K161795 · Siemens Medical Solutions USA, Inc. · LNH · Nov 18, 2016 · Radiology
Device Facts
Record ID
K161795
Device Name
MAGNETOM ESSENZA
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Nov 18, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MAGNETOM ESSENZA with syngo MR E11Q is indicated for use as magnetic resonance diagnostic devices (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM ESSENZA may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.
Device Story
MAGNETOM ESSENZA is an MRI system utilizing magnetic resonance to acquire transverse, sagittal, coronal, and oblique cross-sectional images and spectroscopic data. Operated by healthcare professionals in clinical settings, the system processes signals to display internal anatomy and function. The syngo MR E11Q software update introduces migrated features including MyoMaps for tissue composition, Quiet Suite for noise reduction, various Dot Engines for optimized exam workflows (e.g., Spine, Breast, Neuro), and advanced packages for fMRI, DTI, and liver evaluation. Hardware updates include a modified Measurement and Reconstruction System (MaRS), updated Acquisition Workplace (MRAWP), and a successor Electrical Optical Signal Converter (EOSC). Output is interpreted by physicians to assist in diagnosis and guide interventional procedures using MR-safe tools. Benefits include improved workflow efficiency, quantitative tissue analysis, and high-quality diagnostic imaging.
Clinical Evidence
No clinical tests or animal testing were conducted. Substantial equivalence is supported by bench testing, including system and integration testing of updated hardware components and verification/validation of all software features to ensure performance as intended.
Technological Characteristics
MR diagnostic system; hardware includes modified MaRS, MRAWP, and EOSC. Software: syngo MR E11Q. Conforms to IEC 62304 (software lifecycle), IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-33 (MR safety), and ISO 14971 (risk management). Connectivity: networked/standalone. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for patients requiring MR diagnostic imaging of head, body, or extremities to visualize internal structure/function via cross-sectional or spectroscopic images. Used for diagnostic purposes and interventional procedures with MR-compatible devices. No specific age or gender contraindications listed.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2016
Siemens Medical Solutions USA, Inc. % Mr. Cordell Fields, Esq. Regulatory Affairs Specialist 65 Valley Stream Parkway MALVERN PA 19355
Re: K161795
Trade/Device Name: MAGNETOM ESSENZA Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, LNI, MOS Dated: October 20, 2016 Received: October 21, 2016
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a teal or turquoise shade. The letters are evenly spaced and appear to be horizontally aligned. The background is plain white, which makes the text stand out.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-012<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)<br>K161795 | |
| Device Name<br>MAGNETOM ESSENZA | |
| Indications for Use (Describe) | |
transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM ESSENZA may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.
Type of Use (Select one or both, as applicable)
| <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
| Department of Health and Human Services |
|----------------------------------------------------------------|
| Food and Drug Administration |
| Office of Chief Information Officer |
| Paperwork Reduction Act (PRA) Staff |
| <a href="mailto:PRAStaff@fda.hhs.gov">PRAStaff@fda.hhs.gov</a> |
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The text is a light blue color. The background is white.
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
#### l. General Information
| Establishment | Siemens Medical Solutions USA, Inc.<br>65 Valley Stream Parkway<br>Mail Code 65-1A<br>Malvern, PA 19355, USA |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 2240869 |
| Date Prepared | June 29, 2016 |
| Manufacturer | SIEMENS SHENZHEN MAGNETIC RESONANCE<br>LTD.<br>Siemens MRI Center<br>Hi-Tech Industrial park (middle)<br>Gaoxin C. Ave., 2nd<br>Shenzhen 518057, P.R. CHINA<br>Registration Number: 3004754211<br>Siemens Healthcare GmbH<br>Henkestrasse 127<br>D-91052 Erlangen, Germany<br>Registration Number: 3002808157 |
| Contact Person | Mr. Cordell L. Fields, Esq.<br>Requlatory Affairs Specialist<br>Siemens Healthcare<br>Siemens Medical Solutions USA, Inc.<br>65 Valley Stream Parkway<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Phone: (610) 448-6469<br>Fax: (610) 448-1787 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is displayed in a light blue color. The background is white.
| Device Name | Software syngo MR E11Q for the MAGNETOM<br>ESSENZA |
|-----------------------|----------------------------------------------------|
| Trade Names: | MAGNETOM ESSENZA |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH, Secondary: LNI, MOS |
{5}------------------------------------------------
# SIEMENS
## II. Safety and Effectiveness Information Supporting Substantial Equivalence
The subject device, MAGNETOM ESSENZA with syngo MR E11Q, has the same intended use and indications for use as the legally marketed predicate devices.
