K161848 · Cefla S.C. · FLE · Feb 23, 2017 · General Hospital
Device Facts
Record ID
K161848
Device Name
Family of Steam sterilizers B17 B22 Series
Applicant
Cefla S.C.
Product Code
FLE · General Hospital
Decision Date
Feb 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
B CLASSIC-17 is a steam sterilizer that is intended for use by a heath care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process. The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
Device Story
Table-top steam sterilizers (B17/B22 series) for healthcare facilities; use saturated steam at high pressure/temperature to sterilize re-usable medical instruments. Input: demineralized water; process: fractionated vacuum system removes air from hollow/porous materials; electric heating elements maintain chamber temperature. Output: sterilized instruments; cycle reports (PDF/USB/Ethernet/Wi-Fi). Operated by healthcare staff; output used to confirm sterilization success via cycle outcome (positive/negative) and process parameters (temp/pressure/time). Benefits: ensures sterility of re-usable medical devices, including dental handpieces and porous textiles.
Clinical Evidence
Bench testing only. Validated per AAMI ST55:2010. Biological indicators achieved SAL 10^-6. Chamber temperature verified within -0°C to +3°C of set point. Bowie-Dick test pack showed uniform color change. Air leak rate < 1mmHg. Moisture retention < 0.5% for wrapped instruments and < 2% for textiles.
Technological Characteristics
Table-top steam sterilizer; 304/316L stainless steel chamber; fractionated vacuum system; electric heating foil; 17L or 22L volume; connectivity via Ethernet/Wi-Fi/USB; sterilization temps 121°C/134°C; pressure 1.1/2.1 bar; air filtration 0.027 μm (99.999% efficiency).
Indications for Use
Indicated for use by healthcare providers (hospitals, medical/dental practices, clinics, nursing homes) to sterilize heat/moisture-resistant, steam-compatible re-usable medical products. Contraindicated for fluids, liquids, or pharmaceutical products.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines extending from the bottom profile. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 23, 2017
Cefla S.C. c/o Maurizio Pantaleoni ISEMED srl Via Altobelli Bonetti 3/A Imola, Bologna 40026 ITALY
Re: K161848
Trade/Device Name: Family of Steam Sterilizers B17 B22 Series Regulation Number: 21 CFR 880.6880 Regulation Name: Steam sterilizer Regulatory Class: Class II Product Code: FLE Dated: January 15, 2017 Received: January 18, 2017
Dear Maurizio Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161848
Device Name Family of Steam sterilizers B17 B22 series
Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
B CLASSIC-17 is a steam sterilizer that is intended for use by a heath care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
Program description, cycle times, temperature and dry time are described in the table below:
{3}------------------------------------------------
### B CLASSIC-17 model
| MODEL NAME | CYCLE NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|--------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B Classic-17 | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 12.75 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C2)<br>UNIVERSAL<br>121<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 121°C / 250°F<br>20 minutes | 12.75 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C4) SOLID 134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 12.75 min | 4 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
| | (C5) HOLLOW<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 3 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray) |
| | HELIX/BD<br>TEST<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>35 minutes | 1.5 min | Test device only (without another load) |
| | VACUUM TEST<br>(-0.8 bar) | / | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| | VACUUM +<br>HELIX/BD<br>TEST (****) | / | / | / |
B CLASSIC-17 Factory programmed sterilization cycles and parameters
(*) Default drying time, but the drying time can be manually increased
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
__ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161848
Device Name Family of Steam sterilizers B17 B22 series
Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
B CLASSIC-22 is a steam sterilizer that is intended for use by a heath care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
Program description, cycle times, temperature and dry time are described in the table below:
{6}------------------------------------------------
### B CLASSIC-22 model
| | | | Chamber Dimensions (Depth 450mm / Diameter 250mm / Volume about 22 liters) | |
|--|--|--|----------------------------------------------------------------------------|--|
|--|--|--|----------------------------------------------------------------------------|--|
| MODEL NAME | CYCLE<br>NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|--------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B Classic-22 | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 14.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
| | (C2)<br>UNIVERSAL<br>121<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 121°C / 250°F<br>20 minutes | 14.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
