Huvitz Imaging system (HIS-5000U) is intended for controlling digital imaging devices and for acceptance, transfer, display, storage and digital processing of documentation ophthalmic images and videos, acquired from a slit lamp camera through a direct connection to a PC or to multiple PCs via a local network.
Device Story
Huvitz Imaging System (HIS-5000U) comprises slit lamp camera, communication cable, and HIS-5000N imaging software. Device captures ophthalmic images/videos via direct PC connection or local network. Software manages patient diagnostic data, supports image enhancement (brightness, contrast, sharpness), measurement (pixels/mm), and side-by-side image comparison. Used in clinical settings by eye care professionals to view and store findings; does not provide automated diagnosis. System utilizes MS-SQL server database for patient records and supports data backup/restore. Output is displayed on PC monitor for clinician review to assist in documentation and patient management.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via camera image capture, file saving, color adjustment, measurement accuracy, and data transfer speed tests per manufacturer specifications.
Technological Characteristics
PC-based software (HIS-5000N) and slit lamp camera (HIS-5000L/U). Connectivity via USB 2.0 and local network (TCP/IP). Database management via MS-SQL server. Supports JPEG format. Features include image enhancement, measurement tools, and multi-PC networking. No DICOM interface.
Indications for Use
Indicated for use in ophthalmic clinical settings to manage, store, and process diagnostic images and videos acquired from slit lamp cameras. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Huvitz Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway Houston, TX 77025
Re: K161829
Trade/Device Name: Huvitz Imaging System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: NFJ Dated: March 2, 2017 Received: March 6, 2017
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161829
Device Name Huvitz Imaging System
Indications for Use (Describe)
Huvitz Imaging system (HIS-5000U) is intended for controlling digital imaging devices and for acceptance, transfer, display, storage and digital processing of documentation ophthalmic images and videos, acquired from a slit lamp camera through a direct connection to a PC or to multiple PCs via a local network.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
| Date Prepared:<br>Submitter: | April 4, 2017<br>Huvitz Co., Ltd.<br>298-29, Gongdan-ro, Gunpo-so, Gyeonggi-do, 435-862, Republic of Korea<br>Tel: +82-31-428-9100<br>Establishment Registration Number: 9616199<br>E-mail: heohm@huvitz.com |
|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Dave Kim<br>Mtech Group.<br>8310 Buffalo Speedway, Houston, TX 77025<br>Tel: +713-467-2607<br>Fax: +713-583-8988<br>E-mail: davekim@mtech-inc.net |
| Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class::<br>Product Code: | Huvitz Imaging System (HIS-5000U)<br>System, Image Management, Ophthalmic.<br>21 CFR 892.2050<br>Picture archiving and communications system<br>Class II<br>NFJ |
| Predicate Device:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class::<br>Product Code: | EyeSuite Imaging, K142423<br>21 CFR 892.2050<br>Picture archiving and communications system<br>Class II<br>NFJ |
| Device Description: | Huvitz Imaging System (HIS-5000U) consists of slit lamp camera (HIS-5000L<br>(10.0M), HIS-5000U(1.4M)), Communication Cable and Imaging Software<br>(HIS-5000N). It is a digital imaging system with a high resolution digital<br>camera and PC based patient diagnostic data management software<br>intended for controlling digital imaging devices, acceptance, transfer,<br>display, storage and digital processing of ophthalmic images including<br>videos, acquired from a of a slit lamp camera through direct connection or<br>network.<br>The imaging software (HIS-5000N) supports networking function to share<br>patient information and examination images between multiple PCs existing<br>on the local network or internet via TCP/IP protocol. Also, this version uses<br>MS-SQL server solution as more reliable database, and includes additional<br>HIS Server Management software to support to back up or restore patient<br>and imaging data. HIS-5000N imaging system software is not able and |
{4}------------------------------------------------
intended to provide any diagnosis but help the user to view the visible structures of the eye and store the findings.
- Intended Use: Huvitz Imaging system (HIS-5000U) is intended for controlling digital imaging devices and for acceptance, transfer, display, storage and digital processing of documentation ophthalmic images and videos, acquired from a slit lamp camera through a direct connection to a PC or to multiple PCs via a local network.
#### Technological characteristics:
Huvitz Imaging system (HIS-5000U) has been compared to the predicated device. The following attachment shows differences and similarities among the device.
#### Non-clinical and clinical tests:
Performance testing was conducted on the Huvitz Imaging System (HIS-System and supported instrument testing was completed based 5000U). on product specifications and hazard effects determined from the risk analysis. The Huvitz Imaging System (HIS-5000U) performed as intended and is deemed substantially equivalent to and comparable to the predicate device, an ophthalmic patient database device.
Also, the following tests have been conducted to validate the performance of Huvitz Imaging System (HIS-5000U).
1) Camera Image capture, 2) Saving image file, 3) Adjust color of capture image, 4) Measurement of figures, 5) Camera resolution & data transfer speed as per the manufacturer's performance report.
