Horizontal-Vertical Lumbar Valve Systems, Spetzler Lumbar Peritoneal Shunt Systems
Applicant
Integra LifeSciences Corporation
Product Code
JXG · Neurology
Decision Date
Sep 22, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Horizontal-Vertical Lumbar Valve Systems are implantable devices used in the treatment of patients with communicating hydrocephalus to shunt cerebrospinal fluid (CSF) from the lumbar subarachnoid region to the perioneal cavity. They provide controlled intraventricular pressure and CSF drainage in patients with hydrocephalus. The antechamber can be electively mounted in line with the Valve Unit to allow for CSF sampling or injections in the subarachnoid space. Spetzler Lumbar Peritoneal Shunt Systems Percutaneous lumbar perioneal shunting may be utilized in the treatment of communicating hydrocephalus. It is designed to shunt CSF from the lumbar subarachnoid space to the peritoneal cavity. The shunt may be used for diagnosis, evaluation or treatment of normal pressure communicating hydrocephalus. A percutaneous lumbar peritoneal shunt is also useful in the management of persistent cerebrospinal fluid fistulas, bulging cranial and suboccipital decompressions and in transient CSF absorption defects, e.g. post-memingitic or post-hemorrhagic hydrocephalus. The In-Line Valve, available as a separate component of the system, is indicated for use where added resistance is desired to alleviate symptoms of low pressure in the small percentage of patients who, after normal drainage and in the normal course of treatment, develop such symptoms.
Device Story
Implantable shunt systems for communicating hydrocephalus; diverts CSF from lumbar subarachnoid space to peritoneal cavity. Horizontal-Vertical Lumbar Valve System uses differential pressure valve with two ball-in-cone mechanisms: spring-actuated for horizontal position; gravity-actuated for vertical position; minimizes overdrainage by compensating for hydrostatic pressure changes. Spetzler Lumbar Peritoneal Shunt System uses double-slit valve at peritoneal end combined with small-diameter catheter for pressure control. Percutaneous insertion via Tuohy spinal needle. Optional in-line valve provides added resistance for low-pressure symptom management. Used by neurosurgeons in clinical settings. Benefits include controlled CSF drainage and pressure regulation; management of fistulas and absorption defects.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including magnetically induced displacement force/torque, RF heating, image artifact, and pressure flow testing, as previously cleared under K101381 and updated per K152897.
Indicated for patients with communicating hydrocephalus, normal pressure communicating hydrocephalus, persistent CSF fistulas, bulging cranial/suboccipital decompressions, and transient CSF absorption defects (e.g., post-meningitic or post-hemorrhagic hydrocephalus).
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Horizontal-Ventricular Lumbar Valves and Spetzler Lumbar Peritoneal Shunt Systems (K101381)
Reference Devices
Integra DP Valve Systems (Including Burr Hole Reservoirs and Stainless Steel Connectors), Gravity Compensating Accessory (K152897)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2016
Integra LifeSciences Corporation Ms. Jennifer Siegel Senior Regulatory Affairs Specialist 311 Enterprise Drive Plainsboro, New Jersey 08536
Re: K161992
Trade/Device Name: Horizontal-Vertical Lumbar Valve Systems, Spetzler Lumbar Peritoneal Shunt Systems Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: August 22, 2016 Received: August 23, 2016
Dear Ms. Siegel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Carlos L. Pena -S 同公
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K161992
#### Device Name
Horizontal-Vertical Lumbar Valve Systems, Spetzler Lumbar Peritoneal Shunt Systems
Indications for Use (Describe) Horizontal-Vertical Lumbar Valve Systems
Horizontal-Vertical Lumbar Valve Systems are implantable devices used in the treatment of patients with communicating hydrocephalus to shunt cerebrospinal fluid (CSF) from the lumbar subarachnoid region to the perioneal cavity. They provide controlled intraventricular pressure and CSF drainage in patients with hydrocephalus. The antechamber can be electively mounted in line with the Valve Unit to allow for CSF sampling or injections in the subarachnoid space.
Spetzler Lumbar Peritoneal Shunt Systems
Percutaneous lumbar perioneal shunting may be utilized in the treatment of communicating hydrocephalus. It is designed to shunt CSF from the lumbar subarachnoid space to the peritoneal cavity.
