← Product Code [OUP](/productcode/OUP) · K161885

# The Cellfina System (K161885)

_Ulthera, Inc. · OUP · Oct 5, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161885

## Device Facts

- **Applicant:** Ulthera, Inc.
- **Product Code:** [OUP](/productcode/OUP.md)
- **Decision Date:** Oct 5, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4790
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Cellfina™ System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 3 years of observation.

## Device Story

Cellfina System provides focal release of subcutaneous tissue to improve cellulite appearance. System comprises sterile, single-use disposable kit (CK1) and electromechanical, non-sterile, 50-use motor module (CM1). Used in physician offices/clinics by trained clinicians. Device uses mechanical energy to release fibrous septae causing cellulite dimples. Clinician operates device to perform procedure; output is physical tissue release. Clinical benefit is long-term reduction in cellulite severity, assessed by physician via photographic comparison and patient satisfaction scales. Procedure is minimally invasive, performed under local anesthesia.

## Clinical Evidence

Prospective, multi-center, non-randomized open-label study (IDE G120116) with 55 subjects. Primary endpoint: ≥1 point reduction in Cellulite Severity Scale at 3 years via independent physician photo assessment. Results: 2.0 point average improvement (p<0.0001). Secondary endpoint: >60% of subjects with ≥1 grade improvement; achieved 91.1% (95% CI: 81.6%). 100% of subjects free from serious adverse events at 3 years. 93% of subjects satisfied/very satisfied at 3 years. Data analysis used LOCF for 10 missing subjects.

## Technological Characteristics

System consists of electromechanical motor module and sterile disposable kit. Materials: Makroblend and Wacker Elastosil (non-patient contacting). Energy: Mechanical. Connectivity: None. Sterilization: Ethylene Oxide (EO) for disposable kit (SAL 10^-6). Compliance: IEC 60601-1 (Edition 3.1), IEC 60601-1-2 (3rd Edition).

## Regulatory Identification

A powered surgical instrument for improvement in the appearance of cellulite is a prescription device that is used for the controlled release of subcutaneous tissue for improvement in the appearance of cellulite. The device consists of a cutting tool powered by a motor and a means for instrument guidance to control the areas of subcutaneous tissue cutting underneath the cellulite depressions or dimples.

## Special Controls

In addition to the general controls of the FD&C Act, the Cabochon System is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.
(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.
(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.
(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.
(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.

## Predicate Devices

- The Cellfina System ([K153677](/device/K153677.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 5, 2016

Ulthera, Inc. Ms. Suzon Lommel Vice President, Regulatory & Quality Affairs 1840 S Stapley Drive. Suite 200 Mesa, Arizona 85204

Re: K161885 Trade/Device Name: The Cellfina System Regulation Number: 21 CFR 878.4790 Regulation Name: Powered Surgical Instrument for Improvement in the Appearance of Cellulite Regulatory Class: Class II Product Code: OUP Dated: July 8, 2016 Received: July 11, 2016

Dear Ms. Lommel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv.

# Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) TBD KJ 67886

Device Name Cellfina™ System

Indications for Use (Describe)

The Cellfina™ System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 3 years of observation.

| Type of Use (Select one or both, as applicable)                                                       |
|-------------------------------------------------------------------------------------------------------|
| <label><input checked="checked" type="checkbox"/> Remediation Use (Part 21 CFR 201 Subpart D)</label> |
| <label><input type="checkbox"/> Source Tissue Contracting Use (21 CFR 361 Subpart C)</label>          |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510(k) Summary

This 510(k) Summary for the Cellfina System is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows the Office of Device Evaluation (ODE) guidance concerning the organization and content of a 510(k) summary.

| Applicant:                  | Ulthera, Inc.                                                                     |
|-----------------------------|-----------------------------------------------------------------------------------|
| Address:                    | 1840 South Stapley Drive<br>Suite 200<br>Mesa, AZ 85204                           |
| Contact Person:             | Kristen DeJeu, Regulatory Affairs Specialist                                      |
| Telephone:                  | (408) 822-6507                                                                    |
| Fax:                        | (480) 619-4071                                                                    |
| Submission Date:            | July 08, 2016                                                                     |
| Device Trade Name:          | The Cellfina System                                                               |
| Common Name:                | Powered surgical instrument for improvement in the appearance<br>of cellulite.    |
| Classification:             | Regulatory Class II                                                               |
| Classification Name:        | Powered surgical instrument for improvement in the appearance<br>of cellulite.    |
| Regulation Number:          | 21CFR 878.4790                                                                    |
| Product Code:               | OUP                                                                               |
| Legally Marketed Predicate: | The Cellfina System, 510(k): K153677                                              |
| Applicable Guidance:        | The following guidance special controls are applicable to the<br>Cellfina System: |

# Class II (special controls). The special controls for this device are:

(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.

