K161974 · Bio-Medical Research, Ltd. · NGX · Nov 1, 2016 · Physical Medicine
Device Facts
Record ID
K161974
Device Name
SLENDERTONE® Connect Abs, Type 570
Applicant
Bio-Medical Research, Ltd.
Product Code
NGX · Physical Medicine
Decision Date
Nov 1, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SLENDERTONE® Connect Abs, Type 570 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. It is indicated for the improvement of abdominal muscle tone, for the strengthening of the abdominal muscles and for the development of a firmer abdomen.
Device Story
Portable neuromuscular electrical stimulation (NMES) system; delivers electrical pulses to abdominal muscles via adhesive gel pads. System includes control unit, abdominal garment, and electrodes. Wireless communication via Bluegiga BLE113 Bluetooth module enables control and program management through iOS and Android mobile applications. User-operated; intended for home use. Device provides pre-installed toning program; users can download additional programs via app. Output parameters are strictly limited to ensure safety. Provides muscle stimulation to facilitate performance and improve tone. Benefits patient through non-invasive abdominal muscle conditioning.
Clinical Evidence
Bench testing only. Device performance evaluated against IEC 60601-1, IEC 60601-2-10, IEC 60601-1-11, IEC 60601-1-6, EN 60601-1-2, ISO 10993-5, ISO 10993-10, and IEC 62133. Wireless co-existence testing performed in environments with Bluetooth/Wi-Fi interference. No clinical data presented.
Indicated for healthy individuals seeking to improve abdominal muscle tone, strengthen abdominal muscles, and develop a firmer abdomen.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 1, 2016
Bio-Medical Research Ltd Anne-Marie Keenan Regulatory Affairs Specialist Parkmore Business Park West Galway, H91 NHT7 Ireland
Re: K161974
Trade/Device Name: SLENDERTONE® Connect Abs, Type 570 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: October 3, 2016 Received: October 5, 2016
Dear Anne-Marie Keenan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -A
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161974
Device Name SLENDERTONE® Connect Abs, Type 570
#### Indications for Use (Describe)
The SLENDERTONE® Connect Abs, Type 570 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. It is indicated for the improvement of abdominal muscle tone, for the abdominal muscles and for the development of a firmer abdomen.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k SUMMARY
| I. SUBMITTER | |
|-------------------------|------------------------------------------------------|
| Name: | Anne-Marie Keenan |
| Address: | Bio-Medical Research Ltd., |
| | Parkmore Business Park West, |
| | Galway, IRELAND |
| Telephone: | +353 91 774316 |
| Fax: | +353 91 774301 or +353 91 774302 |
| E-Mail: | akeenan@bmr.ie |
| Prepared: | July 14, 2016 |
| II. DEVICE | |
| Trade Name of Device: | SLENDERTONE® Connect Abs, Type 570 |
| Common Name: | Powered muscle stimulator |
| Regulation Number: | 21 CFR 890.5850 |
| Regulation Description: | Stimulator, muscle, powered, for muscle conditioning |
| Product Code: | NGX |
| Device Class: | 2 |
| III. PREDICATE DEVICES | |
| 510(k) Number: | K151903 |
| Manufacturer: | Bio-Medical Research Ltd. |
| Trade Name: | SLENDERTONE® Connect Abs, Type 570 |
The listed predicate has not been subject to a design-related recall.
# IV. DEVICE DESCRIPTION
The SLENDERTONE® Connect Abs, Type 570 is a portable, neuromuscular, electrical stimulation system intended to deliver electrical stimulation to the abdominal muscles. The system includes a control unit, abdominal garment, 3 adhesive gel pads (electrodes), USB cable, pouch and instructions for use. One toning program is pre-installed in the unit. Wireless communication is enabled by a Bluegiga BLE113 Bluetooth module. The modified Slendertone Connect Abs device supports a range of mobile devices via both iOS and android mobile applications.
{4}------------------------------------------------
# V. INDICATIONS FOR USE
The SLENDERTONE® Connect Abs, Type 570 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. It is indicated for the improvement of abdominal muscle tone, for the strengthening of the abdominal muscles and for the development of a firmer abdomen.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The following tables summarizes the similarities and differences between the technological characteristics of the modified device versus the predicate device;
Table I Unit Characteristics
Table II Output Characteristics (values representing maximum power limit of 180mW)
Users may securely download additional treatment programs through the mobile application. The program parameters and resulting outputs for any such program are limited in accordance with the Table II.
