Solution Set with 1.2 Micron Air Eliminating Filter
K161808 · Baxter Healthcare Corporation · FPA · Jun 21, 2017 · General Hospital
Device Facts
Record ID
K161808
Device Name
Solution Set with 1.2 Micron Air Eliminating Filter
Applicant
Baxter Healthcare Corporation
Product Code
FPA · General Hospital
Decision Date
Jun 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For administering fluids from container to a patient's vascular system through a vascular access device. Used for the removal of particulate matter and elimination of air from infusion fluids while administering. May be used with TPN solutions containing lipid emulsions.
Device Story
Single-use disposable intravascular administration set; removes particulate matter and eliminates air from infusion fluids. Components include 1.2-micron polysulfone membrane filter, Teflon air vent, acrylic housing, and PVC tubing. Used in clinical settings for fluid delivery; operated by healthcare professionals. Connects fluid container to patient vascular access device. Output is filtered infusion fluid. Benefits include reduced risk of air embolism and particulate infusion. Modifications include alternative non-DEHP PVC tubing (using Di (2-Ethylhexyl) terephthalate plasticizer), sterile packaging, and updated labeling.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included solvent bond strength, bubble tests, clamp shut-off/stability, occlusion tests, flow rate accuracy, and TPN resistance. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemocompatibility. Sterilization validated per ISO 11137-2.
Technological Characteristics
Materials: Polysulfone membrane, Teflon air vent, acrylic housing, ABS spike, PVC tubing (non-DEHP). Dimensions: 1.2-micron filter rating, 45 psi max pressure. Connectivity: Luer lock connectors. Sterilization: Radiation (SAL 10^-6). Standards: ISO 594 (Luer), ISO 10993 (Biocompatibility), ISO 11137-2 (Sterilization), USP <661> (Physiochemical).
Indications for Use
Indicated for patients requiring fluid administration from a container to the vascular system via a vascular access device, including TPN solutions with lipid emulsions. No specific age or gender contraindications stated.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Solution Set with 1.2 Micron Air Eliminating Filter (K952074)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in all caps.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 21, 2017
Baxter Healthcare Corporation Mr. John Lamela Regulatory Affairs Specialist 32650 North Wilson Road Round Lake, Illinois 60073
Re: K161808
Trade/Device Name: Solution Set with 1.2 Micron Air Eliminating Filter Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: May 19, 2017 Received: May 22, 2017
Dear Mr. Lamela:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S6
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K161808
Device Name
Solution Set with 1.2 Micron Air Eliminating Filter
Indications for Use (Describe)
For administering fluids from container to a patient's vascular system through a vascular access device. Used for the removal of particulate matter and elimination of air from infusion fluids while administering. May be used with TPN solutions containing lipid emulsions.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary - K161808
June 19, 2017
## OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
# CONTACT PERSON:
John Lamela Regulatory Specialist, Regulatory Affairs 32650 N. Wilson Road Round Lake, IL 60073 Telephone: (224) 270-2850 Fax: (224) 270-4119
# IDENTIFICATION OF THE DEVICE:
Common Name: Intravascular Administration Set Trade Name: Solution Set with 1.2 Micron Air Eliminating Filter Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Regulation Name: Set, Administration, Intravascular Regulatory Class: Class II Product Code: FPA
#### Table 1. Product Codes for Solution Sets with 1.2 Micron Air Eliminating Filter
| Code Number | Name |
|-------------|------------------------------------------------------------------------------|
| 2C1103 | INTERLINK System Non-DEHP Extension Set with 1.2 Micron Downstream<br>Filter |
| 2H8603 | CLEARLINK System Non-DEHP Extension Set |
| 2H8486 | CLEARLINK System Non-DEHP Solution Set |
| 2R8486 | CLEARLINK System Non-DEHP Solution Set |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, with the letters slanting slightly to the right. The word is the logo for the healthcare company Baxter International.
# PREDICATE DEVICE:
| Device | Company | Predicate 510(k) | Clearance Date |
|--------------------------------------------------------|----------------------------------|------------------|----------------|
| Solution Set with 1.2 Micron<br>Air Eliminating Filter | Baxter Healthcare<br>Corporation | K952074 | July 27, 1995 |
Table 2. Predicate Device(s)
## DESCRIPTION OF THE DEVICE:
The Solution Set with 1.2 Micron Air Eliminating Filter product line consists of single use disposable device for administering fluids from container to a patient's vascular system through a vascular access device. Used for the removal of particulate matter and elimination of air from infusion fluids while administering. The filter consists of a 1.2 micron filter consisting of polysulfone solution membrane and a Teflon air vent media enclosed in an acrylic housing and may be used with TPN solutions containing lipid emulsions. The 1.2 micron filter has a maximum pressure of 45 psi (2241 kPa). The Solution Set with 1.2 Micron Air Eliminating Filter was previously cleared under 510(k) premarket notification K952074 (cleared July 27, 1995).
