K161813 · William Cook Europe Aps · BTR · May 19, 2017 · Anesthesiology
Device Facts
Record ID
K161813
Device Name
Frova Intubating Introducer
Applicant
William Cook Europe Aps
Product Code
BTR · Anesthesiology
Decision Date
May 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Frova Intubating Introducer is intended to facilitate endotracheal intubation in patients where the visualization of the glottis in inadequate. The 14 French catheter has been designed for placement of a single lumen endotracheal tube whose inner diameter is 6 mm or larger.
Device Story
The Frova Intubating Introducer is a 14 French, 70 cm radiopaque catheter used to facilitate endotracheal intubation when glottis visualization is inadequate. The device features a blunt, curved tip for blind tracheal passage and two distal sideports for airflow. It may include a stiffening stylet for columnar support and Rapi-Fit adapters for ventilator connection. Used by clinicians in clinical settings, the device acts as a guide for placing a single lumen endotracheal tube (ID ≥ 6 mm). Centimeter markings on the shaft assist in depth placement. The device provides a pathway for oxygenation (mean delivered minute volume 8.6 L/min) during the intubation procedure. It benefits patients by enabling successful airway management in difficult intubation scenarios.
Clinical Evidence
Bench testing only. Performance verified via tensile strength (EN/ISO 10555-1:2013), kink radius (ISO 5361:2012), Luer Lock connection integrity (ISO 594-2), and 15mm adapter compliance (ISO 5356-1:2015). Simulated use testing in domestic swine confirmed radiopacity and ink adherence. Biocompatibility testing conducted per ISO 10993-1:2009 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity). VOC and particulate matter emission testing demonstrated clinically insignificant health risks. Accelerated aging (3 years) confirmed structural and functional integrity.
Technological Characteristics
14 French, 70 cm radiopaque catheter; blunt curved tip; two distal sideports. Materials tested for biocompatibility per ISO 10993-1. Connectivity via Luer Lock and 15mm adapters. Mechanical testing per EN/ISO 10555-1, ISO 5361, ISO 594-2, and ISO 5356-1. Sterile barrier packaging tested per ASTM F1886, F2096, and F1929.
Indications for Use
Indicated for patients requiring endotracheal intubation where glottis visualization is inadequate; facilitates placement of single lumen endotracheal tubes (ID ≥ 6 mm).
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2017
William Cook Europe ApS Henriette Christiansen Director of Regulatory Affairs Sandet 6, 4632 Bjaeverskov Denmark
Re: K161813
Trade/Device Name: Frova Intubating Introducer Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: Class II Product Code: BTR Dated: April 19, 2017 Received: April 21, 2017
Dear Henriette Christiansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 11.0 INDICATIONS FOR USE
#### INDICATIONS FOR USE
510(k) Number (if known): K161813
Device Name: Frova Intubating Introducer
#### Indications for Use:
The Frova Intubating Introducer is intended to facilitate endotracheal intubation in patients where the visualization of the glottis in inadequate. The 14 French catheter has been designed for placement of a single lumen endotracheal tube whose inner diameter is 6 mm or larger.
| Prescription Use | X |
|-----------------------------|------------------------|
| | Over-The-Counter Use |
| (Part 21 CFR 801 Subpart D) | (21 CFR 801 Subpart C) |
AND/OR
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 12.0 510(k) SUMMARY
# Frova Intubating Introducer 21 CFR 807.92 Date Prepared: May 17, 2017
| Submitted By: | |
|-----------------------|---------------------------------|
| Applicant: | William Cook Europe ApS |
| Contact: | Henriette S. Christiansen |
| Applicant Address: | William Cook Europe ApS |
| | Sandet 6 |
| | 4632 Bjaeverskov, Denmark |
| Contact Phone Number: | +45 56 86 87 56 |
| Contact Fax Number: | +45 56 86 86 96 |
| Device Information: | |
| Trade Name: | Frova Intubating Introducer |
| Common Name: | Intubating Introducer |
| Classification Name: | Tube, Tracheal (W/Wo Connector) |
| Regulation: | 21 CFR 868.5730 |
| Product Code: | BTR |
### Predicate Devices:
The predicate is the Endotracheal Tube Exchanging Guide from Sheridan Catheter Corp. (K862347) used to facilitate airway management, by assisting in placement of an endotracheal tube while maintaining airflow.
### Device Description:
The Frova Intubating Introducer is a 14 French, 70 cm long radiopaque catheter introducer with centimeter markings and a blunt, curved tip that can be passed blindly into the trachea, when visualization of the glottis is inadequate and only visualization of the epiglottis can be confirmed by laryngoscopy. The catheter introducer has a tubular design with two distal sideports to ensure adequate airflow. It can be supplied with or without a stiffening stylet and/or two Rapi-Fit adapters for connection to a ventilator device. Pressure and flow specifications for the device when used for oxygenation are as follows: mean delivered minute volume is 8.6 L/min and mean measured average maximum airway pressure is 12.1 cm H2O (when input pressure is set to 50 psi, lung compliance is 100 mL/cm H2O, and resistance is 3 cm H2O/L/s). Incremental centimeter markings are placed on the shaft of the catheter introducer, with the first measurement marking designating 10 centimeters from the distal tip. The centimeter marks along the catheter shaft facilitate accurate depth placement of the catheter introducer into the airway.