{5}------------------------------------------------
| Table I<br>Basic Unit Characteristics | Modified Device<br>Slendertone Connect Abs | Predicate Device<br>Slendertone Connect Abs | Justification |
|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| 1. 510(k) Number | K161974 | K151903 | N/A |
| 2. Device Name, Model | Slendertone® Connect<br>Abs,<br>Type 570 | Slendertone® Connect Abs,<br>Type 570 | Same |
| 3. Manufacturer (Contract) | JDI Electronics Factory<br>Cima Industrial Center<br>Chang Ping, Dongguan<br>Guangdong, CHINA<br><br>Owner Operator: Blue<br>Ocean Innovation Ltd<br>RM 1813, Fo Tan<br>Industrial Centre<br>26-28 Au Pui Wan Street<br>Fo Tan, Hong Kong | Blue Ocean Innovation Ltd<br>RM 1813, Fo Tan Industrial Centre<br>26-28 Au Pui Wan Street<br>Fo Tan, Hong Kong | |
| 4. Power Source | 3.7V Lithium Polymer<br>Single Cell Rechargeable | 3.7V Lithium Polymer Single Cell<br>Rechargeable | Same |
| - Method of line Isolation | No line connection<br>possible when connected<br>to body | No line connection possible when<br>connected to body | Same |
| - Patient Leakage Current | Not applicable, no line<br>connection, no AC<br>charger connection or<br>operation. Connection<br>method does not allow<br>AC charger connection to<br>Patient. | Not applicable, no line connection,<br>no AC charger connection or<br>operation. Connection method<br>does not allow AC charger<br>connection to Patient. | Same |
| 5. No. of Output Modes | 1 (Symmetric, Pulsed,<br>Biphasic) | 1 (Symmetric, Pulsed, Biphasic) | Same |
| 6. Number of Output<br>Channels | 2 | 2 | Same |
| - Synchronous/Alternating? | Synchronous | Synchronous | Same |
| - Method of channel<br>isolation | Transistor | Transistor | Same |
| 7. Regulated Current or<br>Regulated Voltage | Constant Current | Constant Current | Same |
| 8. Software/Firmware/Micro<br>processor Control? | Yes | Yes | Same |
| 9. Mobile Apps | iOS + Android | iOS only | Different, but does not<br>adversely impact the safety<br>and effectiveness of the<br>subject device |
| 10. Automatic overload<br>Trip? | Yes | Yes | Same |
| 11. Automatic No-Load | Yes | Yes | Same |
| Table I | Modified Device | Predicate Device | Justification |
| Basic Unit Characteristics<br>Trip? | Slendertone Connect Abs | Slendertone Connect Abs | |
| 12. Automatic Shut Off | Yes | Yes | Same |
| 13. Patient Override<br>Control? | Yes, pause button stops treatment immediately. | Yes, pause button stops treatment immediately. | Same |
| 14. Indicator Display<br>- On/Off Status?<br>- Low Battery?<br>- Voltage/Current Level? | Yes, Unit LED and Smart device<br>Yes, Unit LED and Smart device<br>Yes, via Smart device | Yes, Unit LED and Smart device<br>Yes, Unit LED and Smart device<br>Yes, via Smart device | Same |
| 15. Timer range (minutes) | 20-30 minutes | 20-30 minutes | Same |
| 16. Compliance with<br>Voluntary Standards? | IEC 60601-1: 2005 & A1:2012<br>IEC 60601-2-10:2012<br>EN 60601-1-2: 2007<br>IEC 60601-1-11:2010<br>IEC 60601-1-6:2010<br>IEC 62133:2012<br>FCC Rule Part 15.247:2012 | IEC 60601-1: 2005 & A1:2012<br>IEC 60601-2-10:2012<br>EN 60601-1-2: 2007<br>IEC 60601-1-11:2010<br>IEC 60601-1-6:2010<br>IEC 62133:2012<br>FCC Rule Part 15.247:2012 | Same |
| 17. Compliance with CFR<br>21 898? | Yes | Yes | Same |
| 18. Weight (unit) | 37g (including batteries) | 37g (including batteries) | Same |
| 19. Dimensions (un.)<br>{W x H x D} | 70 x 50 x 14 mm approx. | 70 x 50 x 14 mm approx. | Same |
| 20. Housing Materials and<br>Construction | Injection molded thermosetting plastic | Injection molded thermosetting plastic | Same |
| Table II | Modified Device | Predicate Device | Justification |
