The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. The Miethke proSA Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
Device Story
Programmable/adjustable shunt systems (proGAV, proGAV 2.0, proSA) manage hydrocephalus by shunting CSF from lateral ventricles to peritoneum; valve mechanisms (leaf spring/ball) controlled by internal magnets; titanium housing. Physician uses external manual tools (verification compass, adjustment disc, Check-mate) to set/verify pressure levels pre- and post-operatively. Subject devices remove requirement for radiographic (X-ray) confirmation of pressure settings, relying instead on non-invasive verification tools. Benefits include reduced radiation exposure for patients during routine pressure adjustments.
Clinical Evidence
Bench testing only. Verification studies compared non-invasive measurement tools (Verification Tool, Verification Compass) against radiographic (X-ray) confirmation. Results demonstrated measurement agreement and met all acceptance criteria, supporting the removal of mandatory X-ray verification.
Technological Characteristics
Titanium alloy (Ti4Al6V) housing; Neodym Ferrite Boron magnets; Alpha Sapphire ball; leaf spring mechanism. Adjustable differential pressure (proGAV/proGAV 2.0) or gravitational (proSA) valves. Dimensions 1.5 x 1.5 mm. Sterile, single-use, steam sterilized. Manual external adjustment/verification tools.
Indications for Use
Indicated for patients requiring cerebrospinal fluid (CSF) shunting from the lateral ventricles of the brain into the peritoneum.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Miethke proGAV Programmable Shunt System (K062009 / K103003)
Miethke proGAV 2.0 Adjustable Shunt System (K141687)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, arranged in a cascading manner, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2017
Aesculap, Inc. Kathy A. Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K161853
Trade/Device Name: Miethke proGAV Programmable Shunt System, Miethke proGAV 2.0 Adjustable Shunt System, Miethke proSA Adjustable Shunt System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: January 20, 2017 Received: January 23, 2017
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Carlos L. Pena-S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161853
#### Device Name
Miethke proGAV Programmable Shunt System Miethke proGAV 2.0 Adjustable Shunt System Miethke proSA Adjustable Shunt System
Indications for Use (Describe)
The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
The Miethke proSA Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
| | Miethke proGAV Programmable Shunt System<br>February 15, 2017 |
|--------------------|--------------------------------------------------------------------------------------------------------------------|
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com |
| TRADE NAME: | Aesculap Miethke proGAV Programmable Shunt System |
| COMMON NAME: | Hydrocephalus Shunt System |
| CLASSIFICATION: | Class II |
| | CLASSIFICATION NAME: Shunt, Central Nervous System and Components |
| REGULATION NUMBER: | 882.5550 |
| PRODUCT CODE: | JXG |
## DEVICE DESCRIPTION
The proGAV is a "programmable" shunt that can be set for a range of pressures. The valve in the proGAV is a leaf spring and ball mechanism that is mechanically controlled by internal magnets. The outer case for the device is made of titanium. The shunt comes with a manual device to verify the pressure setting and another to set or re-set the pressure. These manual accessories are both for external use by the physician. Various Miethke shunt system accessories such as shunt assistants, catheters, connectors, deflectors and reservoirs are also offered with the proGAV .
## INDICATIONS FOR USE
The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
## SUBSTANTIAL EQUIVALENCE
Comparative information presented in the 510(k) supports the substantial equivalence of the subject proGAV Programmable Shunt System to the predicate proGAV Programmable Shunt System (K062009 / K103003). The proGAV Programmable Shunt System is substantially equivalent to the identified predicate with respect to performance characteristics, design, principles of operation and materials. The difference between the subject device and the
{4}------------------------------------------------
predicate is the removal of the required radiographic verification stated in the indications for use and labeling.
# TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The Aesculap Miethke proGAV Programmable Shunt System is substantially equivalent to the predicate Miethke proGAV Programmable Shunt System. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate device through comparison in design, principles of operation, and materials. The proGAV device characteristics are summarized below.
| | Miethke proGAV<br>Programmable Shunt System<br>K161853 | Miethke proGAV<br>Programmable Shunt System<br>K103003/K062009 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication: | The Miethke proGAV<br>Programmable Shunt System is<br>intended to shunt cerebrospinal<br>fluid (CSF) from the lateral<br>ventricles of the brain into the<br>peritoneum. | The Miethke proGAV<br>Programmable Shunt System is<br>intended to shunt cerebrospinal<br>fluid (CSF) from the lateral<br>ventricles of the brain into the<br>peritoneum. Adjustments of the<br>proGAV shunt can be verified by<br>using the verification instrument<br>but must be confirmed by<br>radiograph (X-ray). |
| Adjustable | Same | Yes |
| ValveType | Same | Adjustable differential pressure |
| Material: | | |
| Housing | Same | Titanium Alloy Ti4Al6V |
| Magnets | Same | Neodym Ferrite Boron |
| Ball | Same | Alpha Sapphire |
| Spring | Same | Titanium Alloy Ti4Al6V |
| Design | Same | Circular |
| Magnet: | | |
| # of magnets | Same | 2 |
| Dimensions | Same | 1.5 x 1.5 mm |
| Volume | Same | 5.3 mm3 |
| Pressure levels: | Same | Adjustable<br><br>0 - 20 cmH2O |
| Sterile/ Single Use | Same | Yes |
| Sterilization | Same | Steam |
| Packaging: | Same | Double Peel Pouch |
{5}------------------------------------------------
| | Miethke proGAV<br>Programmable Shunt System<br>K161853 | Miethke proGAV<br>Programmable Shunt System<br>K103003/K062009 |
|-----------------------------------------------------------------------------------------------|--------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Tools: | | |
| Verification tool<br>Masterdisc<br>Verification Compass<br>Adjustment tool<br>Adjustment disc | Same | Manually wipe with an alcohol<br>base cleaner (>75% alcohol). May<br>not be re-sterilized. |
| Check-mate | Same | For use in a sterile surgical field.<br>Sterile, reusable, may be re-<br>sterilized |
| Tool settings and<br>readings: | Same | 0 - 20 cmH2O |
# PERFORMANCE DATA
Bench testing on the proposed device, proGAV Programmable Shunt System, included the following:
- Verification testing assessing the measurement agreement between X-ray confirmation . and the proGAV Verification Tool when used with the proGAV Programmable Shunt System
- . Verification testing assessing the measurement between X-ray confirmation and the proGAV Verification Compass when used with the proGAV Programmable Shunt System
Test results demonstrated the substantial equivalence between the required radiographic verification method and the non-invasive method with the proGAV Verification Tool and proGAV Verification Compass. The test results met the acceptance criteria. Results of verification testing have demonstrated that the proposed proGAV Programmable Shunt System is substantially equivalent to the predicate proGAV Programmable Shunt System.
{6}------------------------------------------------
Page 1 of 3
## 510(k) SUMMARY
| Miethke proGAV 2.0 Adjustable Shunt System | |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| February 15, 2017 | |
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com |
| TRADE NAME: | Aesculap Meithke proGAV 2.0 Programmable Shunt System |
| COMMON NAME: | Hydrocephalus Shunt System |
| CLASSIFICATION: | Class II |
| CLASSIFICATION NAME: | Shunt, Central Nervous System and Components |
| REGULATION NUMBER: | 882.5550 |
| PRODUCT CODE: | JXG |
## DEVICE DESCRIPTION
proGAV 2.0 is an adjustable differential pressure valve that can be set for a range of pressures. The proGAV 2.0 valve is comprised of a titanium housing that contains a leaf spring and ball mechanism that is mechanically controlled by internal magnets. Manual devices are available to locate, verify the pressure setting and to set or re-set the pressure pre and postoperatively. These manual accessories are for external use by the physician. The device will be distributed by itself or in combination with the ShuntAssistant valve or proSA valve. The proGAV 2.0 adjustable differential pressure valve includes the same legally marketed accessories that are available with the Miethke Shunt Systems.
## INDICATIONS FOR USE
The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
## SUBSTANTIAL EQUIVALENCE
Comparative information presented in the 510(k) supports the substantial equivalence of the subject proGAV 2.0 Adjustable Shunt System to the predicate proGAV 2.0 Adjustable Shunt
{7}------------------------------------------------
System (K141687). The proGAV 2.0 Adjustable Shunt System is substantially equivalent to the identified predicate with respect to performance characteristics, design, principles of operation and materials. The difference between the subject device and the predicate is the removal of the required radiographic verification stated in the indications for use and labeling.
## TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The Aesculap Miethke proGAV 2.0 Adjustable Shunt System is substantially equivalent to the predicate Miethke proGAV 2.0 Adjustable Shunt System. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate device through comparison in design, principles of operation, and materials. The proGAV 2.0 device characteristics are summarized below.
| | Miethke proGAV 2.0 Adjustable<br>Shunt System<br>K161853 | Miethke proGAV 2.0 Adjustable<br>Shunt System<br>K141687 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication: | The Miethke proGAV 2.0<br>Adjustable Shunt System is<br>intended to shunt cerebrospinal<br>fluid (CSF) from the lateral<br>ventricles of the brain into the<br>peritoneum. | The Miethke proGAV 2.0<br>Adjustable Shunt System is<br>intended to shunt cerebrospinal<br>fluid (CSF) from the lateral<br>ventricles of the brain into the<br>peritoneum. Adjustments of the<br>proGAV 2.0 shunt can be verified<br>by using the verification<br>instrument but must be confirmed<br>by radiograph (X-ray). |
| Adjustable | Same | Yes |
| ValveType: | Same | Adjustable differential pressure |
| Material: | | |
| Housing | Same | Titanium Alloy Ti4Al6V |
| Magnets | Same | Neodym Ferrite Boron |
| Ball | Same | Alpha Sapphire |
| Spring | Same | Titanium Alloy Ti4Al6V |
| Design | Same | Circular |
| Magnet: | | |
| # of magnets | Same | 4 |
| Dimensions | Same | 1.5 x 1.5 mm |
| Volume | Same | 10.6 mm3 |
| Pressure levels: | Same | Adjustable |
| | Same | 0 - 20 cmH2O |
| Sterile/ Single Use | Same | Yes |
| Sterilization | Same | Steam |
| Packaging: | Same | Double Peel Pouch |
{8}------------------------------------------------
| | Miethke proGAV 2.0 Adjustable<br>Shunt System | Miethke proGAV 2.0 Adjustable<br>Shunt System |
|----------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| | K161853 | K141687 |
| <b>proGAV 2.0 Tools:</b> | | |
| Compass | Same | Manually wipe with an alcohol |
| Adjustment tool | Same | base cleaner (>75% alcohol).<br>May not be re-sterilized. |
| <b>Compatible w/<br/>proGAV Tools:</b> | | |
| | Same | Verification tool, Verification<br>compass, Master disk, Adjustment<br>tool, Adjustment disk, Check-<br>mate |
| Tool settings and<br>readings: | Same | 0 – 20 cmH2O |
## PERFORMANCE DATA
Bench testing on the proposed device, proGAV 2.0 Adjustable Shunt System, included the following:
- Verification testing assessing the measurement agreement between X-ray confirmation . and the proGAV 2.0 Compass when used with the proGAV 2.0 Adjustable Shunt System
- Verification testing assessing the measurement between X-ray confirmation and the proGAV Verification Compass when used with the proGAV 2.0 Adjustable Shunt System
- Verification testing assessing the measurement agreement between X-ray confirmation and the proGAV Verification Tool when used with the proGAV 2.0 Adjustable Shunt System
Test results demonstrated the substantial equivalence between the required radiographic verification method and the non-invasive method with the proGAV 2.0 Compass, proGAV Verification Compass and proGAV Verification Tool. The test results met the acceptance criteria. Results of verification testing have demonstrated that the proposed proGAV 2.0 Adjustable Shunt System is substantially equivalent to the predicate proGAV 2.0 Adjustable Shunt System.
{9}------------------------------------------------
#### 510(k) SUMMARY
| Miethke proSA Adjustable Shunt System | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| February 15, 2017 | |
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com |
| TRADE NAME: | Aesculap Miethke proSA Adjustable Shunt System |
| COMMON NAME: | Hydrocephalus Shunt System |
| CLASSIFICATION: | Class II |
| CLASSIFICATION NAME: | Shunt, Central Nervous System and Components |
| REGULATION NUMBER: | 882.5550 |
| PRODUCT CODE: | JXG |
## DEVICE DESCRIPTION
proSA is an adjustable gravitational valve that can be set for a range of pressures. The proSA valve is comprised of a titanium housing that contains a tantalum weight, leaf spring and ball mechanism that is mechanically controlled by internal magnets. Several manual devices are available to verify the pressure setting and to set or re-set the pressure pre and postoperatively. These manual accessories are for external use by the physician. The device will be distributed by itself or in combination with the miniNAV valve or proGAV valve. The proSA adjustable gravitational valve includes the same legally marketed accessories that are available with the Miethke Shunt Systems.
