (neuraxial connectors) and ISO 594-2 (Luer fittings) for functional integrity, including resistance to overriding, separation, and leakage. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity). Sterile barrier integrity validated per ISO 11607.
Indicated for pediatric and adult patients requiring delivery of regional anesthetics or narcotics via neuraxial or regional anesthetic infusion applications.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Pharmacia Deltec, Extension Set with Anti-Siphon Valve (K942046)
Extension Set with Microbore Tubing and Filter (K974013)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 20, 2017
Smiths Medical ASD, Inc. Michael Johnson Sr. Manager, Regulatory Affairs 6000 Nathan Lane North Minneapolis, Minnesota 55442
Re: K162219
Trade/Device Name: CADD® Infusion Disposables Portfolio with NRFit™ connectors Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: FPA, LHI Dated: June 15, 2017 Received: June 19, 2017
Dear Mr. Michael Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Tara A. Ryan -S
for
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Change Control Table, Change History
# Change Control Table
| Version | Document Author | Document Approver | Date Approved |
|---------|---------------------|---------------------|---------------|
| 1.00 | Name, Title, Office | Name, Title, Office | MM/DD/YYYY |
Complete Change Control Table (all versions) retained in SWIFT Docs.
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162219
Device Name
CADD® Infusion Disposables Portfolio with NRFit™ connector
Indications for Use (Describe)
CADD Yellow Medication Cassette Reservoirs with NRFit™ connector
CADD Yellow Medication Cassette Reservoirs with NRFit™ connectors are designed for use only with NRFit™ components for delivery of regional anesthetics or narcotics, and are designed for use with CADD pumps (see CADD pump Operator's Manual for compatibility) and CADD Yellow Extension Sets with NRFit™ connectors.
CADD Filling Adapter with male NRFit™ connector and female Luer
The CADD Filling Adapter with male NRFit™ connector and female Luer is an accessory for use in filling of the CADD Medication Cassette Reservoirs with NRFit™ Connector.
CADD Yellow Extension Sets with NRFit™ connector
CADD Yellow Extension Sets with NRFit™ connectors are designed for use only with CADD Yellow Medication Cassette Reservoirs with NRFit™ connectors for the delivery of regional anesthetics or narcotics.
CADD Yellow Administration Sets with NRFit™ connector
CADD Yellow Administration Sets with NRFit™ connectors are designed for use only with NRFit™ components for delivery of regional anesthetics or narcotics and are designed for use with CADD pumps (see CADD pump Operator's Manual for compatibility).
Male Yellow Cap with NRFit™ connector
The Male Yellow Cap with NRFit™ connector is intended for use as an accessory with the CADD Yellow Medication Cassette reservoirs.
Female Yellow Cap with NRFit™ connector
The Female Yellow Cap with NRFit™ connector is intended for use as an accessory with the CADD Yellow Extension sets with NRFit™ connectors and the CADD Yellow Administration sets with NRFit™ connector.
CADD Infusion Adapter with male NRFit™ connector and female Luer
The CADD Infusion Adapter with male NRFit™ connector and female to be used by clinicians to convert administration sets with Luer connectors into permanently dedicated lines for use only with NRFit connectors. The NRFit™ connector with non-luer taper is intended for the injection of regional anesthetics or narcotics and may help reduce the risk of mis-connection or mis-injection.
The Intended population is pediatrics and adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 1. ADMINISTRATIVE INFORMATION
| 510(k) | K162219 |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Applicant's Name<br>and Address | Smiths Medical ASD, Inc.<br>6000 Nathan Lane North<br>Minneapolis, MN 55442 USA |
| Contact Person | Michael R. Johnson<br>Sr. Regulatory Specialist<br>Phone: 763-383-3341<br>Fax: 763-383-3679<br>Email: mike.johnson@smiths-medical.com |
| Date | July 15, 2017 |
| Regulation No. | 21 CFR 880.5440 |
| Regulation Name | Intravascular administration set |
| Primary Product<br>Code | FPA |
| Secondary Product<br>Code | LHI |
| Trade Name | CADD® Infusion Disposables Portfolio with NRFit™ connectors |
## 2. REASON FOR SUBMISSION
The purpose of this submission is to make a modification to the currently marketed Smiths Medical CADD Yellow Infusion components which are being updated to include an ISO 80369-6 compliant connector for neuraxial applications.