The basis for this premarket notification is modifications to the Solution Set with 1.2 Micron Air Eliminating Filter product line. The modifications consists of the use of a Non-DEHP PVC tubing material that uses Di (2-Ethylhexyl) terephthalate as a plasticizer to provide an alternate to the current Non-DEHP PVC tubing material, conversion to sterile packaging, addition of an expiration date, and updates to the labeling.
| Code and Description | Change Summary |
|---------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| 2C1103 – INTERLINK System Non-DEHP Extension Set<br>with 1.2 Micron Downstream Filter | 1. Label Changes<br>2. Packaging Change<br>3. Addition of Exp. Date |
| 2H8603 - CLEARLINK System Non-DEHP Extension Set | 1. Label Changes<br>2. Packaging Change<br>3. Addition of Exp. Date |
| 2H8486 - CLEARLINK System Non-DEHP Solution Set | 1. Label Changes<br>2. Packaging Change<br>3. Addition of Exp. Date |
| 2R8486 - CLEARLINK System Non-DEHP Solution Set | 1. Label Changes<br>2. Packaging Change<br>3. Addition of Exp. Date |
Table 3. Summary of Proposed Changes per Product Code
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is centered and takes up most of the image.
| Code and Description | Change Summary |
|----------------------|---------------------------------------------------------------------------------------------------------|
| | 4. Use of Non-DEHP PVC Tubing Material<br>that uses Di (2-Ethylhexyl) terephthalate as a<br>plasticizer |
#### Table 3. Summary of Proposed Changes per Product Code
These modifications do not impact the intended use or the fundamental technology of the devices. The product labels are being updated to revise statements regarding latex and pump device references, add the indications for use statement that was previously cleared for this device (K952074 - cleared July 27, 1995), include expiration dating, change from "fluid path is sterile" to "sterile" (sterile packaging), and implement other modifications to comply with Baxter's labeling standards. Note: The indications for use statement from the previously cleared 510(k) was modified, per the FDAs recommendation, to include the description in 21 CFR 880.5440.
Table 4. Set Configuration Using Proposed Non-DEHP PVC Tubing Material
| Device Description | Code Number |
|-----------------------------------------------------------------------------------------|-------------|
| CLEARLINK System Non-DEHP Solution Set<br>Image: CLEARLINK System Non-DEHP Solution Set | 2R8486 |
## Table 5. Set Configuration of Current Codes
| Device Description | Code Number |
|---------------------------------------------------------------------------|-------------|
| INTERLINK System Non-DEHP Extension Set with 1.2 Micron Downstream Filter | 2C1103 |
| CLEARLINK System Non-DEHP Extension Set | 2H8603 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly blurred, giving it a sense of depth. The background is white.
Image /page/6/Figure/1 description: This image shows a table with two columns, "Device Description" and "Code Number". The device description is "CLEARLINK System Non-DEHP Solution Set", and there is a diagram of the device below the text. The code number for this device is "2H8486".
#### Table 5. Set Configuration of Current Codes
# INDICATIONS FOR USE:
For administering fluids from container to a patient's vascular system through a vascular access device. Used for the removal of particulate matter and elimination of air from infusion fluids while administering. May be used with TPN solutions containing lipid emulsions.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed devices have equivalent technological characteristics as Baxter's current legally marketed device cleared under 510(k) premarket notification K952074 (cleared July 27, 1995).
