← Product Code [GWJ](/productcode/GWJ) · K162037

# Eclipse with VEMP (K162037)

_Interacoustics A/S · GWJ · Mar 23, 2017 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162037

## Device Facts

- **Applicant:** Interacoustics A/S
- **Product Code:** [GWJ](/productcode/GWJ.md)
- **Decision Date:** Mar 23, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1900
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Pediatric

## Indications for Use

The Eclipse with VEMP (Vestibular Evoked Myogenic Potential) is intended for vestibular evoked myogenic potential testing to assist in the assessment of vestibular function. The target population for Eclipse with VEMP includes patients aged from 8 years and up. The device is to be used only by qualified medical personnel with prior knowledge of the medical and scientific facts underlying the procedure.

## Device Story

Eclipse with VEMP is a PC-based audiometric system for performing Vestibular Evoked Myogenic Potential (VEMP) testing. It utilizes a hardware platform, preamplifier, stimulation transducers, and recording electrodes to measure evoked potentials from sternocleidomastoid (cVEMP) or inferior oblique (oVEMP) muscles. The system provides auditory stimulation to assess otolith organs (saccule/utricle) and afferent pathways. Operated by qualified medical personnel in hospitals or clinics, the device includes an integrated EMG monitor to guide electrode placement and ensure muscle activation is within range for recording. The software displays waveforms for manual, subjective interpretation by the clinician. By facilitating VEMP testing, the device assists medical practitioners in diagnosing balance disorders. It does not provide an automated diagnosis.

## Clinical Evidence

Two internal clinical evaluations (test-retest reliability and reproducibility) were conducted with 14 normal adults. cVEMP and oVEMP waveforms were recorded using 500Hz tone bursts at 95dBnHL. Test-retest correlation means were 0.9374 (cVEMP) and 0.9095 (oVEMP). Reproducibility study correlation means were 0.8794 (cVEMP) and 0.8923 (oVEMP). Results indicate high reliability and reproducibility, consistent with or exceeding performance metrics reported for the predicate device (K143670).

## Technological Characteristics

PC-based system with external hardware platform, preamplifier, stimulation transducers, and recording electrodes. Connectivity via USB interface. Energy source: AC power. Features integrated EMG monitoring for VEMP data collection. Software-controlled protocols. No changes to electrical safety or EMC compared to the cleared Eclipse platform (K090406).

## Regulatory Identification

An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.

## Predicate Devices

- ICS CHARTR EP 200 WITH VEMP ([K143670](/device/K143670.md))
- Eclipse with EP15, EP25 software modules ([K090406](/device/K090406.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 23, 2017

Interacoustics A/S Mr. Erik Nielsen Director, Regulatory & Compliance Audiometer Allé 1 5500 Middelfart, Denmark

Re: K162037

Trade/Device Name: Eclipse with VEMP Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: Class II Product Code: GWJ Dated: February 08, 2017 Received: February 21, 2017

Dear Mr. Nielsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162037

Device Name Eclipse with VEMP

Indications for Use (Describe)

The Eclipse with VEMP (Vestibular Evoked Myogenic Potential) is intended for vestibular evoked myogenic potential testing to assist in the assessment of vestibular function. The target population for Eclipse with VEMP includes patients aged from 8 years and up.

The device is to be used only by qualified medical personnel with prior knowledge of the medical and scientific facts underlying the procedure.

Type of Use (Select one or both, as applicable)

|  | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
|  | Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(K) SUMMARY

### Eclipse with VEMP

### Submitter Information:

| Company Name          | Interacoustics A/S                 |
|-----------------------|------------------------------------|
| Address               | Audiometer Allé                    |
|                       | 5500 Middelfart                    |
|                       | Denmark                            |
| Phone                 | +45 6371 3555                      |
| e-mail                | erni@dgs.com                       |
| Contact Person        | Erik Nielsen,                      |
|                       | Director, Regulatory & Compliance, |
| Date Summary Prepared | July 18 2016                       |

## Device Identification:

| Trade Name          | Eclipse with VEMP                       |
|---------------------|-----------------------------------------|
| Common Name         | Auditory Evoked Response Stimulator     |
| Classification Name | Stimulator, Auditory, Evoked Response   |
| Product Code Class  | GWJ                                     |
| Panel               | Neurology                               |
| Device Class        | Class II (According to 21 CFR 882.1900) |

#### Predicate Device 1:

Predicate Device Manufacturer 510(k) No. Date Cleared

ICS CHARTR EP 200 WITH VEMP GN OTOMETRICS A/S7 K143670 10/23/2015

## Predicate Device 2:

Predicate Device Manufacturer 510(k) No. Date Cleared

Eclipse with EP15, EP25 software modules Interacoustics A/S K090406 07/22/2009

{4}------------------------------------------------

### Device Description

The Eclipse with VEMP is audiometric equipment intended to perform various Otoacoustic Emissions (OAEs) and Auditory Evoked Potential evaluations. The Eclipse is operated solely from PC based software modules. The Eclipse platform performs the physical measurements. The protocols are created in the software modules. The Eclipse consists of a hardware platform, a preamplifier, stimulation transducers and recording electrodes.

