Z-Box/Z-Tilt Breast and Thoracic Positioning Device, Z-Square Indexing System, Z-Foam, Z-Bar Index Bar, Z-Pulls Patient Hand Grip Overlay
K162002 · Zdi, Solutions, LLC · IYE · Nov 1, 2016 · Radiology
Device Facts
Record ID
K162002
Device Name
Z-Box/Z-Tilt Breast and Thoracic Positioning Device, Z-Square Indexing System, Z-Foam, Z-Bar Index Bar, Z-Pulls Patient Hand Grip Overlay
Applicant
Zdi, Solutions, LLC
Product Code
IYE · Radiology
Decision Date
Nov 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
ZDi Z-Tilt is intended for use by trained medical professionals for patient support and immobilization during external beam radiation therapy in clinics and hospitals. The Z-Tilt and accessories provide patient comfort and stability in various positions for radiation treatment. The Z-Tilt and is used in the treatment of cancer under the supervision and administration of licensed doctors and clinical staff.
Device Story
Z-Tilt is a patient positioning system for external beam radiation and proton therapy; provides support and immobilization. Device consists of a carbon fiber backboard with a scoop seat for hip positioning and thoracic alignment; includes hinged legs for 4, 8, and 12-degree incline adjustments. Accessories include Z-Square indexing system, Z-Bar, Z-Box for custom molds, Z-Pulls for arm positioning, Z-Foam, and Z-Roll. Used in clinics/hospitals by trained medical professionals. Device ensures reproducible patient positioning; minimizes beam attenuation via thin carbon fiber construction. Output is a stable, indexed platform for patient treatment. Benefits include improved patient comfort, stability, and accurate radiation delivery.
Clinical Evidence
Bench testing only. Mechanical weight-bearing tests performed at 4, 8, and 12-degree inclines (250 lbs for 60 minutes). Material testing evaluated radiation beam attenuation (absorption factors) for Co-60, 6MV, 8MV, and 15MV energies. Stopping power calculated for 230meV proton beam. No clinical data provided.
Technological Characteristics
Carbon fiber composite construction; base plate with adjustable legs; thoracic and table indexing systems; arm positioning components. Dimensions: 16-inch width, tapering to 10-inch square. Radiation attenuation: 0.32 cm carbon fiber thickness. Non-invasive, mechanical device. No software or energy source.
Indications for Use
Indicated for patients undergoing external beam radiation therapy, including proton therapy, requiring support and immobilization for cancer treatment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of four stylized human profiles stacked on top of each other. The profiles are connected by a flowing line, creating a sense of unity and collaboration.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 1, 2016
ZDi, Solutions, LLC % Mr. Zachary Dutton President 800 S. Gay Street, Suite 714 KNOXVILLE TN 37929
Re: K162002
Trade/Device Name: Z-Box/Z-Tilt Breast and Thoracic Positioning Device. Z-Square Indexing System. Z-Foam Cushion and Z-Bar Index Bar and Z-Pulls Patient Hand Grip Overlay Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: October 14, 2016 Received: October 20, 2016
Dear Mr. Dutton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162002
### Device Name
Z-Box/Z-Titt Breast and Thoracic Positioning Device and Z-Square Indexing System and Z-Bar Index Bar and Z-Pulls Patient Hand Grip Overlay
### Indications for Use (Describe)
ZDi Z-Tilt is intended for use by trained medical professionals for patient support and immobilization during external beam radiation therapy in clinics and hospitals. The Z-Tilt and accessories provide patient comfort and stability in various positions for radiation treatment. The Z-Tilt and is used in the treatment of cancer under the supervision and administration of licensed doctors and clinical staff.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
D Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ZDi solutions. The letters "ZDi" are in bold, black font, with a light blue plus sign above the "i". The word "solutions" is written in a smaller font below the letters. The logo is surrounded by a blue oval shape.
Patient Positioning Devices for Radiation and Proton Therapy ZDi, Solutions, LLC Performance Data 510(k) - K162002
510(k) Summary
#### l. Submitter
ZDi, Solutions, LLC 800 S Gay St, STE 714 Knoxville, TN 37909
Phone: 865-809-7283
Contact Person: Zach Dutton Date Prepared: July 18, 2016
#### II. Device
Name of Device: Z-Tilt Breast and Thoracic Positioning Device and Accessories Common or Usual Name: Radiation Therapy Patient Positioning Device Classification Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: II Product Code: IYE
#### lll. Predicate Devices
Klarity Medical Products, Inc. Klarity Breast Board (R610), K982624
This predicate device has not been subject to a design-related recall.
#### IV. Device Description
The Z-Tilt is a patient-positioning device designed specifically for applications within Proton Therapy. The Z-Tilt features a thin, low-stopping power backboard able to hold the heaviest of patients treated with proton therapy. The Z-Tilt offers a scoop feature that acts as a seat for patients to rest against. This seat allows an easily reproducible, fixed hip position. The scoop prevents the patient from sliding down the tilt, and also ensures thoracic alignment in the treatment area. The Z-Tilt is sixteen inches wide at its widest point, tapering into a ten-inch square to receive another ZDi device, the Z-Square. The Z-Square easily indexes to the Z-Tilt backboard to securely lock and index the custom foam patient molds to the device. The Z-Tilt is designed to be strong and sturdy, but with minimal amounts of material in critical areas of potential beam passage. The Z-Tilt backboard is attached to a slim, strong base. The base is narrow in the middle to allow beam passage, but wide enough to prevent wobble. The cranial end of the base is also a
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ZDi Solutions. The letters "ZDI" are in a bold, black font, with a blue plus sign above the "i". The word "solutions" is written in a smaller, blue font below the letters. The logo is surrounded by a blue oval shape.
