← Product Code [LDF](/productcode/LDF) · K161873

# Pacel Flow Directed Pacing Catheter (K161873)

_St Jude Medical · LDF · Nov 22, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161873

## Device Facts

- **Applicant:** St Jude Medical
- **Product Code:** [LDF](/productcode/LDF.md)
- **Decision Date:** Nov 22, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3680
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Pacel™ Flow Directed Pacing Catheters are indicated for use in temporary, transvenous, right ventricular pacing.

## Device Story

Pacel™ Flow Directed Pacing Catheter is a temporary transvenous device for right ventricular pacing. It features an inflatable balloon between two platinum electrodes to facilitate flow-directed positioning across the tricuspid valve without fluoroscopy. Once positioned in the right ventricular apex, the balloon is deflated. The device enables ECG monitoring and bipolar pacing via electrical signals transmitted to/from external equipment. Used in clinical settings by healthcare providers; aids in establishing reliable pacing thresholds and verifying lead placement. Benefits include simplified catheter placement and temporary cardiac rhythm management.

## Clinical Evidence

Bench testing included design verification (shelf life, surface appearance, balloon inflation strength, inflation decay) and biocompatibility testing per ISO 10993. A chronic GLP animal study compared the subject device to the predicate, evaluating clinical pathology, thrombogenicity, and histopathology. Results showed the devices were well-tolerated and equivalent in safety profiles.

## Technological Characteristics

Temporary transvenous pacing catheter with platinum electrodes and an inflatable latex balloon. Features two curve styles and a two-pin proximal electrical connector. Adhesive used is Loctite 4311 (visible light-cured cyanoacrylate). Biocompatibility compliant with ISO 10993. No active electronic components; passive signal transmission for pacing and ECG monitoring.

## Regulatory Identification

Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.

## Predicate Devices

- Pacel™ Flow Directed Pacing Catheters ([K914185](/device/K914185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2016

St. Jude Medical Blair Schwartz Regulatory Affairs Specialist 14901 Deveau Place Minnetonka, Minnesota 55345

Re: K161873

Trade/Device Name: Pacel™ Flow Directed Pacing Catheter Regulation Number: 21 CFR 870.3680 Regulation Name: Cardiovascular Permanent or Temporary Pacemaker Electrode Regulatory Class: Class II Product Code: LDF Dated: October 18, 2016 Received: October 19, 2016

Dear Blair Schwartz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mude Jellman
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161873

#### Device Name

Pacel™ Flow Directed Pacing Catheters

Indications for Use (Describe)

Pacel™ Flow Directed Pacing Catheters are indicated for use in temporary, transvenous, right ventricular pacing.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;">☒</span> Exemption Use (Part 21 CFR 601.21, Subpart E) | ☐ Compassionate Use (21 CFR 601.21, Subpart E) |
|-------------------------------------------------------------------------------------------------|------------------------------------------------|
|-------------------------------------------------------------------------------------------------|------------------------------------------------|

|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for St. Jude Medical. The logo consists of a grid of green squares with one square rotated, followed by the text "ST. JUDE MEDICAL" in a serif font. The text is aligned to the right of the grid.

# 510(K) SUMMARY

The 510(k) Summary of the Pacel™ Flow Directed Pacing Catheters, per 21 CFR 807.92, is provided below.

