MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System
K162194 · Medicrea International SA · KWQ · Dec 8, 2016 · Orthopedic
Device Facts
Record ID
K162194
Device Name
MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System
Applicant
Medicrea International SA
Product Code
KWQ · Orthopedic
Decision Date
Dec 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MEDICREA INTERNATIONAL S.A. Anterior Cervical Locking Plate System is intended for anterior interbody screw/ plate fixation of the cervical spine until the bone fusion occurs. The MEDICREA INTERNATIONAL S.A. Anterior Cervical Locking Plate System is indicated for use in the temporary stabilization of the anterior cervical (C2 to T1) spine during development of a solid fusion in patients with the following conditions: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion This implant has to be used with an appropriate bone graft. Patients should have been unresponsive for a minimum six (6) weeks to conservative treatment before the surgery.
Device Story
Anterior cervical locking plate system; provides temporary stabilization/immobilization of cervical spine (C2-T1) to facilitate bone fusion. System comprises titanium alloy plates and bone screws; utilizes mechanical locking mechanism (rotating part) to prevent screw back-out. Used by surgeons in clinical/OR settings; implanted during spinal fusion procedures. Provides mechanical resistance to stabilize spinal segments; supports fusion process in patients with degenerative disc disease, trauma, or deformities. Benefits patient by maintaining alignment and stability during healing.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including static and dynamic compression and static torsion tests per ASTM F1717-15, and pyrogenicity testing.
Technological Characteristics
Titanium alloy (Ti-6Al-4V ELI) per ASTM F136 and ISO 5832-3. System includes cervical plates (1-4 levels) and bone screws (4mm/4.5mm, fixed/variable angles). Mechanical locking mechanism via rotating part. Standalone implant system.
Indications for Use
Indicated for patients with cervical spine conditions (C2-T1) including DDD, spondylolisthesis, trauma, spinal stenosis, deformities, tumor, pseudoarthrosis, or failed fusion. Requires minimum 6 weeks of failed conservative treatment. Must be used with bone graft.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2016
Medicrea International S.A. David Ryan VP Product Development and Marketing 14 Porte Du Grand Lyon Neyron, 01700 France
Re: K162194
Trade/Device Name: MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: November 10, 2016 Received: November 14, 2016
Dear Mr. Ryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Vincent J. Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162194
Device Name
MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System
Indications for Use (Describe)
MEDICREA INTERNATIONAL S.A. Anterior Cervical Locking Plate System is intended for anterior interbody screw/ plate fixation of the cervical spine until the bone fusion occurs.
The MEDICREA INTERNATIONAL S.A. Anterior Cervical Locking Plate System is indicated for use in the temporary stabilization of the anterior cervical (C2 to T1) spine during development of a solid fusion in patients with the following conditions:
· degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies),
- · spondylolisthesis,
- · trauma (i.e., fracture or dislocation),
- · spinal stenosis,
- · deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- · tumor,
- · pseudoarthrosis,
- · and failed previous fusion
This implant has to be used with an appropriate bone graft.
Patients should have been unresponsive for a minimum six (6) weeks to conservative treatment before the surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
### MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System
In accordance with 21 CFR 807.92 of the Federal Code of Requlations the following 510(k) summary is submitted for MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System:
### Date Prepared: 02 August 2016
### 1. Submitter:
### Contact Person:
MEDICREA INTERNATIONAL 14 Porte du Grand Lyon NEYRON 01700 FR Phone Number : +33 4 72 01 87 87
David RYAN MEDICREA INTERNATIONAL 14 Porte du Grand Lyon NEYRON 01700 FR
- 2. Trade name: MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System
### Regulatory Identification/ Classification
SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS Regulation Number: 21CFR 888.3060 Product Code: KWQ Class II
#### 3. Predicate or legally marketed devices which are substantially equivalent:
### Primary predicate:
- . MaxAn Anterior Cervical Plate System, (BIOMET SPINE, K133518)
### Additional predicate:
- Stella Cervical Plate, (SCIENT'X, K052763) ●
- Synthes Spine Anterior CSLP System (SYNTHES SPINE, K030866)
- o Uniplate Anterior Cervical Plate System (DEPUY SPINE, K100070)
No reference devices were used in this submission.
#### Description of the device: ব
MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System is an anterior system, which consists of a variety of cervical plate lengths and a variety of cervical bone screws.
The implants are manufactured in titanium alloy Ti-6Al-4V ELI conforming to ISO 5832-3 specifications and ASTM F136 specifications.
### Never use stainless steel and titanium implant components in the same construct.
{4}------------------------------------------------
To achieve best results, do not use any of the MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System components from any other systems or manufacturer unless specifically labeled to do so in this or another MEDICREA® INTERNATIONAL document.