(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.

(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.

(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.

Image /page/3/Picture/9 description: The image shows the word "ulthera" in a sans-serif font, with the "u" and "l" in a light blue color and the rest of the letters in a darker shade of blue. To the right of the word, there are three yellow circles arranged horizontally, followed by the trademark symbol. The overall design is clean and modern.

{4}------------------------------------------------

(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.

(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.

- Device Description: The Cellfina System is intended to provide precise focal release of subcutaneous tissue for improvement in the appearance of cellulite. The system consists of a sterile, single-use, disposable kit (CK1) and an electromechanical, non-sterile, 50-use motor module (CM1).
- Indications for Use: The Cellfina System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 3 year of observation.

# Substantial Equivalence Comparison:

The subject device and the predicate device are similar in the following ways: same regulation number and product code, anatomical use area, clinical use group (doctor's office, clinic), energy type with the same mechanism of action, packaging materials and configuration are unchanged, there are no differences in patient contacting materials, design of the CK1 Disposable Kit is the same, and the sterilization SAL is the same (EO, SAL 10th. Both devices are compliant to IEC 60601 standards and EMC standards.

The subject device differs from the predicate device in that the subject device is intended for 3 year efficacy whereas the predicate is intended for 2 year efficacy. In addition, the subject device's CK1 Disposable Kit has a shelf life of 24 months, whereas the predicate device had a shelf life of 12 months. The subject device CK1 is now qualified for a fully validated sterilization cycle, whereas the predicate went through a sterilization batch release process. The Subject Motor Module is intended for use up to 50 uses, whereas the predicate was intended for use up to 15 uses. The subject Motor Module is now validated for two types of cleaning methods (IPA based wipes and Chloride based wipes). The housing material for the enclosure and power button also changed from ABS and Elastomer to Makroblend and Wacker Elastosil, respectively in order to include a more robust material type that would withstand the new cleaning method. These are non-patient contacting materials and therefore additional biocompatibility testing was not necessary. The predicate device was only validated for one type of cleaning (Chloride based wipes). The subject device Motor Module includes minor tolerance changes to allow for injection molding, whereas the predicate was machined. Minor changes to location of epoxy, simplification of design to remove metal pin and using screws instead of heat staking were made to accommodate a 50 use "use life". The subject device will have a separate IFU for the CK1 and CM1, whereas the predicate utilized identical IFUs with repetitive information.

### Non-clinical Performance Data:

Bench testing was performed to verify that the modifications to the subject device do not impact the safety or effectiveness of the device. Retesting for EMC and IEC 60601 was completed for the motor module to ensure that the changes in design would have no impact on safety or efficacy. Shelf life testing and use life testing were conducted on the CK1 and CM1 respectively for the changes in shelf and use life, which demonstrated full functionality. Cleaning and disinfection testing was conducted for the addition of the cleaning method for the motor module

Image /page/4/Picture/10 description: The image features the logo for Ulthera, a company specializing in medical aesthetic devices. The logo consists of the word "ulthera" in a lowercase, sans-serif font, with the "u" and "l" in a light blue color and the remaining letters in a darker shade of blue. Above the "thera" portion of the word, there are three orange circles arranged horizontally, adding a visual element to the brand identity.

{5}------------------------------------------------

with no impact to device performance. Sterilization was completed for the CK1 disposable kit in order to remove the requirement for batch release testing. The SAL level and sterilization method (EO) remains the same.

# Clinical Performance Data (In-Vivo):

The safety and effectiveness of the Cellfina System was evaluated in a pivotal clinical study conducted under IDE G120116. The IDE was modified to allow for extended follow-up clinical performance data to be gathered for safety and efficacy at 3 years. No additional treatments were performed.