#### Indications for Use
The MAGNETOM ESSENZA with syngo MR E11Q is indicated for use as magnetic resonance diagnostic devices (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
The MAGNETOM ESSENZA may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.
#### Device Description
The subject device, MAGNETOM ESSENZA with software syngo MR E11Q, is the latest software version for the Siemens MR MAGNETOM ESSENZA. It is modified based on the software version syngo MR E11A, which was cleared with K141977 on November 19, 2014. The software functionality and applications are based on syngo MR E11A SW and have been migrated from this previously cleared software. Only minor adaptations were needed to support the system specific hardware and to optimize the sequence/protocols, also to support the DGSU (Digital Gradient Small signal Unit) option for GPA (Gradient Power Amplifier).
Listed below are the hardware updates to the MAGNETOM ESSENZA system with software syngo MR E11Q:
- a modified MARS(Measurement and Reconstruction System), which is the same . as the MaRS of MAGNETOM Amira system with software syngo MR E11N (K152283, cleared on December 24, 2015)
- . an updated MRAWP (MR Syngo Acquisition Workplace), which is the same as the host platform of reference device MAGNETOM Amira with syngo MR E11N (K152283, cleared on December 24, 2015)
- . a successor EOSC (Electrical Optical Signal Converter) of original COB (CAN Open Bridge)
The MAGNETOM ESSENZA with software version syngo MR E11Q will be offered exfactory (new production) as well as in-field upgrades for the currently installed MAGNETOM ESSENZA systems.
A summary of the software feature updates to the MAGNETOM ESSENZA system with software synqo MR E11Q is provided below. All migrated features listed below have been cleared previously with K141977, K151579, and K152283.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the word "SIEMENS" in a bold, sans-serif font. The color of the text is a light blue or cyan. The word is presented in all capital letters and is the central focus of the image.
## Migrated Software Features:
- MyoMaps
- o Provides quantitative information on tissue composition
- Quiet Suite ●
- o Provides sequences for quiet imaging
- 10-min Exam Dot Engine ●
- o 10 minute exams for the following regions: Head, L-spine, Knee, C-spine, Brain, Shoulder, Spine, T-spine, Hip, Extremities
- Spine Dot Engine ●
- o Delivers optimized cervical, thoracic and lumbar spine imaging
- . Breast Dot Engine
- o Delivers optimized breast imaging
- Large Joint Dot Engine .
- o Delivers optimized large joint imaging
- Neuro fMRI Package ●
- o Package consisting of 3D PACE, MR 3D Inline fMRI, and MR 3D Offline fMRI
- DTI Package ●
- o Allows for acquisition of data sets with multi-directional diffusion weighting to assess anisotropic diffusion properties of brain tissue
- Neuro Perfusion Package
- FREEZEit Body MRI Package ●
- o Facilitates high quality diagnostic MR imaging
- . LiverLab
- o Non-invasive evaluation of liver fat and/or iron.
- BOLD 3D Evaluation .
- o Provides features for clinical fMRI
- DTI Evaluation ●
- o Provides offline post-processing to generate and visualize parametric maps derived from the diffusion tensor to assess diffusion properties of brain tissue
### Technological Characteristics
The subject device, MAGNETOM ESSENZA with syngo MR E11Q and the primary predicate device, MAGNETOM ESSENZA with syngo MR D14, are substantially equivalent with regard to acquiring MR images steps/features and with regard to the operational environment, programming language, operating system and performance.
The subject device, MAGNETOM ESSENZA with syngo MR E11Q and MAGNETOM ESSENZA with software syngo MR D14 both conform to the standard for software medical devices (IEC 62304:2006) and IEC as well as NEMA standards.
{7}------------------------------------------------
# SIEMENS
While there are some different technological characteristics such as some new and modified software applications, and hardware additions these differences have been tested and the conclusions from the non-clinical data suggest that the features (of different technological characteristics with respect to the predicate device) bear an equivalent safety and performance profile as that of the predicate devices.
# Nonclinical Tests
The following performance testing was conducted on the subject device:
- All software features were verified and validated to ensure that they will perform . as intended when in the clinical environment.
- Updated hardware MaRS, MRAWP and EOSC were tested in system and integration tests.
All functionality and performance have been tested in system and integration tests.