The shunt may be used for diagnosis, evaluation or treatment of normal pressure communicating hydrocephalus.
A percutaneous lumbar peritoneal shunt is also useful in the management of persistent cerebrospinal fluid fistulas, bulging cranial and suboccipital decompressions and in transient CSF absorption defects, e.g. post-memingitic or post-hemorrhagic hydrocephalus.
The In-Line Valve, available as a separate component of the system, is indicated for use where added resistance is desired to alleviate symptoms of low pressure in the small percentage of patients who, after normal drainage and in the normal course of treatment, develop such symptoms.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
A summary of 510(k) substantial equivalence information in accordance with the requirements of 21 CFR 807.92
| 807.92(a)(1) – Submitter information | |
|--------------------------------------|--------------------------------------------------------------------------------------|
| Name | Integra LifeSciences Corporation |
| Address | 311 Enterprise Drive Plainsboro, NJ 08536 USA |
| Phone Number | 1-609-275-0500 |
| Fax Number | 1-609-275-5363 |
| Establishment Registration<br>Number | 3003418325 |
| Name of Contact Person | Jennifer Siegel |
| Date Prepared | July 18, 2016 |
| 807.92(a)(2) – Name of device | |
| Trade or Propriety Name | Horizontal-Vertical Lumbar Valve Systems<br>Spetzler Lumbar Peritoneal Shunt Systems |
| Common or Usual Name | Hydrocephalus Shunts |
| Classification Name | Shunt, Central Nervous System And Components |
| Classification Panel | Neurology |
| Regulation | Class II, under 21 CFR 882.5550 |
| Product Code(s) | JXG |
# 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed
Predicate device: Horizontal-Ventricular Lumbar Valves and Spetzler Lumbar Peritoneal Shunt Systems; K101381
Reference device: Integra DP Valve Systems (Including Burr Hole Reservoirs and Stainless Steel Connectors), Gravity Compensating Accessory; K152897
807.92(a)(4) - Device description
{4}------------------------------------------------
Horizontal-Vertical Lumbar Valve Systems
The Horizontal-Vertical Lumbar Valve System is an implantable device that provides controlled intraventricular pressure and cerebrospinal fluid (CSF) drainage from the lumbar subarachnoid region to the peritoneal cavity. The system minimizes overdrainage by automatically compensating for changes in CSF hydrostatic pressure when the patient changes from a recumbent to an upright position.
## Spetzler Lumbar Peritoneal Shunt Systems
The Spetzler Lumbar Peritoneal Shunt Systems are designed to shunt cerebrospinal fluid (CSF) from the lumbar subarachnoid space to the peritoneum. Due to the small diameter of the catheter tubing, shunting may be accomplished by percutaneous techniques. The lumbar catheter is inserted into the lumbar subarachnoid space through a 8.89cm (3½-inch), 14- gauge Tuohy spinal needle with a Huber tip. A catheter passer and peritoneal trocar may be used to bring the peritoneal catheter section around the flank and into the peritoneal cavity.
A variety of devices are available: a one-piece model without a reservoir; separate large or small reservoirs that contain a one-way valve, which may be used when a flushing capability is required; and a separate miter valve, available in low, medium, or high pressure, which may be used when added resistance is desired.
