← Product Code [LNH](/productcode/LNH) · K161795

# MAGNETOM ESSENZA (K161795)

_Siemens Medical Solutions USA, Inc. · LNH · Nov 18, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161795

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Nov 18, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MAGNETOM ESSENZA with syngo MR E11Q is indicated for use as magnetic resonance diagnostic devices (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM ESSENZA may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.

## Device Story

MAGNETOM ESSENZA is an MRI system utilizing magnetic resonance to acquire transverse, sagittal, coronal, and oblique cross-sectional images and spectroscopic data. Operated by healthcare professionals in clinical settings, the system processes signals to display internal anatomy and function. The syngo MR E11Q software update introduces migrated features including MyoMaps for tissue composition, Quiet Suite for noise reduction, various Dot Engines for optimized exam workflows (e.g., Spine, Breast, Neuro), and advanced packages for fMRI, DTI, and liver evaluation. Hardware updates include a modified Measurement and Reconstruction System (MaRS), updated Acquisition Workplace (MRAWP), and a successor Electrical Optical Signal Converter (EOSC). Output is interpreted by physicians to assist in diagnosis and guide interventional procedures using MR-safe tools. Benefits include improved workflow efficiency, quantitative tissue analysis, and high-quality diagnostic imaging.

## Clinical Evidence

No clinical tests or animal testing were conducted. Substantial equivalence is supported by bench testing, including system and integration testing of updated hardware components and verification/validation of all software features to ensure performance as intended.

## Technological Characteristics

MR diagnostic system; hardware includes modified MaRS, MRAWP, and EOSC. Software: syngo MR E11Q. Conforms to IEC 62304 (software lifecycle), IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-33 (MR safety), and ISO 14971 (risk management). Connectivity: networked/standalone. Sterilization: N/A (non-invasive).

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- MAGNETOM ESSENZA with syngo MR D14 ([K130262](/device/K130262.md))
- MAGNETOM Aera with syngo MR E11A ([K141977](/device/K141977.md))

## Reference Devices

- MAGNETOM Amira with syngo MR E11N ([K152283](/device/K152283.md))
- MAGNETOM Aera with syngo MR E11B ([K151579](/device/K151579.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 18, 2016

Siemens Medical Solutions USA, Inc. % Mr. Cordell Fields, Esq. Regulatory Affairs Specialist 65 Valley Stream Parkway MALVERN PA 19355

Re: K161795

Trade/Device Name: MAGNETOM ESSENZA Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, LNI, MOS Dated: October 20, 2016 Received: October 21, 2016

Dear Mr. Fields:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a teal or turquoise shade. The letters are evenly spaced and appear to be horizontally aligned. The background is plain white, which makes the text stand out.

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-012<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)<br>K161795                                                            |                                                                                                  |
| Device Name<br>MAGNETOM ESSENZA                                                                |                                                                                                  |
| Indications for Use (Describe)                                                                 |                                                                                                  |

transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

The MAGNETOM ESSENZA may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.

Type of Use (Select one or both, as applicable)

| <span>                     ☒ Prescription Use (Part 21 CFR 801 Subpart D)                 </span> | <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:
 

| Department of Health and Human Services                        |
|----------------------------------------------------------------|
| Food and Drug Administration                                   |
| Office of Chief Information Officer                            |
| Paperwork Reduction Act (PRA) Staff                            |
| <a href="mailto:PRAStaff@fda.hhs.gov">PRAStaff@fda.hhs.gov</a> |

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
 information unless it displays a currently valid OMB number."*

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The text is a light blue color. The background is white.

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.

#### l. General Information

| Establishment       | Siemens Medical Solutions USA, Inc.<br>65 Valley Stream Parkway<br>Mail Code 65-1A<br>Malvern, PA 19355, USA                                                                                                                                                                                                  |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 2240869                                                                                                                                                                                                                                                                                                       |
| Date Prepared       | June 29, 2016                                                                                                                                                                                                                                                                                                 |
| Manufacturer        | SIEMENS SHENZHEN MAGNETIC RESONANCE<br>LTD.<br>Siemens MRI Center<br>Hi-Tech Industrial park (middle)<br>Gaoxin C. Ave., 2nd<br>Shenzhen 518057, P.R. CHINA<br>Registration Number: 3004754211<br>Siemens Healthcare GmbH<br>Henkestrasse 127<br>D-91052 Erlangen, Germany<br>Registration Number: 3002808157 |
| Contact Person      | Mr. Cordell L. Fields, Esq.<br>Requlatory Affairs Specialist<br>Siemens Healthcare<br>Siemens Medical Solutions USA, Inc.<br>65 Valley Stream Parkway<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Phone: (610) 448-6469<br>Fax: (610) 448-1787                                                            |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is displayed in a light blue color. The background is white.

| Device Name           | Software syngo MR E11Q for the MAGNETOM<br>ESSENZA |
|-----------------------|----------------------------------------------------|
| Trade Names:          | MAGNETOM ESSENZA                                   |
| Classification Name:  | Magnetic Resonance Diagnostic Device (MRDD)        |
| Classification Panel: | Radiology                                          |
| CFR Code:             | 21 CFR § 892.1000                                  |
| Classification:       | Class II                                           |
| Product Code:         | Primary: LNH, Secondary: LNI, MOS                  |

{5}------------------------------------------------

# SIEMENS

## II. Safety and Effectiveness Information Supporting Substantial Equivalence

The subject device, MAGNETOM ESSENZA with syngo MR E11Q, has the same intended use and indications for use as the legally marketed predicate devices.