| | <b>Subject Device</b> | <b>Predicate Device</b> |
|------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | CADD Yellow Medication Cassette<br>Reservoir with NRFitTM connector | CADD Medication Cassette Reservoir |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 |
| Regulation Name | Intravascular administration set | Intravascular administration set |
| Regulatory Class | II | II |
| Product Code | FPA | FPA |
| 510(k) | K162219 | K040636 (100ml), K081156 (250ml) |
| | <b>Subject Device</b> | <b>Predicate Device</b> |
| Trade Name | CADD Filling Adapter with NRFitTM<br>connector | CorrectInjectTM Safety System<br>Infusion Adapter Component: The<br>Infusion Adapter is a component of the<br>CorrectInject Safety System which can<br>be considered a transfer device or<br>accessory to a syringe. |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 868.5140 |
| Regulation Name | Intravascular administration set | Anesthesia Conduction Kit |
| Regulatory Class | II | II |
| Product Code | LHI | CAZ |
| 510(k) | K162219 | K110053 |
| | Subject Device | Predicate Device |
| Trade Name | CADD Infusion Adapter with male<br>NRFitTM connector and female Luer | CADD Administration Set with Flow<br>Stop, Medication cassette reservoir<br>with flow stop and extension set |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 |
| Regulation Name | Intravascular administration set | Intravascular administration set |
| Regulatory Class | II | II |
| Product Code | FPA | FPA |
| 510(k) | K162219 | K040636 |
| Trade Name | CADD Yellow Extension Set with<br>NRFitTM connectors | Pharmacia Deltec, Extension Set with<br>Anti-Siphon Valve |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 |
| Regulation Name | Intravascular administration set | Intravascular administration set |
| Regulatory Class | II | II |
| Product Code | FPA | FPA |
| 510(k) | K162219 | K942046 |
| Trade Name | CADD Yellow Extension Set with<br>NRFitTM connectors and Air-<br>Eliminating Filter | Extension Set with Microbore Tubing<br>and Filter |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 |
| Regulation Name | Intravascular administration set | Intravascular administration set |
| Regulatory Class | II | II |
| Product Code | FPA | FPA |
| 510(k) | K162219 | K974013 |
| Trade Name | CADD Yellow Administration Sets<br>with NRFitTM connector | CADD Administration Set |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 |
| Regulation Name | Intravascular administration set | Intravascular administration set |
| Regulatory Class | II | II |
| Product Code | FPA | FPA |
| 510(k) | K162219 | K040636 |
| Trade Name | Yellow Cap with NRFitTM connector | CADD Medication Cassette Reservoir |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 |
| Regulation Name | Intravascular administration set | Intravascular administration set |
| Regulatory Class | II | II |
| Product Code | FPA | FPA |
| 510(k) | K162219 | K040636 |
# 3. DEVICE INFORMATION
{6}------------------------------------------------
smiths medical
bringing technology to life
## 4. DEVICE DESCRIPTION
The CADD® Infusion devices with NRFit™ connectors are part of the CADD System. The CADD System is defined as a CADD pump with an attached CADD Medication Cassette Reservoir with an integral free-flow protection feature and CADD Extension Set with an Integral Anti-Siphon Valve; or as a CADD pump with an attached CADD Administration Set with an integral free-flow protection feature. The CADD Administration Set will be connected to a medication bag to allow for infusion.
The devices are designed to deliver local or regional anesthetics, narcotics indicated for neuraxial or regional anesthetic infusion applications. The NRFit™ connectors conform to ISO 80369-6, Small
{7}------------------------------------------------
bore connectors for liquids and gases in healthcare applications -- Part 6: Connectors for neuraxial applications. The connectors are not compatible with standard luer connectors which is intended to reduce the risk of misconnection that may result in the infusion of medications not intended for neuraxial or regional anesthetic use.
