← Product Code [LFD](/productcode/LFD) · K162167

# Caphosol Artifical Saliva Tablets (K162167)

_Eusa Pharma (Uk) Limited · LFD · Mar 31, 2017 · DE · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162167

## Device Facts

- **Applicant:** Eusa Pharma (Uk) Limited
- **Product Code:** [LFD](/productcode/LFD.md)
- **Decision Date:** Mar 31, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Caphosol Dispersible is indicated for dryness of the mouth and throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent. Caphosol Dispersible is also indicated as an adjunct to standard oral care in the treatment of the mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain.

## Device Story

Caphosol Dispersible is a supersaturated calcium phosphate electrolyte mouth rinse formulated as a tablet; tablet dissolved in water by patient prior to use. Device replaces ionic and pH balance in oral cavity; relieves oral mucosal dryness; ameliorates pain associated with xerostomia and mucositis. Used in home or clinical settings; patient-operated. Output is a liquid rinse applied to oral mucosa. Benefits include relief of dry mouth symptoms and adjunct treatment for chemotherapy/radiation-induced mucositis.

## Clinical Evidence

No clinical performance testing conducted. Biocompatibility assessment performed in accordance with ISO 10993-1. Literature search indicated no clinically significant adverse effects.

## Technological Characteristics

Supersaturated calcium phosphate electrolyte mouth rinse. Components: sodium phosphate dibasic, sodium phosphate monobasic, calcium chloride, sodium hydrogen carbonate, sodium carbonate, maltodextrin, citric acid, sorbitol. Form factor: 1.75 mm diameter tablet. Non-sterile. Standalone use; requires dissolution in water. Biocompatibility testing per ISO 10993-1.

## Predicate Devices

- Caphosol ([K030802](/device/K030802.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 31, 2017

Eusa Pharma (UK) Limited % Mr. J. Harvey Knauss Consultant Delphi Consulting Group 11874 South Evelyn Circle Houston, Texas 77071-3404

Re: K162167

Trade/Device Name: Caphosol Artificial Saliva Tablets Regulatory Class: Unclassified Product Code: LFD Dated: March 7, 2017 Received: March 7, 2017

Dear Mr. Knauss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Image /page/1/Picture/7 description: The image shows a signature that reads "Susan Runno, DDS, MA". The signature is written in cursive and appears to be handwritten. The letters are connected and flow together, creating a fluid and legible signature. The text is black against a white background.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known). K162167

Device Name CAPHOSOL Artificial Saliva

#### Indications for Use (Describe)

Caphosol Dispersible is indicated for dryness of the mouth and throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent. Caphosol Dispersible is also indicated as an adjunct to standard oral care in the treatment of the mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

ੁਕ

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary, K162167, GENERAL INFORMATION:

| Sponsor/Submitter: | EUSA Pharma<br>Breakspear Park, Breakspear Way<br>Hemel Hempstead<br>HP2 4TZ United-Kingdom<br>011 44 1732 832 317 voice<br>Registration # 10049906<br>Mr. Paul Davisson |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Common/Usual Name | None                       |
|-------------------|----------------------------|
| Proprietary Name  | Caphosol Artificial Saliva |
| Regulation Number | Unclassified               |
| Product Code      | LFD                        |
| Predicate Devices | K030802 Primary            |

## Description of the device and its intended application:

Caphosol is a supersaturated calcium phosphate, electrolyte mouth rinse used as artificial saliva, designed in part to replace ionic and pH balance in the oral cavity, and formulated as a tablet to be dissolved in water prior to use.

{4}------------------------------------------------

The following table lists the Caphosol components:

| List of components                          | Function in<br>Caphosol<br>Dispersible | Function in Caphosol |
|---------------------------------------------|----------------------------------------|----------------------|
| Sodium phosphate dibasic<br>anhydrous       | Active component                       | Active component     |
| Sodium phosphate<br>monobasic anhydrous     | Active component                       | Active component     |
| Calcium chloride                            | Active component                       | Active component     |
| Sodium hydrogen carbonate                   | Effervescent Agent                     | Stabilizer           |
| Sodium carbonate<br>anhydrous               | Effervescent Agent                     | none                 |
| Maltodextrin                                | Diluent                                | none                 |
| Granulate Citric acid<br>anhydrous/sorbitol | Effervescent Agent                     | none                 |
| Water                                       |                                        | Active component     |

# Caphosol Statement of Intended Use:

Caphosol Dispersible is indicated for dryness of the mouth and throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent. Caphosol Dispersible is also indicated as an adjunct to standard oral care in the treatment of the mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain.

