Rotograph Prime (under trade mark Villa Sistemi Medicali), I-MAX (under trade mark Owandy Radiology)
K162190 · Villa Sistemi Medicali S.P.A. · MUH · Jul 6, 2017 · Dental
Device Facts
Record ID
K162190
Device Name
Rotograph Prime (under trade mark Villa Sistemi Medicali), I-MAX (under trade mark Owandy Radiology)
Applicant
Villa Sistemi Medicali S.P.A.
Product Code
MUH · Dental
Decision Date
Jul 6, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Rotograph Prime is an extra-oral dental panoramic X-ray unit to radiograph teeth, jaw and oral structures. The device is operated and used by dentists, radiologists and other legally qualified health care professionals. It can be used with both pediatric and adult patients.
Device Story
Extra-oral dental panoramic X-ray system; captures panoramic, sinus, and TMJ images using high-frequency X-ray generator and CCD sensor. Operated by dentists/radiologists in clinics; wall-mounted unit. Patient positioning via laser pointers, temple clamps, bite block, and chin support. System includes optional Extended Projection Package (XP) for segmented, low-dose, and bitewing projections. Images acquired via CCD sensor, processed, and displayed on external PC (not included). Output used by clinicians for diagnostic assessment of teeth/jaw structures. Benefits include improved diagnostic accuracy for interproximal decay and reduced radiation via specific projection modes. Device features motorized height adjustment and face-to-face patient orientation for improved positioning accuracy.
Clinical Evidence
Bench testing only. Evidence includes electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), radiation protection (IEC 60601-1-3), and performance/image quality testing per FDA guidance for solid-state X-ray imaging devices. Biocompatibility of patient-contacting components (temple support, chin supports, bite sticks, handles) assessed per ISO 10993. Software validated per FDA guidance for moderate level of concern.
Technological Characteristics
Extra-oral digital X-ray system; high-frequency generator; 0.5 mm focal spot (IEC 60336). Detector: FFT-CCD area sensor, 48 um pixel size (96 um binning), 4 lp/mm resolution. Wall-mounted form factor. Connectivity: Ethernet to external PC. Materials: dielectric oil, copper, iron, aluminum, glass, tungsten, lead, non-biodegradable plastics. Software: moderate level of concern, microprocessor-controlled exposure. Sterilization: N/A (surface contact only).
Indications for Use
Indicated for extra-oral dental panoramic radiography of teeth, jaw, and oral structures in pediatric and adult patients. Operated by dentists, radiologists, and qualified healthcare professionals.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 6, 2017
VILLA SISTEMI MEDICALI S.P.A. PAOLO SANTIN QA MANAGER VIA DELLE AZALEE 3 BUCCINASCO I-20090 IT
Re: K162190
Trade/Device Name: Rotograph Prime (under Trade Mark Villa Sistemi Medicali), I-max (under Trade Mark Owandy Radiology) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: June 5, 2017 Received: June 8, 2017
Dear Mr. Santin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162190
Device Name
Rotograph Prime (under Trade Mark Villa Sistemi Medicali), I-MAX (under Trade Mark Owandy Radiology)
Indications for Use (Describe)
Rotograph Prime and I-MAX are extra-oral dental panoramic X-ray units to radiograph teeth, jaw and oral structures.
The devices are operated and used by dentists, radiologists and other legally qualified health care professionals.
They can be used with both pediatric and adult patients.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 20px;"> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Villa. The logo consists of a blue square with a white letter "N" inside. Below the square is the word "VILLA" in white, block letters. The letter "N" is stylized with a sharp, angular design.
# 510(k) Summary - rev. 1
This summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Villa Sistemi Medicali S.p.A.<br>via delle Azalee 3<br>20090 Buccinasco (MI)<br>ITALY<br>Phone: +39 02 488591<br>Registration # 8021091 | |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person: | Paolo Casagrande Santin<br>QA Manager<br>Phone: +39 02 48859 239<br>Fax: +39 02 48859 303<br>Email: p.casagrande@villasm.com | |
| Designated Agent: | Ernesto Filizzola<br>Villa Radiology Systems<br>91 Willenbrock Rd. B-1<br>Oxford, CT, 06478<br>ph 203 2628836<br>e-mail ernesto@villaus.com | |
| Date Prepared: | June 27, 2017 | |
| Device Name: | Rotograph Prime (under trade mark Villa Sistemi Medicali)<br>I-MAX (under Trade mark Owandy Radiology as Private Labeler) | |
| Device Type: | System, x-ray, extraoral source, digital | |
| Regulation Number: | 21 CFR 872.1800 | |
| Regulation Name: | Extraoral Source X-Ray System | |
| Regulatory Class: | Class II | |
| Product Code: | MUH | |
| Predicate Device: | The Rotograph Prime (and I-MAX under trade mark Owandy Radiology) is compared<br>with the following predicate device: | |
| | Villa Sistemi Medicali Rotograph EVO D (K090749)<br>- Regulation Number: 21 CFR 872.1800<br>- Regulation Name: Extraoral Source X-Ray System<br>- Regulatory Class: Class II<br>- Product Code: MUH | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo with a blue square background. Inside the square is a white stylized letter "N" that is tilted slightly to the right. Below the square is a white rectangle with the word "VILLA" in a bold sans-serif font.