#### Indications for Use

The MAGNETOM ESSENZA with syngo MR E11Q is indicated for use as magnetic resonance diagnostic devices (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.

The MAGNETOM ESSENZA may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.

#### Device Description

The subject device, MAGNETOM ESSENZA with software syngo MR E11Q, is the latest software version for the Siemens MR MAGNETOM ESSENZA. It is modified based on the software version syngo MR E11A, which was cleared with K141977 on November 19, 2014. The software functionality and applications are based on syngo MR E11A SW and have been migrated from this previously cleared software. Only minor adaptations were needed to support the system specific hardware and to optimize the sequence/protocols, also to support the DGSU (Digital Gradient Small signal Unit) option for GPA (Gradient Power Amplifier).

Listed below are the hardware updates to the MAGNETOM ESSENZA system with software syngo MR E11Q:

- a modified MARS(Measurement and Reconstruction System), which is the same . as the MaRS of MAGNETOM Amira system with software syngo MR E11N (K152283, cleared on December 24, 2015)
- . an updated MRAWP (MR Syngo Acquisition Workplace), which is the same as the host platform of reference device MAGNETOM Amira with syngo MR E11N (K152283, cleared on December 24, 2015)
- . a successor EOSC (Electrical Optical Signal Converter) of original COB (CAN Open Bridge)

The MAGNETOM ESSENZA with software version syngo MR E11Q will be offered exfactory (new production) as well as in-field upgrades for the currently installed MAGNETOM ESSENZA systems.

A summary of the software feature updates to the MAGNETOM ESSENZA system with software synqo MR E11Q is provided below. All migrated features listed below have been cleared previously with K141977, K151579, and K152283.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the word "SIEMENS" in a bold, sans-serif font. The color of the text is a light blue or cyan. The word is presented in all capital letters and is the central focus of the image.

## Migrated Software Features:

- MyoMaps
	- o Provides quantitative information on tissue composition
- Quiet Suite ●
	- o Provides sequences for quiet imaging
- 10-min Exam Dot Engine ●
	- o 10 minute exams for the following regions: Head, L-spine, Knee, C-spine, Brain, Shoulder, Spine, T-spine, Hip, Extremities
- Spine Dot Engine ●
	- o Delivers optimized cervical, thoracic and lumbar spine imaging
- . Breast Dot Engine
	- o Delivers optimized breast imaging
- Large Joint Dot Engine .
	- o Delivers optimized large joint imaging
- Neuro fMRI Package ●
	- o Package consisting of 3D PACE, MR 3D Inline fMRI, and MR 3D Offline fMRI
- DTI Package ●
	- o Allows for acquisition of data sets with multi-directional diffusion weighting to assess anisotropic diffusion properties of brain tissue
- Neuro Perfusion Package
- FREEZEit Body MRI Package ●
	- o Facilitates high quality diagnostic MR imaging
- . LiverLab
	- o Non-invasive evaluation of liver fat and/or iron.
- BOLD 3D Evaluation .
	- o Provides features for clinical fMRI
- DTI Evaluation ●
	- o Provides offline post-processing to generate and visualize parametric maps derived from the diffusion tensor to assess diffusion properties of brain tissue

### Technological Characteristics

The subject device, MAGNETOM ESSENZA with syngo MR E11Q and the primary predicate device, MAGNETOM ESSENZA with syngo MR D14, are substantially equivalent with regard to acquiring MR images steps/features and with regard to the operational environment, programming language, operating system and performance.

The subject device, MAGNETOM ESSENZA with syngo MR E11Q and MAGNETOM ESSENZA with software syngo MR D14 both conform to the standard for software medical devices (IEC 62304:2006) and IEC as well as NEMA standards.

{7}------------------------------------------------

# SIEMENS

While there are some different technological characteristics such as some new and modified software applications, and hardware additions these differences have been tested and the conclusions from the non-clinical data suggest that the features (of different technological characteristics with respect to the predicate device) bear an equivalent safety and performance profile as that of the predicate devices.

# Nonclinical Tests

The following performance testing was conducted on the subject device:

- All software features were verified and validated to ensure that they will perform . as intended when in the clinical environment.
- Updated hardware MaRS, MRAWP and EOSC were tested in system and integration tests.

