← Product Code [HWC](/productcode/HWC) · K162153

# Integra® CAPTURE™ Screw System & Integra® Ti6® Internal Fixation System (K162153)

_Ascension Orthopedics · HWC · Jul 14, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162153

## Device Facts

- **Applicant:** Ascension Orthopedics
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Jul 14, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The CAPTURE™ Screw System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses, and osteotomies of the small bones in the hand and foot. The implants are intended for single use only. The Ti6® Internal Fixation System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

## Device Story

The Integra® CAPTURE™ Screw System and Integra® Ti6® Internal Fixation System consist of self-drilling and self-tapping bone screws. These devices are used by surgeons for the fixation of fractures, non-unions, arthrodeses, and osteotomies in small bones of the hand and foot. The system includes cannulated (Digital Fusion) and non-cannulated (QuickSnap) screw designs. The screws are manufactured from Ti-6Al-4V titanium alloy and feature an anodized surface treatment. The devices are intended for single use only.

## Clinical Evidence

No clinical data was necessary to demonstrate substantial equivalence. Non-clinical performance analysis was conducted per ASTM F543, including testing for torsional strength, driving torque, and axial pullout strength.

## Technological Characteristics

Materials: Ti-6Al-4V titanium alloy. Design: Self-drilling and self-tapping bone screws; cannulated (Digital Fusion) and non-cannulated (QuickSnap) configurations. Surface treatment: Anodization. Testing standards: ASTM F543 (torsional strength, driving torque, axial pullout strength).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Smith & Nephew, Inc. VLP Foot Plating, Screw System and Accessories ([K090675](/device/K090675.md))
- Wright Medical Technology, Inc. WRIGHTTM Compression Screw ([K082320](/device/K082320.md))
- Koby Surgical. Koby Surgical Internal Fixation System ([K060026](/device/K060026.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 14, 2017

Ascension Orthopedics Blesson Abraham Senior Regulatory Affairs Specialist 8700 Cameron Road Austin, Texas 78754

Re: K162153

Trade/Device Name: Integra® CAPTURE™ Screw System & Integra® Ti6® Internal Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: June 14, 2017 Received: June 15, 2017

Dear Blesson Abraham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Vincent J. Devlin -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K162153

Device Name

Integra® CAPTURETM Screw System & Integra® Ti6® Internal Fixation System

## Indications for Use (Describe)

The proposed indications for use for the Integra® CAPTURE™ Screw System are as follows:

"The CAPTURE™ Screw System implants (screws) are intended for fixation of fractures, and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

The proposed indications for use for the Integra® Ti6® Internal Fixation System are as follows: "The Ti6@ Internal Fixation System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

| Type of Use (Select one or both, as applicable) | <span style="text-decoration: underline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------

{3}------------------------------------------------

K162153

Page 1 of 2

# 510(k) Summary

| Sponsor:                           | Ascension Orthopedics<br>8700 Cameron Road<br>Austin, TX 78754-3832, USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                    | Blesson Abraham<br>Senior Regulatory Affairs Specialist<br>512-852-3942<br>512-368-1423                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date:                              | August 1, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Establishment Registration Number: | 1651501                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Trade Name:                        | Integra® CAPTURE™ Screw System &<br>Integra® Ti6® Internal Fixation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Common Name:                       | Bone Screw                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code:                      | HWC – Screw, Fixation, Bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification:                    | 21 CFR §888.3040 – Smooth or threaded metallic bone<br>fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification Panel:              | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Devices:                 | Smith & Nephew, Inc. VLP Foot Plating, Screw System and<br>Accessories, K090675, cleared on June 4, 2009.                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                    | Wright Medical Technology, Inc. WRIGHTTM Compression<br>Screw, K082320, cleared on November 5, 2008.                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                    | Koby Surgical. Koby Surgical Internal Fixation System,<br>K060026, cleared on January 26, 2006.                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Description:                | The Integra® CAPTURETM Screw System and Integra® Ti6®<br>Internal Fixation System consists of bone screws of various<br>designs and sizes intended to fixate bones in cases of<br>fractures, osteotomies, or fusions. The screws are self- drilling<br>and self-tapping, and are manufactured from Ti-6Al-4V<br>titanium alloy. This submission presents new cannulated<br>(Digital Fusion) and non-cannulated (QuickSnap) screw<br>designs that are line extensions to the current Integra®<br>CAPTURE™ Screw System and Integra® Ti6® Internal Fixation<br>System. |
| Intended Use:                      | The proposed indications for use for the Integra®<br>CAPTURE™ Screw System are as follows:<br>"The CAPTURE™ Screw System implants (screws) are<br>intended for fixation of fractures, non-unions, arthrodeses, and<br>osteotomies of the small bones in the hand and foot. The<br>implants are intended for single use only.                                                                                                                                                                                                                                         |

{4}------------------------------------------------

Image /page/4/Picture/3 description: The image shows the logo for INTEGRA. The word "INTEGRA" is in large, gray, sans-serif font. Below the word "INTEGRA" is the phrase "LIMIT UNCERTAINTY" in a smaller, green, sans-serif font. To the right of the word "INTEGRA" is a green graphic consisting of several squares of different sizes.

The proposed indications for use for the Integra® Ti6® Internal Fixation System are as follows:

"The Ti6" Internal Fixation System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses and osteotomies of the small bones in the hand and foot. The implants are intended for single use only.

The additional Integra® CAPTURE™ Screw System and Basis of Substantial Equivalence: Integra® Ti6® Internal Fixation System QuickSnap and Digital Fusion screws have the same intended use and fundamental scientific technology as the predicate devices. Sizing differences between the subject and predicate devices have been assessed and do not present new issues of safety or efficacy. Anodization surface treatment of the subject devices does not present new issues to safety or efficacy. Therefore, the proposed Integra® CAPTURE™ Screw System and Integra® Ti6® Internal Fixation System QuickSnap and Digital Fusion screws, anodization, are substantially equivalent to the predicate devices.

### Non-Clinical Performance Data: The following non-clinical performance analysis was conducted per ASTM F543 in order to establish substantial equivalence.

- 1. Torsional Strength
- 2. Driving Torque
- Axial Pullout Strength 3.

The Digital Fusion and OuickSnap performed substantially equivalent or better when compared to the predicate device.

#### Clinical Performance Data: Clinical performance data was not necessary to demonstrate substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K162153](https://fda-staging.innolitics.com/device/K162153)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).