### Product Description: Rotograph Prime (and I-MAX under trade mark Owandy Radiology) is a complete panoramic X-ray system. The basic version performs:
- -Panoramic adult or child exams, with 3 sizes and 3 tvpes of biting for a total of 18 combinations with automatic parameter selection; with manual selection, it is possible to select a high voltage between 60kV and 70kV, in 2kV steps and anodic current from 2 mA to 7.1 mA in the R20 scale steps.
- -Sinus mode to take exams of the paranasal sinuses with front projection (postero/anterior);
- -TMJ (closed/open mouth) exams in lateral projection.
XP (Extended Projection Package) optional function allows to carry out the following additional exams:
- -Right or left half-panoramic, to be used when the patient is known to have a problem only on one side of the arch, in order to reduce radiation;
- -Low dose Panoramic, which reduces the dose radiated by excluding the TMJ's ascending rami from the radiograph;
- Frontal dentition, for a radiograph of the front part (roughly from canine to canine); -
- -Ortho Rad Panoramic which reduces teeth overlap, thereby improving the diagnosis of interproximal decay:
- Bitewing left or right for lateral dentition (generally from eighth to fourth) with a trajectory that reduces teeth overlap;
- -Bilateral Bitewing (left and right), which sequentially performs both bitewings, showing them on the same image.
The images are acquired by a CCD sensor and are displayed on a monitor, and image manipulation, archiving and communication are performed via a computer (not included in the device). In particular, Rotograph Prime is equipped with a PAN sensor: it is suitable for Panoramic-type imaging, all Panoramic, i.e. all images with about 14cm-high field; all Panoramic. TMJ, and Sinus images belong to this type. Here below a list of the main characteristics of the digital sensor:
146x6 mm
3072x128
- -Detector type:
FFT-CCD area image sensor
48 um (96 um in 2X2 binning mode)
- Sensible area (HxL) PAN Sensor: -
- -Pixel dimension:
- -Number of pixels:
- -Spatial resolution:
- 4 lp/mm
- -Sensor cover attenuation equivalent: < 0.4 mm Al eq.
### Installation mode
The x-ray system is sold only in a wall mounted version.
### Accessories and components
The device can be equipped with different accessories to fulfill different diagnostic needs, such as standard chin support with bite stick, chin support for edentulous patients and Xray push button with extensible cable.
The device contains the following materials and/or components:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Villa. The logo features a white stylized letter "N" on a blue square background. Below the square is the word "VILLA" in white, block letters.
- -Tube-head: dielectric oil, copper, iron, aluminum, glass, tungsten.
- -Collimator: lead
- Other parts of the device: non-biodegradable plastic materials, metal materials, printed circuits, iron-plastic materials, lead
The patient contacting components are:
- temple support rods
- chin supports
- bite sticks
- handles
Chin supports and bite sticks are unchanged from predicate device. Handles and temple support rods are specific for the subject device.
All parts in contact with patient have been assessed according to ISO 10993.
Contact duration is in the range of seconds to few minute and contact type is surfacecontacting.
### Indication for Use: Rotograph Prime (and I-MAX under trade mark Owandy Radiology) is an extra-oral dental panoramic X-ray unit to radiograph teeth, jaw and oral structures. The device is operated and used by dentists, radiologists and other legally qualified health care professionals. It can be used with both pediatric and adult patients.
### Rationale for Substantial Equivalence
Here below a summing up the main features of the proposed and predicate devices in order to point out the significant similarities and differences.
