K161907 · Xiamen Senyang Co., Ltd. · IRP · Feb 23, 2017 · Physical Medicine
Device Facts
Record ID
K161907
Device Name
Pressure Therapy System
Applicant
Xiamen Senyang Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Feb 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is indicated for use by medical professionals and patient at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, edema following trauma and sport injuries, post immobilization edema, venous insufficiencies, lymphedema.
Device Story
Pressure Therapy System (Pt 1002) is a pneumatic compression device for home or clinical use. System comprises air pump, hoses, and 4-chambered leg sleeves. Pump inflates/deflates chambers sequentially from limb end to body center to promote blood circulation and prevent clots. Microprocessor-controlled system monitors pressure via sensors to cycle airflow. User selects between two compression modes (A or B) and therapy duration (up to 30 minutes). Output is sequential pneumatic compression; healthcare providers use device to manage edema and venous conditions. Benefits include improved circulation and reduced swelling. Device operates on 110V/60Hz power.
Clinical Evidence
Bench testing only. Device tested for electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5 for cytotoxicity and ISO 10993-10 for irritation/sensitization). No clinical data provided.
Technological Characteristics
Pneumatic compression system; 4-chambered thermoplastic urethane sleeves. Microprocessor-controlled air pump; 0-250 mmHg pressure range; 30s cycle time. 110V/60Hz power source; 30W consumption. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for medical professionals and patients at home under medical supervision to treat primary lymphedema, post-trauma/sport injury edema, post-immobilization edema, and venous insufficiencies.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Predicate Devices
Compressible Limb Therapy System (Power-Q1000 Premium) (K113275)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2017
Xiamen Senyang Co., Ltd. % Raymond Luo Technical Manager Shanghai SUNGO Management Consulting Company Limited 4th Floor 1500# Central Avenue Shanghai, Shanghai 200122 CN
Re: K161907
Trade/Device Name: Pressure Therapy System Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: January 26, 2017 Received: January 26, 2017
Dear Mr. Luo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Michael J. Hoffmann -S" in a simple, sans-serif font. The text is horizontally aligned and appears to be the primary focus of the image. The letters are clear and legible, with consistent spacing between them.
- for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161907
Device Name Pressure Therapy System
Indications for Use (Describe)
The device is indicated for use by medical professionals and patient at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, edema following trauma and sport injuries, post immobilization edema, venous insufficiencies, lymphedema.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
## K161907
## A. Applicant:
Name: XIAMEN SENYANG CO., LTD. Address: 4-5 FLOOR, XINGBEI INDUSTRY, NO 95-99, WEST 2 ROAD, JIU TIANHU, XINGLIN XIAMEN, 361000, P.R. China
Official Contact Person Information Name: Raymond Luo Tel: 0086-21-68828050 Mail: fda.sungo@gmail.com
### B. Subject device:
| Trade name: | Pressure Therapy System |
|------------------------------|-----------------------------------|
| Common name: | Powered inflatable Tube Massager |
| Classification name: | Massager, Powered Inflatable Tube |
| Regulation Medical Specialty | Physical Medicine |
| Regulation Number | 890.5650 |
| Product Code | IRP |
| Classification | Class II |
## c. Predicate device:
K113275 - Compressible Limb Therapy System (Power-Q1000 Premium)
### D. Indications for Use:
The device is indicated for use by medical professionals and patient at home, who are under medical supervision, in treating many conditions, such as: Primary lymph edema following trauma and sport injures, Post immobilization edema, Venous insufficiencies, Lymph edema.
### E. Device Description:
The system consists of an air pump, leg sleeves and hoses working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses; one sleeve has 4 compression chambers. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor. This software is to control the timing and the pressure reflected by the sensor, it cycles the airflow to reach the function of cycling compression of body parts.
