K161878 · Baylis Medical Company, Inc. · HRX · Sep 26, 2016 · Orthopedic
Device Facts
Record ID
K161878
Device Name
PORTAGE System
Applicant
Baylis Medical Company, Inc.
Product Code
HRX · Orthopedic
Decision Date
Sep 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Indications for Use
The PORTAGE System is indicated for visualization of the surgical field in any area of the body cut open during a surgical procedure. When used in the cervical, thoracic, or lumbar spine either from an anterior or posterior direction, for example, the PORTAGE Endoscope and accessories are intended to aid the surgeon's visualization of the surgical area and allow him/her to perform any type of surgical spinal procedure such as herniated disc repair, visualization of the circumferential decompression of the nerve roots, aiding in the search and removal of nucleus material, spinal fusion, or insertion of spinal implants. Other examples of generic surgical use of the PORTAGE System would be for use in the knee, ankle, shoulder, hand, wrist, and temporomandibular joint (TMJ).
Device Story
PORTAGE System consists of reusable rigid endoscope and manual arthroscopic instruments; used for visualization of surgical field during spinal or arthroscopic procedures. Endoscope utilizes fiber optic cable for light transmission from external source to tip; rigid rod lens system transmits images back to eyepiece. System connects to external light guide, light source, video camera, monitor, and printer. Accessories include retractors, dilators, rongeurs, forceps, suture passers, cutters, pushers, probes, curettes, and gouges. Operated by surgeons in clinical settings; provides visual aid for procedures like herniated disc repair, nerve root decompression, and spinal fusion. Reusable components undergo manual cleaning and steam sterilization; single-use accessories supplied sterile. Output is visual image of surgical site; assists surgeon in performing minimally invasive or open surgical interventions.
Clinical Evidence
Bench testing only. Verification and validation testing included endoscope field and direction of view, diopters, eccentricity, size of view, vignetting, biocompatibility, system compatibility, and usability. All test requirements were met.
Technological Characteristics
Rigid endoscope with fiber optic light transmission and rod lens image transmission. 90° short configuration. Manual accessories (retractors, dilators, rongeurs, etc.). Reusable components validated for manual cleaning and steam sterilization. No software or electronic processing components.
Indications for Use
Indicated for visualization of surgical fields in any body area opened during surgery, including cervical, thoracic, or lumbar spine procedures (e.g., herniated disc repair, decompression, spinal fusion, implant insertion) and joint procedures (knee, ankle, shoulder, hand, wrist, TMJ).
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 26, 2016
Baylis Medical Company, Inc. Ms. Meghal Khakhar Director, Regulatory and Scientific Affairs 2645 Matheson Blvd., East Mississauga, Ontario, Canada L4W 5S4
Re: K161878
Trade/Device Name: PORTAGE System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: August 12, 2016 Received: August 15, 2016
Dear Ms. Khakhar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth
{1}------------------------------------------------
in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161878
Device Name PORTAGE™ System
#### Indications for Use (Describe)
The PORTAGE System is indicated for visualization of the surgical field in any area of the body cut open during a surgical procedure. When used in the cervical, thoracic, or lumbar spine either from an anterior or posterior direction, for example, the PORTAGE Endoscope and accessories are intended to aid the surgeon's visualization of the surgical area and allow him/her to perform any type of surgical spinal procedure such as herniated disc repair, visualization of the circumferential decompression of the nerve roots, aiding in the search and removal of nucleus material, spinal fusion, or insertion of spinal implants. Other examples of generic surgical use of the PORTAGE System would be for use in the knee, ankle, shoulder, hand, wrist, and temporomandibular joint (TMJ).
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 7. 510(k) Summary (K161878)
### Submitter Information
- A. Company Name: Baylis Medical Company Inc.