# DEVICE COMPARISON TABLE:
The comparison table below identifies similarities and differences of the proposed devices to the current legally marketed devices to which substantial equivalency is claimed.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. The background is plain white.
| | Features | Predicate Device<br>(K952074) | Proposed Devices | Assessment of<br>Differences |
|----------------------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Indications for use | The Baxter solution<br>administration sets with<br>1.2 micron Supor IV-3<br>filter will be used for the<br>removal of particulate<br>matter and elimination of<br>air from infusion fluids.<br>The 1.2 micron filter sets<br>may be used with TPN<br>solutions containing lipid<br>emulsions. | For administering fluids<br>from a container to a<br>patient's vascular system<br>through a vascular access<br>device. Used for the<br>removal of particulate<br>matter and elimination of<br>air from infusion fluids<br>while administering. May<br>be used with TPN<br>solutions containing lipid<br>emulsions. | Alignment of language<br>provided in 21 CFR<br>880.5440.<br>In addition, reference to<br>trade name (Supor IV-3)<br>filter was removed since<br>this name is no longer<br>used (filter remains the<br>same).<br>These administrative<br>changes have no impact<br>to safety or effectiveness<br>of the device. |
| | Sterile | Yes | Same | No difference |
| | Non-Pyrogenic | Yes | Same | No difference |
| | Single Use | Yes | Same | No difference |
| | Materials | | | |
| | Solution<br>Membrane | Polysulfone | Same | No difference |
| 1.2<br>Micron<br>Filter | Air Vent<br>Membrane | Teflon | Same | No difference |
| | Housing | Acrylic<br>(Polymethyl<br>Methacrylate) | Same | No difference |
| | Spike | | Acrylonitrile Butadiene<br>Styrene | Same |
| | Non-DEHP Drip<br>Chamber | Polyvinyl Chloride | Same | No difference |
| Features | Predicate Device<br>(K952074) | Proposed Devices | Assessment of<br>Differences | |
| Female Luer Lock<br>Connector | N/A | Blend of Copolyester and<br>Hytrel | The predicate device does<br>not have a female luer<br>lock connector. Design<br>control activities<br>confirmed there was no<br>impact to the safety or<br>effectiveness of the<br>device due to this change. | |
| Non-DEHP Tubing1<br>(used on codes<br>2H8486, 2C1103,<br>and 2H8603) | Material: Polyvinyl<br>Chloride<br>Plasticizer: Tris (2-<br>Ethylhexyl) Trimellitate | Same | No difference | |
| Non-DEHP Tubing2<br>(used on code<br>2R8486) | N/A | Material: Polyvinyl<br>Chloride<br>Plasticizer: Di (2-<br>Ethylhexyl) terephthalate | This is the basis of this<br>submission: modification<br>consisting of the use of a<br>Non-DEHP PVC tubing<br>material that uses Di (2-<br>Ethylhexyl) terephthalate<br>as a plasticizer.<br>Design control activities<br>have confirmed there is<br>no impact to the safety or<br>effectiveness for the new<br>material. | |
| Interlink Y-Site | N/A | Copolyester (Housing)<br>Synthetic polyisoprene<br>(Septum) | The predicate device does<br>not have an Interlink Y-<br>Site. Design control<br>activities confirmed there<br>was no impact to the<br>safety or effectiveness of<br>the device due to this<br>change. | |
| Clearlink Y-Site | N/A | Silicone Rubber (Gland)<br>Polycarbonate (Housing,<br>Housing Inlet, and Center<br>Post) | The predicate device does<br>not have a Clearlink Y-<br>Site. Design control<br>activities confirmed there<br>was no impact to the<br>safety or effectiveness of<br>the device due to this<br>change. | |
| Features | Predicate Device<br>(K952074) | Proposed Devices | Assessment of<br>Differences | |
| Two-Piece Male<br>Luer Lock<br>Connector | N/A | Acrylonitrile Butadiene<br>Styrene | The predicate device does<br>not have a Two-Piece<br>Male Luer Lock<br>Connector. Design control<br>activities confirm the<br>device meets the<br>functional test<br>requirements when used<br>in accordance with the<br>instructions for use. | |
| One-Piece Male<br>Luer Lock<br>Connector | Acrylonitrile Butadiene<br>Styrene | Same | No difference | |
# Table 6. Device Comparison
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif and the letters are closely spaced together.
# Table 6. Device Comparison
<sup>1</sup> Non-DEHP tubing material for current codes (2H8486, 2C1103, and 2H8603).
<sup>2</sup> Non-DEHP tubing material for proposed code (2R8486).
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is a dark blue color. The letters are slightly slanted to the right, giving the word a sense of movement. The background is white.
#### Table 6. Device Comparison
# DISCUSSION OF NONCLINICAL TESTS:
Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.