The CADD Infusion devices with NRFit™ connectors are color-coded yellow to indicate medication intended for neuraxial or regional anesthetic delivery.
This 510(k) includes various configurations of CADD Infusion devices with NRFit ™ connectors. A description of each configuration is provided in the table below.
| Configuration | Description |
|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CADD Yellow Medication Cassette<br>Reservoirs with NRFit™ connector | Self-contained, proprietary medication cassette reservoirs<br>for use only with NRFit ™ components for delivery of<br>regional anesthetics or narcotics, and are designed for use<br>with CADD pumps. Available in 100ml or 250ml. |
| CADD Filling Adapter with male<br>NRFit™ connector and female Luer | An accessory used to fill the CADD Medication Cassette<br>Reservoirs with NRFit ™ connectors. |
| CADD Infusion Adapter with male<br>NRFit™ connector and female<br>Luer | Used to convert administration sets or extension sets with<br>Luer connectors into permanently dedicated line for use only<br>with NRFit ™ connectors. |
| CADD Yellow Extension Sets with<br>NRFit™ connector | Used only with CADD Yellow Medication Cassette<br>Reservoirs with NRFit ™ connectors for the delivery of<br>regional anesthetics or narcotics. Used to extend the distance<br>between pump and patient. The extension set is configured<br>with or without an air-eliminating filter. |
| CADD Yellow Administration Sets<br>with NRFit™ connector | Dedicated administration set for use only with NRFit ™<br>components for delivery of regional anesthetics or narcotics,<br>and are designed for use with CADD pumps. The Yellow<br>Administration sets are configured with or without an air-<br>eliminating filter. |
| Male Yellow Cap with NRFit™<br>connector/<br>Female Yellow Cap with NRFit™<br>connector | The cap has an NRFit ™ thread design and is sold<br>individually as an accessory that can be used to cap the<br>Medication Cassette or Administration Set. The cap is<br>available as a male yellow cap intended to fit the female<br>NRFit ™ side or female yellow cap intended to fit the male<br>NRFit™ side. |
# 5. INDICATIONS FOR USE
| Configuration | Indication for Use |
|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CADD Yellow Medication Cassette<br>Reservoirs with NRFitTM connector | CADD Yellow Medication Cassette Reservoirs with NRFitTM connectors are designed for use only with NRFitTM components<br>for delivery of regional anesthetics or narcotics, and are designed<br>for use with CADD pumps (see CADD pump Operator's Manual<br>for compatibility) and CADD Yellow Extension Sets with<br>NRFitTM connectors. |
| CADD Filling Adapter with male<br>NRFitTM connector and female Luer | CADD Filling Adapter with male NRFitTM connector and female<br>Luer is an accessory for use in filling of the CADD Medication<br>Cassette Reservoirs with NRFitTM connector. |
{8}------------------------------------------------
bringing technology to life
| Configuration | Indication for Use |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CADD Infusion Adapter with male NRFitTM connector and female Luer | The CADD Infusion Adapter with male NRFitTM connector and female Luer is intended to be used by clinicians to convert administration sets with Luer connectors into permanently dedicated lines for use only with NRFitTM connectors. The NRFitTM connector with non-Luer taper is intended for the injection or infusion of regional anesthetics or narcotics and may help reduce the risk of mis-connection or mis-injection. |
| CADD Yellow Extension Sets with NRFitTM connector | CADD Yellow Extension Sets with NRFitTM connectors are designed for use only with CADD Yellow Medication Cassette Reservoirs with NRFitTM connectors for the delivery of regional anesthetics or narcotics. |
| CADD Yellow Administration Sets with NRFitTM connector | CADD Yellow Administration Sets with NRFitTM connectors are designed for use only with NRFitTM components for delivery of regional anesthetics or narcotics, and are designed for use with CADD® pumps (see CADD pump Operator's Manual for compatibility). |
| Male Yellow Cap with NRFitTM connector/<br>Female Yellow Cap with NRFitTM connector | The Male Yellow Cap with NRFitTM connector is intended for use as an accessory with the CADD Yellow Medication Reservoirs.<br>The Female Yellow Cap with NRFitTM connector is intended for use as an accessory with the CADD Yellow Extension sets with NRFitTM connectors and the CADD Yellow Administration sets with NRFitTM connector. |
# 6. SUBSTANTIAL EQUIVALENCE DISCUSSION
The Smiths Medical CADD Infusion Disposables Portfolio with NRFit™ connectors have the same technological characteristics as the predicate devices with the exception of the NRFit ™ Connectors.