| | (C4) SOLID<br>134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 14.5 min | 4.50 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
| | (C5)<br>HOLLOW 134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 3 min | 4.50 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray) |
| | HELIX/BD<br>TEST<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>3.5 minutes | 1.5 min | Test device only (without another load) |
| | VACUUM<br>TEST<br>(-0.8 bar) | / | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| | VACUUM +<br>HELIX/BD<br>TEST (****) | / | / | / |
B CLASSIC-22 Factory programmed sterilization cycles and parameters
(*) Default drying time, but the drying time can be manually increased
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
{7}------------------------------------------------
Type of Use (Select one or both, as applicable)
__ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161848
Device Name Family of Steam sterilizers B17 B22 series
Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
B FUTURA-17 is a steam sterilizer that is intended for use by a heath care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
Program description, cycle times, temperature and dry time are described in the table below:
{9}------------------------------------------------
### B FUTURA-17 model
Chamber Dimensions (Depth 350mm / Diameter 250mm / Volume about 17 liters)
| MODEL NAME | CYCLE<br>NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|-------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B Futura-17 | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 9.67 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C2)<br>UNIVERSAL<br>121<br>(3 prevacuum<br>steps: -0.8 bar<br>each step) | 121°C / 250°F<br>20 minutes | 9.67 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C4) SOLID<br>134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 9.67 min | 4 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
| | (C5)<br>HOLLOW 134<br>(3 prevacuum<br>steps: -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 3 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray) |
| | HELIX/BD<br>TEST<br>(3 prevacuum<br>steps: -0.8 bar<br>each step) | 134°C / 273°F<br>3.5 minutes | 1.33 min | Test device only (without another load) |
| | VACUUM<br>TEST<br>(-0.8 bar) | / | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| | VACUUM +<br>HELIX/BD<br>TEST (****) | / | / | |
B FUTURA-17 Factory programmed sterilization cycles and parameters
(*) Default drying time, but the drying time can be manually increased
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
{10}------------------------------------------------
Type of Use (Select one or both, as applicable)
__ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{11}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161848
Device Name Family of Steam sterilizers B17 B22 series
Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
B FUTURA-22 is a steam sterilizer that is intended for use by a heath care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
Program description, cycle times, temperature and dry time are described in the table below:
{12}------------------------------------------------
# B FUTURA-22 model
Chamber Dimensions (Depth 450mm / Diameter 250mm / Volume about 22 liters)
| MODEL NAME | CYCLE<br>NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|-------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B Futura-22 | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 11.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
| | (C2)<br>UNIVERSAL<br>121<br>(3 prevacuum<br>steps: -0.8 bar<br>each step) | 121°C / 250°F<br>20 minutes | 11.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
| | (C4) SOLID<br>134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 11.5 min | 4.50 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
| | (C5)<br>HOLLOW 134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 3 min | 4.50 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray) |
| | HELIX/BD<br>TEST<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>3.5 minutes | 1.33 min | Test device only (without another load) |
| | VACUUM<br>TEST<br>(-0.8 bar) | / | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| | VACUUM +<br>HELIX/BD<br>TEST | / | / | / |
B FUTURA-22 Factory programmed sterilization cycles and parameters
(*) Default drying time, but the drying time can be manually increased
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
{13}------------------------------------------------
Type of Use (Select one or both, as applicable)
__ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{14}------------------------------------------------
# 510(k) Summary
### Family of Steam sterilizers B17 B22 series
## K161848
#### 2.1. General Information
Submitter:
CEFLA S.C. Via Selice Provinciale, 23/A 40026 - IMOLA (BO) ITALY
Establishment Registration Number: 3006610845
| Contact Person in Italy: | Maurizio Pantaleoni<br>ISEMED srl<br>Via Altobelli Bonetti 3/a<br>40026 Imola (BO)<br>Italy<br>Mob.phone: +39-348 4435155<br>Telephone: +39-542 683803<br>Fax: +39-542 698456<br>Email: regulatory@isemed.eu |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary Preparation Date: | February 15, 2017 |
