The introducer sheath is used to facilitate placing a catheter through the skin and into a vein or artery, including, but not limited to, the radial artery. The micropuncture needle or IV catheter is used to gain access to a vein or artery, including, but not limited to, the radial artery, for placement of the mini-guidewire. The mini-guidewire is used for placement of the sheath into a vein or artery, including, but not limited to, the radial artery.
Device Story
InTRAkit Access Kit facilitates percutaneous vascular access for cardiac catheterization. Kit includes micropuncture needle or IV catheter, mini-guidewire, introducer sheath, and dilator. Physician uses needle/catheter to gain initial vascular access; inserts mini-guidewire; advances sheath/dilator over wire to establish access site. Used in clinical settings (e.g., cath lab). Single-use, sterile device. Provides conduit for subsequent catheter placement. Benefits patient by enabling minimally invasive vascular entry.
Clinical Evidence
No clinical data. Evidence consists of bench testing and biocompatibility testing. Design verification performed per ISO 11070:2014. Biocompatibility testing performed per ISO 10993-1:2009/Cor 1:2010, including cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, genotoxicity, hemolysis, complement activation, partial thromboplastin time, and in-vivo thromboresistance.
Technological Characteristics
Components: 4F-6F introducer sheath (7-23cm length), 18G-21G micropuncture needles, 20G-22G IV catheter needles, 0.021" diameter mini-guidewires (45-80cm length). Sterile, single-use. Complies with ISO 11070:2014 and ISO 10993-1:2009/Cor 1:2010.
Indications for Use
Indicated for patients requiring percutaneous vascular access into a vein or artery, including the radial artery, for cardiac catheterization procedures.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2016
Medtronic Vascular Nisarg Shah Senior Regulatory Affairs Specialist 37A Cherry Hill Drive Danvers, Massachusetts 01923
Re: K162097
Trade/Device Name: InTRAkit Access Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: October 18, 2016 Received: October 19, 2016
Dear Nisarg Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/7 description: The image shows the name "Brian D. Pullin -S" in a clear, sans-serif font. The text is horizontally oriented and appears to be part of a larger document or label. The background is a light color, providing good contrast for the dark text.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162097
Device Name InTRAkit™ Access Kit
#### Indications for Use (Describe)
The introducer sheath is used to facing a catheter through the skin and into a vein or artery, including, but not limited to, the radial artery. The micropuncture needle or IV catheter is used to gain access to a vein or artery, including, but not limited to, the radial artery, for placement of the mini-guidewire is used for placement of the sheath into a vein or artery, including, but not limited to, the radial artery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary - K162097
| Submitter: | Medtronic Vascular<br>37A Cherry Hill Drive,<br>Danvers, Massachusetts 01923, USA | | |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-----------------------|
| Contact Person: | Nisarg Shah<br>Senior Regulatory Affairs Specialist<br>37A Cherry Hill Drive<br>Danvers, MA 01923, USA<br>Phone: (978) 739-6632<br>Fax: (978) 750-8204<br>Email: nisarg.g.shah@medtronic.com | | |
| Date Prepared: | November 16, 2016 | | |
| Trade Name(s): | InTRAkit™ Access Kit | | |
| Common Name: | Catheter Introducer | | |
| Classification Name: | Catheter Introducer<br>Class II per 21 CFR §870.1340<br>Product Code: DYB | | |
| Predicate Device: | Device Name | Manufacturer | 510(k)<br>clearance # |
| | Glidesheath® Catheter<br>Introducer | Terumo Medical<br>Corporation | K082644 |
Medtronic's InTRAkit™ device is a Catheter Introducer (PCI) Device Description: which is intended to facilitate percutaneous vascular access into a vein or artery, performed as part of the Cardiac catheterization procedure. The InTRAkit™ Catheter Introducers are designed to facilitate access into the vasculature from different access sites including but not limited to the radial arteries. InTRAkit™ Catheter Introducers are single-use use devices sold in a sterile condition. The InTRAkit™ Access Kit includes four main type of components required for percutaneous vascular access procedure:
{4}------------------------------------------------
| | 1) Needles<br>2) Mini-Guidewires<br>3) Introducer Sheath<br>4) Dilator |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Statement of<br>Intended Use: | The introducer sheath is used to facilitate placing a catheter<br>through the skin and into a vein or artery, including, but not<br>limited to, the radial artery. The micropuncture needle or IV<br>catheter is used to gain access to a vein or artery, including,<br>but not limited to, the radial artery, for placement of the mini-<br>guidewire. The mini-guidewire is used for placement of the<br>sheath into a vein or artery, including, but not limited to, the<br>radial artery. |
| Summary of<br>Technological<br>Characteristics: | Medtronic's InTRAkit™ device components comprise of<br>following technological characteristics: |
| | Introducer Sheath and Dilator: Sheath size: 4French to 6French Sheath length: 7 to 23cm Needles: Micropuncture Needles Gauge size: 18 to 21G IV Catheter Needles Gauge size: 20 to 22G Mini-Guidewires Diameter size: 0.021" Length: 45- 80cm |
| Comparison to the<br>predicate devices: | The following information outlines the differences and<br>similarities between the subject device and the predicate<br>device: Similar Intended Use/ Indication for Use Similar Device Design Component/ Construction Different device materials |
| | Medtronic's InTRAkit™ Access Kit is substantially |
equivalent to the predicate device based on similarities in intended use and technological characteristics. The testing
{5}------------------------------------------------
on the InTRAkit™ Catheter performed Introducers demonstrates that the technological differences do not raise any new concerns or risks relevant to the intended use of the device.
The following non-clinical testing was performed on Summary of Non-Medtronic's InTRAkit™ device: clinical Data:
## 1. Design Verification Testing/ In-vitro bench testing:
Design Verification (DV) testing was completed on the InTRAkit™ device in accordance with the requirements of the ISO 11070: 2014 - Sterile single-use intravascular introducers, dilators and guidewires.
### 2. Biocompatibility Testing:
The following Biocompatibility Testing was completed on InTRAkit™ device in compliance with the the requirements of ISO 10993-1: 2009/ Cor 1: 2010-Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
- Cytotoxicity testing ●
- Sensitization testing ●
- Intracutaneous reactivity testing ●
- Acute systemic toxicity testing ●
- Pyrogenicity testing
- Genotoxicity testing ●
- Hemolysis testing
- C3a and SC5b-9 Complement Activation study ●
- Partial Thromboplastin Time study ●
- In-vivo Thromboresistance testing ●
No new risks or concerns relevant to the device's intended use were raised during the testing. The bench testing qualification and biocompatibility testing demonstrated that InTRAkit™ is substantially equivalent to the predicate device.
Summary of No clinical investigation was performed on the subject device Clinical Data: (InTRAkit™).
{6}------------------------------------------------
| Conclusion from | The data provided in this 510(k) premarket notification |
|-----------------|---------------------------------------------------------|
| Data: | demonstrated that the subject device is substantially |
| | equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.