All functionality and performance have been tested in system and integration tests.

The results from each set of tests demonstrate that the device performs as intended and is therefore substantially equivalent to the predicate devices to which it has been compared.

### Clinical Tests

No clinical tests were conducted to support the subject device and the substantial equivalence argument.

No animal testing has been performed on this device.

### Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in software development, SW testing and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the operators are healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.

The MAGNETOM ESSENZA with software syngo MR E11Q conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document. The following standards are conformed to:

| 19-4   | General   | C1:2009/(R)2012 and<br>A2:2010/(R)2012<br>(Consolidated Text) Medical                                                                                                                                       | ES60601-<br>1:2005/(R)20<br>12 and        | AAMI ANSI |
|--------|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------|
|        |           | electrical equipment – Part<br>1: General requirements for<br>basic safety and essential<br>performance (IEC 60601-<br>1:2005, MOD)                                                                         | A1:2012                                   |           |
| 19-1   | General   | Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility -<br>Requirements and tests | 60601-1-2<br>Edition<br>3:2007-03         | IEC       |
| 12-207 | Radiology | Medical electrical<br>equipment - Part 2-33:<br>Particular requirements for<br>the basic safety and<br>essential performance of<br>magnetic resonance<br>equipment for medical<br>diagnostic                | 60601-2-33<br>Edition 3.0<br>2010-03      | IEC       |
| 5-40   | General   | Medical devices -<br>Application of risk<br>management to medical<br>devices                                                                                                                                | 14971<br>Second<br>edition 2007-<br>03-01 | ISO       |
| 5-89   | General   | Medical electrical<br>equipment -- Part 1-6:<br>General requirements for<br>basic safety and essential<br>performance -- Collateral<br>Standard: Usability                                                  | 60601-1-6<br>Edition 3.1<br>2013-10       | IEC       |
| 13-8   | Software  | Medical device software -<br>Software life cycle<br>processes                                                                                                                                               | 62304 First<br>edition 2006-<br>05        | IEC       |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the word "SIEMENS" in a bold, sans-serif font. The word is displayed in a light blue or teal color. The letters are evenly spaced and appear to be part of a logo or brand name.

### Substantial Equivalence

While the new and modified software and hardware features give the subject device greater capabilities than the primary predicate device, MAGNETOM ESSENZA with software syngo MR D14 (K130262; March 1, 2013), these additional capabilities are currently cleared features of secondary predicate device, MAGNETOM Aera with software syngo MR E11A. (K141977; November 19, 2014). Therefore, we believe the subject device to be substantially equivalent to the predicate devices.

{9}------------------------------------------------

# SIEMENS

| Predicate Device Name                                               | Clearance                              | Product code  | Manufacturer                                   |
|---------------------------------------------------------------------|----------------------------------------|---------------|------------------------------------------------|
| MAGNETOM ESSENZA with<br><i>syngo</i> MR D14<br>(Primary Predicate) | K130262<br>cleared on<br>Mar. 1, 2013  | LNH           | Siemens Shenzhen<br>Magnetic Resonance<br>Ltd. |
| MAGNETOM Aera with<br><i>syngo</i> MR E11A<br>(Secondary Predicate) | K141977<br>cleared on<br>Nov. 19, 2014 | LNH, LNI, MOS | Siemens Healthcare<br>GmbH                     |

## Reference Device:

| Reference Device Name                | Clearance                                   | Product<br>code  | Manufacturer                                   |
|--------------------------------------|---------------------------------------------|------------------|------------------------------------------------|
| MAGNETOM Amira with<br>syngo MR E11N | K152283<br>cleared on<br>Dec 24, 2015       | LNH              | Siemens Shenzhen<br>Magnetic Resonance<br>Ltd. |
| MAGNETOM Aera with<br>syngo MR E11B  | K151579<br>cleared<br>September 29,<br>2015 | LNH, LNI,<br>MOS | Siemens Healthcare<br>GmbH                     |

# Conclusion as to Substantial Equivalence

There are no changes to the Indications for Use for the subject device, compared to that of the legally marketed predicate device MAGNETOM ESSENZA with software syngo MR D14.

While the updated software provides the user with additional capabilities compared to the primary predicate device, MAGNETOM ESSENZA with software version syngo MR D14, these software features have been cleared with secondary predicate device MAGNETOM Aera with software syngo MR E11A (K141977, cleared on November 19, 2014). The different technological characteristics of the new software do not raise new questions of safety and effectiveness. All features have been tested and verified and validated to be effective.

Therefore, Siemens believes that the subject device, software version syngo MR E11Q for MAGNETOM ESSENZA, is substantially equivalent to the predicate devices mentioned above.

---

**Source:** [https://fda-staging.innolitics.com/device/K161795](https://fda-staging.innolitics.com/device/K161795)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).