K162131 · Respiratory Motion, Inc. · BZK · May 9, 2017 · Anesthesiology
Device Facts
Record ID
K162131
Device Name
ExSpiron 1Xi
Applicant
Respiratory Motion, Inc.
Product Code
BZK · Anesthesiology
Decision Date
May 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1850
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162131 · May 9, 2017
ExSpiron 1Xi
Respiratory Motion, Inc.
Historical clinical cohort data; Multi-year patient population respiratory data
The device's proprietary, non-linear, regressive model was trained and refined using a large historical cohort of respiratory data collected over three years across diverse patient populations to improve accuracy.
Algorithm training; Historical cohort; Retrospective data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Historical Cohort Training Dataset; Retrospective analysis of historical clinical data; Follow-up/Duration: 3 years; Study Period: 3 years
Diverse patient populations; Sample Size: Thousands of data points
Not applicable for this study
Algorithm accuracy for respiratory parameters (MV, TV, RR)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Minute Ventilation
Non-linear regressive model
—
11.5% (Overall Accuracy)
Large historical cohort and thousands of additional data points
—
Clinical study in Waltham, MA: 20 subjects
—
Tidal Volume
Non-linear regressive model
—
11.4% (Overall Accuracy)
Large historical cohort and thousands of additional data points
—
Clinical study in Waltham, MA: 20 subjects
—
Respiratory Rate
Non-linear regressive model
—
0.08% (Overall Accuracy)
Large historical cohort and thousands of additional data points
—
Clinical study in Waltham, MA: 20 subjects
—
Indications for Use
ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients. ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of: * Minute Ventilation (MV) * Tidal volume (TV) * Respiratory rate (RR) ExSpiron 1Xi measurements are used as an adjunct to other clinical information.
Device Story
ExSpiron 1Xi is a non-invasive respiratory monitoring system for use in healthcare facilities by clinicians. It utilizes thoracic bioimpedance to measure respiratory parameters. A single-patient use electrode padset applied to the patient's torso delivers a low-current (50 kHz) signal and records impedance changes. The monitor enclosure contains a high-frequency current generator and signal conditioning circuitry; a tablet PC performs signal processing, calibration, and GUI display. The device provides real-time graphical traces of lung volume over time and scalar values for minute ventilation, tidal volume, and respiratory rate. These measurements serve as an adjunct to other clinical information, assisting healthcare providers in assessing patient respiratory status. The system eliminates the need for patient-specific calibration required by the predicate, improving usability in basic monitoring modes.
Clinical Evidence
Clinical study (n=20, age 22-80, BMI 19-42) compared ExSpiron 1Xi to the predicate device. Subjects performed 20 breathing tests in two sessions (24 hours apart). Primary endpoints included bias, precision, and accuracy for Minute Ventilation, Tidal Volume, and Respiratory Rate. Overall accuracy was 11.5% for MV, 11.4% for TV, and 0.08% for RR. Results demonstrate clinically relevant accuracy over a 24-hour period.
Technological Characteristics
Thoracic bioimpedance sensing; 50 kHz low-current signal. Formed ABS plastic housing. Single-patient use biocompatible electrode padset. Windows tablet PC for processing/GUI. Compliant with IEC 60601-1 (safety), IEC 60601-1-2 (EMC/EMI), IEC 60601-1-6/62366 (usability), and IEC 60601-1-8 (alarms).
Indications for Use
Indicated for adult patients (≥21 years) in healthcare facilities (e.g., post-operative, critical care) for non-invasive monitoring of breathing parameters.
Regulatory Classification
Identification
A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 9, 2017
Respiratory Motion, Inc. Susan Hamann Director of Quality and Regulatory 411 Waverly Oaks Road, Building 1, Suite 150 Waltham, Massachusetts 02452
Re: K162131
Trade/Device Name: ExSpiron™ 1Xi Regulation Number: 21 CFR 868.1850 Regulation Name: Monitoring Spirometer Regulatory Class: Class II Product Code: BZK, BZQ Dated: April 5, 2017 Received: April 10, 2017
Dear Susan Hamann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162131
Device Name ExSpiron 1Xi
Indications for Use (Describe)
ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients.
ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of:
* Minute Ventilation (MV)
* Tidal volume (TV)
* Respiratory rate (RR)
ExSpiron 1Xi measurements are used as an adjunct to other clinical information.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows a pair of human lungs against a black background. The lungs are translucent and colored in shades of blue, giving them a somewhat ethereal appearance. The anatomical structure of the lungs is visible, including the lobes and some of the branching airways. The lighting and color choice create a visual effect that highlights the shape and form of the lungs.
## 510(k) Summary – Traditional 510(k)
| 510(k) Owner: | Respiratory Motion, Inc. |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 411 Waverley Oaks Road<br>Building 1, Suite 150<br>Waltham, Massachusetts 02452 |
| Phone: | 781-373-1636 |
| Fax: | 781-373-1653 |
| Contact<br>person: | Jenny Freeman, MD |
| Date 510(k)<br>Summary prepared and<br>type of 510(k): | May 9, 2017 – Traditional 510(k) |
| Trade name: | ExSpironTM 1Xi |
| Common<br>name: | Respiratory Monitoring System |
| Classifications: | Regulation Number: 21 CFR 868.1850<br>Regulation Name: Monitoring Spirometer<br>Regulatory Class: II<br>Product Code: BZK, BZQ |
| Predicate<br>devices: | ExSpiron 1Xi Respiratory Monitor, marketed by Respiratory Motion, Inc, Waltham, MA (K130170). |
| Device<br>Description: | The ExSpiron 1Xi is a noninvasive respiratory monitoring system that graphically displays<br>lung volume against time and reports Minute Ventilation, Tidal Volume and Respiratory Rate.<br><br>The ExSpiron 1Xi system consists of:<br>Monitor:<br>The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure.<br>Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC. Computer: A tablet PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate. |
{4}------------------------------------------------
| | A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It<br>delivers current to, and records impedance measurements from, the skin. The Electrode<br>PadSet for the ExSpiron 1Xi is identical to that cleared for the Predicate device. | | | | |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Intended use: | ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as<br>post-operative care and critical care units, to monitor breathing in adult (at least 21 years old)<br>patients.<br>ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and<br>reports an approximate value of:<br>* Minute Ventilation (MV)<br>* Tidal volume (TV)<br>* Respiratory rate (RR)<br>ExSpiron 1Xi measurements are used as an adjunct to other clinical information. | | | | |
{5}------------------------------------------------
| Comparison of<br>technological<br>characteristics: | Characteristic | ExSpiron 1 (Proposed Device) | ExSpironTM1Xi<br>(Predicate) K130170 | Comment |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| | Intended Use | ExSpiron 1Xi is indicated for use<br>by healthcare professionals in<br>healthcare facilities, such as post-<br>operative care and critical care<br>units, to monitor breathing in adult<br>(at least 21 years old) patients | ExSpiron 1Xi is indicated for use by<br>healthcare professionals in<br>healthcare facilities, such as post-<br>operative care and critical care units,<br>to monitor breathing in adult (at<br>least 21 years old) patients | Both the proposed device and the predicate have an identical intended<br>use. |
| | | Each ExSpiron 1Xi is a<br>noninvasive system that<br>graphically displays lung volume<br>against time and reports an<br>approximate value of:<br>• Tidal volume<br>• Respiratory rate, and<br>• Minute ventilation. | Each ExSpiron 1Xi is a noninvasive<br>system that graphically displays<br>lung volume against time and<br>reports an approximate value of:<br>• Tidal volume<br>• Respiratory rate, and<br>• Minute ventilation. | |
