← Product Code [DGC](/productcode/DGC) · K161899

# ImmunoCAP Total IgE Calibrator Strip/Total IgE Curve Control Strip, ImmunoCAP Total IgE Calibrators/Total IgE Curve Controls, ImmunoCAP Specific IgE Calibrator Strip/Specific IgE Curve Control Strip, ImmunoCAP Specific IgE Calibrators/Specific IgE Curve Controls (K161899)

_Phadia AB · DGC · Jul 28, 2016 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161899

## Device Facts

- **Applicant:** Phadia AB
- **Product Code:** [DGC](/productcode/DGC.md)
- **Decision Date:** Jul 28, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 866.5510
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K161899 · Jul 28, 2016 | ImmunoCAP Total IgE Calibrator Strip/Total IgE Curve Control Strip, ImmunoCAP Total IgE Calibrators/Total IgE Curve Controls, ImmunoCAP Specific IgE Calibrator Strip/Specific IgE Curve Control Strip, ImmunoCAP Specific IgE Calibrators/Specific IgE Curve Controls | Phadia AB | Clinical negative and positive patient samples | Clinical patient samples were evaluated to compare the performance of the device using the new reference material (3rd WHO IRR) against the predicate device (2nd WHO IRR). | Clinical samples; Performance verification; Reference material change |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Evaluation of clinical negative and positive patient samples in ImmunoCAP Specific IgE; Comparative performance evaluation | Clinical patients (negative and positive for IgE) | Predicate device (IgE calibrators traceable to 2nd WHO IRR) | Concentration determination/performance comparison |

## Indications for Use

ImmunoCAP Specific IgE is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum or plasma (EDTA or Na-Heparin). ImmunoCAP Specific IgE is to be used with instruments Phadia 100, Phadia 250, Phadia 1000, Phadia 2500 and Phadia 5000. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings, and is to be used in clinical laboratories. ImmunoCAP Total IgE is an in vitro test system for the quantitative measurement of circulating total IgE in human serum and plasma. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings, and is to be used in clinical laboratories. ImmunoCAP Total IgE is to be used with the instruments Phadia 100, Phadia 250, Phadia 1000,

## Device Story

ImmunoCAP Specific and Total IgE assays are in vitro immunofluorescence assays for quantitative measurement of IgE in human serum or plasma. Device operates on Phadia instrument platforms (100, 250, 1000, 2500, 5000) using Phadia Information Data Manager software. Modification involves updating calibrator traceability from 2nd WHO IRP (75/502) to 3rd WHO IRP (11/234) for serum IgE. System uses specific conjugates, calibrators, and controls to produce quantitative results (kUA/L or kU/L). Used by clinical laboratory professionals to assist physicians in diagnosing IgE-mediated allergic disorders. Output provides objective IgE levels to support clinical decision-making alongside other patient findings.

## Clinical Evidence

Bench testing only. Verification studies compared IgE calibrators traceable to the 3rd WHO IRR against the predicate traceable to the 2nd WHO IRR. Studies included concentration determination of stock solutions used for calibrator production and evaluation of clinical negative and positive patient samples.

## Technological Characteristics

Immunofluorescence assay system. Analyte: IgE. Instrumentation: Phadia 100, 250, 1000, 2500, 5000. Software: Phadia Information Data Manager. Traceability: 3rd WHO IRP (11/234). Quantitative output.

## Regulatory Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

## Predicate Devices

- ImmunoCAP Specific IgE (k051218)
- ImmunoCAP Total IgE (k133404)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(k): Device Modification OIR Decision Memorandum

To: THE FILE

RE: K161899

This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II devices requiring 510(k). The following items are present and acceptable:

1. The name and 510(k) number of the SUBMITTER'S previously cleared device. The ImmunoCAP Specific IgE and Total IgE assays were previously cleared under K051218 (for Specific IgE) and K133404 (for Total IgE).

2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes directions for use for the ImmunoCAP Specific IgE and ImmunoCAP Total IgE assays on the cleared instruments.