Patient Positioning Devices for Radiation and Proton Therapy ZDi, Solutions, LLC
Performance Data 510(k) - K162002
ten-inch square, holding hinged support legs to brace the backboard at various positions equaling four, eight, and twelve degrees; the most commonly needed patient tilt positions. The Z-Tilt base is designed with indexing holes to index to any standard index bar used on treatment tables. It is designed to remain within sixteen inches wide to prevent interference with table edges, table tracks, and CT bores. The Z-Tilt will have marks on each hole for accurate reproducibility of the Z-Square indexing system and for the leg supports for the incline. The patient will be in contact with their custom mold used on top of the Z-System.
The associated accessories include:
- Z-Square Indexing System for patient mold reproducibility ●
- Z-Bar Index Bar for Z-Square mounting ●
- Z-Box for custom formation of patient molds
- Z-Pulls for arms-down reproducible arm positioning
- . Z-Foam for various foam patient supports under the custom patient molds
- Z-Scoop for thoracic positioning on the Z-Tilt and table
- Z-Roll for lateral positioning of the Z-Tilt ●
#### V. Indications for Use:
These devices are intended for use by trained medical professionals for patient support and immobilization during external beam radiation therapy in clinics and hospitals. The devices provide patient comfort and stability in various positions for radiation treatment. The devices are used in the treatment of cancer under the supervision and administration of licensed doctors and clinical staff.
#### VI. Comparison of Technological Characteristics with the Predicate Device
Patient positioning for Radiation Therapy treatments is the technological principle for both the subject and predicate devices. The devices must support the patient in various positions to ensure the accurate delivery of a radiation therapy treatment. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Base Plate with adjustable legs for backboard support
- Backboard for overall patient support
- Arm positioning system
- Thoracic Indexing System
- Table Indexing System
- Alignment marks
- Mainly carbon fiber design, some plastic and metal components
The following technological differences exist between the subject and predicate devices, but in no way alter the safety or effectiveness of the subject device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ZDI solutions. The letters ZDI are in bold black font, with a light blue plus sign above the I. The word "solutions" is written in a smaller font below the letters ZDI. The logo is surrounded by a blue oval shape.
Patient Positioning Devices for Radiation and Proton Therapy ZDi, Solutions, LLC Performance Data 510(k) - K162002
- Seat-like thoracic positioner
- Winged arm supports
- Foam Accessories ●
- . Custom Molding accessories
#### VII. Performance Data
Both the subject and the predicate consist of carbon fiber components which allow minimal radiation beam attenuation. With both devices possessing the same dimensions and characteristics, the beam influence will be the same.
Mechanical Testing
The subject device underwent a weight test at each of the three degrees of incline. Each angle held the manufacturer recommendation of 250 pounds for a 60 minute time frame. Each accessory was able to hold their respective manufacturer suggested weights of 50 lbs each for 60 minutes.
Material Testing
These are the standard dose effects on various beam energies for carbon fiber, the main component of the subject and predicate devices.
| | | Absorption Factor % | | | |
|-----------|--------------|---------------------|-----|-----|------|
| Thickness | Material | Co-<br>60 | 6MV | 8MV | 15MV |
| 20mm | Carbon Fiber | 2.6 | 2 | 1.7 | 1.5 |
Carbon Fiber thickness (Minimum and Maximum) in Z-System - .32 cm
.32 cm Carbon Fiber WET for 230 meV (max me\/)= 0.12 cm
Stopping Power ("attenuation" in proton beam) = 3.657e/cm2 for a 230meV proton beam in Carbon Fiber
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ZDi solutions. The logo features the letters "ZDi" in a bold, black font, with a light blue plus sign above the "i". The word "solutions" is written in a smaller font below the letters. The logo is surrounded by a blue oval shape.
Patient Positioning Devices for Radiation and Proton Therapy ZDi, Solutions, LLC Performance Data 510(k) - K162002
# Biocompatibility
The subject and predicate devices are both carbon fiber composites. Per article – carbon fiber, in vivo and in vitro, has no cytotoxic effects. Furthermore, the subject device will be used in conjunction with a custom patient mold. The expandable foam and vacuum locking molds are already approved devices offered by Med-Tech, Inc, K951808 and K935300, respectively. The Z-System is an aide for making and supporting these other devices.
## Effectiveness
The subject and predicate devices are both similarly designed. Clinics using the predicate device have had successful outcomes for treatments. The device is not responsible for the actual treatment, rather it is only used as an aid for aligning the patient and supporting the patient while receiving their treatment. Both devices are non-invasive and the subject device, considering all features, is no less effective than the predicate device.
#### VIII. Conclusions
Since the predicate device was cleared based on substantial equivalence to previously cleared, comparable devices, using the same materials, design, and having the same intended uses and manufacturing processes, the ZDi, Solutions, LLC Z-Tilt and accessories should be cleared based on their similarities to the predicate device and other devices. Technological, safety, effectiveness, biological, and all other factors of the subject Z-Tilt device are all substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.