VEMP evaluations are tests of the vestibular portion of the inner ear and acoustic nerve, evoked with an auditory stimulation. The evoked response results in a potential recorded from the sternocleidomastoid (neck) muscles or the inferior oblique (eye) muscles. VEMP is not a test of the neck or eye musculature directly; the clinician is interested in the vestibular anatomy which triggers the response. The cervical Vestibular Evoked Myogenic Potential (cVEMP) is an evoked potential measured from the sternocleidomastiod (SCM) muscle and the ocular VEMP (oVEMP) is an evoked potential measured from the inferior oblique muscle. Both tests are used to assess the otolith organs (saccule and utricle) of the vestibular system and their afferent pathways and assist medical practitioners in the diagnosis of various balance disorders.

Summary: VEMP is Auditory Evoked Potentials like ABR obtained using any commercially available EP system. The addition of the VEMP module to Eclipse will make it possible for clinicians to conduct VEMP tests while using EMG (electromyography) monitoring and scaling. The VEMP function of the Eclipse with VEMP does not make a diagnosis. It only assists the medical professional.

## Indications for Use

The Eclipse with VEMP (Vestibular Evoked Myogenic Potential) is intended for vestibular evoked myogenic potential testing to assist in the assessment of vestibular function. The target population for Eclipse with VEMP includes patients aged from 8 years and up.

## Technological characteristics

The Eclipse with VEMP consists of a hardware platform, a preamplifier, stimulation transducers and recording electrodes. The Eclipse VEMP is solely operated from a PC with VEMP software

{5}------------------------------------------------

### Product Comparisons charts summary

We have chosen to compare Eclipse with VEMP with the currently cleared Eclipse with EP15, EP25 and the currently cleared Chartr EP200 for the following reasons.

- Chartr EP200 has the same medical purpose as Eclipse with VEMP and is intended as the . primary predicate for VEMP functionality
- . Eclipse with EP15, EP25 supports identical hardware as Eclipse with VEMP and provide identical technical characteristics