2.2. Names
| Device Name: | Family of Steam sterilizer B17 B22 series |
|----------------------|-------------------------------------------|
| Classification Name: | Steam Sterilizer |
| Product Code: | FLE |
| Regulation number: | 880.6880 |
| Class: | II |
#### Predicate Devices 2.3.
The Family of steam sterilizers B17 B22 series is substantially equivalent to the following legally marketed predicate device:
| Applicant | Carlos Concession Comers of Concession Comers of Concession Comers of Concession Comers of Concession Comers of Concession Comers of Concession Comers of Concession Comers of<br>Device name | 510(k) Number |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| CEFLA S.C. | Family of steam sterilizers B28 series | K151597 |
{15}------------------------------------------------
#### 2.4. Indications for Use
B CLASSIC-17 is a steam sterilizer that is intended for use by a health care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
#### B CLASSIC-17 model
Chamber Dimensions (Depth 350mm / Diameter 250mm / Volume about 17 liters)
| MODEL NAME | CYCLE NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|--------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B Classic-17 | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 12.75 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C2)<br>UNIVERSAL<br>121<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 121°C / 250°F<br>20 minutes | 12.75 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C4) SOLID 134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 12.75 min | 4 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
| | (C5) HOLLOW<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 3 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray) |
| | HELIX/BD<br>TEST<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>3.5 minutes | 1.5 min | Test device only (without another load) |
| | VACUUM TEST<br>(-0.8 bar) | / | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| | VACUUM +<br>HELIX/BD<br>TEST | / | / | / |
Table 1 - B CLASSIC-17 Factory programmed sterilization cycles and parameters
(*) Default drying time, but the drying time can be manually increased
{16}------------------------------------------------
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(***) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
B CLASSIC-22 is a steam sterilizer that is intended for use by a health care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
## B CLASSIC-22 model
Chamber Dimensions (Depth 450mm / Diameter 250mm / Volume about 22 liters)
| MODEL NAME | CYCLE<br>NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|--------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 14.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
| B Classic-22 | (C2)<br>UNIVERSAL<br>121<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 121°C / 250°F<br>20 minutes | 14.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
| | (C4) SOLID<br>134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 14.5 min | 4.50 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
| | (C5)<br>HOLLOW 134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 3 min | 4.50 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray) |
| | HELIX/BD<br>TEST<br>(3 prevacuum<br>steps; -0.8 bar | 134°C / 273°F<br>3.5 minutes | 1.5 min | Test device only (without another load) |
{17}------------------------------------------------
| each step) | | |
|-------------------------------------|---|-------------------------------------------------------|
| VACUUM<br>TEST<br>(-0.8 bar) | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| VACUUM +<br>HELIX/BD<br>TEST (****) | / | / |
Table 3 - B CLASSIC-22 Factory programmed sterilization cycles and parameters
(*) Default drying time, but the drying time can be manually increased
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
B FUTURA-17 is a steam sterilizer that is intended for use by a health care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
#### B FUTURA-17 model
Chamber Dimensions (Depth 350mm / Diameter 250mm / Volume about 17 liters)
| MODEL NAME | CYCLE<br>NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|-------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B Futura-17 | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 9.67 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C2)<br>UNIVERSAL<br>121<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 121°C / 250°F<br>20 minutes | 9.67 min | 4 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.6 kg unwrapped porous and fabric textiles<br>0.5 kg single package porous and fabric textiles<br>0.4 kg double package porous and fabric textiles |
| | (C4) SOLID<br>134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 9.67 min | 4 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
{18}------------------------------------------------
| (C5)<br>HOLLOW 134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C/273°F<br>4 minutes | 3 min | 4 kg unwrapped metal or polymer instruments and dental handpieces (max 1.33 kg per tray) |
|---------------------------------------------------------------------|----------------------------|----------|------------------------------------------------------------------------------------------|
| HELIX/BD<br>TEST<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C/273°F<br>3.5 minutes | 1.33 min | Test device only (without another load) |
| VACUUM<br>TEST<br>(-0.8 bar) | / | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| VACUUM +<br>HELIX/BD<br>TEST (****) | / | / | / |
#### Table 2 - B FUTURA-17 Factory programmed sterilization cycles and parameters
(*) Default drying time, but the drying time can be manually increased
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
B FUTURA-22 is a steam sterilizer that is intended for use by a health care provider (i.e. hospitals, medical and dental practices, clinics, and nursing homes) to sterilize re-usable medical products that are heat and moisture resistant and compatible with the steam sterilization process.