| Output Characteristics | Slendertone Connect Abs | Slendertone Connect Abs | |
| Waveform | Pulsed, Symmetrical, Biphasic | Pulsed, Symmetrical, Biphasic | Same |
| Shape | Rectangular, with interphase<br>interval | Rectangular, with interphase<br>interval | Same |
| Maximum Output<br>Voltage (RMSV) (+/-10%) | 9.47V @ 500Ω<br>17.5V @ 2kΩ<br>6.2V @ 10kΩ | 9.47V @ 500Ω<br>17.5V @ 2kΩ<br>6.2V @ 10kΩ | Same |
| $ Vp2x2xPW √1/freq $ | | | |
| Maximum Output<br>Current (RMSA) (+/-10%) | 18.9mA @ 500Ω<br>8.75mA @ 2kΩ<br>620µA @ 10kΩ | 18.9mA @ 500Ω<br>8.75mA @ 2kΩ<br>620µA @ 10kΩ | Same |
| Pulse Width | 900 µs | 900 µS | Same |
| Baseline to peak current<br>@500Ω | 80mA | 80mA | Same |
| Frequency (Hz) | 10 to 80 Hz | 10 to 80 Hz | Same |
| - Phase Duration | 100 - 400µS | 100 - 400µS | Same |
| Net Charge (µC per<br>pulse) | 0@500Ω<br>Symmetric, biphasic and<br>leading polarity alternates for<br>each successive pulse | 0@500Ω<br>Symmetric, biphasic and<br>leading polarity alternates for<br>each successive pulse | Same |
| Maximum Phase Charge<br>(µC)<br>C= Ip*PW | 1 phase 32 µC @500Ω<br>2 phase 64 µC @500Ω | 1 phase 32 µC @500Ω<br>2 phase 64 µC @500Ω | Same |
| Maximum Current<br>Density (mA/cm²) | 0.28 mA/cm² @500Ω | 0.28 mA/cm² @500Ω | Same |
| Maximum Power Density<br>(W/ cm²)<br>Using smallest electrode<br>conductive surface area | 2.57 mW/ cm² @500Ω | 2.57 mW/ cm² @500Ω | Same |
| Contraction Time | 0.5 - 5 sec | 0.5 - 5 sec | Same |
| Relaxation Time | 0.5 - 6 sec | 0.5 - 6 sec | Same |
| Additional Features (if<br>applicable) | N/A | N/A | Same |
{6}------------------------------------------------
{7}------------------------------------------------
# VII. PERFORMANCE DATA
Performance testing was conducted in accordance with the following international standards for safety:
IEC 60601-1: 2005/A1:2012 Medical electrical equipment. General requirements for basic safety and essential performance
{8}------------------------------------------------
- Medical electrical equipment Part 1-6: General IEC 60601-1-6:2010 requirements for basic safety and essential performance -Collateral standard: Usability Medical electrical equipment - part 1-2: general EN 60601-1-2:2007 requirements for for safety - collateral standard:
- electromagnetic compatibility requirements and tests IEC 60601-2-10:2012 Medical Electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators IEC 60601-1-11:2010 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- Biological evaluation of medical devices. Tests for irritation ISO 10993-10:2010 and skin sensitization
The performance of SLENDERTONE® Connect Abs for wireless co-existence was evaluated in an environment with equipment operating in the ISM band i.e. Bluetooth and Wi-Fi devices, cellphones, cordless phones etc.. The device met all specified requirements.
BLE module testing was conducted in accordance with EN 60950-1:2006+A11:2009 +A1:2010+A12:2011+A2:2013. Safety of Technology Equipment and FCC Rule Part 15.247:2012.
Battery testing was conducted in accordance with IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them, for use in portable applications.
# VIII. CONCLUSION
The modified SLENDERTONE® Connect Abs, Type 570 has
{9}------------------------------------------------
- . The same principles of operation as the predicate device with no differences in technological characteristics.
- . The same Intended Use and Indications for Use as the predicate device.
Modifications to the SLENDERTONE® Connect Abs has not resulted in any new issues of safety or effectiveness, and is therefore substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.