## Performance Data:
The following bench tests (Table 7) were conducted to evaluate the effect of using a proposed Non-DEHP PVC tubing material that uses Di (2-Ethylhexyl) terephthalate as a plasticizer. All evaluations were performed on the product code set configuration (Code: 2R8486) that is using the proposed Non-DEHP PVC tubing material.
| Test | Acceptance Criteria |
|----------------------------------------------------|-------------------------|
| Solvent Bond Tensile Strength Test | Per Baxter Test Method. |
| Solvent Bond Air Pressure Test | Per Baxter Test Method. |
| Continuous Bubble Test (Pump Compatibility Test) | Per Baxter Test Method. |
| Accumulated Bubble Test (Pump Compatibility Test) | Per Baxter Test Method. |
| Slide Clamp Shut-Off Test | Per Baxter Test Method. |
| Slide Clamp Shut-Off Test-Post 24 Hour | Per Baxter Test Method. |
| Slide Clamp 7 Day Subsystem and Tubing Damage Test | Per Baxter Test Method. |
| Slide Clamp Flow Stability Test | Per Baxter Test Method. |
| Slide Clamp Tug Stability Test | Per Baxter Test Method. |
Table 7. Performance Testing Summary
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is a dark blue color. The background is white. The word is slightly blurred.
| Test | Acceptance Criteria |
|---------------------------------------------------------------------------------------------|-------------------------|
| Upstream Occlusion Test (Pump Compatibility Test) | Per Baxter Test Method. |
| Downstream Occlusion Test (Pump Compatibility Test) | Per Baxter Test Method. |
| Roller Clamp Force Test | Per Baxter Test Method. |
| Roller Clamp Shut-Off Test | Per Baxter Test Method. |
| Roller Clamp 7 Day Subsystem and Tubing Damage Test | Per Baxter Test Method. |
| Roller Clamp Flow Stability Test | Per Baxter Test Method. |
| Roller Clamp Tug Stability Test | Per Baxter Test Method. |
| Total Parenteral Nutrition (TPN) Resistance Test | Per Baxter Test Method. |
| Roller Clamp Flow Control Test | Per Baxter Test Method. |
| Administration Set Integrity Test After Maximum Fluid Delivery<br>(Pump Compatibility Test) | Per Baxter Test Method. |
| Flow Rate Accuracy Test (Pump Compatibility Test) | Per Baxter Test Method. |
| ISO Luer Tests on Male Luer Lock Connectors | Per ISO Standard 594 |
Table 7. Performance Testing Summary
Note: All pump testing performed with the Sigma Spectrum Pump.
All tests met the acceptance criteria.
#### Biocompatibility:
Biocompatibility assessments were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for prolonged duration, external communicating device, indirect blood path, and Blue Book Memorandum G95-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing", as recommended in the IV Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]". The Biocompatibility Tests and the applicable standards are listed in Table 8.
Table 8. Biocompatibility Tests and Applicable Standards
| Biocompatibility Test | Applicable Standard |
|----------------------------------------------|---------------------|
| In Vitro Cytotoxicity | ISO10993-5 |
| Sensitization | ISO10993-10 |
| Intracutaneous (Irritation) Reactivity Assay | ISO10993-10 |
| Systemic Injection (AcuteToxicity) | ISO10993-11 |
| Sub-Chronic Toxicity (Repeat Dose) | ISO10993-11 |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif, and the letters are closely spaced together.
| Biocompatibility Test | Applicable Standard |
|----------------------------------------|---------------------|
| Material Mediated Pyrogen | ISO10993-11 |
| Hemocompatibility (In Vitro Hemolysis) | ISO10993-4 |
| USP Physiochemical tests | USP <661> |
Table 8. Biocompatibility Tests and Applicable Standards
Based upon the results of this prolonged duration, external communicating, indirect blood path testing, the Solution Sets with 1.2 Micron Air Eliminating Filter have been shown to be biocompatible and appropriate for their intended use.
## Sterilitv
The Solution Set with 1.2 Micron Air Eliminating Filter product line is sterilized with radiation. The Minimum Sterilizing Dose (MSD) required to provide a 10 ° Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility using Method No. 1 as described in ISO 11137-2, "Sterilization of health care products- Radiation-Part 2: Establishing the Sterilization Dose."
These products are labeled "Sterile, Nonpyrogenic." Package Verification testing is based on Visual Inspection, ASTM F88 Seal Strength, and ASTM F2096 Bubble Leak.
## Shelf Life
Baxter has provided aging testing to support a shelf-life claim of twenty-four (24) months.
# CONCLUSION:
The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate devices that are currently marketed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.