Rotograph Prime (and I-MAX under trade mark Owandy Radiology) is based on a well-known technology and it has the same indication for use as the predicate device as reported here below.
| | Villa Sistemi Medicali - Rotograph Prime<br>and<br>Owandy Radiology - I-MAX<br>510(k) number tbd | Villa Sistemi Medicali<br>Rotograph EVO D<br>510(k) number K090749 |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Rotograph Prime is an extra-oral dental panoramic<br>X-ray unit to radiograph teeth, jaw and oral<br>structures.<br><br>The device is operated and used by dentists,<br>radiologists and other legally qualified health care<br>professionals.<br><br>It can be used with both pediatric and adult<br>patients. | Extraoral source x-ray system, which are<br>intended for dental radiographic examination of<br>the teeth, jaw, and oral structures, specifically<br>for panoramic examinations and implantology<br>and for TMJ studies and cephalometry |
Rotograph Prime (and I-MAX under trade mark Owandy Radiology) has the same functions in the same environment as the predicate device. In particular, the proposed devices perform the same exams for the examination of teeth, jaw and oral structures, as the Rotograph EVO D device.
The proposed devices also use a high frequency x-ray generator with the same x-ray exposure time control and focal spot value of the predicate device.
In the subject devices, patient positioning is assured through two laser pointers that allow to locate the reference planes, as in Rotograph EVO D predicate device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features a logo with a stylized white letter 'N' set against a blue square background. Below the square, the word 'VILLA' is written in white, bold letters. The logo appears clean and modern, with a simple color scheme.
The subject device detector is exactly the same model as the detector. The equivalence of the image quality of panoramic radiographies of the subject device has been assessed with image comparison tests.
For all these reasons, Rotograph Prime (and I-MAX under trade mark Owandy Radiology) share the fundamental technological characteristics as the predicate device, even if there are some minor differences, not affecting the safety and effectiveness, between the subject and predicate devices.
The following tables list the similarities and the differences between the subject device (Rotograph Prime under trade mark Villa Sistemi Medicali and I-MAX under trade mark Owandy Radiology) and the predicate device (Rotograph EVO D, manufactured by Villa Sistemi Medicali and cleared by FDA with 510(k) number K090749).
| | Villa Sistemi Medicali - Rotograph Prime<br>and<br>Owandy Radiology - I-MAX<br>510(k) number K162190 | Villa Sistemi Medicali<br>Rotograph EVO D<br>510(k) number K090749 |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Versions | | |
| Digital | YES | YES |
| Exam projections | | |
| Pan | YES | YES |
| Improved orthogonality<br>dentition | YES | YES |
| Segmented Pan | YES | YES |
| Frontal dentition | YES | YES |
| Low dose panoramic | YES | YES |
| TMJ | YES | YES |
| Sinus | YES | YES |
| Size of pixel (acquired) | 96 μm | 96 μm |
| Magnification (PAN) | 1,23 constant | 1,23 constant |
| Source to image distance | 500 mm | 500 mm |
| Panoramic max image size | equivalent to 15x30 cm film | equivalent to 15x30 cm film |
| | Villa Sistemi Medicali - Rotograph Prime<br>and<br>Owandy Radiology - I-MAX<br>510(k) number K162190 | Villa Sistemi Medicali<br>Rotograph EVO D<br>510(k) number K090749 |
| Standard PAN image resolution | 1536 x 2805 pixels | 1536 x 2805 pixels |
| Dynamic (acquisition) | 12 bit | 12 bit |
| Generator / tube | | |
| X-ray generator | High frequency | High frequency |
| Focal spot value | 0.5 mm (IEC 60336) | 0.5 mm (IEC 60336) |
| Anode type | fixed | fixed |
| X-ray exposure time control | Automatic - pre-programmed Microprocessor<br>Controlled | Automatic - pre-programmed Microprocessor<br>Controlled |
| Independent kV-mA regulation | YES | YES |
| DAP Software | YES | YES |
| Patient positioning | | |
| Height adjustment | motorized | motorized |
| Positioning lights | 2 laser pointers | 2 laser pointers |
| Patient position | standing | standing |
| Patient positioning tools | temple clamps, bite block, chin support | temple clamps, bite block, chin support |
| Focal layer adjustment<br>(prognatism compensation) | Electronic, three positions, no patient movement | Electronic, three positions, no patient<br>movement |
| User interface | | |
| Real time visualization | YES | YES |
| PC connection | Ethernet | Ethernet |
| Installation | | |
| Telescopic column | YES | YES |
| Power supply voltage | 110-120 V, 50/60 Hz | 110-120 V, 50/60 Hz |
| | Villa Sistemi Medicali - Rotograph Prime<br>and<br>Owandy Radiology - I-MAX<br>510(k) number K162190 | Villa Sistemi Medicali<br>Rotograph EVO D<br>510(k) number K090749 |
| Versions | | |
| Analog | NO | YES |
| Exam projections | | |
| Ceph | NO | YES |
| Adult PAN scan time | 14.4 sec | 13.8 sec |
| Child PAN scan time | 13.3 sec | 13.8 sec |
| Generator / tube | | |
| kV Range | 60-70 kV step 2kV | 60-86 kV step 2kV |
| mA range | 2-7.1 mA | 6-10 mA |
| Total filtration | 2.0 mm Al eq | 2.5 mm Al eq |
| Collimator | Fixed | Automatic |
| Patient positioning | | |