The Smiths Medical CADD Infusion Disposables Portfolio with NRFit™ connectors and predicate devices are both designed to deliver local or regional anesthetics, narcotics indicated for neuraxial or regional anesthetic infusion applications. They are both made of the same materials, have the same chemical composition, and have the same design features excluding the NRFit ™ connector design.
A comparative analysis is provided in the following Tables;
| Characteristic | Predicate (K040636) | Subject (K162219) | Discussion |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indication for<br>Use | The CADD Medication<br>Cassette Reservoir with<br>Flow Stop is designed for<br>use with the CADD pumps<br>(except CADD-Micro and<br>CADD-TPN) for delivery<br>of medications and fluids.<br>The CADD administration<br>Set with Flow Stop is<br>designed for use with the<br>CADD pumps (except<br>CADD-Micro and CADD-<br>TPN) for delivery of<br>medications and fluids. | CADD Yellow Medication<br>Cassette Reservoirs with<br>NRFitTM connectors are<br>designed for use only with<br>NRFitTM components for<br>delivery of regional<br>anesthetics or narcotics,<br>and are designed for use<br>with CADD pumps (see<br>CADD pump Operator's<br>Manual for compatibility)<br>and CADD Yellow<br>Extension Sets with<br>NRFitTM connectors. | Similar |
| Extension Set<br>Length | 8 inches | Same | N/A |
| Priming Volume | 0.2 mL | Same | N/A |
Table 1: CADD Yellow Medication Cassette Reservoir, 100mL with NRFit™ connector
{9}------------------------------------------------
# smiths medical
bringing technology to life
| Characteristic | Predicate (K040636) | Subject (K162219) | Discussion | |
|----------------|-----------------------------------------------|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Connector | ISO 594 Luer | ISO 80369-6 NRFit™ | ISO 80369-6 NRFit™ connector intended to reduce risk of misconnections. | |
| Materials | Female Connector;<br>Polyvinyl chloride (PVC) | Female Connector;<br>Polycarbonate | Subject device materials meet functional and biocompatibility requirements. Yellow indicates neuraxial or regional anesthetic application. | |
| | Male Connector;<br>Polypropylene | Male Connector;<br>Polybutylene terephthalate (PBT) | | |
| | Housing;<br>Polypropylene | Housing;<br>Polypropylene | | |
| | Packaging | Tyvek Pouch | Same | N/A |
| | Sterility | Sterile, EO | Same | N/A |
| | Use | Single Use Disposable | Same | N/A |
# Table 2: CADD Yellow Medication Cassette Reservoir, 250mL with NRFit™ connector
| Characteristic | Predicate (K081156) | Subject (K162219) | Discussion | | | |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------|-----|
| Indication for<br>Use | The CADD Medication<br>Cassette reservoirs are<br>intended for the delivery of<br>medications and fluids for<br>subcutaneous,<br>intramuscular, intravenous,<br>intra-arterial,<br>intraperitoneal, or<br>intraspinal infusion. | CADD Yellow Medication<br>Cassette Reservoirs with<br>NRFitTM connectors are<br>designed for use only with<br>NRFitTM components for<br>delivery of regional<br>anesthetics or narcotics,<br>and are designed for use<br>with CADD pumps (see<br>CADD pump Operator's<br>Manual for compatibility)<br>and CADD Yellow<br>Extension Sets with<br>NRFitTM connectors. | Similar | | | |
| | Extension Set<br>Length | 8 inches | Same | N/A | | |
| | Priming Volume | 0.2 mL | Same | N/A | | |
| | Connector | ISO 594 Luer | ISO 80369-6 NRFitTM | ISO 80369-6 NRFitTM<br>connector intended to<br>reduce risk of<br>misconnections | | |
| | Materials | Female Connector;<br>Polyvinyl chloride (PVC) | Female Connector;<br>Polycarbonate | Subject device materials<br>meet functional and<br>biocompatibility<br>requirements. Yellow<br>indicates neuraxial or<br>regional anesthetic<br>application. | | |
