The Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist. It can be used at medical facilities or at home. The intended wrist circumference is 13.5-19.5 cm. The patient population does not include adolescents aged 12 to <18 years of age. The patient population includes transition adolescent B (18 to < 22 years of age but treated like adults (at least 22 years of age).
Device Story
Battery-powered automatic wrist blood pressure monitor; uses oscillometric technique to measure systolic/diastolic blood pressure and pulse rate. Device consists of LCD, keys, cuff, MCU, pump, and transducer. Used in medical facilities or home settings by adults and transition adolescents. User wraps inflatable cuff around wrist; device automatically inflates and performs measurement. Output displayed on LCD; allows selection of mmHg or Kpa units. Data storage provided for two-patient mode. Healthcare providers use output for routine blood pressure monitoring; assists in clinical decision-making regarding hypertension management. Benefits include convenient, non-invasive, self-administered blood pressure tracking.
Clinical Evidence
Bench testing only. Compliance verified with IEC 60601-1 (safety), IEC 60601-1-11 (home healthcare), IEC 60601-1-2 (EMC), IEC 80601-2-30 (automated sphygmomanometers), and ISO 81060-2:2013 (clinical validation of automated measurement). Biocompatibility testing per ISO 10993 confirmed no cytotoxicity, irritation, or sensitization.
Indicated for adults (at least 22 years old) and transition adolescents (18 to <22 years old) for non-invasive measurement of systolic/diastolic blood pressure and pulse rate via wrist-worn oscillometric cuff. Intended for medical facilities or home use. Wrist circumference 13.5-19.5 cm. Contraindicated for adolescents aged 12 to <18 years.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
PG-800A Series Electronic Blood Pressure Monitor Models: PG-800A5, PG-800A5D (K131569)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the United States Department of Health and Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a stylized human profile. The logo is black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2017
Shenzhen Pango Electronic Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, CN 200120
Re: K161845
Trade/Device Name: Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: July 3, 2017 Received: June 27, 2017
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Hilleman
for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K161845
#### Device Name Electronic Blood Pressure Monitor
#### Indications for Use (Describe)
The Electronic Blood Pressure Monitor is intended to measure the systolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist. It can be used at medical facilities or at home. The intended wrist circumference is 13.5-19.5 cm. The patient population does not include adolescents aged 12 to <18 years of age. The patient population includes transition adolescent B (18 to < 22 years of age but treated like adult) and adults (at least 22 years of age).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K161845
- 1. Date of Preparation: 7/11/2017
- 2. Sponsor Identification
### Shenzhen Pango Electronic Co.,Ltd.
No.25, 1st Industrial Park, Fenghuang Road, Xikeng, Henggang, Longgang District, Shenzhen, Guangdong, 518115, China.
Establishment Registration Number: 3006792041
Contact Person: Ms. Xiaoyun Yang Position: Vice General Manager Tel: +86-755-33825988 Fax: +86-755-33825989 Email:sales@pan-go.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
# Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
- 4. Identification of Proposed Device
Trade Name: Electronic Blood Pressure Monitor; Common Name: Wrist Blood Pressure Monitor; PG-800A25, PG-800A27, PG-800A31, PG-800A32, PG-800A33, PG-800A35, Models: PG-800A36 and PG-800A37
{4}------------------------------------------------
Regulatory Information Classification Name: Noninvasive blood pressure measurement system Classification: 2 Product Code: DXN; Regulation Number: 21 CFR 870.1130; Review Panel: Cardiovascular;
#### Intended Use Statement:
The Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist. It can be used at medical facilities or at home. The intended wrist circumference is 13.5-19.5 cm.
The patient population does not include adolescents aged 12 to <18 years of age. The patient population includes transition adolescent B (18 to < 22 years of age but treated like adults (at least 22 years of age).
Device Description:
The proposed device, Electronic Blood Pressure Monitor, is a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation and measurement, which can measure systolic and diastolic blood pressure and pulse rate of the adult person at wrist within its claimed range and accuracy via the oscillometric technique. User can select the unit of the measurement: mmHg or Kpa.
The proposed electronic blood pressure monitor has eight models, including PG-800A25, PG-800A27, PG-800A31, PG-800A32, PG-800A33, PG-800A36 and PG-800A37. All models follow the same software, measurement principle, algorithm and data storage. The main differences are product appearance.
The proposed device is intended to be used in medical facilities or at home.