| Height of chin support | 975-1635 mm | 920-1755 mm |
| Patient positioning orientation | face to face | lateral with adjustable mirror |
| User interface | | |
| On board user interface | keyboard and virtual control panel | keyboard and O-LED display |
| Installation | | |
| Type of installation | wall mount | floor mount |
| Current rating | 7 A | 15 A |
| Weight (wall mount version) | 62 kg | 157 kg |
| Dimensions (wall mount version) | 1107 x 953 mm | 1260 x 1040 mm |
### Similarities
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Villa. The logo consists of a white stylized letter "N" on a blue square background. Below the square is the word "VILLA" in white, set against a blue background.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image features a logo with a stylized white letter 'N' on a blue square background. Below the square, the word 'VILLA' is written in white, block letters. The overall design is simple and modern, with a clear contrast between the white elements and the blue background.
### Differences
See the next paragraph "Detailed description of the differences" for additional information and details about each difference between subject and predicate device.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for Villa. The logo features a white stylized letter "V" inside of a blue square. Below the square is the word "VILLA" in white, block letters.
### Detailed description of the differences
As detailed described on the following paragraphs, all the differences do not affect the safety and effectiveness of the subject devices.
#### Versions .
The Predicate device is available also in analog version (using film-screen combination as image receptor). The subject device is available only in digital version (using the same digital detector as the predicate device). This choice has been driven by the current market and technology trend. Analog extraoral dental units have substantially disappeared from the market in the last 5 years.
#### . Exam projections
### Ceph
The predicate device includes, as an option, also cephalometric examination of the skull. The subject device doesn't include cephalometric exams. The reason is that the device is dedicated to a market segment in which doctors don't need this exam. The fact that this function is not offered has no impact on the performance of the other available projections neither introduces differences vs pan-only version of the predicate device. The subject device is therefore equivalent to the pan-only version of the predicate device.
### Adult Pan Scan time
During panoramic exams, the machine has to perform movements according to the selected exam specific trajectory; the exam scan time is the time taken by the machine to complete these movements. The panoramic exam trajectory implemented in the subject device is slightly slower from the one on the predicate device in order to adapt to the small geometric differences between the units; for this reason the scan time is increased of about 4%.
### Cihild Pan Scan time
Due the fact that on the subject device the scan time has been slightly increased, it has been implemented a PAN exam specific for children with a reduced scan time.
In fact, the trajectory of child panoramic exam movements is shorter and dedicated to child patients giving a dose reduction vs. the adult exam. This trajectory has been already used on previous Villa panoramic units. It doesn't require different instructions of patient positioning respect to the adult exam.
#### Generator / tube ●
### kV Range and mA range
The radiological parameters ranges have been optimized for the exam projections (on subject device ceph exams that requires higher dose is not present) and receptor type (on subject device film version that requires higher dose is not present) available on the subject device.
### Total filtration
Due to the reduced kV range, the total filtration has been reduced maintaining the compliance with 21CFR 1020.30
### Collimator
Due to the absence of the ceph exams, only one collimator slot is needed on the subject device. This collimator slot corresponds to the predicate devices for the exam projections available on the subject device.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for Villa. The logo features a white stylized letter "V" inside of a blue square. Below the square is the word "VILLA" in white, block letters.
#### Patient positioning ●
### Height of chin support
The Height range of chin support has been reduced based on the following assumption: 975mm correspond to the height of the chin of a 7 years old child, according to the Dreyfuss Tables and 1635mm correspond to the height of the chin of a man of about 190cm of height. According to the Dreyfuss tables this cover about 99% of the adult population.
### Patient Positioning orientation
A difference has been introduced in the patient orientation with respect to the machine and operator. In the predicate device, the patient is facing the column; in order for the operator to have visibility of the patient's face, a mirror is present on the column. In the subject device the patient is facing to the equipment, making it possible for the operator to have a direct face visibility of the patient, therefore improving the positioning accuracy and patient's confidence.