MATERIALS: Titanium Alloy (Ti-6Al-4V) according to the ASTM F136-11 & ISO 5832-3
Function: MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System was developed as an implant:
- · To provide immobilization and stabilization of anterior spinal segments
- to augment the development of a solid spinal fusion
- · to provide stability to ease fusion
- to be mechanically resistant to allow the fusion of the operated level
### 5. Indication for Use
MEDICREA INTERNATIONAL S.A. Anterior Cervical Locking Plate System is intended for anterior interbody screw/plate fixation of the cervical spine until the bone fusion occurs.
The MEDICREA INTERNATIONAL S.A. Anterior Cervical Locking Plate System is indicated for use in the temporary stabilization of the anterior cervical (C2 to T1) spine during development of a solid fusion in patients with the following conditions:
- degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies),
- spondylolisthesis,
- trauma (i.e., fracture or dislocation),
- spinal stenosis, ●
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), ●
- . tumor,
- pseudoarthrosis,
- o and failed previous fusion
This implant has to be used with an appropriate bone graft.
Patients should have been unresponsive for a minimum six (6) weeks to conservative treatment before the surgery.
#### 6. Substantial equivalence claimed to predicate devices
MEDICREA INTERNATIONAL S.A. Anterior Cervical Locking Plate components are technologically similar to the already cleared Stella Cervical Plate: MaxAn Anterior Cervical Plate System, Synthes Spine Anterior CSLP System, and Uniplate Anterior Cervical Plate System from Depuv Spine components in terms of intended use, materials used, mechanical safety and performances.
The table below compares the features and characteristics of MEDICRA INTERNATIONAL Anterior Cervical Locking Plate components to their predicate devices.
{5}------------------------------------------------
| Device | MEDICREA<br>INTERNATIONAL<br>Anterior Cervical<br>Locking<br>Plate<br>System | BIOMET<br>SPINE<br>MaxAn<br>Anterior<br>Cervical<br>Plate<br>System | SYNTHES<br>SPINE<br>Anterior<br>CSLP<br>System | DEPUY<br>SPINE<br>Uniplate<br>Anterior<br>Cervical Plate | SCIENT'X<br>STELLA<br>Cervical Plate |
|----------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| 510(k)<br>number | Unknown | K133518 | K030866 | K100070 | K132012 |
| Intended<br>use | | | | | |
| Anterior<br>Cervical | Yes | Yes | Yes | Yes | Yes |
| Components | | | | | |
| Cervical<br>Plate | 1-Level Plate<br>2-Level Plate<br>3-Level Plate<br>4-Level Plate | 1-Level Plate<br>2-Level Plate<br>3-Level Plate<br>4-Level Plate<br>5-Level Plate | 1-Level Plate<br>2-Level Plate<br>3-Level Plate<br>4-Level Plate | 1-Level Plate<br>2-Level Plate | 1-Level Plate<br>2-Level Plate<br>3-Level Plate<br>4-Level Plate<br>5-Level Plate |
| Cervical<br>Screws | Ø4mm &<br>Ø4.5mm<br>Variable and<br>fixed angles<br>screws available | Ø4mm &<br>Ø4.5mm<br>Variable and<br>fixed angles<br>screws<br>available | Ø4mm &<br>Ø4.35mm &<br>Ø4.5mm<br>Variable and<br>fixed angles<br>screws<br>available | Ø4,6mm &<br>Ø5.2mm | Ø4mm &<br>Ø4.5mm<br>Variable and<br>fixed angles<br>screws<br>available |
| Locking<br>Mechanism | | | | | |
| | Part turned<br>to<br>prevent<br>screw<br>back out | Based on<br>titanium<br>elasticity,<br>ring<br>deformation | Based on<br>Titanium<br>deformation,<br>a locking<br>screw is<br>screwed<br>within the<br>screw head | Part turned<br>to prevent<br>screw back<br>out | Part turned<br>to prevent<br>screw back<br>out |
| Materials | | | | | |
| | Titanium<br>Alloy<br>(Ti-6Al-4V)<br>according<br>to<br>ASTM F136<br>&<br>ISO 5832-3 | Titanium<br>Alloy (Ti-6Al-<br>4V)<br>according to<br>ASTM F136<br>& ISO 5832-<br>3 | Titanium<br>Alloy (Ti-6Al-<br>4V)<br>according to<br>ASTM F136<br>& ISO 5832-3 | Titanium<br>Alloy (Ti-6Al-<br>4V)<br>according to<br>ASTM F136<br>& ISO 5832-3 | Titanium<br>Alloy (Ti-6Al-<br>4V)<br>according to<br>ASTM F136<br>& ISO 5832-<br>3 |
### 7. Non-clinical Test Summary:
The components of MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System were mechanically evaluated in static and dynamic compression and static torsion tests following the ASTM F1717-15.
{6}------------------------------------------------
Pyrogenicity testing was conducted in support to MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System component substantial equivalence.
### 8. Clinical Test Summary
No clinical data was provided
### 9. Conclusions Non clinical and Clinical
MEDICREA INTERNATIONAL Anterior Cervical Locking Plate System is substantially equivalent to legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.