The results from each set of tests demonstrate that the device performs as intended and is therefore substantially equivalent to the predicate devices to which it has been compared.
### Clinical Tests
No clinical tests were conducted to support the subject device and the substantial equivalence argument.
No animal testing has been performed on this device.
### Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in software development, SW testing and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the operators are healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
The MAGNETOM ESSENZA with software syngo MR E11Q conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document. The following standards are conformed to:
| 19-4 | General | C1:2009/(R)2012 and<br>A2:2010/(R)2012<br>(Consolidated Text) Medical | ES60601-<br>1:2005/(R)20<br>12 and | AAMI ANSI |
|--------|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------|
| | | electrical equipment – Part<br>1: General requirements for<br>basic safety and essential<br>performance (IEC 60601-<br>1:2005, MOD) | A1:2012 | |
| 19-1 | General | Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility -<br>Requirements and tests | 60601-1-2<br>Edition<br>3:2007-03 | IEC |
| 12-207 | Radiology | Medical electrical<br>equipment - Part 2-33:<br>Particular requirements for<br>the basic safety and<br>essential performance of<br>magnetic resonance<br>equipment for medical<br>diagnostic | 60601-2-33<br>Edition 3.0<br>2010-03 | IEC |
| 5-40 | General | Medical devices -<br>Application of risk<br>management to medical<br>devices | 14971<br>Second<br>edition 2007-<br>03-01 | ISO |
| 5-89 | General | Medical electrical<br>equipment -- Part 1-6:<br>General requirements for<br>basic safety and essential<br>performance -- Collateral<br>Standard: Usability | 60601-1-6<br>Edition 3.1<br>2013-10 | IEC |
| 13-8 | Software | Medical device software -<br>Software life cycle<br>processes | 62304 First<br>edition 2006-<br>05 | IEC |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the word "SIEMENS" in a bold, sans-serif font. The word is displayed in a light blue or teal color. The letters are evenly spaced and appear to be part of a logo or brand name.
### Substantial Equivalence
While the new and modified software and hardware features give the subject device greater capabilities than the primary predicate device, MAGNETOM ESSENZA with software syngo MR D14 (K130262; March 1, 2013), these additional capabilities are currently cleared features of secondary predicate device, MAGNETOM Aera with software syngo MR E11A. (K141977; November 19, 2014). Therefore, we believe the subject device to be substantially equivalent to the predicate devices.
{9}------------------------------------------------
# SIEMENS
| Predicate Device Name | Clearance | Product code | Manufacturer |
|---------------------------------------------------------------------|----------------------------------------|---------------|------------------------------------------------|
| MAGNETOM ESSENZA with<br><i>syngo</i> MR D14<br>(Primary Predicate) | K130262<br>cleared on<br>Mar. 1, 2013 | LNH | Siemens Shenzhen<br>Magnetic Resonance<br>Ltd. |
| MAGNETOM Aera with<br><i>syngo</i> MR E11A<br>(Secondary Predicate) | K141977<br>cleared on<br>Nov. 19, 2014 | LNH, LNI, MOS | Siemens Healthcare<br>GmbH |
## Reference Device:
| Reference Device Name | Clearance | Product<br>code | Manufacturer |
|--------------------------------------|---------------------------------------------|------------------|------------------------------------------------|
| MAGNETOM Amira with<br>syngo MR E11N | K152283<br>cleared on<br>Dec 24, 2015 | LNH | Siemens Shenzhen<br>Magnetic Resonance<br>Ltd. |
| MAGNETOM Aera with<br>syngo MR E11B | K151579<br>cleared<br>September 29,<br>2015 | LNH, LNI,<br>MOS | Siemens Healthcare<br>GmbH |
# Conclusion as to Substantial Equivalence
There are no changes to the Indications for Use for the subject device, compared to that of the legally marketed predicate device MAGNETOM ESSENZA with software syngo MR D14.
While the updated software provides the user with additional capabilities compared to the primary predicate device, MAGNETOM ESSENZA with software version syngo MR D14, these software features have been cleared with secondary predicate device MAGNETOM Aera with software syngo MR E11A (K141977, cleared on November 19, 2014). The different technological characteristics of the new software do not raise new questions of safety and effectiveness. All features have been tested and verified and validated to be effective.
Therefore, Siemens believes that the subject device, software version syngo MR E11Q for MAGNETOM ESSENZA, is substantially equivalent to the predicate devices mentioned above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.