The device shall not be used for the sterilization of fluids, liquids or pharmaceutical products.
#### B FUTURA-22 model
Chamber Dimensions (Depth 450mm / Diameter 250mm / Volume about 22 liters)
| MODEL NAME | CYCLE<br>NAME | STERILIZATION<br>TEMPERATURE /<br>EXPOSURE TIME | DRYING<br>TIME (*) | DESCRIPTION OF MAXIMUM LOAD AND TYPE<br>(**) (***) |
|--------------------|---------------------------------------------------------------------------|-------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>B Futura-22</b> | (C1)<br>UNIVERSAL<br>134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 11.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
| | (C2)<br>UNIVERSAL<br>121 | 121°C / 250°F<br>20 minutes | 11.5 min | 4.5 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments |
{19}------------------------------------------------
| (3 prevacuum<br>steps; -0.8 bar<br>each step) | | | and dental handpieces (max 1.33 kg per tray)<br>1.25 kg double package metal or polymer instruments<br>and dental handpieces (max 1.33 kg per tray)<br>0.75 kg unwrapped porous and fabric textiles<br>0.60 kg single package porous and fabric textiles<br>0.50 kg double package porous and fabric textiles |
|---------------------------------------------------------------------|------------------------------|----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (C4) SOLID<br>134<br>(1 prevacuum<br>step; -0.8 bar) | 134°C / 273°F<br>4 minutes | 11.5 min | 4.50 kg unwrapped metal or polymer instruments<br>(max 1.33 kg per tray)<br>2.25 kg single package metal or polymer instruments<br>(max 1.33 kg per tray) |
| (C5)<br>HOLLOW 134<br>(3 prevacuum<br>steps; -0.8 bar<br>each step) | 134°C / 273°F<br>4 minutes | 3 min | 4.50 kg unwrapped metal or polymer instruments and<br>dental handpieces (max 1.33 kg per tray) |
| (3 prevacuum<br>steps; -0.8 bar<br>each step)<br>HELIX/BD<br>TEST | 134°C / 273°F<br>3.5 minutes | 1.33 min | Test device only (without another load) |
| (3 prevacuum<br>steps; -0.8 bar<br>each step)<br>VACUUM<br>TEST | / | / | Empty chamber<br>Vacuum -0.8 bar<br>Test Time: 27 min |
| (-0.8 bar)<br>VACUUM +<br>HELIX/BD<br>TEST (****) | / | / | / |
Table 4 - B FUTURA-22 Factory programmed sterilization cycles and parameters (*) Default drying time, but the drving time can be manually increased
(**) Cycles C1, C2 are validated to sterilize up to 6 single sided lumened devices with inner diameter/length of the lumen: ≥1.25 mm/≤118 mm (corresponding to "hollow A" cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750).
Cycles C4 and C5 are validated to sterilize up to 6 double sided lumened devices with inner diameter/length of the lumen: ≥10 mm / ≤115 mm
(***) If the material is sterilized "unwrapped" the sterilized material shall used immediately after sterilization (the cycle shall be considered as immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
#### 2.5. Device Description
The Family of steam sterilizers B17 B22 series includes table-top steam sterilizers that use saturated steam at high pressures and temperatures to kill infectious bio-organisms. The Classic and B Futura models differ only for some minor characteristics such as color touchscreen, optional Wi-FI connection, illumination of the working area and Ethernet connection.
The steam is generated by a steam generator connected to the sterilization chamber without disconnecting devices. An electric heating element surrounds the chamber for preheating of the chamber and maintaining the temperature. The subject family is characterized by a fractionated vacuum system for the complete air removal also from hollow and porous materials as well as by a final vacuum drying phase able to eliminate any trace of condensation from the load.
{20}------------------------------------------------
| Features | Subject Device | | Predicate Device<br>(K151597) | Substantially<br>Equivalent |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| | B FUTURA-17<br>B CLASSIC-17 | B FUTURA-22<br>B CLASSIC-22 | B FUTURA-28<br>B CLASSIC-28 | |
| Regulation<br>number | 880.6880 – Steam Sterilizers | 880.6880 – Steam Sterilizers | 880.6880 – Steam Sterilizers | Same |
| Class | II | II | II…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.