### F. Technical Characteristic:
| Name | Pt 1002 |
|------|---------|
|------|---------|
{4}------------------------------------------------
| Pressure (mmHg) | 0~250 mmHg |
|-----------------|--------------------------------------------------|
| Mode | Mode A and Mode B selection and the initial is A |
| Cycle time | 30s |
| Pressure Time | Time 0-30 min, selection (10, 20, 30min) |
## G. Substantial Equivalence Table:
| Device | Subject Device (K161907) | Predicate Device (K113275) |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | XIAMEN SENYANG CO., LTD. | BSR KOREA Corp. |
| Model Name | Pt 1002 | Power Q1000 Premium |
| Classification | Class II Device, IRP (21 CFR890.5650) | Class II Device, IRP (21 CFR890.5650) |
| Indications for Use | The device is indicated for use by<br>medical professionals and patient at<br>home, who are under medical<br>supervision, in treating many<br>conditions, such as: Primary<br>lymphedema, edema following<br>trauma and sport injures,<br>Post-immobilization edema, Venous<br>insufficiencies, Lymphedema. | The device is indicated for use by medical<br>professionals and patient at home, who are under<br>medical supervision, in treating many conditions,<br>such as: Primary lymphedema, edema following<br>trauma and sport injures, Post-immobilization<br>edema, Venous insufficiencies, Lymphedema. |
| Description | Pt 1002 is a pneumatic pressure<br>treatment system that repeats<br>expansion of sleeves to help blood<br>circulation and prevent blood clots or<br>clogs. | Power Q1000 Premium is a pneumatic pressure<br>treatment system that repeats expansion of<br>sleeves to help blood circulation and prevent<br>blood clots or clogs. |
| Standard | IEC 60601-1 & IEC 60601-1-2<br>ISO10993-5 & ISO10993-10 | EN 60601-1 & EN 60601-1-2 |
| Indications | Primary lymphedema, edema<br>following trauma and sport injures,<br>Post-immobilization edema, Venous<br>insufficiencies, Lymphedema | Primary lymphedema, edema following trauma<br>and sport injures, Post-immobilization edema,<br>Venous insufficiencies, Lymphedema |
| Mode of<br>Compression | Sequential | Sequential |
| Power Source | 110 V, 60Hz | 110V, 60Hz |
| Therapy Time | 0 – 30 mins | 0 - 99 mins |
| Max Pressure | 0-250mm Hg | 0-240mm Hg |
| Min Pressure | | |
| Number of<br>Chambers | 4 Chambers for each unit | 4 Chambers for each unit |
| Compression<br>Applicator<br>Garments | Thermoplastic Urethane | Nylon |
| Sleeve Material | | |
| Power | 30W | 30W |
| consumption | | |
| Cycle time | 30s | 0, 5, 10, 15, 20, 25, 30s selection<br>Initial is 0s |
| Size | Size 260*170*130mm<br>Image: White device with black control panel | Size:150*210*210mm<br>Image: White and purple device with two knobs |
| Body area<br>specific cuffs | Small Leg Cuff LXW:90X30cm<br>Large Leg Cuff LXW:110X30cm<br>Image: Blue leg cuff | Leg:<br>Image: Silver leg cuff |
| Preprogrammed<br>modes | Model A<br>Image: Diagram of leg cuff model A | Model A<br>Image: Diagram of leg cuff model A |
| | Model B<br>Image: Diagram of leg cuff model B | Model C<br>Image: Diagram of leg cuff model C |
| | | |
{5}------------------------------------------------
{6}------------------------------------------------
From the comparison in the form above, the intended use, indication for use and the Mode of Compression,
Power Source, Power consumption of the two devices are same. There are some factors that are different:
- 1) Biocompatibility: The subject device has different sleeve material than the predicate device but subject device complies to ISO 10993-5 and ISO 10993-10 standards.
- 2) Therapy Time: For one time use, the max time for this device is 30 min. This is safer than the predicate device's max time of 90 min. If the user wants, he/she can repeat the 30 min cycle.
- 3) Min Pressure and Max Pressure of the subject device is 250 mmHg which is similar to the max pressure of 240mm Hg in the predicate device.
- 4) Preprogrammed modes: The subject device only has 2 selections which are within the scope of the predicate device. Model A of the subject device is identical to Model A of the predicate device. Model B of the subject device is identical to Model C of the predicate device.
#### H. Performance characteristic
The Pt1002 pressure therapy system has been tested and met the following standards: IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
ISO10993-5 Biological evaluations of medical devices -- Part 5: Tests for In Vitro cytotoxicity
ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
#### -Conclusion
The Pt1002 Pressure Therapy System has substantially equivalent intended use as the cleared Power Q1000 Premium Compression Limb Therapy System and has substantially equivalent technological and performance characteristics. After analyzing laboratory testing to applicable standards, it is concluded that Pt1002 Pressure Therapy System is as safe and effective as the predicate device, has few technological differe are no new indications for use and does not raise any new safety and/or effectiveness concerns. Consequently, it is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.