| | | ExSpiron 1Xi measurements are<br>used as an adjunct to other clinical<br>information sources. | ExSpiron 1Xi measurements are<br>used as an adjunct to other clinical<br>information sources. | |
| | Technology | Measurement is by thoracic<br>bioimpedance. | Measurement is by thoracic<br>bioimpedance. | Both the proposed device and the predicate have identical technology. |
| | Volume<br>Measurements | Tidal volume<br>Minute volume<br>Volume vs. time chart | Tidal volume<br>Minute volume<br>Volume vs. time chart | Both the proposed device and the predicate measure the same<br>respiratory volumes. |
| | Rate<br>Measurements | Respiratory rate (breaths/min) | Respiratory rate (breaths/min) | Both the proposed device and the predicate measure respiratory rate. |
| | Safety | IEC 60601-1, including electrical<br>and mechanical safety | IEC 60601-1, including electrical<br>and mechanical safety | Both the proposed device and the predicate have the same electrical<br>safety profile and are compliant with IEC 60601-1. |
| | Energy Source | The bioimpedance output signal is<br>generated internally to the<br>ExSpiron Monitor. It is a low<br>current signal at approximately 50<br>kHz. | The bioimpedance output signal is<br>generated internally to the ExSpiron<br>Monitor. It is a low current signal at<br>approximately 50 kHz. | Both the proposed device and the predicate have the same energy<br>source. |
| Algorithm | A software algorithm calculates<br>respiratory parameters using a<br>proprietary, non-linear, regressive<br>model trained with a dataset<br>containing respiratory data from a<br>large historical cohort. The<br>algorithm was refined for the<br>ExSpiron 1Xi- using thousands of<br>additional data points collected<br>over the last 3 years in a wide<br>variety of patient populations.<br>With the additional training data,<br>the algorithm now provides<br>accurate respiratory volumes<br>without patient-specific | A software algorithm calculates<br>respiratory parameters using a<br>proprietary, non-linear, regressive<br>model trained with a dataset<br>containing respiratory data from a<br>large historical cohort. | Both the proposed device and the predicate have essentially the same<br>firmware, except the proposed device has been refined with thousands<br>of additional data points that result in a more accurate estimate of<br>respiratory parameters than the predicate. | |
| Enclosure | calibration.<br>Formed plastic (ABS) housing,<br>flame-retardant. | Formed plastic (ABS) housing,<br>flame-retardant. | Both the proposed device and the predicate have the same enclosure<br>material. | |
| EMC/EMI<br>Compliance | IEC 60601-1-2 compliant | IEC 60601-1-2 compliant | Both the proposed device and the predicate are 60601-1-2 compliant. | |
| Electrode<br>PadSet | Single-patient use, biocompatible,<br>printed padset. | Single-patient use, biocompatible,<br>printed padset. | Both the proposed device and the predicate use the same electrode<br>padset. | |
| Impedance<br>Measurement<br>Range | 15 Ohms to 180 Ohms | 15 Ohms to 180 Ohms | Both the proposed device and the predicate have the same<br>measurement range. | |
| Tablet<br>Computer | Windows tablet computer. | Windows tablet computer. | Following FDA clearance of the predicate (K130170), a "letter to file"<br>change was made to substitute an equivalent Windows tablet<br>computer in the device (See "Letter to File" changes; Section 21b in<br>this 510(k) filing). | |
| Biocompatibility | Neither the Monitor nor the<br>Patient Cable are intended for<br>patient contact. The Electrode<br>PadSet is biocompatible. | Neither the Monitor nor the Patient<br>Cable are intended for patient<br>contact. The Electrode PadSet is<br>biocompatible. | Both the proposed device and the predicate have identical<br>biocompatibility. | |