## 807.92(a)(5) - Intended use of the device
{5}------------------------------------------------
| Indications for Use | Horizontal-Vertical Lumbar Valve Systems |
|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Horizontal-Vertical Lumbar Valve Systems are implantable<br>devices used in the treatment of patients with communicating<br>hydrocephalus to shunt cerebrospinal fluid (CSF) from the<br>lumbar subarachnoid region to the peritoneal cavity. They<br>provide controlled intraventricular pressure and CSF drainage<br>in patients with hydrocephalus. The antechamber can be<br>electively mounted in line with the Valve Unit to allow for<br>CSF sampling or injections in the subarachnoid space. |
| | Spetzler Lumbar Peritoneal Shunt Systems |
| | Percutaneous lumbar peritoneal shunting may be utilized in the<br>treatment of communicating hydrocephalus. It is designed to<br>shunt CSF from the lumbar subarachnoid space to the<br>peritoneal cavity. |
| | The shunt may be used for diagnosis, evaluation or treatment<br>of normal pressure communicating hydrocephalus. |
| | A percutaneous lumbar peritoneal shunt is also useful in the<br>management of persistent cerebrospinal fluid fistulas, bulging<br>cranial and suboccipital decompressions and in transient CSF<br>absorption defects, e.g. post-meningitic or post-hemorrhagic<br>hydrocephalus. |
| | The In-Line Valve, available as a separate component of the<br>system, is indicated for use where added resistance is desired to<br>alleviate symptoms of low pressure in the small percentage of<br>patients who, after normal drainage and in the normal course of<br>treatment, develop such symptoms. |
| 807.92(a)(6) Summary of the technological characteristics of the device compared to the<br>predicate | |
{6}------------------------------------------------
The proposed Horizontal-Vertical Lumbar Valve Systems and Spetzler Lumbar Peritoneal Shunt Systems have the same technological characteristics compared to the predicate devices of the same name (K101381) as shown in the table below. The update of the MRI safety information in the labeling to increase the maximum spatial gradient and to align with the FDA's guidance document "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment to the labeling does not alter the indications for use, intended use, materials of composition, manufacturing and sterilization process, or the fundamental scientific technology of the devices. This update to the labeling is consistent with that cleared under K152897, which serves as a reference device for this submission.
| Feature | Technological Characteristics as Compared to Predicate of Same Name<br>Horizontal-Vertical Lumbar<br>Valve Systems | Spetzler Lumbar Peritoneal<br>Shunt Systems |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indications for Use | Same – see 807.92(a)(5) | Same – see 807.92(a)(5) |
| Product Classification | Same – Class II | Same – Class II |
| Product Code | Same – JXG | Same – JXG |
| Design | Same – Differential pressure<br>valve with two balls in cone<br>valve mechanisms: a spring-<br>actuated (lower) pressure<br>mechanism and a gravity-<br>actuated (higher) pressure<br>mechanism. | Same – Pressure control is<br>achieved through a<br>combination of double slit<br>valve at the peritoneal end<br>and the small inner diameter<br>catheter |
| Performance Specifications | Same – Six pressure ranges<br>with closing pressures<br>between 50 and 125 mmH2O<br>in the horizontal position and<br>between 170 and 445<br>mmH2O in the vertical<br>position. | Same – Six systems with<br>pressure ranges with closing<br>pressures between 50 and<br>400 mmH2O |
| MRI Status | Same – MR Conditional | Same – MR Conditional |
| Sterilization | Same - Ethylene Oxide/ SAL<br>= 10-6 | Same - Ethylene Oxide/ SAL<br>= 10-6 |
| Biocompatible | Same – Yes | Same – Yes |
| Non-Pyrogenic | Same – Yes | Same – Yes |
| Packaging | Same – sterile, double<br>packaged system | Same – sterile, double<br>packaged system |
Technological Characteristics as Compared to Predicate of Same Name
{7}------------------------------------------------
No additional non-clinical testing was performed for this submission. Non-clinical testing was performed to support the initial MRI Labeling for Integra Shunts and Implanted Accessories, which was cleared under K101381. Testing included:
- Magnetically Induced Displacement Force and Torque Test ●
- . RF Heating Test
- Image Artifact Test ●
- Pressure Flow Test ●
The non-clinical testing has since been amended to include updated calculations for the maximum spatial gradient based on ASTM F2052-15 and clinical injury threshold considerations, which were cleared under K152897.
## 807.92(b)(3) – Conclusions drawn from non-clinical and clinical data
The proposed Horizontal-Vertical Lumbar Valve Systems and Spetzler Lumbar Peritoneal Shunt Systems are substantially equivalent to the currently marketed Horizontal-Vertical Lumbar Valve Systems and Spetzler Lumbar Peritoneal Shunt Systems which were previously cleared to market by the United States Food and Drug Administration (FDA) on July 7, 2011 under K101381.
The update of the MRI safety information to the labeling does not alter the indications for use, intended use, materials of composition, manufacturing and sterilization process, or the fundamental scientific technology of the devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.