| 510(k) Summary                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                                   | K161873                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                     |
| Submitter Information:                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
| Date Prepared:                                                  | July 06, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                     |
| Submitter Name &<br>Address:                                    | St. Jude Medical<br>14901 DeVeau Place<br>Minnetonka, MN 55345<br>Establishment Registration Number: 2182269                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                     |
| Contact Person:                                                 | Blair Schwartz<br>Regulatory Affairs Specialist<br>5050 Nathan Lane N<br>Plymouth MN 55426<br>Phone (651) 756-2706<br>BSchwartz@sjm.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                     |
| Device Information:                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
| Trade Name:                                                     | Pacel™ Flow Directed Pacing Catheters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                     |
| Common Name:                                                    | Electrode, Pacemaker, Temporary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
| Class                                                           | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                     |
| Classification Name:                                            | 21CFR § 870.3680 – Cardiovascular permanent or temporary<br>pacemaker electrode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
| Predicate Device:                                               | Pacel™ Flow Directed Pacing Catheters (K914185)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
|                                                                 | 510(k) Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                     |
| Device Description:                                             | This catheter is designed to establish temporary right ventricular pacing with or without fluoroscopy guidance for placement. Catheters are depth marked as an aid in catheter placement under fluoroscopy. Electrodes at the catheter tip enable ECG monitoring and bipolar pacing. An inflatable balloon, located between two platinum electrodes, allows for flow directed positioning of the catheter in the right ventricle without fluoroscopy guidance. Once inflated, the balloon aids the operator in crossing the tricuspid valve. After the catheter has crossed the valve, the balloon is deflated and the catheter is advanced to the right ventricular apex in the usual manner. Electrograms are monitored to verify proper positioning in the right ventricular apex and pacing thresholds are obtained to confirm both proper location and reliable pacing.<br>The catheters are available in two curve styles. The distal end of the catheter includes a platinum electrical connector ring and a tip electrode. The proximal end of the catheter includes a two pin electrical connector. The electrical signals can be transmitted to external equipment or from external equipment to the heart. |                                                                                                                                                                                                                                                                                                                                     |
| Intended Use:<br>(Indications for Use)                          | Pacelt™ Flow Directed Pacing Catheters are indicated for use in temporary, transvenous, right ventricular pacing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                     |
| Description of Change and<br>Comparison to Predicate<br>Devices | The Pacel™ Flow Directed Pacing Catheter that is the subject of this application remains substantially equivalent to the predicate device. There have been no changes to the packaging, sterilization, final device specifications or labeling. The proposed change is to switch the current moisture cured cyanoacrylate adhesives, Loctite 4013 which is used to bond the latex balloon to the catheter and Loctite 4031 which is used to seal/trim around the metal electrode ring edges to a visible light cured cyanoacrylate adhesive, Loctite 4311                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                     |
| 510(k) Summary                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
| Summary on Non-Clinical<br>Testing                              | Design Verification:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Design verification testing has been completed for the<br>Pacel Flow Direct catheter to show that the proposed<br>device will meet Product Specification through the<br>labelled Shelf Life of the device (T=18 mo). Design<br>verification testing is complete and passed all<br>acceptance criteria per the DV requirements.      |
|                                                                 | Testing Performed:<br>Surface Appearance: Adhesive on the balloon<br>and electrodes.<br>Results: Conforming   Balloon Repeat Inflation Strength  Results: Conforming   Inflation Decay  Results: Conforming                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                     |
|                                                                 | Biocompatibility:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Biocompatibility testing was performed in accordance<br>with ISO 10993. The proposed adhesive, Loctite 4311,<br>used to secure the balloon to the catheter shaft<br>demonstrated an acceptable biocompatibility profile.<br>Biocompatibility testing passed all acceptance criteria.                                                |
|                                                                 | Chronic GLP Animal Study:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Both the test and control devices were considered<br>substantially equivalent regarding clinical pathology,<br>thrombogenicity and pathology/histopathology. The use<br>of both devices was well tolerated. This data<br>demonstrates that the proposed Pacel Flow Direct<br>catheter's safety is comparable to the control device. |
| 510(k) Summary                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                     |
| Statement of Equivalence                                        | The Pacel         TM Flow Directed Pacing Catheters that are subject<br>of this application have the same indications for use and<br>technological characteristics as the predicate device. Based on<br>this and the data provided in this pre-market notification, the<br>subject device remains substantially equivalent to the<br>predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                     |

{4}------------------------------------------------

Pacel Pacing Catheter
Premarket Notification 510(k) K161873

Image /page/4/Picture/1 description: The image contains the logo for St. Jude Medical. The logo consists of a green square grid with one square rotated. To the right of the grid is the text "ST. JUDE MEDICAL" in a serif font. The text is aligned horizontally with the grid.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a green square grid on the left side and the text "St. Jude Medical" on the right side. The text is in a serif font, with "St." and "Jude" in a slightly larger font size than "Medical".

{6}------------------------------------------------

Pacel Pacing Catheter
Premarket Notification 510(k) K161873

Image /page/6/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a green square made up of nine smaller squares, with one of the squares rotated. To the right of the square is the text "St. Jude Medical" in a serif font.

---

**Source:** [https://fda-staging.innolitics.com/device/K161873](https://fda-staging.innolitics.com/device/K161873)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