#### . User interface
The user interface of the subject device is made by a keyboard on the machine to drive movements and lasers plus a virtual control panel on PC to select exams and their parameters. The user interface of the predicate device is on board but its digital version has also a virtual control panel on the PC with which exams and their parameters can be selected. The reason of this difference is related to the fact that the subject device, being only digital, always requires a PC so the on board keyboard could be removed while the predicate device needed the on-board keyboard for exam selection for the analog version that doesn't require a PC to work.
#### Installation .
### Type of installation
The weight reduction of the subject device respect to the predicate device made it possible to produce a fully wall-mounted version which leaves the floor space free from support structures and more comfortable for the patient, especially in case of particular needs, such as wheelchairs.
### Current rating
Due to the different radiological parameters ranges respect to the current rating of the subject device is also reduced
### Weight (wall mount version)
Due to the absence of analoq version and cephalometric device, and thanks to the use of light-weight materials and optimized components design, the weight of subject device has been significantly reduced compared to the predicate device. The weight reduction doesn't affect the robustness of the device
### Dimensions (wall mount version)
Due to the optimized components design, the dimension of subject device has been reduced compared to the predicate device
#### Regarding software ●
The elimination of the analog (film) version, the cephalometric option and the automatic contributed to the simplification both of the overall electronic and of the software architecture. The changes to the virtual control panel on PC - to select exams and their parameters - are limited just only to the cosmetic graphics elements. No other significant change is present in the software.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image is a logo for Villa. The logo features a stylized white letter "N" on a blue square background. Below the square is the word "VILLA" in a sans-serif font, also in white.
The review of the technological differences between the subject device and the predicate device shows that they are due only to technological evolution and removal of not needed functions and/or version so they do not raise any new questions regarding safety or effectiveness of the Rotograph Prime (and I-MAX under trade mark Owandy Radiology). Basing on the above considerations, we can conclude that the Rotograph Prime (and I-MAX under trade mark Owandy Radiology) is as safe and effective as the predicate device.
### PERFORMANCE DATA AND TESTING EVIDENCE:
Electrical safety, EMC/EMI testing, biocompatibility consideration, performance and image quality testing, verification and validation testing were performed to support the substantial equivalence determination. All standards applied were FDA recognized international standards. The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained
in Medical Devices" (May 11, 2005). The device software is considered a "Moderate Level of Concern".
Rotograph Prime has been designed to be in conformity with the following international standards:
- -IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007): Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- -IEC 60601-1-2:2007: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- -IEC 60601-1-3:2008: Medical electrical equipment - Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray Equipment
- -IEC 60601-2-63:2012: Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment
- -IEC 62304:2006: Medical device software - Software life-cycle processes
- -ANSI/AAMI ES60601-1: 2005 / A2:2010 - Medical electrical equipment, Part 1: General Requirements
- CAN/CSA-C22.2 No. 60601-1:08: Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- -IEC 60601-1-6:2010: Medical electrical equipment - Part 1-6: General requirements for safety - Collateral Standard: Usability
- -IEC 62366:2007: Medical devices - Application of usability engineering to medical devices
- -ISO 10993-2:2006: Biological evaluation of medical devices- part 2: animal welfare requirements
- -ISO 10993-5:2009: Biological evaluation of medical devices- part 5: tests for in vitro cytotoxicity
- -ISO 10993-10:2010: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- -ISO 10993-12:2012: Biological evaluation of medical devices- part 12: sample preparation and reference materials
In the submission are provided all the test reports including test protocol, methods, pass/fail criteria, and results in conformity with the standards here above.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image is a logo for Villa. The logo features a white stylized letter "V" on a blue square background. Below the square is the word "VILLA" in a sans-serif font.
A risk analysis was performed to analyze the hazards associated with the changes.
Non-clinical considerations according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State Xray Imaging Devices" were performed in order to assess the image quality and the relevant analysis has been supplied (see annex 7_Image_Quality_Comparison).
The intended use of the proposed device is the same of the predicate device. Concerning Conclusion: the technical characteristics, it has been demonstrated that Rotograph Prime (and I-MAX under trade mark Owandy Radiology) does not introduce innovative aspects and does not arise different considerations regarding the safety and the relevant risks linked to the device. Despite some minor technological differences, Rotograph Prime (and I-MAX under trade mark Owandy Radiology) can be considered as safe and as effective as the predicate device and therefore substantially equivalent to the predicate Rotograph EVO D.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.