{4}------------------------------------------------
The stiffening stylet is only provided with the catalogue number C-CAE-14.0-70-FII. The stiffening stylet is 10 cm shorter than the catheter introducer and provides columnar stiffness to the catheter introducer when needed. The Rapi-Fit adapters, comprised of a 15 mm connector and a Luer Lock connector, are only provided with the catalog numbers C-CAE-14.0-70-FI and C-CAE-14.0-70-FII.
Catalogue numbers C-CAE-14.0-70-FIC and C-CAE-14.0-70-FIC-SPOPS consist of the catheter introducer only. The C-CAE-14.0-70-FIC is supplied as a box of ten catheter introducers. The C-CAE-14.0-70-FIC-SPOPS is supplied in single units in a Tyvek peelpouch package for easy transportation during special operations. The catheter introducer is curved in the middle to fit in a coat pocket or back-pack.
### Intended Use:
The Frova Intubating Introducer is intended to facilitate endotracheal intubation in patients where the visualization of the glottis in inadequate. The 14 French catheter has been designed for placement of a single lumen endotracheal tube whose inner diameter is 6 mm or larger.
### Comparison to Predicates:
The Frova Intubating Introducer and the predicate device, the Sheridan Endotracheal Tube Exchanging Guide, are substantially equivalent in that these devices have similar design and indications for use. Both are tubular devices of similar size and length and are intended for transient use to guide an endotracheal tube into its correct position within the trachea while providing continuous air flow. Additionally, the Frova Intubating Introducer is similar in technological characteristics, and methods of construction to the Sheridan Endotracheal Tube Exchanging Guide indicated for respiratory use.
# Technological Characteristics:
The Frova Intubating Introducer was subjected to the following tests to assure design and performance under the specified testing parameters:
Time Zero Testing
- . Tensile strength, catheter introducer - The minimum tensile force must be larger than or equal to 15 N in accordance with EN/ISO 10555-1:2013. The acceptance criterion was met.
- . Tensile strength, stiffening stylet - The minimum tensile force must be larger than or equal to 15 N in accordance with EN/ISO 10555-1:2013. The acceptance criterion was met.
- Tensile strength, catheter introducer/Rapi-Fit adapter connection - The minimum tensile force must be larger than or equal to 15 N in accordance with EN/ISO 10555-1:2013. The acceptance criterion was met.
- Kink radius, catheter introducer The catheter introducer must not kink when ● tested at a 40 mm Radius of Curvature according to the test method described in Annex H of ISO 5361:2012. The acceptance criterion was met.
{5}------------------------------------------------
- . Unscrewing torque, Luer Lock - When the fitting is tested in accordance with ISO 594-2. Section 5.5. the female Luer connection shall remain attached to the reference fitting. The acceptance criterion was met.
- Resistance to overriding, Luer Lock When the fitting is tested in accordance . with ISO 594-2, Section 5. 7, the reference fitting shall not override the threads of the conical fitting. The acceptance criterion was met.
- Separation Force, Luer Lock - When the fitting is tested in accordance with ISO 594-2, Section 5.4, it shall remain attached to the reference fitting. The acceptance criterion was met.
- . Compliance verification, 15mm adapter - When the fitting is tested in accordance with ISO 5356-1:2015, Section 3.1.2, the leading edge of the 15 mm Rapi-Fit adapter shall lie between the minimum and maximum diameter step gauges. The acceptance criteria were met.
- . Simulated Use - Each test article shall receive a rating of 2 or 3 on 3-point scale for each performance parameter. The acceptance criterion was met.
- . Oxygenation - A characterization study was performed to 1) establish objective evidence that the introducer catheter is able to insufflate oxygen under a specified pressure, 2) characterize the delivered minute volume through the introducer catheter during simulated high pressure oxygenation insufflation, and 3) establish objective evidence that the introducer catheter may successfully couple with the Luer Lock Rapi-Fit adapter during simulated clinical use.
- Transportation - Testing was divided into three parts:
- Part 1 visual inspection test of the shipping/product box. Damages are accepted on the Shipping boxes, as long as the product boxes are without holes/loose pieces or damages due to e.g. humidity etch. Small visible damages as pressure marks are allowed on the product boxes, as long as the sterile barrier and the Frova catheter is intact.
- -Part 2 inspection of sterile barrier (bubble leak/dye test if applicable). No breaches in the sterile barrier must be seen when the entire seal area is inspected. No visual tears or pinholes must be seen.