| Usability | ISO 60601-1-6 compliant | ISO 60601-1-6 compliant | Both the proposed device and the predicate have the same usability<br>when used in the Volume Synchronization and Percent Baseline<br>modes of operation. In the Basic Monitoring mode that has been<br>introduced in the proposed device, the calibration step that was<br>required for the predicate device has been eliminated. | |
| Accuracy | MV - 11.5%<br>TV - 11.4%<br>RR - 0.1% | MV - 10.7%<br>TV - 10.4%<br>RR - 0.1% | Details of the accuracy of the proposed device versus the predicate are<br>included in Section 20: Performance Testing - Clinical. | |
{6}------------------------------------------------
T
{7}------------------------------------------------
| Respiratory Motion, Inc. | | | 09 May 2017 | | |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-----------------------|-----------------|---------------------|
| Nonclinical<br>testing: | Nonclinical testing was performed to demonstrate the equivalence of the ExSpiron 1Xi to<br>the predicate. The FDA recognized tests that were completed are listed below. See<br>Section 9.0; Declarations of Conformity and Summary Reports for details of the testing. | | | | |
| | - Basic Safety and Essential Performance (IEC 60601-1; Ed. 3.0) - Pass,<br>- Immunity/Emissions (IEC 60601-1-2; Ed. 3.0) - Pass,<br>- Usability (IEC 60601-1-6; 2010, A1:2013 and 62366;2007, A1:2014) –<br>Pass, and<br>- Alarms (IEC 60601-1-8; 2006) - Pass. | | | | |
| | Other tests which are not FDA recognized were also completed. The following non-FDA<br>recognized tests were performed and each achieved favorable results: | | | | |
| | - Degrees of protection provided by enclosures (IEC code) (IEC 60529;<br>2004)<br>- Packaged products weighing 150 lbs (68 kg) or less (ISTA Procedure<br>1A; 2014)<br>- ECG Trunk Cables and patient lead wires (ANSI EC53; 2013) | | | | |
| | The results of this testing demonstrate that the proposed device is equivalent to the<br>predicate in safety and essential performance. | | | | |
| Clinical<br>performance<br>testing: | A clinical study was conducted in Waltham MA, USA to compare simultaneous<br>measurements from the ExSpiron 1Xi and the predicate device. Twenty subjects<br>representing a broad range of intended patients participated in the study. (Age range: 22-<br>80, BMI range: 19-42 with 9 female, 11 male). The study involved two sessions for each<br>subject, an initial session in which electrodes were applied and each subject performed<br>twenty breathing tests. Tidal volume, minute ventilation, and respiratory rate were<br>measured and processed by both the Predicate and the Proposed device. Each subject<br>returned twenty-four hours after the first session with the original electrodes still attached.<br>A second set of twenty breathing tests were performed. | | | | |
| | The results of the clinical study were: | | | | |
| | | Proposed ExSpiron 1Xi /Predicate<br>ExSpiron 1Xi Comparison | Minute<br>Ventilation | Tidal<br>Volume | Respiratory<br>Rate |
| | Bias | Overall | 2.2% | 2.2% | 0.01% |
| | | Day 1 | 3.4% | 3.4% | 0.02% |
| | | Day 2 | 1.1% | 1.1% | 0.00% |
| | Precision | Overall | 7.0% | 6.9% | 0.08% |
| | | Day 1 | 5.2% | 5.1% | 0.11% |
| | | Day 2 | 5.1% | 5.1% | 0.00% |
| | Accuracy | Overall | 11.5% | 11.4% | 0.08% |
| | | Day 1 | 10.1% | 10.1% | 0.11% |
| | | Day 2 | 12.3% | 12.3% | 0.00% |
| | The results indicate clinically relevant accuracy over a 24-hour period. | | | | |
| Conclusions<br>regarding<br>safety and<br>effectiveness: | Based on the comparison of the intended use of the predicate and the proposed device, and<br>on the results of nonclinical and clinical testing, the proposed ExSpiron 1Xi is substantially<br>equivalent to the predicate and is as safe and as effective as the predicate. | | | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.