| Description              | Predicative K143670<br>Chartr EP 200 with VEMP                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Modified K162037<br>Eclipse with VEMP                                                                                                                                                                                                                                                      | Equivalence                                                                                                                                                                                                                                                        |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use   | The ICS Chartr EP 200 with VEMP is<br>indicated for auditory evoked<br>potential testing as an aid in<br>assessing hearing loss and lesions in<br>the auditory pathway. The<br>Vestibular Evoked Myogenic<br>Potential is indicated for vestibular<br>evoked potential testing as an aid in<br>assessing vestibular function in<br>adult patients. The device is to be<br>used only by qualified medical<br>personnel with prior knowledge of<br>the medical and scientific facts<br>underlying the procedure. | The Eclipse with VEMP (Vestibular<br>Evoked Myogenic Potential) is<br>intended for vestibular evoked<br>myogenic potential testing to assist in<br>the assessment of vestibular function.<br>The target population for Eclipse with<br>VEMP includes patients aged from 8<br>years and up. | Same intended<br>use of assisting<br>in the<br>assessment of<br>vestibular<br>function, but<br>the Eclipse with<br>VEMP is<br>indicated for<br>children 8 years<br>and older and<br>substantial<br>equivalence<br>was established<br>using clinical<br>literature. |
| Tests<br>Performed       | Auditory Brainstem Response (ABR)<br>Electrocochleography (ECochG)<br>Auditory Middle Latency Response<br>(AMLR)<br>Auditory Late Response (ALR)<br>P300 - optional<br>Auditory Steady State Response<br>(ASSR) — Optional<br>Vestibular Evoked Myogenic<br>Potential (VEMP) — optional                                                                                                                                                                                                                        | Auditory Brainstem Response (ABR)<br>Electrocochleography (ECochG)<br>Auditory Middle Latency Response<br>(AMLR)<br>Auditory Late Response (ALR)<br>P300/Mismatch Negativity (MMN)<br>Auditory Steady State Response (ASSR)<br>Vestibular Evoked Myogenic Potential<br>(VEMP) — optional   | Same<br>The addition of<br>VEMP is the<br>subject of this<br>510(k                                                                                                                                                                                                 |
| Anatomical<br>sites      | Examination of Ear and hearing<br>nerves and vestibular functions                                                                                                                                                                                                                                                                                                                                                                                                                                              | Examination of Ear, hearing nerves and<br>vestibular functions                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                               |
| Environment<br>of use    | Hospitals and clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Hospitals and clinics                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                               |
| Evaluation of<br>results | Manually / subjective                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Manually / subjective                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                               |
| Description              | Predicative K143670<br>Chartr EP 200 with VEMP                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Modified K162037<br>Eclipse with VEMP                                                                                                                                                                                                                                                      | Equivalence                                                                                                                                                                                                                                                        |
| Photo                    | Image: Chartr EP 200 with VEMP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Image: Eclipse with VEMP                                                                                                                                                                                                                                                                   | Similar<br>The physical<br>differences are<br>not appraised<br>to raise any<br>issues about<br>safety and<br>efficiency                                                                                                                                            |
| System                   | Chartr EP 200 connected to a PC<br>(PC-based system with external<br>hardware platform and peripherals<br>(USB interface))                                                                                                                                                                                                                                                                                                                                                                                     | Eclipse platform connected to PC (PC-<br>based system with external hardware<br>platform and peripherals (USB<br>interface))                                                                                                                                                               | Similar<br>The physical<br>differences are<br>not appraised<br>to raise any<br>issues about<br>safety and<br>efficiency                                                                                                                                            |
| VEMP<br>Monitor          | Image: VEMP Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | A visual or acoustical VEMP monitor is<br>available. The visual VEMP monitor is<br>displayed on an external screen and<br>indicates below range (low), in range<br>for recording (good) and outside range<br>(high).                                                                       | Similar.<br>Both methods<br>have three<br>ranges for<br>assessing the<br>EMG<br>monitoring. We<br>appraise the<br>difference in<br>presentation<br>does not raise<br>any issues<br>about safety<br>and efficiency                                                  |
| Electrode<br>montage     | 2x on each SCM muscle one for<br>recording and one for EMG<br>monitoring on external device, 1 at<br>clavicle junction, 1 low forehead.<br>In total 6 electrodes                                                                                                                                                                                                                                                                                                                                               | 1 on each SCM which is for both<br>recording and monitoring via the built<br>in software EMG monitor, 1 at clavicle<br>junction, 1 at low forehead.<br>In total 4 electrodes                                                                                                               | Similar.<br>Both methods<br>measures at the<br>same anatomic<br>place. The<br>difference in<br>configurations<br>are not<br>appraised to<br>raise any issues<br>about safety<br>and efficiency                                                                     |
| Description              | Predicative K143670<br>Chartr EP 200 with VEMP                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Modified K162037<br>Eclipse with VEMP                                                                                                                                                                                                                                                      | Equivalence                                                                                                                                                                                                                                                        |
| Preamplifier             | 2 channels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 1 or 2 channels                                                                                                                                                                                                                                                                            | Similar<br>(subset)<br>The differences<br>are appraised<br>not to raise any<br>issues about<br>safety and<br>efficiency                                                                                                                                            |

{6}------------------------------------------------

{7}------------------------------------------------

| Description          | Predictive K090406<br>Eclipse with EP15, EP25                                                                                | Modified K162037<br>Eclipse with VEMP                                                                                        | Equivalence                                                                           |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Target<br>population | All ages and genders                                                                                                         | From age 8 and up both genders.                                                                                              | Similar<br>(Subset)<br>The VEMP<br>functionality<br>has a more<br>narrow age<br>range |
| Stimulus             | Click, Tone Burst, NB CE-Chirp, CE-<br>Chirp                                                                                 | Click, Tone Burst, NB CE-Chirp, CE-<br>Chirp                                                                                 | Same                                                                                  |
| Energy<br>Source     | AC Power                                                                                                                     | AC Power                                                                                                                     | Same                                                                                  |
| Photo                | Image: Predictive K090406 Eclipse with EP15, EP25                                                                            | Image: Modified K162037 Eclipse with VEMP                                                                                    | Same                                                                                  |
| System               | Eclipse platform connected to PC (PC-<br>based system with external hardware<br>platform and peripherals (USB<br>interface)) | Eclipse platform connected to PC (PC-<br>based system with external hardware<br>platform and peripherals (USB<br>interface)) | Same                                                                                  |