| | | Male Connector;<br>Polypropylene | Male Connector;<br>Polybutylene terephthalate<br>(PBT) | | | |
| | | Housing;<br>Polypropylene | Housing;<br>Polypropylene | | | |
| | | Packaging | Tyvek Pouch | | Same | N/A |
| | | Sterility | Sterile, EO | | Same | N/A |
| Use | | Single Use Disposable | Same | | N/A | |
{10}------------------------------------------------
# al
bringing technology to life
| Characteristic | Predicate (K110053) | Subject (K162219) | Discussion |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | The CorrectInjectTM Safety<br>System is intended for the<br>injection of local or<br>regional anesthetics,<br>narcotics or other<br>medications indicated for<br>neuraxial injection. The<br>system consists of<br>components that have a<br>unique non-Luer taper that<br>allows connection of<br>compatible CorrectInjectTM<br>components that, when<br>used together as a system,<br>help reduce the risk of mis-<br>connection that may result<br>in the injection of<br>medications not intended<br>for neuraxial use. | CADD Filling Adapter<br>with male NRFit TM<br>connector and female Luer<br>is an accessory for use in<br>filling of the CADD<br>Medication Cassette<br>Reservoirs with NRFit TM<br>connector. | Similar. The predicate<br>Infusion Adapter is a<br>component of the<br>CorrectInject Safety<br>System which can be<br>considered a transfer<br>device or accessory to a<br>syringe. The subject<br>Filling Adapter is an<br>accessory used to fill the<br>CADD Medical Cassette<br>Reservoir and is<br>considered a transfer<br>device. |
| Connector | ISO 594 Luer | ISO 80369-6 NRFitTM | ISO 80369-6 NRFitTM<br>connector intended to<br>reduce risk of<br>misconnections. |
| Materials | Polybutylene terephthalate<br>(PBT) | Polycarbonate - Bayer<br>Polycarbonate Makrolon;<br>Acrylonitrile butadiene<br>styrene (ABS) – Terlux<br>Colorant: Orange, Clariant<br>and Clariant | Subject device materials<br>meet functional and<br>biocompatibility<br>requirements. Yellow<br>indicates neuraxial or<br>regional anesthetic<br>application. |
| Packaging | Tyvek Pouch | Same | N/A |
| Sterility | Sterile, EO | Same | N/A |
| Use | Single Use Disposable | Same | N/A |
#### Table 3: CADD Filling Adapter with male NRFit ™ connector and female Luer
# Table 4: CADD Infusion Adapter with male NRFit ™ connector and female Luer
| Characteristic | Predicate (K040636) | Subject (K162219) | Discussion |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indication for<br>Use | The CADD Medication<br>Cassette Reservoir with<br>Flow Stop is designed for<br>use with the CADD pumps<br>(except CADD-Micro and<br>CADD-TPN) for delivery<br>of medications and fluids.<br>The CADD administration<br>Set with Flow Stop is<br>designed for use with the<br>CADD pumps (except<br>CADD-Micro and CADD-<br>TPN) for delivery of<br>medications and fluids. | The CADD Infusion<br>Adapter with male NRFitTM<br>connector and female Luer<br>is intended to be used by<br>clinicians to convert<br>administration sets with<br>Luer connectors into<br>permanently dedicated lines<br>for use only with NRFitTM<br>connectors. The NRFitTM<br>connector with non-Luer<br>taper is intended for the<br>injection or infusion of<br>regional anesthetics or<br>narcotics and may help<br>reduce the risk of<br>misconnection or<br>misinjection. | Similar |
{11}------------------------------------------------
# smiths medical
bringing technology to life
| Characteristic | Predicate (K040636) | Subject (K162219) | Discussion |