# Comparison of Device Technology:

| COMPARING                                    | CAPHOSOL® Dispersible<br>K162167                                                                                                                                                                                                                                                                                                                                                                                                                                         | CAPHOSOL®<br>Predicate<br>K030802                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                                 | LFD                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | LFD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation Number                            | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Indication for Use                           | Caphosol Dispersible is indicated for dryness of the mouth and throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent. Caphosol Dispersible is also indicated as an adjunct to standard oral care in the treatment of the mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain. | Caphosol is indicated for dryness of the mouth or throat (hypo salivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent. Caphosol is indicated for relief of dryness of the oral mucosa when hypo salivation results from the following: surgery, radiotherapy near the salivary glands, chemotherapy; infection or dysfunction of the salivary glands; inflammation of the mouth or throat; fever, emotional factors such as fear or anxiety; obstruction of the salivary ducts; Sjogren's syndrome; and Bell's Palsy. |
| COMPARING                                    | CAPHOSOL® Dispersible<br>K162167                                                                                                                                                                                                                                                                                                                                                                                                                                         | CAPHOSOL®<br>Predicate<br>K030802                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Caphosol is also indicated for<br>dryness of the oral mucosa due to<br>drugs such as antihistamines or<br>atropine or other anticholinergic<br>agents that suppress salivary<br>secretion. It may be used as part of<br>an oral hygiene program for patients<br>with dry mouth. Caphosol provides<br>intensive hygiene of the oral cavity,<br>and may be used to help relieve bad<br>taste and to relieve offensive nasal<br>discharge and crusting.                                                                                                                              |
| Packaging                                    | Tablet solid form, 1.75 MM<br>diameter.<br>Image: Tablet                                                                                                                                                                                                                                                                                                                                                                                                                 | • Two 15 ml solutions plastid (LDPE)<br>ampules<br>• Solution A (blue ampule) –<br>(phosphate solution)                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Solution B (clear ampule) - (calcium<br>solution)<br>Image: Ampules                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Method of use                                | Ready to use, mix with water.                                                                                                                                                                                                                                                                                                                                                                                                                                            | Mix solutions A & B                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Area of use                                  | Oral cavity.                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Oral cavity.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Presentation                                 | Non-Sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Non-Sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Interaction with other<br>medicinal products | There are no known interactions<br>with medicinal or other products.                                                                                                                                                                                                                                                                                                                                                                                                     | There are no known interactions with<br>medicinal or other products.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{5}------------------------------------------------

## Indication for Use Differences:

The indication for use statement of the predicate device is more specific than the indication for use statement of the subject device. The differences do not raise concerns as the intended use is the same.

## Technology Differences:

The CAPHOSOL® Artificial Saliva Tablets and Caphosol® Liguid Artificial Saliva active chemical components are identical, however the CAPHOSOL® Artificial Saliva Tablets formulation includes a diluent and effervescent agents, and does not contain water. The available physical form of the subject device is a dry tablet, whereas the predicate device is composed of two ampules containing different solutions to be mixed prior to use. The

{6}------------------------------------------------

excipients in the subject device formulation result in effervescence and complete dissolution of the table to obtain the final solution prepared after dissolution of a tablet or after mixing 2 ampules are identical in quantity of the active ingredients (Calcium and Phosphate).

The CAPHOSOL® Artificial Saliva Tablets is substantially equivalent to the predicate devices with regard to operating principle and function.

### Performance Testing:

Biocompatibility assessment conducted in accordance with ISO - 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process.

### Non-Clinical and Clinical Studies:

Clinical performance testing was not conducted for the subject device.

A literature search has not shown any clinically significant adverse effects being reported.

### Conclusion:

Based on the similarities in intended use, technology, and performance testing, the CAPHOSOL Artificial Saliva Effervescent Tablets is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K162167](https://fda-staging.innolitics.com/device/K162167)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com