No clinical test has been conducted.
Conclusions: Based on non-clinical tests, the subject device, Huvitz Imaging System (HIS-5000U) introduces no new concern for safety and effectiveness. Therefore it is the sponsor's opinion that Huvitz Imaging System (HIS-5000U) is substantially equivalent to the predicate device.
{5}------------------------------------------------
### Attachment
### < Comparison Table of Technical Characteristics >
| Characteristic | Subject Device<br>Huvitz Imaging System<br>HIS-5000U | Predicate Device<br>Haag-Streit, EyeSuit Imaging |
|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | New (K161829) | K142423 |
| Manufacturer | Huvitz Co.,Ltd | Haag-Streit |
| Device Name | Huvitz Imaging System<br>HIS-5000U | EyeSuit Imaging |
| Intended Use | Huvitz Imaging system (HIS-5000U) is intended for<br>controlling digital imaging devices and for acceptance,<br>transfer, display, storage and digital processing of<br>documentation ophthalmic images and videos, acquired<br>from a slit lamp camera through a direct connection to a<br>PC or to multiple PCs via a local network. | EyeSuit Imaging is a software program intended for use for<br>controlling digital imaging devices and for acceptance,<br>transfer, display, storage and digital processing of<br>documentational ophthalmic images and videos, acquired<br>from computerized diagnostic instruments, through direct<br>connection or through networks.<br>EyeSuite Basic is a patient and examination management<br>system, acting as a base and communication platform for<br>other EyeSuite software components |
| Memory | 2,048MB | 1GByte |
| Graphic Card | 512MB | 120MB |
| Operating System | Window 7 | Windows Vista, Window XP |
| Interface | USB 2.0 | DICOM PACS |
| Monitor resolution | 1920 x 1080 | 1280 x 800 |
| How to store patient data<br>and diagnostic documents | Be comprised of a local database to store patient data and<br>diagnostic documents | Be comprised of a central database to store patient data and<br>diagnostic documents |
| Type of images and video<br>format | JPEG | JPEG, TIFF, MJPEG |
| Ability to control camera | Be capable of controlling supported camera system and | Be capable of controlling supported camera systems and |
| Characteristic | Subject Device<br>Huvitz Imaging System<br>HIS-5000U | Predicate Device<br>Haag-Streit, EyeSuit Imaging |
| systems and exposure<br>parameters like signal<br>gain, exposure time or<br>white balance (by default<br>or an option) | exposure parameters like signal gain, exposure time or white<br>balance. | exposure parameters like signal gain, exposure time or white<br>balance |
| Ability to provide the<br>management, storage and<br>processing and display of<br>patient, diagnostic video<br>and image data | USB Camera has unique power requirement that varies on<br>image resolution and data bandwidth needed to transport<br>image data | Provide the management, storage and processing and display<br>of patient, diagnostic, video and image data |
| Type of system that<br>provide a software<br>application (client) to view<br>or modify the stored data<br>in the database (server) | Software should be installed in a computer using an installer<br>file. After the software and camera driver installation, HIS-<br>5000U program can check patient image data and current<br>database information on a local computer. | Client-server systems that provide a software application<br>(client) to view or modify the stored data in the database<br>(server) |
| Functions to add test<br>notes to medical records | Diagnosis management can add or edit diagnostic<br>information. | Provide a function to add text notes to medical records |
| Function to highlight<br>features in images | User can select a thumbnail image by clicking it, which is<br>highlighted as bright color, and it is able to select several<br>images from different exams of the same patient. The<br>selections are cleared when the processing on image are<br>completed. | Provide a function to highlight features in images |
| Function to enhance<br>image brightness, contrast<br>and sharpness | Provide functions to enhance image brightness, contrast<br>and sharpness. | Provide functions to enhance image brightness, contrast and<br>sharpness |
| Characteristic | Subject Device<br>Huvitz Imaging System<br>HIS-5000U | Predicate Device<br>Haag-Streit, EyeSuit Imaging |
| Function to measure<br>pixels or mm, | Function to measure pixels and mm | Provides the measurement functionality in pixels and<br>measurements in mm |
| Function to search to<br>retrieve medical records<br>lined to a specific patient | All patient data is stored in a database file, and a search<br>function can retrieve patient database transaction history<br>stored in the Database Log File. | a search function to retrieve medical records linked to a<br>specific patient |
| Function to export and<br>import data | Include a function to export and import data | Include a function to export and import data |
| Function to print out the<br>stored data | Can produce printouts of the stored data. | Can produce printouts of the stored data |
| Function to compare<br>images by displaying<br>images side by side | Function to compare images by displaying images side by<br>side<br>More than two images can be selected from Thumbnail List<br>for comparison. | Comprise an image compare function by displaying images<br>side by side |
| Function to connect to<br>LAN and have DICOM<br>interface | Can connect to LAN but No DICOM interface | Can connect to LAN and have DICOM interface (by default or<br>as a separate option) |
{6}------------------------------------------------
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.