- B. Company Address: 2645 Matheson Blvd. East Mississauga, Ontario L4W 5S4 Canada
- C. Company Phone: (905) 602-4875
- D. Company Facsimile: (905) 602-5671
- E. Contact Person: Meghal Khakhar, Director of Regulatory & Scientific Affairs
- F. Summary Prepared on: 07-Jul-2016
#### Device Identification
- A. Device Trade Name: PORTAGE™ System
- B. Device Common Name: Arthroscope
- C. Classification Name: CFR 888.1100 Arthroscope
- D. Product Code: HRX
- E. Device Class: Class II
#### Identification of Predicate Device
| Trade Name | Manufacturer | 510(k) |
|--------------|-------------------------|---------|
| METRx System | Medtronic Sofamor Danek | K002931 |
## Indications for Use
The PORTAGE System is indicated for visualization of the surgical field in any area of the body cut open during a surgical procedure. When used in the cervical, thoracic, or lumbar spine either from an anterior or posterior direction, for example, the PORTAGE Endoscope and accessories are intended to aid the surgeon's visualization of the surgical area and allow him/her to perform any type of surgical spinal procedure such as herniated disc repair,
{4}------------------------------------------------
| PORTAGE™ System |
|------------------------|
| Baylis Medical Company |
visualization of the circumferential decompression of the nerve roots, aiding in the search and removal of nucleus material, spinal fusion, or insertion of spinal implants. Other examples of generic surgical use of the PORTAGE System would be for use in the knee, ankle, shoulder, hand, wrist, and temporomandibular joint (TMJ).
## Device Description
The PORTAGE System is comprised of an endoscope and manual athroscopic instruments.
The PORTAGE Endoscope is a reusable rigid endoscope, comprised of a fiber optic cable and sensitive image transmission system with eyepiece. The PORTAGE Endoscope is used for visualization of the surgical field during arthroscopic or spinal procedures and may be used in conjunction with a separately cleared and commercially available light quide, light source, video camera, monitor and printer. Light that is created by the external light source is transmitted from the PORTAGE Endoscope light quide connector through the endoscope itself to the tip via a fiber optic system. Images are transferred the other way back through a rigid rod lens system. The subject endoscope is manufactured in a 90° short confiquration and is used with a compatible bracket. The PORTAGE Endoscope is supplied non-sterile.
The PORTAGE System accessories are manual instruments that include retractors, dilators, rongeurs, forceps, suture passers, cutters, pushers, probes, curettes and gouges. They enable the physician to perform any type of surgical spinal procedure. The manual instruments are either single-use and supplied sterile, or reusable and supplied non-sterile.
The reusable PORTAGE System components have undergone complete reprocessing validations, including manual cleaning and steam sterilization.
## Comparison to Predicate Device
The indications for use of the PORTAGE System are identical to that of the predicate METRx System (K002931). The subject and predicate devices also share the same fundamental technology and key device characteristics, including principles of light and image transmission and connection to external components. Like the predicate METRx System, the PORTAGE Endoscope is manufactured in a 90° short configuration and used in conjunction with an endoscope bracket (Table 7.1). Differences in design and technological characteristics between the subject and predicate devices do not raise any new types of questions of safety and effectiveness. The results of verification and validation testing provide reasonable assurance of substantial equivalence of the PORTAGE System to its predicate device.
{5}------------------------------------------------
| Table 9.1: Comparison of Subject and Predicate Devices | |
|--------------------------------------------------------|-------------------------------------------------------------------|
| Characteristic | Subject Device Compared to<br>Predicate METRx System<br>(K002931) |
| Intended Use | Identical |
| Indications for Use | Identical |
| Fundamental scientific technology | Identical |
| Operating principles | Identical |
| Mechanism of action | Identical |
| Technological aspects | Similar |
| System components | Similar |
Table 7.1: Comparison of Subiect and Predicate Devices
# Performance Testing
Performance testing has been completed for the PORTAGE System to demonstrate substantial equivalence to the predicate METRx System (K002931). As applicable, the system components have been subjected to the following verification and validation testing: general performance including endoscope field and direction of view, diopters, eccentricity, size of view, and vignetting, biocompatibility, system compatibility and usability. All test requirements were met as specified by applicable standards and the test protocols.
# Conclusions
The subject PORTAGE System and predicate METRx System (K002931) share the same intended use and indications for use. Furthermore, the subject and predicate devices share the same fundamental technology and key device characteristics.
The results of verification testing demonstrate that the subject device meets its predetermined design criteria with respect to general performance aspects, including endoscope field and direction of view, diopters, eccentricity, size of view, and vignetting and biocompatibility requirements. The results of bench top validation testing demonstrate that the PORTAGE System specifications adequately meet user needs and intended uses and usability requirements. Differences in design and technological characteristics between the subject and predicate devices do not raise any new questions of safety and effectiveness. The results of testing demonstrate the subject device is as safe and effective as the predicate METRx System (K002931) when used for its intended purpose and provide reasonable assurance of the substantial equivalence of the subject device to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.