← Product Code [EIH](/productcode/EIH) · K161922

# XP202 (K161922)

_3M Deutschland GmbH · EIH · Sep 1, 2016 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161922

## Device Facts

- **Applicant:** 3M Deutschland GmbH
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Sep 1, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Single crowns Bridges with a maximum of one pontic between two crowns Inlays, onlays, and veneers

## Device Story

XP202 zirconia blanks used for fabrication of esthetic zirconia dental restorations; available in various heights and shades. Dental CAD/CAM workflow: restorations designed via CAD software; data converted to milling paths by CAM software; blanks processed in milling units suitable for pre-sintered zirconia. Milled restorations require final sintering in furnace per specified cycle. Restorations display gradient shading and inherent fluorescence post-sintering. Used in dental clinics/labs by dental professionals. Benefits include esthetic, durable all-ceramic restorations.

## Clinical Evidence

Bench testing only. In vitro testing conducted per ISO 6872. Performance compared to Katana STML and Lava Plus regarding flexural strength, chemical solubility, and linear thermal expansion. Biocompatibility assessment performed by board-certified toxicologist per ISO 10993 and ISO 7405.

## Technological Characteristics

Zirconia (ZrO2) mill blanks; pre-shaded. Complies with ISO 6872. Dimensions vary by height. Processing requires CAD/CAM milling and furnace sintering. No electronic components or software embedded in the blank itself.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- Katana STML ([K143439](/device/K143439.md))

## Reference Devices

- Lava Plus - Mill Blanks and Dyeing Liquids ([K011394](/device/K011394.md))
- Lava Plus Effect Shades ([K120011](/device/K120011.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a human face in profile, with three overlapping profiles to represent the department's focus on health, human services, and the well-being of the nation. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

3M Deutchland GmbH Ruediger Franke Regulatory Affairs Specialist EPE Platz Seefeld, Bavaria 82229 GERMANY

Re: K161922

Trade/Device Name: XP202 Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: Class II Product Code: EIH Dated: August 26, 2016 Received: August 29, 2016

Dear Mr. Ruediger Franke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

September 1, 2016

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Susan Runno DDS, MA

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name XP202

Indications for Use (Describe)

· Single crowns

· Bridges with a maximum of one pontic between two erowns

· Inlays, onlays, and veneers

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect. of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@tda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

{3}------------------------------------------------

Carl-Schurz-Straße 1 41453 Neuss Germany

☎+49 (0)8152 700-0 됨 +49 (0)8152 700-1366 Internet: www.3MESPE.de E-Mail: info3MESPE@3M.com WEEE-Reg.-Nr. DE 36963167 VAT-ID: DE 120679179

# 510(k) Summary K161922

| Submitter: | 3M Deutschland GmbH                        |
|------------|--------------------------------------------|
|            | ESPE Platz                                 |
|            | 82229 Seefeld                              |
|            | Germany                                    |
|            | Establishment Registration Number: 9611385 |

| Contact Person | Ruediger Franke                |
|----------------|--------------------------------|
|                | Regulatory Affairs Specialist  |
|                | Phone: +49-8152-700 1802       |
|                | Fax: +49-8152-700 1869         |
|                | e-mail: ruediger.franke@3M.com |
| Date:          | August 23, 2016                |

Trade Name: ......................................................................................................................... XP202

Common Name: ................................................................................................................................................................

Classification Name:..................................... Porcelain powder for clinical use ..............................................................................................................................................................................

Device Class: ................................................................................................................................................................

Primary Predicate Device ........ Katana STML (K143439)

Reference Predicate Devices ... Lava Plus - Mill Blanks and Dyeing Liquids (K011394) ..............................................................................................................................................................................

### Description of Device

XP202 zirconia blanks are used for the fabrication of esthetic zirconia restorations. The blanks are available in various heights and shades based on the Vita™ Classical shade guide. After sintering, restorations display a gradient shading and inherent fluorescence. The restorations are designed using dental CAD software and the data is converted into milling paths by CAM software. The blanks can be processed in milling units suitable for pre-sintered zirconia. Milled restorations must be final sintered in a furnace suitable for zirconia per the cycle designated for XP202.