## INDICATIONS FOR USE
The Miethke proSA Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
## SUBSTANTIAL EQUIVALENCE
Comparative information presented in the 510(k) supports the substantial equivalence of the subject proSA Adjustable Shunt System to the predicate proSA Adjustable Shunt System
{10}------------------------------------------------
(K120559). The proSA Adjustable Shunt System is substantially equivalent to the identified predicate with respect to performance characteristics, design, principles of operation and materials. The difference between the subject device and the predicate is the removal of the required radiographic verification stated in the indications for use and labeling.
## TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The Aesculap Miethke proSA Adjustable Shunt System is substantially equivalent to the predicate Miethke proSA Adjustable Shunt System. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate device through comparison in design, principles of operation, and materials. The proSA device characteristics are summarized below.
| | Miethke proSA Adjustable<br>Shunt System<br>K161853 | Miethke proSA Adjustable<br>Shunt System<br>K120559 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication: | The Miethke proSA Adjustable<br>Shunt System is intended to shunt<br>cerebrospinal fluid (CSF) from<br>the lateral ventricles of the brain<br>into the peritoneum. | The Miethke proSA Adjustable<br>Shunt System is intended to shunt<br>cerebrospinal fluid (CSF) from<br>the lateral ventricles of the brain<br>into the peritoneum. Adjustments<br>of the proSA shunt can be verified<br>by using the verification<br>instrument but must be confirmed<br>by radiograph (X-ray). |
| Adjustable | Same | Yes |
| ValveType | Same | Adjustable Gravitational valve |
| Material: | | |
| Housing | Same | Titanium Alloy Ti4Al6V |
| Magnets | Same | Neodym Ferrite Boron |
| Ball | Same | Alpha Sapphire |
| Spring | Same | Titanium Alloy Ti4Al6V |
| Weight | Same | Tantalum |
| Design | Same | Circular |
| Magnet: | | |
| # of magnets | Same | 2 |
| Dimensions | Same | 1.5 x 1.5 mm |
| Volume | Same | 5.3 mm3 |
| Pressure levels: | Same | Adjustable |
| | | |
| | Same | 0 - 40 cmH2O |
| Sterile/ Single Use | Same | Yes |
| Sterilization | Same | Steam |
| Packaging: | Same | Double Peel Pouch |
{11}------------------------------------------------
| | Miethke proSA Adjustable<br>Shunt System | Miethke proSA Adjustable<br>Shunt System |
|-----------------------------------------------------------------------------------------------|------------------------------------------|--------------------------------------------------------------------------------------------|
| | K161853 | K120559 |
| <b>Tools:</b> | | |
| Verification tool<br>Masterdisc<br>Verification Compass<br>Adjustment tool<br>Adjustment disc | Same | Manually wipe with an alcohol<br>base cleaner (>75% alcohol). May<br>not be re-sterilized. |
| Check-mate | Same | For use in a sterile surgical field.<br>Sterile, reusable, may be re-<br>sterilized |
| Tool settings and<br>readings: | Same | 0 - 40 cmH2O |
# PERFORMANCE DATA
Bench testing on the proposed device, proSA Adjustable Shunt System, included the following:
- Verification testing assessing the measurement between X-ray confirmation ● and the proSA Verification Tool when used with the proSA Adjustable Shunt System
- Verification testing assessing the measurement between X-ray confirmation and the proSA Verification Compass when used with the proSA Adjustable Shunt System
Test results demonstrated the substantial equivalence between the required radiographic verification method and the non-invasive method with the proSA Verification Tool and proSA Verification Compass. The test results met the acceptance criteria. Results of verification testing have demonstrated that the proposed proSA Adjustable Shunt System is substantially equivalent to the predicate proSA Adjustable Shunt System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.