|----------------|----------------------------------|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Connector | ISO 594 Luer | ISO 80369-6 NRFitTM | ISO 80369-6 NRFitTM connector intended to reduce risk of misconnections. |
| Materials | Polybutylene terephthalate (PBT) | Polybutylene Terephthalate (PBT)<br>Arnite;<br>Polypropylene<br>Polycarbonate (PC) | Subject device materials meet functional and biocompatibility requirements. Yellow indicates neuraxial or regional anesthetic application. |
| Packaging | Tyvek Pouch | Form Fill Seal | Form Fill Seal packaging is an appropriate alternative to the pouch packaging configuration. |
| Sterility | Sterile, EO | Same | N/A |
| Use | Single Use Disposable | Same | N/A |
#### Table 5: CADD Yellow Extension Set with NRFit ™ connectors
| Characteristic | Predicate (K942046) | Subject (K162219) | Discussion |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | The Extension Set with<br>Anti-Siphon Valve is an<br>accessory for use with<br>administration sets that are<br>used in conjunction with<br>electromechanical infusion<br>pumps that do not have<br>anti-siphon valves. It is<br>intended to protect against<br>unregulated gravity<br>infusion ("free-flow") from<br>improperly attached<br>administration sets. | CADD Yellow Extension<br>Sets with NRFit ™<br>connectors are designed for<br>use only with CADD<br>Yellow Medication<br>Cassette Reservoirs with<br>NRFit ™ connectors for the<br>delivery of regional<br>anesthetics or narcotics. | Similar |
| Set Length | 90 inches | Same | N/A |
| Priming Volume | 2.4 mL | Same | N/A |
| Connector | ISO 594 Luer | ISO 80369-6 NRFit™ | ISO 80369-6 NRFit ™<br>connector intended to<br>reduce risk of<br>misconnections. |
| Materials | Connectors;<br>Polycarbonate<br>Anti-Siphon Valve;<br>Silicone-Elastosil | Connectors;<br>Polycarbonate<br>Anti-Siphon Valve;<br>Silicone-Elastosil | Subject device materials<br>meet functional and<br>biocompatibility<br>requirements. Yellow<br>indicates neuraxial or<br>regional anesthetic<br>application. |
| Packaging | Tyvek Pouch | Same | N/A |
| Sterility | Sterile, EO | Same | N/A |
| Use | Single Use Disposable | Same | N/A |
{12}------------------------------------------------
| Characteristic | Predicate (K974013) | Subject (K162219) | Discussion |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Extension Set with microbore tubing attaches<br>to the Micro Medication Reservoir for use with the<br>CADD-Micropump. The Extension Set with Anti-<br>Siphon Valve must be used with the Medication<br>Cassette reservoir to protect against unregulated gravity<br>infusion that can result from an improperly<br>attached reservoir. | CADD Yellow Extension<br>Sets with NRFitTM<br>connectors are designed for<br>use only with CADD<br>Yellow Medication<br>Cassette Reservoirs with<br>NRFitTM connectors for the<br>delivery of regional<br>anesthetics or narcotics. | Similar |
| Set Length | 90 inches | Same | N/A |
| Priming Volume | 2.4 mL | Same | N/A |
| Connector | ISO 594 Luer | ISO 80369-6 NRFitTM | ISO 80369-6 NRFitTM<br>connector intended to<br>reduce risk of<br>misconnections. |
| Materials | Female Connector;<br>Polyvinyl chloride (PVC)<br>Male Connector;<br>Acrylonitrile butadiene<br>styrene (ABS)<br>Housing and Valve;<br>Polycarbonate, Valve<br>Membrane; Elastosil | Female Connector;<br>Polycarbonate<br>Male Connector;<br>Acrylonitrile butadiene<br>styrene (ABS)<br>Housing and Valve;<br>Polycarbonate,<br>Valve Membrane; Elastosil | Subject device materials<br>meet functional and<br>biocompatibility<br>requirements. Yellow<br>indicates neuraxial or<br>regional anesthetic<br>application. |
| Packaging | Tyvek Pouch | Same | N/A |
| Sterility | Sterile, EO | Same | N/A |
| Use | Single Use Disposable…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.