During the course of the 3 year Clinical study, 10 of the 55 subjects were lost to follow up. The reason for loss to follow up ranged from inability to contact the subject after multiple attempts to the subject removing herself from the study. Imputation analyses were performed for the 10 missing subjects from the 3 year report and carried out using the last observation carried forward (LOCF) for both the primary and 1st secondary endpoints using a paired t-test as well as multiple imputation for the primary effectiveness endpoint only, which met the study endpoint. The overall study success was met, despite the reduction in follow up participation at 3 years with a clinically significant, long term improvement in the appearance of cellulite in the treated subjects. The table below is based on the LOCF data and accounts for the 10 missing subjects:

| Clinical Study Design Characteristics |                                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Study Design:                         | Prospective, multi-center, non-randomized open label, safety and effectiveness study                                                                                                                                                                                                                                                                               |
| Sample Size                           | 55 Subjects participated in the baseline assessment and treatment                                                                                                                                                                                                                                                                                                  |
| Follow up (N)                         | 3 Day (55), 14 Day (54), 1 Month(54), 3Month (55), 6 Month (52), 1 year (50), 2 year (52), 3 year (45)                                                                                                                                                                                                                                                             |
| Inclusion Criteria                    | Female between the ages of 18 and 55, BMI less than 35                                                                                                                                                                                                                                                                                                             |
| Exclusion Criteria                    | No prior drug treatment for contouring or cellulite during prior 90 days; no prior liposuction of thighs or buttocks, subject has had >10 % weight loss in prior 6 months; subject has known difficulty with local anesthesia, subject is pregnant, other minor criteria                                                                                           |
| Primary Endpoint result at 3 years    | The primary endpoint was met: achievement of ≥1 point average reduction in the 0-5 point Cellulite Severity Scale as determined by independent physician assessment of subject photographs taken before and (3 years) after treatment.<br>• The average improvement was 2.0 points (p<0.0001) with the 97.5% confidence limit of a 1.7 point improvement           |
| Secondary Efficacy result at 3 years  | The powered secondary endpoint was met: improvement of one grade or more in severity (none, mild, moderate, severe) in >60% of treated subjects as determined by independent physician assessment of subject photographs taken before and 3 years after treatment.<br>• 91.1% of treated subjects had improvement of 1 grade with a 95% confidence limit of 81.6%. |
| Primary Safety Result                 | The safety endpoint was met: 100% of subjects were free from Serious Adverse Events (SAE) directly attributable to the Cellfina System or procedure at 3 years                                                                                                                                                                                                     |
| Additional                            | All additional secondary measures were also achieved or acceptable at the 3                                                                                                                                                                                                                                                                                        |

Image /page/5/Picture/6 description: The image shows the word "ulthera" in a sans-serif font. The letters are light blue, and there are three yellow dots above the "e" in "ulthera". To the right of the word is a small "TM" symbol, indicating that the word is a trademarked brand name. The overall impression is of a clean and modern logo.

{6}------------------------------------------------

| Measures | year time point:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|          | The average rate of correct selection was 97% with a 95% confidence limit of 92.6%. 100% of the subjects had noticeable improvement by the GAIS and 56% were characterized as having marked improvement or better. 93% of the subjects were either satisfied or very satisfied as evaluated by a 5 point Likert type scale at the 3 year follow-up. 71% of subjects rated pain $\leq$ 3/10 at 3 days, and over 95% rated pain $\leq$ 3/10 thereafter. There were no subjects who reported pain at the 3 year follow-up. |

# Conclusion:

This Traditional 510(k) Premarket Notification is for an expansion to the indications for use (from 2 years to 3 years). Additionally, the motor module (CM1) use life has been extended from 15 uses to 50 uses and the shelf life of the disposable kit (CK1) has been extended from a 12 month shelf life to a 24 month shelf life. Labeling changes have also been implemented to reflect the above described changes. Additional testing was performed for compliance to Amendment 1:2012 (Edition 3.1) of 60601-1 and EMC 60601-1-2:2007 3rd Edition.

These changes do not pose any new questions of safety, product output or efficacy. As demonstrated through bench testing and clinical data, the subject device Cellfina System is as safe, as effective, and performs as well as the legally marketed predicate device (K153677), and as such, is substantially equivalent.

Image /page/6/Picture/5 description: The image shows the logo for Ulthera. The logo is in a sans-serif font and is a light blue color. There are three yellow dots to the right of the word "ulthera". The letters are all lowercase.

---

**Source:** [https://fda-staging.innolitics.com/device/K161885](https://fda-staging.innolitics.com/device/K161885)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