3. A description of the device MODIFICATION(S) in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

This change was for the introduction of a new Reference Material (International Reference Preparation, IRP) for standardization of ImmunoCAP Specific and Total IgE assays, incorporating the 3rd WHO International Standard (3rd WHO IRP) for serum IgE (11/234).

No change of Intended Use/Indications for Use or fundamental scientific technology of the assay system has been made. No change of specifications or manufacturing methods for any assay reagents has been made.

Directions for Use for ImmunoCAP Specific and ImmunoCAP Total IgE will be updated to reflect the change. The update will be phased in over time. During a transition period Phadia intends to supply an intermediate version of labelling containing traceability information to either the old and new reference material..

4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, Intended Use, physical characteristics, and traceability

The similarities and differences between the Predicate Device, ImmunoCAP Specific IgE as cleared under k051218, and the New Device, ImmunoCAP Specific IgE, are listed below.

Similarities ImmunoCAP Specific IgE

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Predicate | Candidate  |
|  Intended Use | ImmunoCAP Specific IgE is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum or plasma (EDTA or Na-Heparin). ImmunoCAP Specific IgE is to be used with instruments Phadia 100, Phadia 250, Phadia 1000, Phadia 2500 and Phadia 5000. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings, and is to be used in clinical laboratories. | Same  |
|  Sample matrix | Human serum or plasma (EDTA or Na-Heparin) | Same  |
|  Assay system and reagents | ImmunoCAP Specific IgE Conjugate, Calibrators, Controls, ImmunoCAP Allergens | Same  |

{1}

2

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Predicate | Candidate  |
|  Assay test principle | Immunofluorescence assay | Same  |
|  Instrumentation and Software | Phadia 100, Phadia 250, Phadia 1000 or Phadia 2500/ 5000, Phadia Information Data Manager | Same  |
|  Reporting of results | Quantitative, kUA/L | Same  |
|  Analytical sensitivity (LoD/LoQ) | 0.1 kUA/L | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Predicate | Candidate  |
|  Traceability | The IgE calibrators are traceable (via an unbroken chain of calibrations) to the 2nd International Reference Preparation (IRP) 75/502 of Human Serum Immunoglobulin E from World Health Organization (WHO). | The IgE calibrators are traceable (via an unbroken chain of calibrations) to the 3rd International Reference Preparation (IRP) 11/234 of Human Serum Immunoglobulin E from World Health Organization (WHO)  |

The similarities and differences between the Predicate Device, ImmunoCAP Total IgE as cleared under k133404, and the New Device, ImmunoCAP Total IgE, are listed below.

Similarities ImmunoCAP Total IgE

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Predicate | Candidate  |
|  Intended Use | ImmunoCAP Total IgE is an in vitro test system for the quantitative measurement of circulating total IgE in human serum and plasma. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings, and is to be used in clinical laboratories. ImmunoCAP Total IgE is to be used with the instruments Phadia 100, Phadia 250, Phadia 1000, | Same  |
|  Sample matrix | Human serum or plasma (EDTA or Na-Heparin) | Same  |
|  Assay system and reagents | ImmunoCAP Total IgE Conjugate, Calibrators, Controls, Anti-IgE | Same  |
|  Assay test principle | Immunofluorescence assay | Same  |
|  Instrumentation and Software | Phadia 100, Phadia 250, Phadia 1000 or Phadia 2500/ 5000, Phadia Information Data Manager | Same  |
|  Reporting of results | Quantitative, kU/L | Same  |
|  Analytical sensitivity (LoD/LoQ) | 2 kU/L | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Predicate | Candidate  |
|  Traceability | The IgE calibrators are traceable (via an unbroken chain of calibrations) to the 2nd International Reference Preparation (IRP) 75/502 of Human Serum Immunoglobulin E from World Health Organization (WHO). | The IgE calibrators are traceable (via an unbroken chain of calibrations) to the 3rd International Reference Preparation (IRP) 11/234 of Human Serum Immunoglobulin E from World Health Organization (WHO)  |

5. A Design Control Activities Summary which includes:

a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis

{2}

b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.

3

---

**Source:** [https://fda-staging.innolitics.com/device/K161899](https://fda-staging.innolitics.com/device/K161899)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).