- Part 3 inspection of device integrity after transportation. Catheter introducer shall have a curved tip after simulated shipping test. No kinks or damages must be found/seen on the catheter introducer. Additional acceptance criteria for C-CAE-14.0-70-FII and C-CAE-14.0-70-FII: The adapter shall be able to be attached and locked to the catheter introducer according to the IFU and not be able to be taken off by pulling in the adapter when locked. The adapter shall be able to be removed from the catheter introducer according to the IFU.
All acceptance criteria were met.
- . Ink adherence and radiopacity - Testing was performed as simulated use testing in domestic swine, following which each test article was quantitatively evaluated on a 3-point scale with respect to legibility of the printed markings and radiopacity.
{6}------------------------------------------------
D-4
Each test article shall receive a rating of 2 or 3 to pass. All acceptance criteria were met.
Accelerated Age Testing (equivalent of three years real time aging)
- Tensile strength, catheter introducer The minimum tensile force must be larger ● than or equal to 15 N in accordance with EN/ISO 10555-1:2013. The acceptance criterion was met.
- . Tensile strength, stiffening stylet - The minimum tensile force must be larger than or equal to 15 N in accordance with EN/ISO 10555-1:2013. The acceptance criterion was met.
- . Tensile strength, catheter introducer/Rapi-Fit adapter connection - The minimum tensile force must be larger than or equal to 15 N in accordance with EN/ISO 10555-1:2013. The acceptance criterion was met.
- Kink radius, catheter introducer The catheter introducer must not kink when . tested at a 40 mm Radius of Curvature according to the test method described in Annex H of ISO 5361:2012. The acceptance criterion was met.
- Unscrewing torque, Luer Lock When the fitting is tested in accordance with ISO . 594-2, Section 5.5, the female Luer connection shall remain attached to the reference fitting. The acceptance criterion was met.
- Resistance to overriding, Luer Lock When the fitting is tested in accordance . with ISO 594-2, Section 5. 7, the reference fitting shall not override the threads of the conical fitting. The acceptance criterion was met.
- . Separation Force, Luer Lock - When the fitting is tested in accordance with ISO 594-2, Section 5.4, it shall remain attached to the reference fitting. The acceptance criterion was met.
- . Compliance verification, 15mm adapter - When the fitting is tested in accordance with ISO 5356-1:2015, Section 3.1.2, the leading edge of the 15 mm Rapi-Fit adapter shall lie between the minimum and maximum diameter step gauges. The acceptance criteria were met.
- . Sterile barrier and dimensional verification (accelerated aging) - Testing was divided into three parts. Part 1 visual inspection test of the Tyvek pouch in accordance with ASTM F1886-16, part 2 bubble leak test of Tyvek pouch in accordance with ASTM F2096-11 (dye test if applicable in accordance with F1929-15) and part 3 inspection of device integrity after accelerated aging. The acceptance criteria were as follows:
- -The sterile barrier shall be intact after accelerated ageing (verified with a bubble leak test in accordance with ASTM F2096-11, dye test according to ASTM F1929-15)
- The product shall appear intact as per visual inspection -
- -The angle of the tip of the catheter must not exceed the tolerances according to the applicable drawing
{7}------------------------------------------------
- -The length of the catheter must be 70 cm ± 3mm according to the applicable drawing and Incoming Quality Control Instruction
- -The ID and OD of the catheter must be according to the applicable Incoming Quality Control Instruction:
- ID: 3.124 mm ± 0.05 mm. O
- OD: 4.674 mm ± 0.05 mm O
- The print must be clear and visible according to the applicable manufacturing instructions and within the tolerances specified on the applicable drawing
- -The Rapi-Fit adapters shall be able to be attached and locked to the catheter introducer according to the IFU.
All acceptance criteria were met.
- Ink adherence and radiopacity Testing was performed as simulated use testing in . domestic swine, following which each test article was quantitatively evaluated on a 3-point scale with respect to legibility of the printed markings and radiopacity. Each test article shall receive a rating of 2 or 3 to pass. All acceptance criteria were met.
Biocompatibility
- . Biocompatibility testing was performed in accordance with ISO 10993-1:2009 and FDA guidance on Use of International Standard ISO 10993-1. Testing for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity and material mediated pyrogenicity was performed to ensure the biocompatibility of the devices. All biocompatibility tests showed passed results.
- Testing for volatile organic compounds was performed on the dry gas pathway of the ● device. No VOCs were detected greater than the quantification limits set for the specific group of compounds that were assayed and no additional VOCs were detected. The subsequent risk assessment determined the likelihood of a toxic effect from VOCs from the device negligible.
- Characterization of particulate matter emission was also performed on the dry gas pathway of the device. Results demonstrate that the Frova Intubating Introducer presents a clinically insignificant health risk to patients on which the device is used.
### Conclusion:
The results of the testing provide reasonable assurance that the Frova Intubating Introducer has been designed such that it will function as intended. The devices do not raise different questions of safety or effectiveness as compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.