#### Identification of Predicate Device 5.
510(k) Number: K131569 Product Name: PG-800A Series Electronic Blood Pressure Monitor Models: PG-800A5, PG-800A5D Manufacturer: Shenzhen Pango Electronic Co., Ltd
{5}------------------------------------------------
#### Non-Clinical Test Conclusion 6.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2005+ A1:2002, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-11:2010, Medical electrical equipment - Part 1-11: General requirements for basic safety, and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility
IEC 80601-2-30:2009, Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
ISO 81060-2:2013, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation of Automated Measurement Type.
| Table 1 Substantially Equivalent Comparison | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ITEM | Proposed Device | Predicate Device<br>K131569 |
| Models | PG-800A25, PG-800A27, PG-800A31,<br>PG-800A32, PG-800A33, PG-800A35,<br>PG-800A36 and PG-800A37 | PG-800A5, PG-800A5D |
| Product Code | DXN | DXN |
| Class | II | II |
| Intended Use | The Electronic Blood Pressure Monitor is<br>intended to measure the systolic and<br>diastolic blood pressure as well as the<br>pulse rate of adult person via non-invasive<br>oscillometric technique in which an<br>inflatable cuff is wrapped around the<br>wrist. It can be used at medical facilities or<br>at home. The intended wrist circumference<br>is 13.5-19.5 cm.<br>The patient population does not include | The PG-800A Series Electronic Blood Pressure<br>Monitor is intended to measure diastolic, systolic blood<br>pressure as well as the pulse rate of adult person via<br>non-invasive technique in which an inflatable cuff is<br>wrapped around the wrist.<br>It can be used at medical facilities or at home.<br>The intended wrist circumference is 13.5cm-19.5cm. |
| adolescents aged 12 to <18 years of age.<br>The patient population includes transition<br>adolescent B (18 to < 22 years of age but<br>treated like adult) and adults (at least 22<br>years of age). | | |
| Measurement Type | Wrist | Wrist |
| Patient Population | Adult | Adult |
| Measurement Item | Systolic Pressure, Diastolic Pressure, Pulse Rate | Systolic Pressure, Diastolic Pressure, Pulse Rate |
| Principle | Oscillometric | Oscillometric |
| Component | LCD / Key / Cuff / MCU / Pump /Transducer/ Batteries | LCD / Key / Cuff / MCU / Pump /Transducer/ Batteries |
| Blood Pressure | 30 ~ 280 mmHg | 30 ~ 280 mmHg |
| Range | | |
| Pulse Rate | 40-199 bpm | 40-199 bpm |
| Range | | |
| Intended wrist<br>circumference | 13.5cm~19.5cm | 13.5cm~19.5cm |
| Records Quantity | Double patients mode: 60/60 records | Single patient mode: 90 records for PG-800A5<br>Double patient mode: 60/60 records for PG-800A5D |
| | | |
| Operating condition | +5 °C~+40°C<br>30%RH~80%RH;<br>Atomospheric pressure | +5 °C~+40°C<br>30%RH~80%RH;<br>Atomospheric pressure |
| | | |
| | | |
| Storage condition | -20°C~+55°C;<br>10%RH~93%RH;<br>Atomospheric pressure | -20°C~+55°C;<br>10%RH~93%RH;<br>Atomospheric pressure |
| | | |
| | | |
| Patient Contact | Cuff – Nylon | Cuff – Nylon |
| Material | Enclosure - ABS<br>Key - ABS | Enclosure - ABS<br>Key - ABS |
| Bicompatibility | Comply with ISO10993 series standards<br>No cytotoxicity;<br>No irritation to skin;<br>No significant evidence of sensitization | Comply with ISO10993 series standards<br>No cytotoxicity:<br>No irritation to skin;<br>No significant evidence of sensitization |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2:2007 | Comply with IEC 60601-1-2:2007 |
| Particular<br>Performance | Comply with IEC<br>80601-2-30:2009 and ISO 81060-2:2013 | ANSI/AAMI SP10 |
#### Substantially Equivalent 7.
{6}------------------------------------------------
# Product Comparison Summary
The propoed device and its predicate devices have the similar indented use, mechanism of action, principle of operation, algorithm and cuff. The proposed devices are different than the predicate device in that the proposed device integrates the Key function of both PG-800A5 and PG-800A5D, same Record quantity as that of PG-800A5D, different Product appareance, modified product label
{7}------------------------------------------------
and software which is modified to applicable to the proposed devices; addtitionally the proposed device details the patient populations. The difference between the proposed device and the predicate device do not raise any question regarding its safety and effectiveness.
- 8. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.