{8}------------------------------------------------

#### Brief discussion of non-clinical tests

Bench verification and testing demonstrates compliance with system and software specifications. There are no changes to the hardware including electrical safety or EMC compared to the already cleared Eclipse platform (K090406). Software verification including risk analysis was performed to demonstrate compliance with the specifications

#### Brief discussion of clinical evaluations

VEMP, like ABR, is an evoked potential obtained using any commercially available EP system. The Eclipse with VEMP module has the ability to perform EMG monitoring during VEMP data collection. The VEMP module of the Eclipse does not make a diagnosis is made by a qualified medical professional. Testing the vestibular evoked myogenic potential (VEMP) is welldocumented in the literature.

Two internal clinical evaluations have been conducted to show the test-reliability and the reproducibility of the VEMP using the Eclipse with VEMP module

A summary of the studies follows:

Testing was conducted at two sites (one in Denmark and one in the United States). Test subjects included 14 normal adults for cVEMP and oVEMP.

| Min number of runs per ear            | 2 repeatable waveforms                             |
|---------------------------------------|----------------------------------------------------|
| Transducer                            | Insert phones                                      |
| Stimulus                              | 500Hz tone burst (Blackman)                        |
| Stimulus rate                         | 5.1/second                                         |
| Stimulus level                        | 95dBnHL                                            |
| High Pass Filter                      | 10Hz                                               |
| Low Pass Filter                       | 1000Hz                                             |
| # of sweeps                           | ~150 samples per waveform                          |
| EMG controlled stimulus (c VEMP only) | 50.0uV RMS to 150uV RMS (subject to aim for 100uV) |

Collection parameters for both the cVEMP and oVEMP tests:

The correlation results for the test-retest reliability study were as follows:

#### Correlation of the cVEMP waveform (5 – 35ms)

| Mean                         | 0.9374 |
|------------------------------|--------|
| Median                       | 0.9562 |
| IQR (75th – 25th percentile) | 0.0547 |

#### Correlation of the oVEMP waveform (4 – 20ms)

| Mean                         | 0.9095 |
|------------------------------|--------|
| Median                       | 0.9280 |
| IQR (75th – 25th percentile) | 0.0738 |

{9}------------------------------------------------

Mean and median values indicate a high degree of test-retest repeatability of the cVEMP and oVEMP measurements. The IQR (75th – 25th percentile) values are low, indicating a large number of the correlation coefficients are clustered near the median. In practical terms, the spread in correlation coefficients is not so large indicating, alongside the high median value, that test-retest measurements are typically highly correlated with each other for both the cVEMP and oVEMP measurements.

The correlation results for the reproducibility study were as follows:

| Mean                         | 0.8794 |
|------------------------------|--------|
| Median                       | 0.9235 |
| IQR (75th – 25th percentile) | 0.0955 |

Correlation of the cVEMP waveform (5 – 35ms)

Correlation of the oVEMP waveform (4 – 20ms)

| Mean                         | 0.8923 |
|------------------------------|--------|
| Median                       | 0.9100 |
| IQR (75th - 25th percentile) | 0.1112 |

Mean and median values indicate a high degree of reproducibility for the cVEMP and oVEMP measurements recorded on different days. The IQR (75th – 25th percentile) values are higher than for the test-retest cVEMP and oVEMP measurements. They are still relatively low, indicating a large number of the correlation coefficients are clustered near the median. The reason for the increase in IQR is probably to be expected, as each of the cVEMP and oVEMP measurements were made on different days with slightly different electrode placements, by different testers.

In the GN Otometrics K143670 (510(k)) document, mean correlation values for both cVEMPs and oVEMPs are reported for a reproducibility study (see table below). The mean correlation values recorded with the Eclipse with VEMP module for both the test-retest reliability study and the reproducibility study are similar to or higher than those reported by GN Otometrics in their 510(k).

| Correlation values reported in K143670 (510(K)) |                         |                      |
|-------------------------------------------------|-------------------------|----------------------|
|                                                 | cVEMP (5 - 35ms)        | oVEMP (4 - 20 ms)    |
| Mean                                            | 0.9146 (R) / 0.9162 (L) | 0.926 (R) / 0.93 (L) |

## Conclusion

We appraise that the results of this comparison, the clinical evaluations and literature review, demonstrates that the Eclipse with VEMP (the modified indications for use) is safe and efficient and is substantially equivalent to the marketed predictive devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K162037](https://fda-staging.innolitics.com/device/K162037)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).