The SICONUS™ SI Joint Fixation System is intended to provide fixation of large bones, including the sacrum and ilium. It is intended for use in skeletally mature patients as an adjunct to sacroiliac joint fusion in the treatment of degenerative sacroiliitis, or sacroiliac joint disruptions.
Device Story
Siconus™ SI Joint Fixation System is a partially threaded, fully cannulated metallic bone fixation device; features threaded distal end with buttress-style profile for cancellous bone fixation; proximal head has conical shape for axial compression across sacroiliac joint; smooth shaft connects head and distal thread; large version includes axial slot for bone graft packing to facilitate fusion. Used by orthopedic surgeons in clinical/surgical settings to stabilize sacroiliac joint; device provides mechanical fixation to promote fusion; benefits patient by stabilizing joint and treating degenerative sacroiliitis or disruptions.
Clinical Evidence
Bench testing only. Mechanical testing included static torsion, static bending, dynamic bending, and screw thread pullout per ASTM F2193-02 and ASTM F543-07. Results demonstrate performance equivalent to or exceeding predicate acceptance criteria.
Technological Characteristics
Partially threaded, fully cannulated metallic bone fixation fastener. Features buttress-style distal thread, conical proximal head, and smooth shaft (with axial slot in large version). Materials are biocompatible orthopedic-grade metals. Testing performed per ASTM F2193-02 and ASTM F543-07.
Indications for Use
Indicated for skeletally mature patients requiring fixation/stabilization of sacrum and ilium as an adjunct to sacroiliac joint fusion for degenerative sacroiliitis or sacroiliac joint disruptions.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Xtant (X-Spine) Silex™ SI Joint Fusion System (K123702)
Zyga SImmetry Sacroiliac Joint Fusion System (K102907)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a central emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three figures in profile facing right. The figures are stacked vertically, suggesting a sense of community and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 18, 2017
Camber Spine Technologies % Mr. Justin Eggleton Senior Director, Spine Regulatory Affairs MCRA, LLC 1331 H St NW, 12th Fl Washington, District of Columbia 20005
Re: K162121
Trade/Device Name: Siconus™ SI Joint Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: January 11, 2017 Received: January 13, 2017
Dear Mr. Eggleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
## Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K162121
Device Name Camber Spine Technologies SICONUSTM SI Joint Fixation System
#### Indications for Use (Describe)
The SICONUS™ SI Joint Fixation System is intended to provide fixation of large bones, including the sacrum and ilium. It is intended for use in skeletally mature patients as an adjunct to sacroiliac joint fusion in the treatment of degenerative sacroiliitis, or sacroiliac joint disruptions.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Trade Name | Siconus™ SI Joint Fixation System |
|-----------------------|---------------------------------------------------------|
| Common Name | SI Joint Screw System / SI Joint Fixation System |
| Manufacturer | Camber Spine Technologies |
| | 418 E. Lancaster Ave. |
| | Wayne, PA 19087 |
| Contact | Justin Eggleton |
| | Senior Director, Spine Regulatory Affairs |
| | Musculoskeletal Clinical Regulatory Advisers, LLC |
| | 1331 H Street NW, 12th Floor |
| | Washington, DC 20005 |
| | Phone: (202) 552-5800 |
| | jeggleton@mcra.com |
| Prepared By | Mr. Damian Heinz, PE |
| | Senior Principal Engineer |
| | Institute of Musculoskeletal Science and Education |
| | 418 E. Lancaster Ave. |
| | Wayne, PA 19087 |
| | Phone: 484.420.4286 x307 |
| | Fax: 484.318.8031 |
| Date Prepared | January 17, 2017 |
| Device Class | Class II |
| Classification Name | Smooth or threaded metallic bone fixation fastener |
| Classification Number | 21 CFR §888.3040 |
| Classification Panel: | Orthopedic |
| Product Code | OUR |
| Reason for 510(k) | New Device |
| Predicate Device | Xtant (X-Spine) Silex™ SI Joint Fusion System (K123702) |
| | Synthes Cannulated Screw (K021932) |
| | Zyga SImmetry Sacroiliac Joint Fusion System (K102907) |
### Indications:
The Siconus™ SI Fixation System is intended to provide fixation and stabilization of large bones, including the sacrum and ilium. It is intended for use in skeletally mature patients as an adjunct
{4}------------------------------------------------
to sacroiliac joint fusion in the treatment of the degenerative sacroiliitis, or sacroiliac joint disruptions.
### Device Description:
The Camber Spine Technologies Siconus™ SI Joint Fixation System is a partially threaded, fully cannulated, bone fixation device that has a threaded distal end with a buttress-style profile in order to secure the device in cancellous bone. The proximal head of the device has a conical shape that provides axial compression of the bones across the sacroiliac joint in order to promote fusion. The head of the device also contains a tapered that allows the device to advance into hard cortical bone by displacing material, while also maintaining compression and preventing movement of the device. The proximal head and distal thread is connected by a smooth shaft. In the large version of the device, this shaft contains a slot along the axis to permit packing with bone graft to facilitate fusion.
#### Predicate Device:
The predicate devices selected for this 510(k) are the X-spine Silex™ SI Joint Fusion System (K123702), Synthes Cannulated Screw (K021932), and Zyga SImmetry Sacroiliac Joint Fusion System (K102907). The subject Siconus™ SI Fixation device is substantially equivalent to the predicate devices with respect to indications, intended use, design, function, and performance.
### Substantial Equivalence:
The Siconus™ SI Joint Fixation System is considered to be equivalent in safety and effectiveness (i.e., "substantially equivalent") as the predicates, the X-spine Silex™ Sl Joint Fusion System (K123702), Synthes Cannulated Screw (K021932), and Zyga SImmetry Sacroiliac Joint Fusion System (K102907). All systems utilize comparable surgical techniques. All systems include implants and instruments that are manufactured using biocompatible materials with a long history of use in orthopedic surgery.
{5}------------------------------------------------
### Performance Testing:
Mechanical tests performed indicate that the Siconus™ SI Joint Fixation System performs equally or better as compared to the predicate devices. Testing included static torsion, static bending, dynamic bending, and screw thread pullout per ASTM F2193-02 and ASTM F543-07. The results of these tests demonstrate that the subject device is equivalent or exceeds the acceptance criteria defined by equivalent tests performed on the predicate devices.
### Conclusion:
Camber Spine Technologies has provided sufficient information to demonstrate the Siconus™ SI Joint Fixation System is substantially equivalent to the predicate X-Spine Silex™ Sacroiliac Fusion Device (K123702), Synthes Cannulated Screw (K021932), and Zyga SImmetry Sacroiliac Joint Fusion System (K102907) with respect to indications, intended use, design, function, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.