K162125 · Surgtech, Inc. · JDI · May 10, 2017 · Orthopedic
Device Facts
Record ID
K162125
Device Name
MALUC™ Total Hip Arthroplasty System
Applicant
Surgtech, Inc.
Product Code
JDI · Orthopedic
Decision Date
May 10, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SurgTech MALUC™ System is intended for use for cases of severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis or congenital hip dysplasia, avascular necrosis of the femoral head, acute traumatic fracture of the femoral head or neck, failed previous hip surgery (including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty and/or total arthroplasty), and certain cases of ankylosis.
Device Story
MALUC™ Total Hip Arthroplasty System is a prosthetic hip replacement system for primary total hip arthroplasty. Components include monolithic cemented and press-fit femoral stems, modular CoCr and BIOLOX® delta femoral heads, uncemented acetabular shells, and conventional polyethylene liners. Accessory components include distal centralizers, cancellous bone screws, and cement restrictors. The device is implanted by orthopedic surgeons in a clinical/surgical setting to replace a diseased or damaged hip joint. The system restores joint function, reduces pain, and improves mobility for patients suffering from degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through non-clinical mechanical testing in accordance with ISO 21535, ISO 7206-4, ISO 7206-6, ISO 7206-10, ASTM F2996, ASTM F2009, ASTM F1820, and ASTM F543, and bacterial endotoxin testing per USP <85> and ANSI/AAMI ST72.
Technological Characteristics
System includes CoCr and BIOLOX® delta femoral heads, uncemented acetabular shells, and conventional polyethylene liners. Materials and mechanical performance validated per ISO 21535, ASTM F2996, ISO 7206-4/6/10, ASTM F2009, ASTM F1820, and ASTM F543. Sterilization and endotoxin testing per USP <85> and ANSI/AAMI ST72.
Indications for Use
Indicated for skeletally mature patients with severely painful/disabled hip joints due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, congenital hip dysplasia, avascular necrosis, acute traumatic fracture, failed previous hip surgery, or ankylosis.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract human figures, possibly representing individuals or families, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 10, 2017
SurgTech, Inc., % Kellen Hills Quality And Regulatory Consultant Orchid Design 4600 E Shelby Dr Memphis. Tennessee 38118
Re: K162125
Trade/Device Name: MALUCTM Total Hip Arthoplasty System Regulation Number: 21 CFR 888.3350 Regulation Name: Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JDI, LPH, LZO Dated: April 7, 2017 Received: April 11, 2017
Dear Mr. Hills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: QMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known)
#### K 162125 (Page 1 of 1)
#### Device Name
MANDO Fotal Fotal Bo
#### Indications for Use (Describe)
The Surg Tech MALL C ™ System is intended for use for cases of severely painful and or disabled joint from osteoarthritis, traumatic arthritis or congenital hip dysplasia, avascular necrosis of the femoral head acute traumatic fracture of the femoral head or neck, failed previews hip surgery (including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty and or total arthroplasty), and certain cases of ankylos
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D).
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## A STATE FOR FDA USE ONLY CARRET A BREAK A
oncurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| (a)(1) | Submitted By: | SurgTech, Inc.<br>24600 Center Ridge Road, Suite 195<br>Westlake, OH 44145 |
|--------|------------------------------------|----------------------------------------------------------------------------|
| | Phone: | 901-433-1990 |
| | Fax: | 901-433-1989 |
| | Date: | April 7, 2017 |
| | Contact Persons | |
| | Primary: | Kellen Hills (Orchid Design Consulting) |
| | Secondary: | Brian Hewko (SurgTech, Inc.) |
| (a)(2) | Proprietary Name: | MALUC™ Total Hip Arthroplasty System |
| | Common Name: | Total Hip Prosthesis |
| | Classification Name and Reference: | 21CFR 888.3350, 888.3358, 888.3353 |
| | Product Code: | JDI, LPH, LZO |
- (a)(3) Predicate Devices: Primary:
Additional:
DePuy SUMMIT (K023453);
Stryker Exeter V40 (K011623); Stryker Accolade TMZF (K032300); Stryker Trident (K040412, K991952, K022077, K010757, K070885); Depuy PINNACLE with GVF Liner (K000306);
- (a)(4) Device Description:
The MALUC™ Total Hip Arthroplasty System is used for primary total hip replacement in skeletally mature individuals. The system consists of monolithic cemented and press-fit femoral stems, modular CoCr and BIOLOX® delta femoral heads in various sizes and offsets, uncemented acetabular shells and conventional polyethylene liners. Accessory components include distal centralizers, cancellous bone screws and cement restrictors. Instrumentation necessary for proper implantation is also included.
The purpose of this submission is to gain initial marketing authorization in the United States.
(a)(5) Indications for Use:
> The SurgTech MALUC™ System is intended for use for cases of severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis or congenital hip dysplasia, avascular necrosis of the femoral head, acute traumatic fracture of the femoral head or neck, failed previous hip surgery (including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty and/or total arthroplasty), and certain cases of ankylosis.
- Comparison of Technological Characteristics: (a)(6) The MALUC™ Total Hip Arthroplasty System is substantially equivalent to the previously cleared predicate devices based on similarities in intended use, design, materials, packaging, sterilization and
{4}------------------------------------------------
mechanical performance. The technological characteristics do not raise any new questions of safety and efficacy.
- (b)(1) Non-clinical testing:
Mechanical testing in accordance with the following standards was performed and the subject MALUC™ THA System met all predetermined acceptance criteria: ISO 21535, ASTM F2996, ISO 7206-4, ISO 7206-6, ISO 7206-10, ASTM F2009, ASTM F1820, ASTM F543.
Bacterial endotoxin testing was performed in accordance with USP <85> and ANSI/AAMI ST72 and met the predetermined acceptance criteria.
#### (b)(2) Clinical testing:
Clinical testing was not required to demonstrate substantial equivalence in this premarket notification.
#### (b)(3) Conclusions:
Based on the information provided in this premarket notification, we believe that the subject MALUC™ Total Hip Arthroplasty System demonstrates substantial equivalence to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.