Image /page/3/Picture/16 description: The image shows a logo and the text "Page 6 of 29". The logo is circular with a star in the middle. The text indicates that this is page 6 of a document that is 29 pages long.

{4}------------------------------------------------

# Applicable Standards for Product Tests

- ISO 6872: Dentistry Ceramic materials .
# Indications for Use for XP202

- . Single crowns
- . Bridges with a maximum of one pontic between two crowns
- Inlays, onlays, and veneers .

# Comparison

Chemical composition, performance, fundamental technology, and intended use of XP202 have been compared to the predicate devices.

The tables below summarize the indications and technology of XP202 and predicate devices:

| XP202                                                                                                                     | Katana STML<br>(K143439)                                                                                                                                            | Lava Plus<br>Blanks and<br>Dyeing Liquids<br>(K011394)                                                                                                                                                 | Lava Plus<br>Effect Shades<br>(K120011)                                                                                                                                                                               |
|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • Single crowns<br>• Bridges with a<br>maximum of one<br>pontic between two<br>crowns<br>• Inlays, onlays, and<br>veneers | KATANA Zirconia is<br>used for the<br>fabrication of the all-<br>ceramic restorations<br>(frameworks, FCZ<br>crowns, FCZ bridges,<br>Inlays, onlays and<br>veneers) | The LAVA™ system is<br>intended for CAD/CAM<br>fabrication of all-ceramic<br>dental restorations.<br>The system is used for<br>the manufacturing of<br>inlays, onlays, veneers,<br>crowns and bridges. | Lava Plus Effect<br>Shades are suited<br>for more intensive<br>coloring of Lava Plus<br>zirconia frameworks<br>and Lava Plus all-<br>zirconia restorations<br>after basic dyeing<br>using lava Plus<br>Dyeing Liquids |

Table Comparison of indications

| Technology           | XP202            | Katana STML<br>(K143439) | Lava Plus<br>(K011394)<br>(K120011)                                                                                     |
|----------------------|------------------|--------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Zirconia Mill Blanks | ×                | ×                        | ×                                                                                                                       |
| Shading              | Preshaded Blanks | Preshaded Blanks         | Unshaded Blanks<br>(K011394) with dyeing<br>liquids (K011394) and<br>effect shades<br>(K120011) as shading<br>solutions |

Table Comparison of technology

XP202, Kanata STML and Lava Plus are zirconia mill blanks with ZrO2 as the mean component. 3M Deutschland GmbH provided information to FDA about the composition of XP202 and compared this with chemical analysis of the predicate devices.

{5}------------------------------------------------

In vitro testing was conducted to show that XP202 fulfils the requirements of FDA recognized standard ISO 6872. Additionally, the performance of XP202 was compared to the predicate devices Katana STML and Lava Plus regarding flexural strength, chemical solubility and linear thermal expansion. The results of XP202 are similar to Katana STML. The difference between XP202 and Lava Plus are their flexural strength and is addressed by limiting the indications to three unit bridges according ISO 6872. In summary, 3M Deutschland GmbH concludes that XP202 is substantially equivalent to the predicate devices regarding performance and physical properties.

### Biocompatibility

The biocompatibility assessment for this product was conducted in accordance with the following quidance:

1) Testing guidelines outlined in the FDA General Program Memorandum G95. 2) ISO 10993-1 :2009(E) Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process; in addition, relevant detailed guidance in ISO Standards 10993-3:2014 (Tests for genotoxicity, carcinogenicity and reproductive toxicity), 10993-5:2009 (Tests for in vitro cytotoxicity), 10993-10:2010 (Tests for irritation and skin sensitization); and 10993-11:2006 (Tests for systemic toxicity) was considered;

3) ISO 7405:2008/ Amd 1: 2013 Dentistry-- Evaluation of Biocompatibility of Medical Devices in Dentistry;

The biocompatibility of XP202 has been assessed by a board-certified toxicologist according to recommendations in FDA guidance and internationally recognized standards for medical and dental devices. The conclusion of the assessment is that XP202 is safe for its intended.

### Conclusion .

.

Comparisons of the indications for use/intended use, composition, technology, and physical properties showed that XP202 does not raise any new questions about safety and effectiveness and is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K161922](https://fda-staging.innolitics.com/device/K161922)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).